OPHIR mks VEGA User manual

4.5. Energy or Average Power Measurement
Warning:
Do not exceed maximum sensor limits for power, energy, power density
and energy density as listed in tables 5 and 6 in section 10.2 of the manual,
Sensor Specifications. Otherwise, there is a risk of damaging the absorber.
With the pyroelectric sensor, you have been supplied a test slide with the
same coating as on your pyroelectric detector. You can also obtain this slide
from your dealer. You should use this slide to test the damage threshold with
your laser pulses. If the slide shows damage, then either enlarge your beam
or lower the laser energy until damage is no longer seen.
4.6. To choose Energy or Average Power Measurement
To switch from the main energy measurement screen to power measurement,
press “Power”. To switch from the main power measurement screen to energy
measurement, press “Energy”.
Vega Quick Reference
VEGA
Quick Reference
Ophir Optronics Solutions LTD.
Science Based Industry Park P.O.B. 45021, Jerusalem 9145001, Israel
Phone +972-2-5484444
https://www.ophiropt.com/photonics
rev 1.84-3/17.Feb.19/be Ophir P/N 1J06048
1Getting Started
The Vega is equipped with “soft keys”. That is, the functions of
the keys change as indicated by the legend above each key. It
is also equipped with 4 navigation keys and a round enter key in
the center.
To connect sensor to the Vega meter:
Insert the 15 pin D type connector of the measuring sensor
cable into the socket marked “Head Input” on the rear panel of
the Vega meter.
To switch the Vega on:
Briefly press the on/off switch (bottom-most key). The unit will
switch on, and the display will appear. Note that the sensor
must be plugged in before the unit is switched on. The backlight
for the Vega’s LCD can be configured to toggle between full
and partial illumination. Similarly, the keypad lighting can be
configured to toggle between on and off. This toggling will be
performed by briefly pressing the on/off switch after the Vega
has been switched on. To switch the Vega off, press the on/off
switch and hold it for ~ 3 seconds until the display blanks.
If you wish to save the current Vega configuration, use the
“Configure” function. (See section 2.3.2.).
To set the display modes:
1. Disconnect the sensor and switch off then on again. “Head
Disconnected” will appear.
2. Select “Display Mode” and enter. Choose the desired mode,
press save. The Vega will now be in this display mode for
all sensors.
3. Select “Date” and press the enter button. Using the
navigation keys set to the correct date. Press save.
4. Select “Time” and enter. Using the navigation keys, set to
the correct time. Press save.
5. For adjusting RS232 baud rate and An. Output voltage see
the main manual.
4
Pyroelectric or Photodiode Energy Sensors
4.1. Zeroing Instrument against Sensor
For most accurate calibration, you should zero the pyroelectric sensor
against the Vega it is being used with. Proceed as follows:
Make sure the sensor is in a quiet environment and not subject to pulsed
radiation. Plug sensor into Vega and turn on. Press “Menu”, select “Zero”
and enter. Press “Start”. When “ZEROING COMPLETED” appears, press
“Save”.
4.2. Changing Chosen Wavelengths (metallic type only)
1. Select “Laser” and Enter.
2. Select the wavelength you wish to use and enter. To save this
wavelength as the startup default, press “Save” before exiting.
3. To modify one of the wavelengths, select the closest wavelength and
press the right navigation key.
4. Using the up/down keys to change the numbers and the right/left keys
to move to the next number, modify to the wavelength you want.
5. When finished, press “Save” and exit. If you only want a temporary
change, exit without “Save”.
4.3. To Set Type of Laser Being Used
From main measurement screen, select “Laser” and enter. Choose the
desired laser type or wavelength. To save this as the startup wavelength,
press “Save” before exiting.
4.4. Setting Startup Configuration
1. Turn on the Vega with the sensor connected, select “Menu” and Enter.
Select “Configure”.
2. With the navigation keys select “Startup Mode”, select “Power” or
“Energy” as the startup mode. Press the Enter key to exit.
3. With the navigation keys select “Display Mode”. Press the Enter
key and select the type of screen you wish to start up with. Press
“Save” and exit. If you do not want to save this as the startup mode
but only change for now, exit without save. (This choice is saved in
the instrument and not the sensor so the Vega will start up from the
screen selected for all sensors).
4. Select “Energy Range”, press Enter and choose the manual energy
range you wish to be the default.
5. Select “Average” to select the average power period over or select the
default, “NONE”.
6. If the sensor is a diffuser sensor, set diffuser to IN or OUT.
7. Select “Max Pulse Len” and choose the shortest pulse length setting
that is longer than your laser’s pulse length.
Warning: Incorrect readings will result if pulse length is not set up correctly.
8. Select “Laser” and select the laser type you wish to be the default.
9. Press “Save” to save all present settings then “Exit” to get back to the
main measurement screen.
Note: Some sensors have only the long pulse setting for all pulse lengths.
In that case N/A appears. For further details on configuration see Sections
6.2 and 6.3 of the manual.

then move to the wavelength you wish to change or add. Press the
right navigation key.
3. Using the up/down keys to change each number and the right left keys
to move to the next number, key in the desired wavelength. Press enter
twice. To save this new wavelength as a startup default, press “Save”,
then press Enter.
4. Follow the directions on configuration in section 3.2 below.
To choose manual or automatic ranging or dBm in power measurement:
1. Press “Range” and then select the appropriate manual range,
autorange or dBm (logarithmic scale).
2. Press Enter to return to the main measurement screen. To save the
selected range as the startup default, press “Save” before exiting.
3.2. Setting the Startup Configuration
1. Select “Menu” and press Enter. Select “Configure” and Enter.
2. Select “Filter” and Enter. Choose filter in or out to be the default.
3. Select “Display Mode” and choose which screen you want the
instrument to start up in. Press Enter. (Note that this is an instrument
setting and will be valid for all sensors.)
4. Select “Power Range” and choose the manual power range you wish
to be the default or choose autorange or dBm.
5. Select “Average” and choose which period to average power over or
leave the default “NONE”.
6. Select “Laser” to choose the startup laser wavelength. When you are
finished, to save this as the default startup, press “Save”. If you only
want to change for now, exit without saving. For further details on
configuration see Section 5.3 of the main manual.
3.3. Power Measurement
Warning:
Do not exceed maximum sensor limits for power, energy, power density
and energy density as listed in tables 5 and 6 in section 10.2 of the manual,
Sensor Specifications. Otherwise, there is a risk of damaging the absorber.
To expand the bargraph scale ±5x about the present reading:
1. From the main power measurement screen, press the “Zoom” button.
2. Press the “Zoom” button again to return to the ordinary bargraph.
To subtract background and set current reading to zero:
1. From the main power measurement screen press the “Offset” button.
2. Press “Offset” again to cancel. See Section 5.4.5 of the main manual
for details.
To use the Vega to fine-tune laser power:
1. From the main power measurement screen select “Menu” and Enter.
Select “Tune” and Enter.
2. Set the percentage range of the power scale to be displayed by
repeatedly pressing the ±50% key.
3. Now select the horizontal sweep time you desire by selecting “Graph”
and choosing the time scale.
4. You may now use the screen to tune the laser power. Press “Power” to
exit to the main measurement screen. See Section 4.4.2.3 of the manual
for details.
2.3. Power or Single Shot Energy Measurement
Warning:
Do not exceed maximum sensor limits for power, energy, power density and
energy density as listed in tables 5 and 6 in section 10.2. of the main manual.
Otherwise, there is a risk of damaging the absorber.
To use the Vega to measure laser power:
Verify that the display shows “Menu: Power” in the upper right hand side of
the display. If the display shows “Menu: Energy” and units of J, mJ etc. then
press “Power” to switch to the power measurement mode.
To select an analog needle type display for the Vega:
Select “Menu” and enter. Select “Needle” and enter. The display will now
show a simulated analog needle type display. In order to expand the display,
press “Expand”.
To expand the bargraph scale ±5x about the present reading:
1. From the main power measurement screen press the “Zoom” button.
2. Press the “Zoom” button again to return to the ordinary bargraph.
To subtract background and set current reading to zero:
1. From the main power measurement screen press the “Offset” button.
2. Press “Offset” again to cancel. See Section 4.4.2.2. of the main manual
for full details.
To use the Vega to fine-tune laser power:
1. From the main power measurement screen select “Menu” and enter.
Now select “Tune” and enter.
2. Set the percentage range of the power scale to be displayed by
repeatedly pressing the ±50% key.
3. Now select the horizontal sweep time you desire by selecting “Graph”
and choosing the time scale.
4. You may now use the screen to tune the laser power. See Section
4.4.2.3. of the main manual for full details.
5. Press “Power” to exit to the main measurement screen.
To use the Vega to measure laser energy:
1. Verify that the display shows “Energy” in the upper right hand side of
the display. If the display shows “Power” and units of W (or mW etc)
then press “Energy” to switch to the energy measurement mode.
2. The energy mode is manual ranging. Select “Range”, enter then use the
navigation keys to select the proper range. Press “Save” if you want
this range to be the startup default. The correct range is the lowest one
that is larger than the pulse energy measured. Note that the range will
be displayed in the lower right corner of the main display.
3. When the Vega screen flashes “READY”, on and off, fire the laser. See
Section 4.5 of the main manual for full details.
3Photodiode Sensors
3.1. Selecting Chosen Wavelengths
1. While the Vega is off, plug in the sensor then switch it on again. From
the main measurement screen, press “Laser” to select the correct laser
wavelength. If you want to save this new wavelength as the startup
default, press “Save” before exiting.
2. If the wavelength you want is not among the wavelengths in the
six wavelengths list and you want to change or add a wavelength,
To zero instrument:
1. Disconnect the sensor, turn off then on again. “Head Disconnected” will
be displayed.
2. Make sure instrument is not in an electrically noisy environment and is
undisturbed.
3. Press “Zero” and “Start”. Wait until “ZEROING COMPLETED” appears.
For more details, see main manual section 3.5.5.
2Thermal Sensors
2.1. Use of Vega with Thermal Type Sensors
To set type of laser being used:
1. While the Vega is off, plug in the sensor then switch it on again.
2. Press the navigation keys until “Laser” is highlighted. Press the enter
key and select the appropriate laser wavelength.
3. If you want to save this new wavelength as the startup default, press
“Save”.
4. Return to main screen by pressing the enter key.
To choose manual or automatic ranging in power measurement:
1. Select “Range” with the navigation keys. Press the Enter key.
2. Select the appropriate manual range or AUTO.
3. If you want to save this new range as the startup default, press “Save”.
4. Press the Enter key to return to the main measurement screen.
To choose power or energy measurement:
To go from the main power measurement screen to the energy
measurement screen, press the lower left soft key “Energy”. To go from the
energy measurement screen to power measurement, press “Power”.
2.2. Setting and Saving the Startup Configuration
1. Turn on the Vega with the sensor connected, select “Menu” and enter.
Select “Configure” and enter.
2. With the navigation keys select “Startup mode”. Press the Enter key
and select “Power” or “Energy” as the startup mode. Press the Enter
key again to exit. Press “Save” to make this change permanent.
3. Select “Display Mode”. Press the Enter key and select the type of
screen you wish to start up with. Press Save to set this as the default
display mode. If you do not want to save this as the startup mode but
only change for now, press the Enter key. (This choice is saved in the
instrument and will apply to all sensors that work with this Vega).
4. Select “Power Range”. Press the Enter key and choose the manual
power range you wish to be the default or choose auto range.
5. Select “Average” to select what period you wish to average power
over or select the default, “NONE”. See section 4.4.2.4 of the main
manual for details.
6. Select “Energy Range” and choose the default energy range.
7. Select “Threshold” if you wish to change the energy threshold. Select
“Laser” to select the laser type you wish to be the default. Press “Save”
if you want to save this configuration, then press Exit to return to the
main measurement screen.
Further details on configuration in section 4.3. of the main manual.
Other OPHIR Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

OHAUS
OHAUS MB90 quick start guide

Gossen MetraWatt
Gossen MetraWatt lunasix Operating instructions manual

Benning
Benning IT 101 operating manual

Absolute Process Instruments
Absolute Process Instruments API 4003 G I quick start guide

SUTO
SUTO S461 Instruction and operation manual

eXact
eXact Eco-Check instruction manual