AMG Physio Logic User manual

Automatic Wrist
Blood Pressure Monitor
TM
ys
Instruction Manual
Features:
• Step Valve Technology
• Memory for up to 30 readings
• One-Touch Operation
• Easy-to-read Display
• Time and Date Display
106-960

Table of Contents
Step valve measuring technology ........................3
About blood pressure .................................3
Blood pressure standard ...............................4
Blood pressure fluctuation .............................4
Before taking a measurement ...........................5
Installing batteries....................................7
Recalling memory....................................8
Time adjustment .....................................8
Statement of accuracy .................................3
Name / function of each part............................6
Carrying case .......................................6
Display explanation ..................................7
Applying the arm cuff.................................9
Correct measuring posture .............................9
Taking a measurement ...............................10
Troubleshooting ....................................11
Cautionary notes ....................................11
Specifications ......................................12
2

3
Step Valve Measuring Technology
Statement of Accuracy
This Physio Logic™ monitor uses the oscillometric method to
detect your blood pressure. Before the cuff starts inflating, the
device will establish a baseline cuff pressure equivalent to the air
pressure. Your Physio Logic™ monitor will determine the
appropriate inflation level based on one’s pressure oscillations,
followed by cuff deflation. Step valve technology uses an
electronic valve system as opposed to the traditional mechanical
valve system, allowing the monitor to analyse oscillations during
measurement and adjust cuff pressure accordingly.Adjustments in
cuff pressure occur in steps and result in an audible ticking as the
valve releases air from the cuff. Releasing pressure in steps allows
the monitor to analyse oscillations in the cuff and evaluate them for
accuracy. During the deflation, the device will detect the amplitude
and slope of the pressure oscillations and determine the systolic,
diastolic, and pulse for you.
Blood pressure measurements determined with Physio Logic™
are equivalent to those obtained by a trained observer using cuff /
stethoscope auscultation method, within the limits prescribed by
the American National Standard for Electronic or Automated
Sphygmomanometers (ANSI /AAMI).
Blood pressure is the pressure exerted on the arterial wall while
blood flows through the arteries. The pressure measured when the
heart contracts and sends blood out of the heart is systolic
(highest). The pressure measured when the heart dilates with blood
flowing back into the heart is called diastolic (lowest) blood
pressure.
About Blood Pressure

4
Blood Pressure Standard
The Joint National Committee on prevention, detection, evalua-
tion, and treatment of high blood pressure has developed a blood
pressure standard, according to which areas of low and high risk
blood pressure are identified. This standard is a guideline as blood
pressure varies among different people and different age groups...
It is important that you consult with your physician regularly.
Your physician will tell you your normal blood pressure range as
well as the point at which you will be considered at risk.
Optimal <120 and <80
Normal <130 and <85
High-normal 130-139 or 85-89
Hypertension Stage 1 140-159 or 90-99
Hypertension Stage 2 160-179 or 100-109
Hypertension Stage 3 >180 or >110
Category Systolic (mm Hg) Diastolic (mm Hg)
Blood Pressure for Adults Age 18 and Older
Blood Pressure Fluctuation
Blood pressure fluctuates all the time! You should not be overly
worried if you encounter two or three measurements at high levels.
Blood pressure changes over the month and throughout the day.
(Case:
Male, 35 years old)
Wake
up
Arrive
office
Leave
office
Dinner
/ Sleep
Telephone
Discussion
Argument
in a meeting
18:00 Midnight
12:00
6:00
50
100
150 mmHg Systolic
Diastolic

5
Before Taking a Measurement
To help you obtain more accurate readings:
• Urinate if necessary.
• Remove any clothing and jewellery from your left arm. Placing
the cuff on anything other than bare skin can result in an
inaccurate reading.
• Sit in a comfortable position with your legs and ankles
uncrossed and your back supported.
• Sit quietly for 15 minutes before measuring your blood pressure.
• If you need to retake a reading, wait at least 5 minutes to allow
your circulation to return to normal.
• Pick a time of day, and take your blood pressure at about the
same time every day. Since your blood pressure varies at
different times of the day, this will make it easier for you to
compare your daily blood pressure readings.A good time to take
your blood pressure is in the morning, before breakfast.
• Don't smoke, eat or exercise for 30 minutes before taking your
blood pressure. These activities can change your blood pressure.
• Do not measure your blood pressure immediately after
consuming a large meal. To obtain more accurate readings,
please wait one hour before measuring.
• Only take your blood pressure in a quiet place and when you are
relaxed. Don't take your blood pressure when you are under
stress.Any stress will raise your blood pressure.
• Position your arm so that the cuff is at the level of your heart.
• If the monitor is stored at very low temperature (near freezing),
place it in a warm location for at least one hour before using it.

6
Name / Function of Each Part
Carrying Case
Recall key
Battery
cover
LCD display Wrist cuff
Add to
memory key
Start key
Power key
2 “AAA” size, 1.5 V Alkaline batteries
Please store your blood pressure monitor in the carrying case as
shown below.
Time set key

7
Display Explanation
Installing Batteries
1. Slide battery cover off.
2. Install lower battery first.
Ensure correct polarity.
Battery type: 2Alkaline 1.5 V
LR03 (AAA) size.
3. Replace the battery cover.
Replace batteries when:
1. Low battery icon appears
on display.
2. Nothing appears in the
display when the power
is switched on.
mmHg
Systolic pressure
Memory mark
Month
Date
Hour
Minute
Diastolic pressure
Pumping mark
Pulse rate
Weak battery mark
Appears when a
measurement value is
stored in memory or is
recalled from memory.
Appears while inflating.
Appears when batteries
should be replaced.
Occurs when a error was
made during measurement.
Shows the pulse rate per
minute.

8
Recalling Memory
Time Adjustment
1. Press and hold the “ ” button for more
than 3 seconds to enter Time Adjustment
Mode. The display will show a blinking
year.
2. Press the “ ” or “ ” to change the year.
3. Press the “ ” button again to confirm the
entry.
4. The display will then show a blinking
Month, followed by the Date, the Hour
and the Minutes. Repeat steps 2 and 3 to
adjust them.
5. A“0” will reappear as
the blood pressure
monitor is ready for
measurement again.
+–
Hold for
more than
3 seconds
1. To recall stored blood pressure readings
from memory, simply press the Recall key.
The last set of memorized readings will be
displayed.
2. Another press of the Recall key will recall
the previous set of readings.
Note: The data will be erased when any of the
batteries are removed for at least 10 seconds
mmHg
r
e
c
a
l
l
r
a
p
p
e
l
a
d
d
a
j
o
u
t

Applying the Arm Cuff
Correct Measuring Posture
1. Remove all watches, jewellery, etc.
prior to attaching the wrist monitor.
Clothing sleeves should be rolled up
and the cuff should be wrapped on bare
skin for correct measurements.
2. Apply cuff to the left wrist with the
palm facing up as in FigureA.
3. Make sure the edge of the cuff is about
1 cm from the palm as in Figure B.
4. Fasten the velcro strap securely around
your wrist so there is no extra space
between the cuff and the wrist as in
Figure C. If the cuff is not wrapped
snugly enough (snug, not tight), the
measurement will be false.
5. If your physician has diagnosed you
with poor circulation in your left arm,
carefully place the cuff around your
right wrist as shown in Fig. D.
1. Place your elbow on a table so that the
cuff is at the same level as your heart.
2. If the cuff is not at the same level as
your heart, use a soft object such as a
folded towel to support your arm.
3. Turn your palm upwards, sit upright in
a chair, and take 5-6 deep breaths.
Avoid leaning back.
9
Fig. A
Fig. B
Fig. C
Fig. D
(X)

Taking a Measurement
1. Wrap the cuff on the wrist. Press the
POWER key. All digits will light up
checking the display functions. This will
take approximately 2 seconds.
2. After all the symbols appear, the display
will show a “0”. The monitor is now ready
to measure.
3. Press the START key. Once the monitor
inflates the cuff to approximately 180
mmHg, the measurement will begin.
4. When the measurement is completed,
Notes: This monitor automatically switches
off approximately 3 minutes after last key
operation.
To interrupt the measurement, simply
press any key. The cuff will deflate
immediately after any key is pressed.
During the measurement, do not talk or
move your arm or hand muscles.
systolic, diastolic and pulse will show
simultaneously on the LCD screen.
Note: The monitor will re-inflate
automatically to approximately 220
mmHg if the system detects that it needs
more pressure to do a measure.
5. To save the values in the memory system,
press theAdd to memory key. The Memory
mark will appear for 1 second. Up to 30
measurements can be saved in the memory.
10
mmHg
mmHg

Troubleshooting
Symptom:
Correction:
Symptom:
Correction:
No display when the Power button is pressed.
If the batteries have run down, replace them with
new alkaline batteries. If the batteries are not in the correct
position, re-insert them with the correct polarities.
mark shown on display or the blood pressure value
is displayed excessively low (high).
Wrap the cuff properly so that it is positioned
correctly. If you talked or moved during the measurement, measure
it again quietly. If the wrist was shaken during the measurement,
repeat measurement.
Note: If the unit does not work, please call our consumer
information line at 1-800-363-2381. Under no circumstance
should you disassemble the unit and attempt repairs.
EE
11
Cautionary Notes
1. Avoid extreme temperatures, humidity, and direct sunlight.
Avoid strongly shocking the main unit, and protect it from dust.
2. Clean the blood pressure monitor body and the cuff carefully
with a slightly damp, soft cloth. Do not press. Do not wash the
cuff or use chemical cleaner.
3. Do not press the Start key when the cuff is not on the wrist.
4. Remove the batteries when the unit is not used for a long time.
5. The unit should not be operated by children.
6. This device might not be suitable for use by people with poor
blood circulation, common arrhythmias, such as atrial or
ventricular premature beats or atrial fibrillation. It is not
designed for ambulatory use.

Specifications
Measurement method : Oscillometric
Measurement range : Pressure: 20 to 280 mmHg ;
Pulse: 40 to 180 beats / min.
Pressure sensor : Semi conductor
Accuracy : Pressure: ± 3 mmHg;
Pulse:±5%ofreading
Inflation : Pump driven
Deflation : Automatic pressure release valve
Memory : Up to 30 readings
Auto-shut-off : 3 minutes after last button
operation
Operation temperature :
Operation humidity : 40 to 85 % HR maximum
Storage temperature :
Storage humidity : 10 to 95 % HR maximum
Power source : 3V DC, two LR03 1.5 V “AAA”
batteries
Dimensions : 81 (L) x 78 (L) x 78 (H) mm
Weight : 115 g (G.W.) (without batteries)
Wrist circumference : 13.5 to 19.5 cm (5.3” to 7.7”)
Minimum age : Over 18 years old
10°C to 40°C (50°F to 104°F)
-5°C to 60°C (23°F to 140°F)
* Specifications are subject to change without notice.
12
Consumer Info-Line
1-800-363-2381
AMG MEDICAL INC.
8505 Dalton, Montreal, QC H4T 1V5
www.amgphysiologic.com
www.amgmedical.com
P/N 803-036 REV: 1.0 1201
Other manuals for Physio Logic
2
Table of contents
Other AMG Blood Pressure Monitor manuals