Amprobe AM-34 User manual

AM-34
3 3/4 DIGIT AUTO DIGITAL MULTIMETER
OPERATION MANUAL
SAFETY RULES
λ This meter is designed and tested in accordance with IEC publication 1010, pollution
degree II and installation category (overvoltage category) III 600 .
λ This meter has been tested according to the following EC Directives
♦ 89/336/EEC (EMC of Nov., 1992, Electromagnetic Compatibility)
♦ 73/23/EEC (Product safety law of June 11, 1979, Low oltage Directive of
February 19, 1973)
λ This meter is designed to be indoor use at temperature 41°F to 104°F (5°C to 40°C)
and altitude up to 2,000m.
λ To ensure that the meter is used safely, follow all safety and operating instructions in
this operation manual. If the meter is not used as described in this operation manual,
the safety features of this meter might be impaired.
λ Do not use the meter if the meter or test leads look damaged, or if you suspect that the
meter is not operating properly.
λ When using the probes, keep your fingers behind the finger guards on the probes.
λ Disconnect the live test lead before disconnecting the common test lead.
λ Make sure power is off before cutting, unsoldering, or breaking the circuit. Small
amount of current can be dangerous.
λ Do not apply more than 600 dc or 600 ac rms between a terminal and earth ground.
λ To avoid electrical shock, use CAUTION when working above 60 dc or 25 ac rms.
Such voltages pose a shock hazard.
λ Never make measurements with the battery cover or bottom case off.
λ To avoid electrical shock or damage to the meter, do not exceed the input limits.
INTERNATION SYMBOLS
Important information Diode
see manual Continuity
AC Ground
DC Double insulation
FEATURES
λ One rotary switch operation.
λ Auto polarity with engineering units annunciator
λ Dual polarity A-D converter.
λ Auto voltage ranges selection: DC : 1m - 600 , AC : 1m - 600
λ Auto ampere ranges selection: DCA: 40mA - 10A, ACA: 40mA - 10A
λ Auto resistance ranges selection: 0.1Ω - 40MΩ
λ Diode & continuity test
3.1 General Specifications
Display : 3 3/4 digit LCD, 3999 counts with decimal point and
engineering units annunciator display
Polarity : Automatic, (-) negative polarity indication.
Zero adjustment : Automatic.
Measuring Method : Dual slop integration A to D converter system.
Over range indication : Only the “OL” is displayed.
Power : Two, standard 1.5 volt (AAA) battery, UM4 or equivalent.
Dimension : 70 (H) x 144 (W) x 40 (D) mm.
Weight : Approx. 280gm. (including batteries and packaging)
3.2 Electrical Specifications
Accuracies are ± (% of reading + number of least significant digits) at 23°C ± 5°C, <75% RH.
DC oltage Range Resolution Accuracy Input
Impedance
Overload
Protection:
4 1m
40 10m ±(0.8%+1) 10MΩ600 DC/AC rms
600 1
AC oltage Range Resolution Accuracy Frequency
(Hz)
Input Impedance:
10MΩ
4 1m Overload
40 10m
±(1.2%+3) 40-400 Protection:
400 - 600 1 ±(1.5%+5) 40-100 600 rms
DC Current Range Resolution Accuracy Input resistor
10A 10mA ±(1.8%+5) 0.01ΩOverload
Protection:
AC Current Range Resolution Accuracy Input resistor
10A 10mA ±(3.0%+5) 0.01Ω10A/600 fuse
Resistance Range Resolution Accuracy Measurement
oltage/Current
Response time:
5 sec. max. on
400Ω100mΩDDA 1.2mA or
less
40MΩ range
3 sec. on all other
4kΩ1Ω0.7 (TYP)
0.7mA or less
range
Temperature range
40kΩ10Ωfor 40MΩ from
400kΩ100Ω0.47 (TYP) 18°C to 30°C
4MΩ1kΩ
±(1.0%+2)
0.7mA or less
40MΩ10kΩ±(1.5%+2)
Diode Test Range Resolution
0~3.999 1m Open short or Normal Range
Continuity Range Resolution
Test 400Ω100mΩBuzzer sound generation resistance
_50Ω
PANEL DESCRIPTIONS
1) 3 3/4 digit LCD display
2) Continuity test
3) Resistance range
4) AC voltage range
5) DC voltage range
6) Power off
7) Rotary switch
8) Data hold / power resume button
9) Case
10) Diode range
11) DC current range
12) AC current range
OPERATION WARNING
1) When measuring voltage ensure that the instrument is not connected or switched to a
current or resistance or temperature or to the diode check/continuity range. Always ensure
that the correct terminals are used for the type of measurement to be made.
2) Use extreme care when measuring voltage above 50 , especially from sources where
high energy exists.
3) Avoid making connections to “live” circuits whenever possible.
4) When making current measurements ensure that the circuit is not “live” before opening it
in order to connect the test leads.
5) Before making resistance measurements or continuity / diode test, ensure that the circuit
under test is de-energised.
6) Always ensure that the correct function and range is selected. If in doubt about the correct
range, start with the highest and work downwards.
7) Extreme care should be taken when using the instrument to conjunction with a current
transformer connected to the terminals. High voltage may be produced at the terminals if
an open circuit occurs.
8) Ensure that the test leads and prods are in good condition with no damage to the
insulation.
9) Take care not to exceed the over-load limits as given in the specifications.
10) Fuse for replacement must be of the correct type and rating.
5.1 DC and AC oltage measurement
1) Connect the black test lead to the “COM” socket and red test lead to the “ ” socket.
2) Set the selector switch to DC or AC position and connect the test prods across the
source or load under measurement.
5.2 DC and AC Current measurement
1) Connect the black test lead to the “COM” socket and red test lead to the “10A” socket for
measurement up to 10A.
2) Set the selector switch to the DC current or AC current position.
3) Connect the prods in series with the current source to be measured.
5.3 Resistance measurement
1) Connect the black test lead to the “COM” socket and red test lead to the “ ” socket.
2) Set the selector switch to resistance Ω position.
3) Connect the prods across the circuit to be tested.
CAUTION: Ensure that the circuit to be tested is “dead”.
Max. input over-load : 600 rms < 10sec.
5.4 Diode test
1) Connect the black test lead to the “COM” socket and red test lead to the “ ” socket.
2) Set the selector switch to position.
3) Connect the black and red test prods to the cathode (-) and anode (+) ends of the diode to
be tested respectively.
4) Read the forward voltage drop (junction) value from the display. If reverse connected the
prods to diode, display shows over-load.
5.5 Continuity test
1) Connect the black test lead to the “COM” socket and red test lead to the “ ” socket.
2) Set the selector switch to position.
3) Connect the prods across the circuit to be tested, if the resistance less than approx. 50Ω,
buzzer will be activated.
5.6 Data hold
A push switch at the middle of rotary switch is used to hold data during measurement.
Pressing the key, reading is held and “H” appears on display. Resume active reading by
pressing switch gently once.
5.7 Auto power off
The meter will turn off automatically to extend the battery life after approx. 15 minutes without
operation. Resume power by pressing data hold push switch gently once.
6. MAINTENANCE
CAUTION
BEFORE ATTEMPTING BATTERY AND FUSE REMO AL OR REPLACEMENT,
DISCONNECT TEST LEADS FROM ANY ENERGISED CIRCUITS TO A OID SHOCK
HAZARD.
6.1 Fitting and replacing the battery and fuse
1) Ensure that the instrument is not connected to any external circuit, set the selector switch
to OFF position and remove the test leads from the terminals.
2) Remove the screw of the back case.
3) Replace the spent battery or fuse with the same type and rating.
4) Reinstate the back case, tighten and securing screw.
6.2 Cleaning
Periodically wipe the case with a soft damp cloth and mild household cleanser. Do not use
abrasives or solvents. Ensure that no water gets inside the equipment to prevent possible
shorts and damage.
FOR TECHNICAL ASSISTANCE,
PLEASE CONTACT:
AMPROBE
Tel : 305-423-7500
Fax : 305-423-7554
Technical Support : 800-327-5060
!
!
!

Other Amprobe Multimeter manuals

Amprobe CR50A User manual

Amprobe

Amprobe CR50A User manual

Amprobe 15XP-A User manual

Amprobe

Amprobe 15XP-A User manual

Amprobe AC71B User manual

Amprobe

Amprobe AC71B User manual

Amprobe PXP-A User manual

Amprobe

Amprobe PXP-A User manual

Amprobe 34XR-A User manual

Amprobe

Amprobe 34XR-A User manual

Amprobe AM-500 User manual

Amprobe

Amprobe AM-500 User manual

Amprobe DM78B User manual

Amprobe

Amprobe DM78B User manual

Amprobe 30XR-A User manual

Amprobe

Amprobe 30XR-A User manual

Amprobe AM-140-A User manual

Amprobe

Amprobe AM-140-A User manual

Amprobe 38XR-A User manual

Amprobe

Amprobe 38XR-A User manual

Amprobe AC75 User manual

Amprobe

Amprobe AC75 User manual

Amprobe PM51A User manual

Amprobe

Amprobe PM51A User manual

Amprobe DM78C User manual

Amprobe

Amprobe DM78C User manual

Amprobe ACDC-600A User manual

Amprobe

Amprobe ACDC-600A User manual

Amprobe 5XP-A User manual

Amprobe

Amprobe 5XP-A User manual

Amprobe PM51A User manual

Amprobe

Amprobe PM51A User manual

Amprobe KWC-2000 User manual

Amprobe

Amprobe KWC-2000 User manual

Amprobe ACD-15 Pro User manual

Amprobe

Amprobe ACD-15 Pro User manual

Amprobe AM-510-EUR User manual

Amprobe

Amprobe AM-510-EUR User manual

Amprobe AM555 User manual

Amprobe

Amprobe AM555 User manual

Amprobe AM-560 User manual

Amprobe

Amprobe AM-560 User manual

Amprobe AM-100 User manual

Amprobe

Amprobe AM-100 User manual

Amprobe AM-510 User manual

Amprobe

Amprobe AM-510 User manual

Amprobe ACDC-400 User manual

Amprobe

Amprobe ACDC-400 User manual

Popular Multimeter manuals by other brands

Chauvin Arnox F 05 user manual

Chauvin Arnox

Chauvin Arnox F 05 user manual

Velleman AVM360 Operation manual

Velleman

Velleman AVM360 Operation manual

Omega HHM14 user manual

Omega

Omega HHM14 user manual

Tenma 72-14630 operating manual

Tenma

Tenma 72-14630 operating manual

3B SCIENTIFIC PHYSICS Escola 30 instruction sheet

3B SCIENTIFIC PHYSICS

3B SCIENTIFIC PHYSICS Escola 30 instruction sheet

PRO-TEK 608 user manual

PRO-TEK

PRO-TEK 608 user manual

SATA 03022 user manual

SATA

SATA 03022 user manual

Brymen BM157 user manual

Brymen

Brymen BM157 user manual

Agilent Technologies L34410A user guide

Agilent Technologies

Agilent Technologies L34410A user guide

APPA 506 user manual

APPA

APPA 506 user manual

Fluke 110 user manual

Fluke

Fluke 110 user manual

Elenco Electronics M-1005K Assembly and instruction manual

Elenco Electronics

Elenco Electronics M-1005K Assembly and instruction manual

Mastech MY-60 manual

Mastech

Mastech MY-60 manual

Associated Equipment 6024 Operator's manual

Associated Equipment

Associated Equipment 6024 Operator's manual

Lovato DMG600 instruction manual

Lovato

Lovato DMG600 instruction manual

Fluke 114 Specifications

Fluke

Fluke 114 Specifications

Hanna HI 96711 instruction manual

Hanna

Hanna HI 96711 instruction manual

Meterman 37XR user manual

Meterman

Meterman 37XR user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.