Apollo Healthcare Technologies Limited APH059 User manual

Apollo Pro™
USER
MANUAL
A World of Healthcare Solutions
DECLARATION OF CONFORMITY
The Company
Apollo Healthcare Technologies Ltd
Holme Street
Liversedge
West Yorkshire
WF15 6JF
Declares that the articles which it sells, called …………................
and related accessories, are classified as Class I non-invasive
Medical Devices according to Medical Directive 93/42/EEC and to
national law no. 46/97 and comply with the European Directive on
CE marking; in addition
• they are not measuring instruments;
• they are not intended for clinical investigations;
• they are sold in non-sterile packaging;
The above mentioned devices meet with the
essential requirements and provisions of
Directive 93/42/EEC.
Apollo Healthcare Technologies Ltd endeavour to provide the
exact specification as printed. Apollo Healthcare Technologies Ltd
reserve the right to change specification without prior notice.
Apollo Healthcare Technologies Limited, Holme Street, Liversedge, West Yorkshire WF15 6JF
Tel: +44 (0) 1924 400900 Fax: +44 (0) 1924 412466 Email: sales@apollo-ht.co.uk www.apollo-ht.co.uk
Dynamic Replacement Mattress
Product Code - APH059

Others
Symbols on the printed label on the outside package box are as below:
Class II equipment
Attention! Consult accompanying documents.
Type BF equipment
Follow operating Instructions
3
...... 5
...... 5
.... 7
...... 7
.... 9
.... 10
.... 11
.... 11
.. 14
.... 14
.... 15
.... 16
.. 17
.. 17
.... 17
.... 17
... 18
.... 18
.... 19
.... 19
TABLE OF CONTENTS
STATEMENTS & SYMBOLS ..................................................................................................... 3
INTRODUCTION ....................................................................................................................... 4
PACKAGE CONTENT ............................................................................................................... 4
PRODUCT FUNCTIONS ........................................................................................................... 6
Pump .................................................................................................................................. 6
Mattress .............................................................................................................................. 7
INSTALLATION ......................................................................................................................... 8
OPERATION .............................................................................................................................. 9
General ............................................................................................................................... 9
CPR function .................................................................................................................. 12
Pressure set up .............................................................................................................. 12
Low pressure warning .................................................................................................... 13
CLEANING ............................................................................................................................ 14
HANDLING AND STORAGE ................................................................................................. 15
MAINTENANCE ..................................................................................................................... 15
General ........................................................................................................................... 15
Low pressure .................................................................................................................. 15
TROUBLESHOOTING ........................................................................................................... 16
DISPOSAL OF THE UNIT. ..................................................................................................... 16
SPECIFICATIONS ................................................................................................................. 17
System ............................................................................................................................ 17
STATEMENTS & SYMBOLS
Note, caution, warning & danger statements
NOTE
Indicates some tips or some information users should be aware of.
CAUTION
Indicates correct operating or maintenance procedure in order to prevent damage to or
destruction of the equipment or other property.
WARNING/DANGER
To reduce the risk of burns, electrocution, fire, or injury to persons:
A product should never be left unattended when plugged in.
Close supervision is necessary when this product is used by, on, or near children or invalids.
Use this product only for its intended use as described in this manual.
Do not use attachments not recommended by the manufacturer.
Never operate this product if it has a damaged cord or plug, if it is not working properly, if it
has been dropped or damaged, or dropped into water. Return the product to a service
center for examination and repair.
Keep the cord away from heated surfaces.
Never block the air openings of the product or place it on a soft surface, such as a bed or
couch, where the air openings may be blocked. Keep the air openings free of lint, hair, and
the like.
Never drop or insert any object into any opening or hose.
Do not use outdoors or operate where aerosol (spray) products are being used or
where oxygen is being administered.
No modification of this equipment is allowed.
3
Others
Symbols on the printed label on the outside package box are as below:
Class II equipment
Attention! Consult accompanying documents.
Type BF equipment
Follow operating Instructions

4
5
INTRODUCTION
This manual should be used for the initial set up of the system and for reference purposes.
General
The Apollo Pro is a high quality and affordable air support surface suitable for medium and
high-risk pressure ulcer treatment. It has been specifically designed for prevention of bedsores
and offers an affordable solution to 24-hour pressure area care.
The Apollo Pro has been tested and certified for the following standards:
EN60601-1:2006 and IEC60601-1-2:2007 and IEC60601-1-11: 2010
Intended Use
The Apollo Pro is intended to reduce the incidence of pressure ulcers while optimising patient
comfort. It also provides the following:
Individual home care setting and long-term care
Pain management as prescribed by a physician.
Contraindication
Patient conditions for which the application of pressure relieving therapy on an alternation
system is contraindicated are as follows:
Cervical or skeletal traction
Unstable spinal cord injuries
PACKAGE CONTENT
Check inside the package. If any of the following items are missing or damaged, please
contact your local dealer or service center for replacement immediately.
Pump
Make sure the type of pump is identical to which you
purchased. Functionalities vary from model to model.
Mattress
Cell type: The mattress comes with a base sheet.
For cell height of more than 8 inches, a foam base is not a must.
Cover Sheet
*Optional
The cover sheet makes cleaning an easy task.
Manual
Always read this manual before using this product.

6
7
STATIC/ALTERNATING switch (6)
Press the STATIC/ALTERNATING switch to switch the air mattress to STATIC mode or
ALTERNATING mode. The amber STATIC indicator (or the green ALTERNATING
Pump
(1)
POWER switch
(2)
MUTE switch
(3)
PANEL LOCK switch
(4)
AUTOFIRM switch
(5)
SEAT switch
(6)
STATIC/ALTERNATE switch
(7)
CYCLE TIME switch
(8)
PRESSURE SELECTION Keys
(9)
POWER FAILURE indicator
(10)
SYSTEM FAILURE indicator
(11)
LOW PRESSURE indicator
POWER switch (1)
The POWER switch is at the lower right side of the pump. Turning the POWER switch
ON/OFF will start/stop the pump. The switch lights up when the power is ON, and
extinguishes when the power is OFF.
MUTE switch (2)
The audible/visible alarm turns on either when the pressure is low or the system fails to
alternate. Press the Mute switch to mute the audible alarm, and the MUTE indicator
(amber) will light up. The alarm indicator will continue flickering. Re-press the Mute
switch to re-activate the audible alarm and to extinguish the MUTE indicator.
indicator) lights up to indicate the correspondent operation mode.
CYCLE TIME switch (7)
The CYCLE TIME switch can be used to select the appropriate cycle time of the inflation
cycle. The 4 different cycle times, 10, 15, 20 and 25 minutes can be selected and are
indicated by the number of indicators (green) that light up.
PRESSURE SELECTION Keys (8)
The PRESSURE SELECTION Keys can be used to decide the pressure in the inflated
cells in the mattress.The 8 pressure levels are indicated by the number of indicators
(green) that light up.
POWER FAILURE indicator (9)
This indicator (red) flickers when there is no power input to pump.
SYSTEM FAILURE indicator (10)
This indicator (red) flickers when the system fails to alternate.
LOW PRESSURE indicator (11)
This indicator (red) flickers when the pressure is below the pre-defined level.
Mattress
The Apollo Pro comes with cell type mattress of different materials (Nylon/PVC or PU),
provided optionally with the pump.
Please refer to the Specifications section for details.
PANEL LOCK switch (3)
Press the PANEL LOCK switch over 3 seconds to lock the panel. Press the PANEL
LOCK switch over 3 seconds again to unlock the panel. The PANEL LOCK indicator
(amber) lights up when the panel is locked. The panel will automatically lock up if no
operations are done for over 2 minutes.
AUTOFIRM switch (4)
Press the AUTOFIRM switch to set AUTOFIRM mode to quickly inflate the air mattress
to the maximum pressure which facilitates nursing and caring. The AUTOFIRM indicator
(amber) lights up when the system is in AUTOFIRM mode. The system will automatically
return to the previous mode 30 minutes after the AUTOFIRM mode is in operation. It is
also possible to cancel AUTOFIRM mode by pressing the AUTOFIRM switch.
SEAT switch (5)
Press the SEAT switch to set SEAT mode to inflates all cells and increase the pressure
by 10mmHg. The SEAT indicator (amber) lights up when the system is in SEAT mode.
To cancel SEAT mode, press the SEAT switch.
9 10 11
7 6 5 4 3 2
PRODUCT FUNCTIONS

8
9
INSTALLATION
Step 1 Place the mattress flatly on the bed frame.
The inflation tube should be towards the foot end so that it can be connected to the
inflation nozzles on the pump. See figure below for reference.
Step 2 Cover with a cotton sheet to avoid direct skin contact and reduce friction.
Step 3 Hang the pump over the frame or board at the foot end of the bed.
Make sure the pump is secured.
Step 4 Connect the inflation tubes from the mattress to the pump’s inflating nozzles.
Make sure they are properly attached.
NOTE!!
Make sure the air hoses are not kinked or tucked under the mattress.
Also check to ensure the CPR valve is properly attached.
Step 5 Plug the power cord into an electrical outlet with AC230V/50Hz output.
The plug is used as the disconnection device from the mains supply. The plug should
be positioned for easy access. The power cord shall be kept away from the patient to
prevent injuries such as strangulation.
@NOTE: BEFORE INSERTING THE PLUG INTO THE OUTLET, MAKE SURE THE VOLTAGE IS
COMPATIBLE.
Step 6 Turn on the power by pressing the power switch at the right side of the pump.
Proceed to the Operation section.
Step 7 Make sure to disconnect the pump by unplugging the power cord when it is not
in use.
OPERATION
General
This product is designed to provide maximum comfort to home care patients. Make sure that
you operate this product in a proper way to optimise its value. Here we provide some general
information you should be aware of.
For products: DO NOT use another pump with different specifications. It is dangerous to
use a pump with pressure capacity greater than 120 mmHg. This may result
in cell-puncture.
DO NOT change any component by yourself. If there is need for
replacement or repair, always contact your local dealer or service center.
For patients: When the Normal Pressure indicator (green) comes on to indicate that the
pressure has been adjusted to a desired level of firmness, the patient can
then lie on the mattress.
Please cover the mattress with a cotton sheet to avoid direct skin contact
and for the patient’s comfort.
A firm surface will make it easier for the patient to transfer or reposition.
Make use of the static mode function. It can support the patient by keeping
the mattress from bottoming out when the patient is in a sitting position. See
figure below.
Hand check: Check if the pressure is properly adjusted by sliding one hand between the
air mattress and the foam base (or bed frame if there is no foam base) to
feel the patient’s buttock.
Users should be able to feel the space in between, and the acceptable
range is approximately 25 to 40 mm (1” to 1-1/2”). This hand check
procedure is issued by AHCPR.
Always read the operating instructions before use.

10
11
Turn the power ON
A BEEP sound will begin the operation. The indicator (green) of the power switch will light up,
and the pump starts to pump air into the mattress. The LOW PRESSURE indicator (red) will
flicker as the inflation of the mattress undergoes.
NOTE!!
The audible alarm would not be heard when inflating up the mattress for the first time.
Automatic STATIC mode
The pump will automatically enter STATIC when it is turned on, and will return to the previous
conditions when it was shut down last time once the pressure reaches 20 mmHg.
NOTE!!
If the pressure does not reach 20 mmHg 40 minutes after the pump is turned on, the audible
alarm will start to beep, and the LOW PRESSURE indicator will continue flickering.
Mode selection
Press the STATIC / ALTERNATING switch to change the function mode from ALTERNATING to
STATIC and vice versa.
NOTE!!
It is recommended to always keep the pump in the ALTERNATING mode when the patient is in
a lying position for more efficient pressure ulcer prevention and treatment.
The pump will automatically return to ALTERNATING mode 2 hours after switching to the
STATIC mode.
NOTE!!
The STATIC mode provides an even support that will make the patient transfer or reposition
easier. The STATIC mode also prevents the patient from bottoming out when in the seating
position.
Pressure adjustment
The pressure of the mattress can be set by pressing the PRESSURE SELECTION keys.
There are totally 8 different pressure levels selectable ranging from 20 to 60 mmHg.
NOTE!!
Pressing the [+] key will bring the pressure one level higher, and the [-] key will bring the pressure
one level lower.
It is recommended to follow the HAND CHECK procedure as demonstrated on page 9 to decide
the appropriate pressure level.
Automatic pressure control
During normal operation, the pump will monitor pressure changes and keep it
constant at the set level. When the pressure is below the set pressure level, the pump will
automatically start to inflate the mattress. The pump will stop when the set pressure level is
reached.
NOTE!!
If the pressure is consistently low, the audible alarm will beep and its indicator will light up to
attract attention.
If there should be obvious leakage, for example caused by loose connection of tubes, the
audible/visible alarm will be activated.
For nursing and caring convenience
Press the AUTOFIRM switch to automatically inflate the mattress to the maximum level (above
70 mmHg) for about 30 minutes. The pump will return to the previous pressure level 30
minutes later.
Placing the patient in the seating position
Press the SEAT switch to for the even support and 10mmHg higher pressure to prevent the
patient from bottoming out when placing the patient in the seating position for feeding or other
applications.

12
13
To silence the alarm
If the pressure drops below the pre-defined level, the LOW PRESSURE indicator will light up
accompanied with a BEEP alarm tone. If the pump fails to alternate in the ALTERNATING
mode, the SYSTEM FAILURE indicator will light up accompanied with a BEEP alarm tone. If
the pump looses power input, the POWER FAILURE indicator will light up accompanied with a
BEEP alarm tone.
In any case, pressing the MUTE switch will silence the alarm, but the indicator will continue
flickering. Re-pressing the MUTE switch will re-activate the audible alarm.
Always read the operating instructions beforeuse.
CPR function
When there is an emergency to perform CPR on the patient, pull the CPR strap immediately to
release the air quickly from the mattress.
The CPR strap is located at the front right-hand side of the mattress.
Pressure set up
Users can adjust the pressure level of the air mattress to a desired firmness by themselves or
according to the suggestion from a health care professional.
NOTE!!
It is recommended that the pressure-adjust knob turned to Firm or press Auto
Firm on the panel when the mattress is first inflated. Users can then easily
adjust the air mattress to a desired firmness according to the patient’s weight
and comfort.
Low pressure warning
When abnormal pressure occurs, the Low Pressure indicator will come on.
The audible alarm will be activated to bring attention to a low pressure condition.
Check if the connections are secure and correctly installed according to the relevant
instructions.
NOTE!!
If the pressure is consistently low, check for any leakage (in tubes or connecting
hoses). If necessary, contact your local dealer to replace any damaged tubes or hoses.
EMC WARNING
During operation, cell phone with a maximum output power of 2W shall be kept at least
3.3 m away.

14
15
CLEANING
In this section, we describe the procedures to clean and decontaminate the pump. It is
important to follow these procedures before using the system on patients.
The cleaning task is required at least once a week to maintain personal hygiene.
Pump
DO NOT immerse or soak the pump.
Check for external damage and move the pump to the cleaning area.
Place the pump on a work surface and spray or wipe the outside of the case wi t h
quaternary ammonium solution.
DO NOT spray any cleaning solution directly on the surface of the pump.
DO NOT use a Hypo carbonate or Phonetic based cleaning solution as this may cause
damage to the case. Allow the solution to incubate for 10 minutes or accordingly as
stated by the cleaning product used.
Wipe case with a clean cloth. Make sure all areas are clean (top and bottom, both
sides).
Spray cloth with cleaning solution and clean faceplate. DO NOT allow excess cl eani ng solution
on faceplate or control panel. (If solution gets inside, damage will occur.) Allow
surface to thoroughly dry after cleaning.
After the pump is thoroughly cleaned and dried, proceed to plug in the pump and test to
see if it runs normally.
Unplug the pump and store with proper identification tag.
Mattress
Brush off or wipe down all surfaces of the cover sheet with soap and water bef or e wetting
with any liquid disinfectant.
Any obvious blood spots should be wet thoroughly with 1:9 Hypochiorite solution
(1 part bleach to 9 parts water) and allow drying for at least 10 minutes. Then blot with
a clean, damp cloth.
Unzip the top cover from the mattress.
Brush or wipe down all surfaces with soap and water before applying any liquid.
Covers are immersed and soaked in disinfectant for the required incubation time.
After pre-soaking, the cover is rinsed through a regular cycle in a washer with no soap then
laundered with mild detergent (wash temperature 93°F, rinse temperature 78°F or on the
coldest setting).
Covers are aerated until they are fully dry. (Drying temperature range 90-120°F or on the
coldest setting.)
The air cells are unsnapped from one side and are sprayed on all sides with a
disinfectant. Let it sit for the required incubation time and wipe down with a clean cloth.
(Make sure to disconnect all the air cells, one by one, and spray the disinfectant on all
sides, including all the connecting tubes and hoses. Let it sit for least 10 minutes. )
If there is a base after you remove all the air cells, the base has to be sprayed down with
the disinfectant, inside and outside. Let it sit for the required incubation time and wipe
down with a cloth.
Repeat the process with the tubing set: spray, incubate, and then wipe clean.
The carrying bag should be turned inside out and completely wiped down using the
disinfectant solution.
Allow it to thoroughly air dry. Once the inside is dry, turn it back: wipe down the outside of
the bag with disinfectant.
Dry the mattress on a SUNLESS area after cleaning.
HANDLING AND STORAGE
Lay the mattress out flat and upside down.
Roll from the foot end towards the head end; the foot-end strap can then be st ret ched
around the rolled mattress to prevent unrolling.
Do not fold, crease or stack the mattress.
Follow the national requirement to dispose the pump.
MAINTENANCE
General
Check the power cord and plug to see if there are abrasions or excessive wear.
Check the mattress cover for signs of wear or damage. Ensure the mattress cover and tubes
are connected together correctly.
Plug in the pump and check the airflow from the hose connection port. The airflow
should alternate between ports every half-cycle time.
Check the air hoses to see if there are kinks or breaks. For replacement, pl ease
contact your local agent or dealer.
Make sure the mattress tube is well connected.
Check the pump and make sure both power and power indicator are off when the
switch is turned off.
Low pressure
Examine if there is air leakage between the pump and the mattress connections or from the
air mattress tubes:
Check connectors between the air mattress and pump. If there is any
disconnection, please reconnect it.
Check the CPR Valves. Ensure their outlets are sealed.
Check the air-connecting tubes. Ensure each single cell is not broken.
Turn the pressure-adjust knob to Firm. Keep the tubes fully inflated and inspect for ai r
leakage.
Check if there is any air leakage from cells. Ensure no leakage occurs. If any leakage
occurs, please contact your local agent or dealer.
Warranty / Servicing
Both pump and mattress carry a 2-year warranty in subject to being serviced in accordance to Apollo
Healthcare Technologies service manual.
SERVICING YOUR SYSTEM
The mattress system should be serviced every 12 months by Apollo Healthcare Technologies or approved
personnel using genuine Apollo Healthcare Technologies spare parts.
Compressor/ Bellows should be changed. Filters should be changed.

16
17
TROUBLESHOOTING
Problems
Solutions
The pump doesn’t work
1. Check if the plug is inserted firmly into the outlet.
2. Turn on the power switch again.
If the power indicator is ON and the pump still doesn’t wor k,
contact your local dealer immediately.
If the power indicator is OFF, there may be a faulty outlet. Try to
connect the power cord to another outlet. If the power indicator
is still OFF, contact a qualified electrician for main power check.
Incomplete inflation
(Low Pressure)
1. For a quick check, adjust the pressure to Firm.
2. Check to see if the tubes connected to the pump are twisted or
there is any leakage occurring.
Always keep the tubes straight.
Change tubes if there is any leakage.
Ensure the CPR valves are closed. Ensure every single cell isnot
broken.
Slow air flow
A dirty filter may decrease the air flow.
Wash the filter with mild detergent to keep it clean.
Check the air filter at the back of the pump at least once
amonth.
DISPOSAL OF THE UNIT
Follow the national requirement to dispose the unit.
SPECIFICATIONS
System
Item
Specifications
Classification
Class II; IP21; AP/APG NO; Type BF
Applied Part
Air mattress
Input Rating
230V 50Hz; 12W, 1A
Fuse Rating
250V, T1AL
T: Time delay
L: Low Breaking Capacity
Pressure Range
20 ~ 60 mmHg
Cycle Time
10, 15, 20 , 25 minutes
Dimensions (Pump)
369 x 135 x 250 mm (14½” x 5¾” x 10”)
Weight (Pump)
4.20 kg (9.3 lbs)
Environment Requirements
Temperature: Operation 41°F~104°F (5~40°C)
Storage -13°F ~158°F (-25°C to 70°C)
Shipping -13°F ~158°F (-25°C to 70°C)
Humidity: Operation less than 93% non-
condensing Storage: less than 93%
non-condensing
Safety Standards
EN/IEC 60601-1
EMC Standards
CISPR 11 Class B
IEC 61000-3-2 IEC 61000-3-3 IEC 61000-4-2
IEC 61000-4-3 IEC 61000-4-4 IEC 61000-4-5
IEC 61000-4-6 IEC 61000-4-8 IEC 61000-4-11
Service Life
2 years
* Due to the continual development of the above products, specifications are subject to change
without prior notice.

18
19
Table 1–MANUFACTURER’S declaration – ELECTROMAGNETIC
EMISSIONS – for all ME EQUIPMENT and ME SYSTEMS
Table 2 –MANUFACTURER’S declaration – electromagnetic
IMMUNITY –for all ME EQUIPMENT and ME SYSTEMS
GUIDANCE AND MANUFACTURER’S DECLARATION –
ELECTROMAGNETIC EMISSIONS
The Air pump is intended for use in the electromagnetic environment specified below.
The customer or the user of the Air pump should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment –
guidance
RF emissions
CISPR 11
Group 1
The Air pump uses RF energy only
for its internal function. Therefore,
its RF emissions are very low and
are not likely to cause any
interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B
The Air pump is suitable for use in
all establishments, including
domestic establishments and those
directly connected to the public
low-voltage power supply network
that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
GUIDANCE AND MANUFACTURER’S DECLARATION –
ELECTROMAGNETIC IMMUNITY
The Air pump is intended for use in the electromagnetic environment specified below.
The customer or the user of the Air pump should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic
environment –
guidance
Electrostatic
6 kV contact
±6 kV contact
Floors should be wood,
discharge (ESD)
8 kV air
±8 kV air
concrete or ceramic tile. If
IEC 61000-4-2
floors are covered with
synthetic material, the
relative humidity should
be at least 30 %.
Electrical fast
2 kV for power
±2 kV for power
Mains power quality
transient/burst
supply lines
supply lines
should be that of a typical
commercial or hospital
IEC 61000-4-4
1 kV for
±1 kV for
environment.
input/output
input/output
lines
lines
Surge
1 kV line(s) to
±1 kV line(s) to
Mains power quality
IEC 61000-4-5
line(s)
line(s)
should be that of a typical
commercial or hospital
2 kV line(s) to earth
±2 kV line(s) to
environment.
earth
interruptions and
<5 % UT
<5 % UT
Mains power quality
voltage
(>95 % dip in UT)
(>95 % dip in UT)
should be that of a typical
variations
for 0,5 cycle
for 0,5 cycle
commercial or hospital
on power supply
environment. If the user of
input lines
40 % UT
40 % UT
the Air pump requires
(60 % dip in UT)
(60 % dip in UT)
continued operation
IEC 61000-4-11
for 5 cycles
for 5 cycles
during power mains
interruptions, it is
70 % UT
70 % UT
recommended that the Air
(30 % dip in UT)
(30 % dip in UT)
pump be powered from an
for 25 cycles
for 25 cycles
uninterruptible power
supply or a battery.
<5 % UT
<5 % UT
(>95 % dip in UT)
(>95 % dip in UT)
for 5 sec
for 5 sec
Power Frequency
Power frequency
(50/60 Hz)
3 A/m
3 A/m
magnetic fields should be
magnetic field
at levels characteristic of
a typical location in a
IEC 61000-4-8
typical commercial or
hospital environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.

20
21
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption
and reflection from structures, objects and people.
TTable
3
–
GGuuiidaanncce
aannd
MANNUUFFAACCTTUURREERR’S
ddeeccllaarraatttiioon
–
eelleeccttroommaaggnneettic
IMMUUNNIITY
–ffor
ME
EQUIIPMENT
and
ME
SYSTEMS
tthhat
are
noot
LLIIFFEE-SSUUPPPPOORRTTING
GUIDANCE AND MANUFFAACTUREERR’S DECLARATIOONN –
EELEECTROMAGGNNEEETTIICC IMMMUUNNIITTYY
The Air ppummp iis iinntteenddeed ffoor uusssee
i
n
t
h
e eelle
ctro
mmaaggnne
t
i
c
ennvviirroonnmmeenn
t
ssppeecciiffiieedd
b
eellooww..
The ccuussttomer or the usser of the
A
i
r
pump
sshh
ou
l
d
assssurre
t
h
a
t
i
t
i
s
u
ssed
i
n
ssuucchh a
n
e
nv
i
rro
nm
e
n
t
.
Immunity
test
IIEC 660060011 tteest
level
CCoommpplliiaanncce
level
EEllecttroommagnneettiicc ennvvironment –
guidance
CC
o
nndd
uct
e
d
RRFF
3 Vrms
3 VVrrm
s
PPoorrttaabbllee aa
nd
mmoobbii
l
e RRFF
IIEECC 661100000-4-6
11550 kkHHzz
t
o
80
cco
mmun
i
ccaatt
i
ons
e
qu
i
pm
e
n
t
MHz
ssshhould be uuussseee
d
no
ccclllo
o
se
r
ttt
o
aann
y
ppa
rt
oo
f
t
h
e AAii
r
ppuummpp, ii
n
cclluu
d
iinn
g
Radiated RF
3 VV//
m
3 VV//
m
ccaa
b
llee
s,
t
haa
n
tthhe rree
co
mmmmeenn
d
eedd
IIEC 6611000000-4-3
80 MHz
t
o
2,,55
ssee
p
aarraattiio
n
dd
i
st
ann
ce
ccaa
l
ccuulla
t
e
d
GHz
f
rroo
m
t
h
e e
qu
att
i
on
a
pp
li
cca
b
l
e ttoo
tthhe ffrreeqq
u
eennccyy
o
f
tthhee ttrraannssmmiitttteerr..
Reccoo
mm
e
nd
e
d
ssee
p
arra
t
i
on
dd
i
st
ann
ce
d
= 11,,2
d
= 11,,2 80 MHz tt
o
800 MHz
d = 2,3 800 MHz to 2,5 GHz
ww
h
eerree P ii
s
tthhe
m
aaxx
i
mmuu
m
oouuttpp
u
t
pp
o
wwee
r
rrraaattii
ng
ooff tthhe ttrraa
n
ssmmii
t
ttee
r
iinn
wwaa
t
ttss ((WW)) aaccccoo
rd
iinn
g
ttoo
t
h
e
t
rra
n
ssm
i
t
ttee
r
m
a
nu
f
acct
u
rre
r
a
nd
d
ii
s
tthhe rreeccoommmmeenn
d
eedd sseeppaarraattiioo
n
dd
i
st
ann
ce
ii
n
m
eettrree
s
((mm).
FFiiee
l
d
st
rennggtthh
s
ffrroo
m
ffii
x
eedd RRFF
t
rra
n
ssm
i
tt
errss,, a
s
d
etterrmm
i
n
e
d
by
a
n
ee
l
eeccttrroo
m
aagg
n
eettii
c
ssiitte ssuurrvveyy,
a
sshh
ou
l
d
b
e
l
esss
t
h
a
n
t
h
e
ccoo
m
pp
l
iiaan
ce
llee
v
eell
i
n
eeaacchh
fffrreeqq
u
eenncccyy rraann
g
ee..
b
IInntteerrffeerreen
n
ce
mmaa
y
ooccccuurr iinn tthhe
v
iiccii
n
iitt
y
ooff eqquu
i
pp
m
eenn
t
mmarrkkee
d
ww
i
t
h
tthhe ffoolllloowwwiinngg
sy
mmbbooll::
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the Air pump is used
exceeds the applicable RF compliance level above, the Air pump should be observed to verify
normal operation. If abnormal performance is observed, additional measures may be
necessary, such as reorienting or relocating the Air pump.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

22
23
Tabllee 4 – RReccoommmenndded ssseeepaarraattiion ddiissttaancceess bbeettwweeenn ppoortabblle anndd moobbiillee NOTES
RF ccoommuniccaations equipment and the ME EEQUIPPMEENT or ME SSYYSSTEEM –
ffoorr MMEE EQUIPMEENNT aannd ME SSYYSSTTEEMMS tthhaatt aarree nnoot LLIIFFEEE-SSUUPPPPOORRTTIINNG
RRECOMMMENDDEEDD SEPARRATTIOON DDIISSTTANNCCES
BBEETTWEEN PPOORRTTAABBLLEE AANND MMOOBBIILLE RRFF
CCOOMMMMUUNNIICCAATTIIOONNSS EQUIPMEENNTT AND TTHHE AAIIR PPUUMP
Thhee AAiir ppuummpp iis iinntteenddeed ffoor uusssee
i
n
aa
n
ee
l
eeccttrroo
m
aagg
n
eettii
c
eenn
v
ii
ro
nnmmenn
t
i
n
ww
h
ii
ch
rraa
d
iiaattee
d
RF
disstturrbbanccees arre ccontrrolled. TThhe ccussttoo
m
e
r
o
r
t
h
e
u
sse
r
o
f
t
h
e AAiirr ppuummpp ccaa
n
h
e
l
p
p
rre
v
e
n
t
eleccttrroomagnetic interrfferreenccee b
y
m
a
i
n
t
a
i
n
i
ng
a
m
i
n
i
mum
d
i
ssta
n
cce
b
e
t
wee
n
po
rrta
b
l
e a
nd
mob
il
e
RRF ccoommmuuniiccaattiionns eeqquiippmeent ((ttrraann
sm
iittttee
rs)
aann
d
tthhe AAii
r
ppuummpp aass rrreeeccoo
m
mmenndde
d
bbee
l
ooww,
accccorrding ttoo the maximum outtpp
u
t
pow
e
r
o
f
t
h
e cco
mmun
i
ccattii
ons
e
qu
i
pm
e
n
t
.
Rated maximum
oouuttppuu
t
pp
o
wwee
r
oo
f
ttrraannssmmiittttee
r
W
SSeepaarraattii
on
dd
i
st
ann
ce
aaccccoorrddii
ng
t
o
ffrree
q
uuee
ncy
ooff ttrraa
n
ssmmiitttteerr
m
1150 kkkHHHzz
t
o
80 MMHHz
d = 1,2
80 MHz to 800 MHz
d = 1,2
88000 MMHHz ttoo 22,,5 GHz
d = 2,3
0,01
0,12
0,12
0,23
0,1
0,38
0,38
0,73
1
11,,2
1,2
2,3
10
3,8
3,8
7,3
100
12
12
23
FFoo
r
t
ra
nnssmmiitttte
rs
ra
t
eedd a
t
a
m
aaxx
i
mm
um
oouuttpp
u
t
pp
o
wwe
r
n
oott
l
iissttee
d
aabb
o
vve, tthhe rree
co
mmmmeenn
d
eedd
ssee
p
aarraattii
on
ddiissttaa
n
ce
d
ii
n
m
eettrree
s
((mm) ccaa
n
b
e eessttii
m
aattee
d
uu
si
nn
g
tthhe ee
q
uuaattii
on
aapp
p
llii
ca
bblle tt
o
tthhe
f
f
rrreeeqqq
u
eennncccyy
o
of
f
t
thhhee
t
trrraaannnsssmm
i
itttttteeerrr,,
wh
hee
re
P i
i
s
t
t
h
h
e
m
aaaxxx
i
m
mum
ooouuutttp
pu
t
p
po
wwee
r
rrraaat
t
iii
ng
oof
f
t
t
h
h
e tttrraannssmmi
i
t
tteeer
r
iin
n
wwa
t
tt
s
((WW)) accccoo
rd
iinn
g
ttoo
t
h
e
t
ra
nnsssmmiitttte
r
m
aann
u
f
a
ctu
rreerr..
NNOOTTE 1
At
880 MMHz ann
d
8800 MMHz, tthhe sseepparraattii
on
ddiissttaa
n
ce
f
o
r
tthhe
h
iig
gh
eerr
f
re
qquuennccyy
ra
nnggee
aa
p
ppll
i
eess..
NOTE 2 T
h
essee
gu
i
d
e
li
n
e
s
m
a
y
no
t
a
pp
l
y
i
n
a
ll
ssi
t
u
a
t
ii
o
nnss..
El
eeccttrroo
m
aag
gn
eettii
c
pprroo
p
aagga
t
i
oo
n
ii
s
aaffffeeccttee
d
bb
y
abb
so
rrppttiioo
n
ann
d
rreefflle
ct
i
oo
n
ffrroo
m
ssttrruuccttuurree
s,
oobbjje
cts
aa
nd
ppee
o
pplle.
*
Due
to
ccoo
n
t
i
nnuua
l
ddee
v
ee
l
oopp
m
eenn
t
oo
f
tthhe abboo
v
e pprroo
d
uucctt
s,
ssppecciiffii
ca
t
i
oonn
s
aa
re
su
bbjjee
ct
tt
o
cchhann
g
e
without p
p
rrriiio
r
noot
t
iiiccceee.
Table of contents
Other Apollo Healthcare Technologies Limited Medical Equipment manuals

Apollo Healthcare Technologies Limited
Apollo Healthcare Technologies Limited Apollo 5 Plus User manual

Apollo Healthcare Technologies Limited
Apollo Healthcare Technologies Limited APH040 User manual

Apollo Healthcare Technologies Limited
Apollo Healthcare Technologies Limited Apollo 5 Plus APHos2 User manual
Popular Medical Equipment manuals by other brands

Patterson Medical
Patterson Medical Midland 4512E Operation and service guide

Pari
Pari SinuStar Instructions for use

INTEGRA Biosciences
INTEGRA Biosciences VacuSafe Operator's manual

Moog
Moog CURLIN INFUSION PainSmart user manual

Med-Mizer
Med-Mizer FlexTilt TF16 user manual

Stryker
Stryker 1079 Operation manual