103N / 105N / 106 EN
Note :
1. BLUE KEY switch and RANGE Switch resets all special 
 modes other than HZ mode .
2. HZ switch is only used to toggle between frequency 
 measurement mode and voltage or current measurement
 mode, and thus can not be nested in any other special 
 modes.
3. The bar graph always shows the current value, regardless 
 of special mode.
7. Light switch — Press the switch to turn on or turn o the 
 back-light.
8. HZ — If “ HZ” switch is pushed in voltage or current 
 measurement mode, the meter enters frequency 
 counter mode with automatic range selection. 
 Therefore, pressing “RANGE” switch in HZ 
 mode does not change the frequency range. 
 However,“RANGE” switch changes the sensitivity 
 of frequency detection. If the input signal has a
 small amplitude, the user shall increase the 
 sensitivity.
Note : Since HZ mode is treated as a temporary frequency 
 measurement in voltage or current mode, pressing 
 “RANGE” switch in HZ mode also changes the full 
 scale range of the original voltage or current mode 
 after VAHZ mode is cancelled.
9. PEAK H— The precision of PEAK hold mode 
 measurement can be enhanced by calibration. Calibration 
 for PEAK hold mode is invoked by pressing “PEAK H” 
 switch for more than 2 seconds until the “caL” is displayed.
 Pressing the “Peak H” switch to toggle between PMAX 
 and PMIN.Pressing “Peak H” switch and held for longer 
 than one second to release the peak hold mode.
10. Range Switch, (Manual Range) — “Range” switch is 
 pressed to select manual ranging and to change ranges.
 When “Range” switch is pressed on “RANGE” annunciator 
 on the LCD appears.
 Press “RANGE” switch to select appropriate range to be 
 used.
 Press “RANGE” switch and hold for 2 seconds to return 
 to Autoranging. In “ HZ” mode “RANGE” is not used 
 to change the full scale frequency range, but change the 
 sensitivity of frequency measurement.
11. Blue Switch — Press the switch to measure AC Voltage 
 / Current or DC Voltage / Current in the voltage / Current 
 mode, or to measure Resistance or continuity or diode in 
    Ω /  / mode or to measure Frequency or RPM in RPM 
 in Hz / PRM mode.