APPA sFLEX-D User manual

Warning
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . .
Safety sheet
. . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . .
. . . . . . . . .
Read First
APPA sFLEX-D Instruction
Flexible Current Meter Card
Safety Information
To ensure safe operation and service of the
meter, follow these instructions.Failure to
observe warnings can result in severe Injury.
Warning
Identify hazardous conditions and actions that
could cause BODILY HARM or DEATH
-Individual protective equipment should be used
if HAZARDOUS LIVE parts in the installation
where measurement is to be carried out could
Be ACCESSIBLE.
-If the equipment is used in a manner not
specified by the manufacturer, the protection
provided by the equipment may be impaired.
-To reduce the risk of fire or electric shock, do
not expose this product to rain or moisture.
-Verify the Meter operation by measuring a
known Current. If in doubt, have the Meter
serviced.
-Do not apply more than the rated voltage/
Current, as marked on Meter.
-To avoid false readings that can lead to electric
shock and injury, replace battery as soon as low
battery indicator.
- Do not use Meter around explosive gas or
vapor.
See instruction manual
Do not apply to or remove from
HAZARDOUS LIVE conductors.
Equipment protected by double or
reinforced insulation
Battery
Earth
Conforms to EU directives
Do not discard this product or throw away
AC measurement
Maintenance
Do not attempt to repair this Clamp Meter.
It contains no user-serviceable parts.
Repair or servicing should only be performed by
qualified personnel.
Cleaning
Periodically wipe the case with a dry cloth and
detergent. Do not use abrasives or solvents.
Power on/off
APPA TECHNOLOGY CORP.
9F, 119-1 Pao-Zong Rd.,
Shin-Tien, Taipei, 23115, Taiwan.
P.O.Box. 12-24 Shin-Tien, Taiwan.
Tel: +886-2-29178820
Fax: +886-2-29170848
http: //www.appatech.com
- Do not use a flexible current sensor if the
inner copper wire of the flexible cord is visible.
- De-energize the installation under test or wear
suitable protective clothing during fitting a
removal of the Flexible Current Probe.
- Do not apply around or remove from UNINSU
LATED HAZARDOUS LIVE conductors, which
may render electric shock, electric burn, or
arc flash.
Caution
Identify conditions and actions that could cause
DAMAGE the meter or equipment under test.
Do not expose Meter to extremes in tempera-
ture of high humidity.
Symbols as marked on the meter and
Instruction card
Press > 2Sec
OFF
Press > 2Sec
OFF
Press > 2Sec
OFF
Press > 2Sec
OFF
Flexible AC current probe
Lock
Lock
Unlock
Unlock
Do not touch the
area above the LCD
during measurement
Auto Power off Function
20 min20 min20 min20 min
The meter automatically turns the power off
after powering up and no operation for
20 minutes ± 30 seconds.
Disable Auto Power off Function
Press > 2Sec
OFF
Press
Press “Power” button, and during the LCD
shows battery capacity, press “backlight”
button until the LCD shows “AoFF” to disable
Auto Power Off Function.

Warning
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . .
Safety sheet
. . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . .
. . . . . . . . .
Read First
APPA sFLEX-D Instruction
Flexible Current Meter Card
Safety Information
To ensure safe operation and service of the
meter, follow these instructions.Failure to
observe warnings can result in severe Injury.
Warning
Identify hazardous conditions and actions that
could cause BODILY HARM or DEATH
-Individual protective equipment should be used
if HAZARDOUS LIVE parts in the installation
where measurement is to be carried out could
Be ACCESSIBLE.
-If the equipment is used in a manner not
specified by the manufacturer, the protection
provided by the equipment may be impaired.
-To reduce the risk of fire or electric shock, do
not expose this product to rain or moisture.
-Verify the Meter operation by measuring a
known Current. If in doubt, have the Meter
serviced.
-Do not apply more than the rated voltage/
Current, as marked on Meter.
-To avoid false readings that can lead to electric
shock and injury, replace battery as soon as low
battery indicator.
- Do not use Meter around explosive gas or
vapor.
See instruction manual
Do not apply to or remove from
HAZARDOUS LIVE conductors.
Equipment protected by double or
reinforced insulation
Battery
Earth
Conforms to EU directives
Do not discard this product or throw away
AC measurement
Maintenance
Do not attempt to repair this Clamp Meter.
It contains no user-serviceable parts.
Repair or servicing should only be performed by
qualified personnel.
Cleaning
Periodically wipe the case with a dry cloth and
detergent. Do not use abrasives or solvents.
Power on/off
APPA TECHNOLOGY CORP.
9F, 119-1 Pao-Zong Rd.,
Shin-Tien, Taipei, 23115, Taiwan.
P.O.Box. 12-24 Shin-Tien, Taiwan.
Tel: +886-2-29178820
Fax: +886-2-29170848
http: //www.appatech.com
- Do not use a flexible current sensor if the
inner copper wire of the flexible cord is visible.
- De-energize the installation under test or wear
suitable protective clothing during fitting a
removal of the Flexible Current Probe.
- Do not apply around or remove from UNINSU
LATED HAZARDOUS LIVE conductors, which
may render electric shock, electric burn, or
arc flash.
Caution
Identify conditions and actions that could cause
DAMAGE the meter or equipment under test.
Do not expose Meter to extremes in tempera-
ture of high humidity.
Symbols as marked on the meter and
Instruction card
Press > 2Sec
OFF
Press > 2Sec
OFF
Press > 2Sec
OFF
Press > 2Sec
OFF
Flexible AC current probe
Lock
Lock
Unlock
Unlock
Do not touch the
area above the LCD
during measurement
Auto Power off Function
20 min20 min20 min20 min
The meter automatically turns the power off
after powering up and no operation for
20 minutes ± 30 seconds.
Disable Auto Power off Function
Press > 2Sec
OFF
Press
Press “Power” button, and during the LCD
shows battery capacity, press “backlight”
button until the LCD shows “AoFF” to disable
Auto Power Off Function.

Warning
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . .
Safety sheet
. . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . .
. . . . . . . . .
Read First
APPA sFLEX-D Instruction
Flexible Current Meter Card
Safety Information
To ensure safe operation and service of the
meter, follow these instructions.Failure to
observe warnings can result in severe Injury.
Warning
Identify hazardous conditions and actions that
could cause BODILY HARM or DEATH
-Individual protective equipment should be used
if HAZARDOUS LIVE parts in the installation
where measurement is to be carried out could
Be ACCESSIBLE.
-If the equipment is used in a manner not
specified by the manufacturer, the protection
provided by the equipment may be impaired.
-To reduce the risk of fire or electric shock, do
not expose this product to rain or moisture.
-Verify the Meter operation by measuring a
known Current. If in doubt, have the Meter
serviced.
-Do not apply more than the rated voltage/
Current, as marked on Meter.
-To avoid false readings that can lead to electric
shock and injury, replace battery as soon as low
battery indicator.
- Do not use Meter around explosive gas or
vapor.
See instruction manual
Do not apply to or remove from
HAZARDOUS LIVE conductors.
Equipment protected by double or
reinforced insulation
Battery
Earth
Conforms to EU directives
Do not discard this product or throw away
AC measurement
Maintenance
Do not attempt to repair this Clamp Meter.
It contains no user-serviceable parts.
Repair or servicing should only be performed by
qualified personnel.
Cleaning
Periodically wipe the case with a dry cloth and
detergent. Do not use abrasives or solvents.
Power on/off
APPA TECHNOLOGY CORP.
9F, 119-1 Pao-Zong Rd.,
Shin-Tien, Taipei, 23115, Taiwan.
P.O.Box. 12-24 Shin-Tien, Taiwan.
Tel: +886-2-29178820
Fax: +886-2-29170848
http: //www.appatech.com
- Do not use a flexible current sensor if the
inner copper wire of the flexible cord is visible.
- De-energize the installation under test or wear
suitable protective clothing during fitting a
removal of the Flexible Current Probe.
- Do not apply around or remove from UNINSU
LATED HAZARDOUS LIVE conductors, which
may render electric shock, electric burn, or
arc flash.
Caution
Identify conditions and actions that could cause
DAMAGE the meter or equipment under test.
Do not expose Meter to extremes in tempera-
ture of high humidity.
Symbols as marked on the meter and
Instruction card
Press > 2Sec
OFF
Press > 2Sec
OFF
Press > 2Sec
OFF
Press > 2Sec
OFF
Flexible AC current probe
Lock
Lock
Unlock
Unlock
Do not touch the
area above the LCD
during measurement
Auto Power off Function
20 min20 min20 min20 min
The meter automatically turns the power off
after powering up and no operation for
20 minutes ± 30 seconds.
Disable Auto Power off Function
Press > 2Sec
OFF
Press
Press “Power” button, and during the LCD
shows battery capacity, press “backlight”
button until the LCD shows “AoFF” to disable
Auto Power Off Function.

Warning
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . .
Safety sheet
. . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . .
. . . . . . . . .
Read First
APPA sFLEX-D Instruction
Flexible Current Meter Card
Safety Information
To ensure safe operation and service of the
meter, follow these instructions.Failure to
observe warnings can result in severe Injury.
Warning
Identify hazardous conditions and actions that
could cause BODILY HARM or DEATH
-Individual protective equipment should be used
if HAZARDOUS LIVE parts in the installation
where measurement is to be carried out could
Be ACCESSIBLE.
-If the equipment is used in a manner not
specified by the manufacturer, the protection
provided by the equipment may be impaired.
-To reduce the risk of fire or electric shock, do
not expose this product to rain or moisture.
-Verify the Meter operation by measuring a
known Current. If in doubt, have the Meter
serviced.
-Do not apply more than the rated voltage/
Current, as marked on Meter.
-To avoid false readings that can lead to electric
shock and injury, replace battery as soon as low
battery indicator.
- Do not use Meter around explosive gas or
vapor.
See instruction manual
Do not apply to or remove from
HAZARDOUS LIVE conductors.
Equipment protected by double or
reinforced insulation
Battery
Earth
Conforms to EU directives
Do not discard this product or throw away
AC measurement
Maintenance
Do not attempt to repair this Clamp Meter.
It contains no user-serviceable parts.
Repair or servicing should only be performed by
qualified personnel.
Cleaning
Periodically wipe the case with a dry cloth and
detergent. Do not use abrasives or solvents.
Power on/off
APPA TECHNOLOGY CORP.
9F, 119-1 Pao-Zong Rd.,
Shin-Tien, Taipei, 23115, Taiwan.
P.O.Box. 12-24 Shin-Tien, Taiwan.
Tel: +886-2-29178820
Fax: +886-2-29170848
http: //www.appatech.com
- Do not use a flexible current sensor if the
inner copper wire of the flexible cord is visible.
- De-energize the installation under test or wear
suitable protective clothing during fitting a
removal of the Flexible Current Probe.
- Do not apply around or remove from UNINSU
LATED HAZARDOUS LIVE conductors, which
may render electric shock, electric burn, or
arc flash.
Caution
Identify conditions and actions that could cause
DAMAGE the meter or equipment under test.
Do not expose Meter to extremes in tempera-
ture of high humidity.
Symbols as marked on the meter and
Instruction card
Press > 2Sec
OFF
Press > 2Sec
OFF
Press > 2Sec
OFF
Press > 2Sec
OFF
Flexible AC current probe
Lock
Lock
Unlock
Unlock
Do not touch the
area above the LCD
during measurement
Auto Power off Function
20 min20 min20 min20 min
The meter automatically turns the power off
after powering up and no operation for
20 minutes ± 30 seconds.
Disable Auto Power off Function
Press > 2Sec
OFF
Press
Press “Power” button, and during the LCD
shows battery capacity, press “backlight”
button until the LCD shows “AoFF” to disable
Auto Power Off Function.

Warning
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . .
Safety sheet
. . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . .
. . . . . . . . .
Read First
APPA sFLEX-D Instruction
Flexible Current Meter Card
Safety Information
To ensure safe operation and service of the
meter, follow these instructions.Failure to
observe warnings can result in severe Injury.
Warning
Identify hazardous conditions and actions that
could cause BODILY HARM or DEATH
-Individual protective equipment should be used
if HAZARDOUS LIVE parts in the installation
where measurement is to be carried out could
Be ACCESSIBLE.
-If the equipment is used in a manner not
specified by the manufacturer, the protection
provided by the equipment may be impaired.
-To reduce the risk of fire or electric shock, do
not expose this product to rain or moisture.
-Verify the Meter operation by measuring a
known Current. If in doubt, have the Meter
serviced.
-Do not apply more than the rated voltage/
Current, as marked on Meter.
-To avoid false readings that can lead to electric
shock and injury, replace battery as soon as low
battery indicator.
- Do not use Meter around explosive gas or
vapor.
See instruction manual
Do not apply to or remove from
HAZARDOUS LIVE conductors.
Equipment protected by double or
reinforced insulation
Battery
Earth
Conforms to EU directives
Do not discard this product or throw away
AC measurement
Maintenance
Do not attempt to repair this Clamp Meter.
It contains no user-serviceable parts.
Repair or servicing should only be performed by
qualified personnel.
Cleaning
Periodically wipe the case with a dry cloth and
detergent. Do not use abrasives or solvents.
Power on/off
APPA TECHNOLOGY CORP.
9F, 119-1 Pao-Zong Rd.,
Shin-Tien, Taipei, 23115, Taiwan.
P.O.Box. 12-24 Shin-Tien, Taiwan.
Tel: +886-2-29178820
Fax: +886-2-29170848
http: //www.appatech.com
- Do not use a flexible current sensor if the
inner copper wire of the flexible cord is visible.
- De-energize the installation under test or wear
suitable protective clothing during fitting a
removal of the Flexible Current Probe.
- Do not apply around or remove from UNINSU
LATED HAZARDOUS LIVE conductors, which
may render electric shock, electric burn, or
arc flash.
Caution
Identify conditions and actions that could cause
DAMAGE the meter or equipment under test.
Do not expose Meter to extremes in tempera-
ture of high humidity.
Symbols as marked on the meter and
Instruction card
Press > 2Sec
OFF
Press > 2Sec
OFF
Press > 2Sec
OFF
Press > 2Sec
OFF
Flexible AC current probe
Lock
Lock
Unlock
Unlock
Do not touch the
area above the LCD
during measurement
Auto Power off Function
20 min20 min20 min20 min
The meter automatically turns the power off
after powering up and no operation for
20 minutes ± 30 seconds.
Disable Auto Power off Function
Press > 2Sec
OFF
Press
Press “Power” button, and during the LCD
shows battery capacity, press “backlight”
button until the LCD shows “AoFF” to disable
Auto Power Off Function.

Environmental Specifications
Indoor Use.
Pollution Degree :2
Maximum operating altitude : 2000m (6562 ft)
Operating temperature & Relative Humidity :
0°C - 30°C, ≦80%RH
30°C - 40°C, ≦75%RH
40°C - 50°C, ≦45%RH
Storage temperature :
-20°C - 60°C, 0 to 80% RH (no batteries).
Temperature coefficient :
0.2 x (Specified accuracy) / °C, < 18°C, > 28°C
Shock vibration : Sinusoidal vibration
MIL-PRF-28800F (5-55 Hz, 3g max.)
Drop Protection : 4 feet drop to hardwood on
concrete Floor
Limited Warranty
This Meter is warranted to the original purchaser
against defects in material and workmanship for
two years from the date of purchase. During this
warranty period, manufacturer will, at its option,
replace or repair the defective unit, subject to
verification of the defect or malfunction. This
warranty does not cover fuses, disposable
batteries, or damage from abuse, neglect,
accident, unauthorized repair, alteration,
contamination, or abnormal conditions of
operation or handling.
Any implied warranties arising out of the sale of
this product, including but not limited to implied
warranties of merchantability and fitness for a
particular purpose, are limited to the above.
The manufacturer shall not be liable for loss of
use of the instrument or other incidental or
consequential damages, expenses, or economic
loss, or for any claim or claims for such damage,
expense or economic loss. Some states or
countries laws vary, so the above limitations or
exclusions may not apply to you.
AC Current
Backlight
PressPress
Data Hold
PressPress
Replacing Batteries
UnlockUnlock
22
11
UnlockUnlock
LockLock
Caution
Remove Meter from conductor and turn off
power before opening the battery cover.
General Specifications
Display : 3000 counts.
Over range Display : "OL"
Dimensions (W x H x D) :
10" coil : Approx. 120mm x 280mm x 25mm
18" coil : Approx. 130mm x 350mm x 25mm
Weight(without Battery) :
10" coil : 170g
18" coil : 200g
Power Requirements : AAA Size Battery x 2
Battery Life : 200 hours ALKALINE Battery
Low Batteries Indication :
Voltage drops below operating voltage,
will flash.
Maximum Conductor Size : coil size
Safety Standard :
EN61010-1, EN61010-2-032, EN61326-1
CAT IV 600V, CAT III 1000V
Conversion Rate : 4 times/second
sFlex-10 Error
Distance from
Optimum (mm)
15 ±2.0%
25 ±2.5%
35 ±3.0%
sFlex-18 Error
Distance from
Optimum (mm)
35 ±1.0%
50 ±1.5%
60 ±2.0%
Range OL
Reading Resolution Accuracy
(45Hz~500Hz)
30.00A 33.00A 0.01A ±(3.0% + 5D)
300.0A 330.0A 0.1A ±(3.0% + 5D)
3000A 3300A 1A ±(3.0% + 5D)
LCD displays 0 counts when the reading < 10
counts.
Frequency Response : 45 - 500Hz (Sine Wave)
AC Function
ACA specifications are ac coupled, true R.M.S.
For non-sinusoidal waveforms, Additional
Accuracy by Crest Factor (C.F.):
Add 3.0% for C.F. 1.0 - 2.0.
Add 5.0% for C.F. 2.0 - 2.5.
Add 7.0% for C.F. 2.5 - 3.0.
Electrical Specifications
Accuracy is given as ±(% of reading + counts of
least significant digit) at 23°C ± 5°C, with
relative humidity Less than 80% R.H.
Accuracy is specified for a period of one year
after calibration.
Position Error of Clamp :
Accuracy and position error assumes central-
ized primary conductor at optimum position, no
external electrical or magnetic field, and within
operating temperature range.

Environmental Specifications
Indoor Use.
Pollution Degree :2
Maximum operating altitude : 2000m (6562 ft)
Operating temperature & Relative Humidity :
0°C - 30°C, ≦80%RH
30°C - 40°C, ≦75%RH
40°C - 50°C, ≦45%RH
Storage temperature :
-20°C - 60°C, 0 to 80% RH (no batteries).
Temperature coefficient :
0.2 x (Specified accuracy) / °C, < 18°C, > 28°C
Shock vibration : Sinusoidal vibration
MIL-PRF-28800F (5-55 Hz, 3g max.)
Drop Protection : 4 feet drop to hardwood on
concrete Floor
Limited Warranty
This Meter is warranted to the original purchaser
against defects in material and workmanship for
two years from the date of purchase. During this
warranty period, manufacturer will, at its option,
replace or repair the defective unit, subject to
verification of the defect or malfunction. This
warranty does not cover fuses, disposable
batteries, or damage from abuse, neglect,
accident, unauthorized repair, alteration,
contamination, or abnormal conditions of
operation or handling.
Any implied warranties arising out of the sale of
this product, including but not limited to implied
warranties of merchantability and fitness for a
particular purpose, are limited to the above.
The manufacturer shall not be liable for loss of
use of the instrument or other incidental or
consequential damages, expenses, or economic
loss, or for any claim or claims for such damage,
expense or economic loss. Some states or
countries laws vary, so the above limitations or
exclusions may not apply to you.
AC Current
Backlight
PressPress
Data Hold
PressPress
Replacing Batteries
UnlockUnlock
22
11
UnlockUnlock
LockLock
Caution
Remove Meter from conductor and turn off
power before opening the battery cover.
General Specifications
Display : 3000 counts.
Over range Display : "OL"
Dimensions (W x H x D) :
10" coil : Approx. 120mm x 280mm x 25mm
18" coil : Approx. 130mm x 350mm x 25mm
Weight(without Battery) :
10" coil : 170g
18" coil : 200g
Power Requirements : AAA Size Battery x 2
Battery Life : 200 hours ALKALINE Battery
Low Batteries Indication :
Voltage drops below operating voltage,
will flash.
Maximum Conductor Size : coil size
Safety Standard :
EN61010-1, EN61010-2-032, EN61326-1
CAT IV 600V, CAT III 1000V
Conversion Rate : 4 times/second
sFlex-10 Error
Distance from
Optimum (mm)
15 ±2.0%
25 ±2.5%
35 ±3.0%
sFlex-18 Error
Distance from
Optimum (mm)
35 ±1.0%
50 ±1.5%
60 ±2.0%
Range OL
Reading Resolution Accuracy
(45Hz~500Hz)
30.00A 33.00A 0.01A ±(3.0% + 5D)
300.0A 330.0A 0.1A ±(3.0% + 5D)
3000A 3300A 1A ±(3.0% + 5D)
LCD displays 0 counts when the reading < 10
counts.
Frequency Response : 45 - 500Hz (Sine Wave)
AC Function
ACA specifications are ac coupled, true R.M.S.
For non-sinusoidal waveforms, Additional
Accuracy by Crest Factor (C.F.):
Add 3.0% for C.F. 1.0 - 2.0.
Add 5.0% for C.F. 2.0 - 2.5.
Add 7.0% for C.F. 2.5 - 3.0.
Electrical Specifications
Accuracy is given as ±(% of reading + counts of
least significant digit) at 23°C ± 5°C, with
relative humidity Less than 80% R.H.
Accuracy is specified for a period of one year
after calibration.
Position Error of Clamp :
Accuracy and position error assumes central-
ized primary conductor at optimum position, no
external electrical or magnetic field, and within
operating temperature range.

Environmental Specifications
Indoor Use.
Pollution Degree :2
Maximum operating altitude : 2000m (6562 ft)
Operating temperature & Relative Humidity :
0°C - 30°C, ≦80%RH
30°C - 40°C, ≦75%RH
40°C - 50°C, ≦45%RH
Storage temperature :
-20°C - 60°C, 0 to 80% RH (no batteries).
Temperature coefficient :
0.2 x (Specified accuracy) / °C, < 18°C, > 28°C
Shock vibration : Sinusoidal vibration
MIL-PRF-28800F (5-55 Hz, 3g max.)
Drop Protection : 4 feet drop to hardwood on
concrete Floor
Limited Warranty
This Meter is warranted to the original purchaser
against defects in material and workmanship for
two years from the date of purchase. During this
warranty period, manufacturer will, at its option,
replace or repair the defective unit, subject to
verification of the defect or malfunction. This
warranty does not cover fuses, disposable
batteries, or damage from abuse, neglect,
accident, unauthorized repair, alteration,
contamination, or abnormal conditions of
operation or handling.
Any implied warranties arising out of the sale of
this product, including but not limited to implied
warranties of merchantability and fitness for a
particular purpose, are limited to the above.
The manufacturer shall not be liable for loss of
use of the instrument or other incidental or
consequential damages, expenses, or economic
loss, or for any claim or claims for such damage,
expense or economic loss. Some states or
countries laws vary, so the above limitations or
exclusions may not apply to you.
AC Current
Backlight
PressPress
Data Hold
PressPress
Replacing Batteries
UnlockUnlock
22
11
UnlockUnlock
LockLock
Caution
Remove Meter from conductor and turn off
power before opening the battery cover.
General Specifications
Display : 3000 counts.
Over range Display : "OL"
Dimensions (W x H x D) :
10" coil : Approx. 120mm x 280mm x 25mm
18" coil : Approx. 130mm x 350mm x 25mm
Weight(without Battery) :
10" coil : 170g
18" coil : 200g
Power Requirements : AAA Size Battery x 2
Battery Life : 200 hours ALKALINE Battery
Low Batteries Indication :
Voltage drops below operating voltage,
will flash.
Maximum Conductor Size : coil size
Safety Standard :
EN61010-1, EN61010-2-032, EN61326-1
CAT IV 600V, CAT III 1000V
Conversion Rate : 4 times/second
sFlex-10 Error
Distance from
Optimum (mm)
15 ±2.0%
25 ±2.5%
35 ±3.0%
sFlex-18 Error
Distance from
Optimum (mm)
35 ±1.0%
50 ±1.5%
60 ±2.0%
Range OL
Reading Resolution Accuracy
(45Hz~500Hz)
30.00A 33.00A 0.01A ±(3.0% + 5D)
300.0A 330.0A 0.1A ±(3.0% + 5D)
3000A 3300A 1A ±(3.0% + 5D)
LCD displays 0 counts when the reading < 10
counts.
Frequency Response : 45 - 500Hz (Sine Wave)
AC Function
ACA specifications are ac coupled, true R.M.S.
For non-sinusoidal waveforms, Additional
Accuracy by Crest Factor (C.F.):
Add 3.0% for C.F. 1.0 - 2.0.
Add 5.0% for C.F. 2.0 - 2.5.
Add 7.0% for C.F. 2.5 - 3.0.
Electrical Specifications
Accuracy is given as ±(% of reading + counts of
least significant digit) at 23°C ± 5°C, with
relative humidity Less than 80% R.H.
Accuracy is specified for a period of one year
after calibration.
Position Error of Clamp :
Accuracy and position error assumes central-
ized primary conductor at optimum position, no
external electrical or magnetic field, and within
operating temperature range.

Environmental Specifications
Indoor Use.
Pollution Degree :2
Maximum operating altitude : 2000m (6562 ft)
Operating temperature & Relative Humidity :
0°C - 30°C, ≦80%RH
30°C - 40°C, ≦75%RH
40°C - 50°C, ≦45%RH
Storage temperature :
-20°C - 60°C, 0 to 80% RH (no batteries).
Temperature coefficient :
0.2 x (Specified accuracy) / °C, < 18°C, > 28°C
Shock vibration : Sinusoidal vibration
MIL-PRF-28800F (5-55 Hz, 3g max.)
Drop Protection : 4 feet drop to hardwood on
concrete Floor
Limited Warranty
This Meter is warranted to the original purchaser
against defects in material and workmanship for
two years from the date of purchase. During this
warranty period, manufacturer will, at its option,
replace or repair the defective unit, subject to
verification of the defect or malfunction. This
warranty does not cover fuses, disposable
batteries, or damage from abuse, neglect,
accident, unauthorized repair, alteration,
contamination, or abnormal conditions of
operation or handling.
Any implied warranties arising out of the sale of
this product, including but not limited to implied
warranties of merchantability and fitness for a
particular purpose, are limited to the above.
The manufacturer shall not be liable for loss of
use of the instrument or other incidental or
consequential damages, expenses, or economic
loss, or for any claim or claims for such damage,
expense or economic loss. Some states or
countries laws vary, so the above limitations or
exclusions may not apply to you.
AC Current
Backlight
PressPress
Data Hold
PressPress
Replacing Batteries
UnlockUnlock
22
11
UnlockUnlock
LockLock
Caution
Remove Meter from conductor and turn off
power before opening the battery cover.
General Specifications
Display : 3000 counts.
Over range Display : "OL"
Dimensions (W x H x D) :
10" coil : Approx. 120mm x 280mm x 25mm
18" coil : Approx. 130mm x 350mm x 25mm
Weight(without Battery) :
10" coil : 170g
18" coil : 200g
Power Requirements : AAA Size Battery x 2
Battery Life : 200 hours ALKALINE Battery
Low Batteries Indication :
Voltage drops below operating voltage,
will flash.
Maximum Conductor Size : coil size
Safety Standard :
EN61010-1, EN61010-2-032, EN61326-1
CAT IV 600V, CAT III 1000V
Conversion Rate : 4 times/second
sFlex-10 Error
Distance from
Optimum (mm)
15 ±2.0%
25 ±2.5%
35 ±3.0%
sFlex-18 Error
Distance from
Optimum (mm)
35 ±1.0%
50 ±1.5%
60 ±2.0%
Range OL
Reading Resolution Accuracy
(45Hz~500Hz)
30.00A 33.00A 0.01A ±(3.0% + 5D)
300.0A 330.0A 0.1A ±(3.0% + 5D)
3000A 3300A 1A ±(3.0% + 5D)
LCD displays 0 counts when the reading < 10
counts.
Frequency Response : 45 - 500Hz (Sine Wave)
AC Function
ACA specifications are ac coupled, true R.M.S.
For non-sinusoidal waveforms, Additional
Accuracy by Crest Factor (C.F.):
Add 3.0% for C.F. 1.0 - 2.0.
Add 5.0% for C.F. 2.0 - 2.5.
Add 7.0% for C.F. 2.5 - 3.0.
Electrical Specifications
Accuracy is given as ±(% of reading + counts of
least significant digit) at 23°C ± 5°C, with
relative humidity Less than 80% R.H.
Accuracy is specified for a period of one year
after calibration.
Position Error of Clamp :
Accuracy and position error assumes central-
ized primary conductor at optimum position, no
external electrical or magnetic field, and within
operating temperature range.

Environmental Specifications
Indoor Use.
Pollution Degree :2
Maximum operating altitude : 2000m (6562 ft)
Operating temperature & Relative Humidity :
0°C - 30°C, ≦80%RH
30°C - 40°C, ≦75%RH
40°C - 50°C, ≦45%RH
Storage temperature :
-20°C - 60°C, 0 to 80% RH (no batteries).
Temperature coefficient :
0.2 x (Specified accuracy) / °C, < 18°C, > 28°C
Shock vibration : Sinusoidal vibration
MIL-PRF-28800F (5-55 Hz, 3g max.)
Drop Protection : 4 feet drop to hardwood on
concrete Floor
Limited Warranty
This Meter is warranted to the original purchaser
against defects in material and workmanship for
two years from the date of purchase. During this
warranty period, manufacturer will, at its option,
replace or repair the defective unit, subject to
verification of the defect or malfunction. This
warranty does not cover fuses, disposable
batteries, or damage from abuse, neglect,
accident, unauthorized repair, alteration,
contamination, or abnormal conditions of
operation or handling.
Any implied warranties arising out of the sale of
this product, including but not limited to implied
warranties of merchantability and fitness for a
particular purpose, are limited to the above.
The manufacturer shall not be liable for loss of
use of the instrument or other incidental or
consequential damages, expenses, or economic
loss, or for any claim or claims for such damage,
expense or economic loss. Some states or
countries laws vary, so the above limitations or
exclusions may not apply to you.
AC Current
Backlight
PressPress
Data Hold
PressPress
Replacing Batteries
UnlockUnlock
22
11
UnlockUnlock
LockLock
Caution
Remove Meter from conductor and turn off
power before opening the battery cover.
General Specifications
Display : 3000 counts.
Over range Display : "OL"
Dimensions (W x H x D) :
10" coil : Approx. 120mm x 280mm x 25mm
18" coil : Approx. 130mm x 350mm x 25mm
Weight(without Battery) :
10" coil : 170g
18" coil : 200g
Power Requirements : AAA Size Battery x 2
Battery Life : 200 hours ALKALINE Battery
Low Batteries Indication :
Voltage drops below operating voltage,
will flash.
Maximum Conductor Size : coil size
Safety Standard :
EN61010-1, EN61010-2-032, EN61326-1
CAT IV 600V, CAT III 1000V
Conversion Rate : 4 times/second
sFlex-10 Error
Distance from
Optimum (mm)
15 ±2.0%
25 ±2.5%
35 ±3.0%
sFlex-18 Error
Distance from
Optimum (mm)
35 ±1.0%
50 ±1.5%
60 ±2.0%
Range OL
Reading Resolution Accuracy
(45Hz~500Hz)
30.00A 33.00A 0.01A ±(3.0% + 5D)
300.0A 330.0A 0.1A ±(3.0% + 5D)
3000A 3300A 1A ±(3.0% + 5D)
LCD displays 0 counts when the reading < 10
counts.
Frequency Response : 45 - 500Hz (Sine Wave)
AC Function
ACA specifications are ac coupled, true R.M.S.
For non-sinusoidal waveforms, Additional
Accuracy by Crest Factor (C.F.):
Add 3.0% for C.F. 1.0 - 2.0.
Add 5.0% for C.F. 2.0 - 2.5.
Add 7.0% for C.F. 2.5 - 3.0.
Electrical Specifications
Accuracy is given as ±(% of reading + counts of
least significant digit) at 23°C ± 5°C, with
relative humidity Less than 80% R.H.
Accuracy is specified for a period of one year
after calibration.
Position Error of Clamp :
Accuracy and position error assumes central-
ized primary conductor at optimum position, no
external electrical or magnetic field, and within
operating temperature range.

Environmental Specifications
Indoor Use.
Pollution Degree :2
Maximum operating altitude : 2000m (6562 ft)
Operating temperature & Relative Humidity :
0°C - 30°C, ≦80%RH
30°C - 40°C, ≦75%RH
40°C - 50°C, ≦45%RH
Storage temperature :
-20°C - 60°C, 0 to 80% RH (no batteries).
Temperature coefficient :
0.2 x (Specified accuracy) / °C, < 18°C, > 28°C
Shock vibration : Sinusoidal vibration
MIL-PRF-28800F (5-55 Hz, 3g max.)
Drop Protection : 4 feet drop to hardwood on
concrete Floor
Limited Warranty
This Meter is warranted to the original purchaser
against defects in material and workmanship for
two years from the date of purchase. During this
warranty period, manufacturer will, at its option,
replace or repair the defective unit, subject to
verification of the defect or malfunction. This
warranty does not cover fuses, disposable
batteries, or damage from abuse, neglect,
accident, unauthorized repair, alteration,
contamination, or abnormal conditions of
operation or handling.
Any implied warranties arising out of the sale of
this product, including but not limited to implied
warranties of merchantability and fitness for a
particular purpose, are limited to the above.
The manufacturer shall not be liable for loss of
use of the instrument or other incidental or
consequential damages, expenses, or economic
loss, or for any claim or claims for such damage,
expense or economic loss. Some states or
countries laws vary, so the above limitations or
exclusions may not apply to you.
AC Current
Backlight
PressPress
Data Hold
PressPress
Replacing Batteries
UnlockUnlock
22
11
UnlockUnlock
LockLock
Caution
Remove Meter from conductor and turn off
power before opening the battery cover.
General Specifications
Display : 3000 counts.
Over range Display : "OL"
Dimensions (W x H x D) :
10" coil : Approx. 120mm x 280mm x 25mm
18" coil : Approx. 130mm x 350mm x 25mm
Weight(without Battery) :
10" coil : 170g
18" coil : 200g
Power Requirements : AAA Size Battery x 2
Battery Life : 200 hours ALKALINE Battery
Low Batteries Indication :
Voltage drops below operating voltage,
will flash.
Maximum Conductor Size : coil size
Safety Standard :
EN61010-1, EN61010-2-032, EN61326-1
CAT IV 600V, CAT III 1000V
Conversion Rate : 4 times/second
sFlex-10 Error
Distance from
Optimum (mm)
15 ±2.0%
25 ±2.5%
35 ±3.0%
sFlex-18 Error
Distance from
Optimum (mm)
35 ±1.0%
50 ±1.5%
60 ±2.0%
Range OL
Reading Resolution Accuracy
(45Hz~500Hz)
30.00A 33.00A 0.01A ±(3.0% + 5D)
300.0A 330.0A 0.1A ±(3.0% + 5D)
3000A 3300A 1A ±(3.0% + 5D)
LCD displays 0 counts when the reading < 10
counts.
Frequency Response : 45 - 500Hz (Sine Wave)
AC Function
ACA specifications are ac coupled, true R.M.S.
For non-sinusoidal waveforms, Additional
Accuracy by Crest Factor (C.F.):
Add 3.0% for C.F. 1.0 - 2.0.
Add 5.0% for C.F. 2.0 - 2.5.
Add 7.0% for C.F. 2.5 - 3.0.
Electrical Specifications
Accuracy is given as ±(% of reading + counts of
least significant digit) at 23°C ± 5°C, with
relative humidity Less than 80% R.H.
Accuracy is specified for a period of one year
after calibration.
Position Error of Clamp :
Accuracy and position error assumes central-
ized primary conductor at optimum position, no
external electrical or magnetic field, and within
operating temperature range.

Environmental Specifications
Indoor Use.
Pollution Degree :2
Maximum operating altitude : 2000m (6562 ft)
Operating temperature & Relative Humidity :
0°C - 30°C, ≦80%RH
30°C - 40°C, ≦75%RH
40°C - 50°C, ≦45%RH
Storage temperature :
-20°C - 60°C, 0 to 80% RH (no batteries).
Temperature coefficient :
0.2 x (Specified accuracy) / °C, < 18°C, > 28°C
Shock vibration : Sinusoidal vibration
MIL-PRF-28800F (5-55 Hz, 3g max.)
Drop Protection : 4 feet drop to hardwood on
concrete Floor
Limited Warranty
This Meter is warranted to the original purchaser
against defects in material and workmanship for
two years from the date of purchase. During this
warranty period, manufacturer will, at its option,
replace or repair the defective unit, subject to
verification of the defect or malfunction. This
warranty does not cover fuses, disposable
batteries, or damage from abuse, neglect,
accident, unauthorized repair, alteration,
contamination, or abnormal conditions of
operation or handling.
Any implied warranties arising out of the sale of
this product, including but not limited to implied
warranties of merchantability and fitness for a
particular purpose, are limited to the above.
The manufacturer shall not be liable for loss of
use of the instrument or other incidental or
consequential damages, expenses, or economic
loss, or for any claim or claims for such damage,
expense or economic loss. Some states or
countries laws vary, so the above limitations or
exclusions may not apply to you.
AC Current
Backlight
PressPress
Data Hold
PressPress
Replacing Batteries
UnlockUnlock
22
11
UnlockUnlock
LockLock
Caution
Remove Meter from conductor and turn off
power before opening the battery cover.
General Specifications
Display : 3000 counts.
Over range Display : "OL"
Dimensions (W x H x D) :
10" coil : Approx. 120mm x 280mm x 25mm
18" coil : Approx. 130mm x 350mm x 25mm
Weight(without Battery) :
10" coil : 170g
18" coil : 200g
Power Requirements : AAA Size Battery x 2
Battery Life : 200 hours ALKALINE Battery
Low Batteries Indication :
Voltage drops below operating voltage,
will flash.
Maximum Conductor Size : coil size
Safety Standard :
EN61010-1, EN61010-2-032, EN61326-1
CAT IV 600V, CAT III 1000V
Conversion Rate : 4 times/second
sFlex-10 Error
Distance from
Optimum (mm)
15 ±2.0%
25 ±2.5%
35 ±3.0%
sFlex-18 Error
Distance from
Optimum (mm)
35 ±1.0%
50 ±1.5%
60 ±2.0%
Range OL
Reading Resolution Accuracy
(45Hz~500Hz)
30.00A 33.00A 0.01A ±(3.0% + 5D)
300.0A 330.0A 0.1A ±(3.0% + 5D)
3000A 3300A 1A ±(3.0% + 5D)
LCD displays 0 counts when the reading < 10
counts.
Frequency Response : 45 - 500Hz (Sine Wave)
AC Function
ACA specifications are ac coupled, true R.M.S.
For non-sinusoidal waveforms, Additional
Accuracy by Crest Factor (C.F.):
Add 3.0% for C.F. 1.0 - 2.0.
Add 5.0% for C.F. 2.0 - 2.5.
Add 7.0% for C.F. 2.5 - 3.0.
Electrical Specifications
Accuracy is given as ±(% of reading + counts of
least significant digit) at 23°C ± 5°C, with
relative humidity Less than 80% R.H.
Accuracy is specified for a period of one year
after calibration.
Position Error of Clamp :
Accuracy and position error assumes central-
ized primary conductor at optimum position, no
external electrical or magnetic field, and within
operating temperature range.

Warning
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . .
Safety sheet
. . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . .
. . . . . . . . .
Read First
APPA sFLEX-D Instruction
Flexible Current Meter Card
Safety Information
To ensure safe operation and service of the
meter, follow these instructions.Failure to
observe warnings can result in severe Injury.
Warning
Identify hazardous conditions and actions that
could cause BODILY HARM or DEATH
-Individual protective equipment should be used
if HAZARDOUS LIVE parts in the installation
where measurement is to be carried out could
Be ACCESSIBLE.
-If the equipment is used in a manner not
specified by the manufacturer, the protection
provided by the equipment may be impaired.
-To reduce the risk of fire or electric shock, do
not expose this product to rain or moisture.
-Verify the Meter operation by measuring a
known Current. If in doubt, have the Meter
serviced.
-Do not apply more than the rated voltage/
Current, as marked on Meter.
-To avoid false readings that can lead to electric
shock and injury, replace battery as soon as low
battery indicator.
- Do not use Meter around explosive gas or
vapor.
See instruction manual
Do not apply to or remove from
HAZARDOUS LIVE conductors.
Equipment protected by double or
reinforced insulation
Battery
Earth
Conforms to EU directives
Do not discard this product or throw away
AC measurement
Maintenance
Do not attempt to repair this Clamp Meter.
It contains no user-serviceable parts.
Repair or servicing should only be performed by
qualified personnel.
Cleaning
Periodically wipe the case with a dry cloth and
detergent. Do not use abrasives or solvents.
Power on/off
APPA TECHNOLOGY CORP.
9F, 119-1 Pao-Zong Rd.,
Shin-Tien, Taipei, 23115, Taiwan.
P.O.Box. 12-24 Shin-Tien, Taiwan.
Tel: +886-2-29178820
Fax: +886-2-29170848
http: //www.appatech.com
- Do not use a flexible current sensor if the
inner copper wire of the flexible cord is visible.
- De-energize the installation under test or wear
suitable protective clothing during fitting a
removal of the Flexible Current Probe.
- Do not apply around or remove from UNINSU
LATED HAZARDOUS LIVE conductors, which
may render electric shock, electric burn, or
arc flash.
Caution
Identify conditions and actions that could cause
DAMAGE the meter or equipment under test.
Do not expose Meter to extremes in tempera-
ture of high humidity.
Symbols as marked on the meter and
Instruction card
Press > 2Sec
OFF
Press > 2Sec
OFF
Press > 2Sec
OFF
Press > 2Sec
OFF
Flexible AC current probe
Lock
Lock
Unlock
Unlock
Do not touch the
area above the LCD
during measurement
Auto Power off Function
20 min20 min20 min20 min
The meter automatically turns the power off
after powering up and no operation for
20 minutes ± 30 seconds.
Disable Auto Power off Function
Press > 2Sec
OFF
Press
Press “Power” button, and during the LCD
shows battery capacity, press “backlight”
button until the LCD shows “AoFF” to disable
Auto Power Off Function.

Warning
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . .
Safety sheet
. . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . .
. . . . . . . . .
Read First
APPA sFLEX-D Instruction
Flexible Current Meter Card
Safety Information
To ensure safe operation and service of the
meter, follow these instructions.Failure to
observe warnings can result in severe Injury.
Warning
Identify hazardous conditions and actions that
could cause BODILY HARM or DEATH
-Individual protective equipment should be used
if HAZARDOUS LIVE parts in the installation
where measurement is to be carried out could
Be ACCESSIBLE.
-If the equipment is used in a manner not
specified by the manufacturer, the protection
provided by the equipment may be impaired.
-To reduce the risk of fire or electric shock, do
not expose this product to rain or moisture.
-Verify the Meter operation by measuring a
known Current. If in doubt, have the Meter
serviced.
-Do not apply more than the rated voltage/
Current, as marked on Meter.
-To avoid false readings that can lead to electric
shock and injury, replace battery as soon as low
battery indicator.
- Do not use Meter around explosive gas or
vapor.
See instruction manual
Do not apply to or remove from
HAZARDOUS LIVE conductors.
Equipment protected by double or
reinforced insulation
Battery
Earth
Conforms to EU directives
Do not discard this product or throw away
AC measurement
Maintenance
Do not attempt to repair this Clamp Meter.
It contains no user-serviceable parts.
Repair or servicing should only be performed by
qualified personnel.
Cleaning
Periodically wipe the case with a dry cloth and
detergent. Do not use abrasives or solvents.
Power on/off
APPA TECHNOLOGY CORP.
9F, 119-1 Pao-Zong Rd.,
Shin-Tien, Taipei, 23115, Taiwan.
P.O.Box. 12-24 Shin-Tien, Taiwan.
Tel: +886-2-29178820
Fax: +886-2-29170848
http: //www.appatech.com
- Do not use a flexible current sensor if the
inner copper wire of the flexible cord is visible.
- De-energize the installation under test or wear
suitable protective clothing during fitting a
removal of the Flexible Current Probe.
- Do not apply around or remove from UNINSU
LATED HAZARDOUS LIVE conductors, which
may render electric shock, electric burn, or
arc flash.
Caution
Identify conditions and actions that could cause
DAMAGE the meter or equipment under test.
Do not expose Meter to extremes in tempera-
ture of high humidity.
Symbols as marked on the meter and
Instruction card
Press > 2Sec
OFF
Press > 2Sec
OFF
Press > 2Sec
OFF
Press > 2Sec
OFF
Flexible AC current probe
Lock
Lock
Unlock
Unlock
Do not touch the
area above the LCD
during measurement
Auto Power off Function
20 min20 min20 min20 min
The meter automatically turns the power off
after powering up and no operation for
20 minutes ± 30 seconds.
Disable Auto Power off Function
Press > 2Sec
OFF
Press
Press “Power” button, and during the LCD
shows battery capacity, press “backlight”
button until the LCD shows “AoFF” to disable
Auto Power Off Function.

Environmental Specifications
Indoor Use.
Pollution Degree :2
Maximum operating altitude : 2000m (6562 ft)
Operating temperature & Relative Humidity :
0°C - 30°C, ≦80%RH
30°C - 40°C, ≦75%RH
40°C - 50°C, ≦45%RH
Storage temperature :
-20°C - 60°C, 0 to 80% RH (no batteries).
Temperature coefficient :
0.2 x (Specified accuracy) / °C, < 18°C, > 28°C
Shock vibration : Sinusoidal vibration
MIL-PRF-28800F (5-55 Hz, 3g max.)
Drop Protection : 4 feet drop to hardwood on
concrete Floor
Limited Warranty
This Meter is warranted to the original purchaser
against defects in material and workmanship for
two years from the date of purchase. During this
warranty period, manufacturer will, at its option,
replace or repair the defective unit, subject to
verification of the defect or malfunction. This
warranty does not cover fuses, disposable
batteries, or damage from abuse, neglect,
accident, unauthorized repair, alteration,
contamination, or abnormal conditions of
operation or handling.
Any implied warranties arising out of the sale of
this product, including but not limited to implied
warranties of merchantability and fitness for a
particular purpose, are limited to the above.
The manufacturer shall not be liable for loss of
use of the instrument or other incidental or
consequential damages, expenses, or economic
loss, or for any claim or claims for such damage,
expense or economic loss. Some states or
countries laws vary, so the above limitations or
exclusions may not apply to you.
AC Current
Backlight
PressPress
Data Hold
PressPress
Replacing Batteries
UnlockUnlock
22
11
UnlockUnlock
LockLock
Caution
Remove Meter from conductor and turn off
power before opening the battery cover.
General Specifications
Display : 3000 counts.
Over range Display : "OL"
Dimensions (W x H x D) :
10" coil : Approx. 120mm x 280mm x 25mm
18" coil : Approx. 130mm x 350mm x 25mm
Weight(without Battery) :
10" coil : 170g
18" coil : 200g
Power Requirements : AAA Size Battery x 2
Battery Life : 200 hours ALKALINE Battery
Low Batteries Indication :
Voltage drops below operating voltage,
will flash.
Maximum Conductor Size : coil size
Safety Standard :
EN61010-1, EN61010-2-032, EN61326-1
CAT IV 600V, CAT III 1000V
Conversion Rate : 4 times/second
sFlex-10 Error
Distance from
Optimum (mm)
15 ±2.0%
25 ±2.5%
35 ±3.0%
sFlex-18 Error
Distance from
Optimum (mm)
35 ±1.0%
50 ±1.5%
60 ±2.0%
Range OL
Reading Resolution Accuracy
(45Hz~500Hz)
30.00A 33.00A 0.01A ±(3.0% + 5D)
300.0A 330.0A 0.1A ±(3.0% + 5D)
3000A 3300A 1A ±(3.0% + 5D)
LCD displays 0 counts when the reading < 10
counts.
Frequency Response : 45 - 500Hz (Sine Wave)
AC Function
ACA specifications are ac coupled, true R.M.S.
For non-sinusoidal waveforms, Additional
Accuracy by Crest Factor (C.F.):
Add 3.0% for C.F. 1.0 - 2.0.
Add 5.0% for C.F. 2.0 - 2.5.
Add 7.0% for C.F. 2.5 - 3.0.
Electrical Specifications
Accuracy is given as ±(% of reading + counts of
least significant digit) at 23°C ± 5°C, with
relative humidity Less than 80% R.H.
Accuracy is specified for a period of one year
after calibration.
Position Error of Clamp :
Accuracy and position error assumes central-
ized primary conductor at optimum position, no
external electrical or magnetic field, and within
operating temperature range.

Warning
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . .
Safety sheet
. . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . .
. . . . . . . . .
Read First
APPA sFLEX-D Instruction
Flexible Current Meter Card
Safety Information
To ensure safe operation and service of the
meter, follow these instructions.Failure to
observe warnings can result in severe Injury.
Warning
Identify hazardous conditions and actions that
could cause BODILY HARM or DEATH
-Individual protective equipment should be used
if HAZARDOUS LIVE parts in the installation
where measurement is to be carried out could
Be ACCESSIBLE.
-If the equipment is used in a manner not
specified by the manufacturer, the protection
provided by the equipment may be impaired.
-To reduce the risk of fire or electric shock, do
not expose this product to rain or moisture.
-Verify the Meter operation by measuring a
known Current. If in doubt, have the Meter
serviced.
-Do not apply more than the rated voltage/
Current, as marked on Meter.
-To avoid false readings that can lead to electric
shock and injury, replace battery as soon as low
battery indicator.
- Do not use Meter around explosive gas or
vapor.
See instruction manual
Do not apply to or remove from
HAZARDOUS LIVE conductors.
Equipment protected by double or
reinforced insulation
Battery
Earth
Conforms to EU directives
Do not discard this product or throw away
AC measurement
Maintenance
Do not attempt to repair this Clamp Meter.
It contains no user-serviceable parts.
Repair or servicing should only be performed by
qualified personnel.
Cleaning
Periodically wipe the case with a dry cloth and
detergent. Do not use abrasives or solvents.
Power on/off
APPA TECHNOLOGY CORP.
9F, 119-1 Pao-Zong Rd.,
Shin-Tien, Taipei, 23115, Taiwan.
P.O.Box. 12-24 Shin-Tien, Taiwan.
Tel: +886-2-29178820
Fax: +886-2-29170848
http: //www.appatech.com
- Do not use a flexible current sensor if the
inner copper wire of the flexible cord is visible.
- De-energize the installation under test or wear
suitable protective clothing during fitting a
removal of the Flexible Current Probe.
- Do not apply around or remove from UNINSU
LATED HAZARDOUS LIVE conductors, which
may render electric shock, electric burn, or
arc flash.
Caution
Identify conditions and actions that could cause
DAMAGE the meter or equipment under test.
Do not expose Meter to extremes in tempera-
ture of high humidity.
Symbols as marked on the meter and
Instruction card
Press > 2Sec
OFF
Press > 2Sec
OFF
Press > 2Sec
OFF
Press > 2Sec
OFF
Flexible AC current probe
Lock
Lock
Unlock
Unlock
Do not touch the
area above the LCD
during measurement
Auto Power off Function
20 min20 min20 min20 min
The meter automatically turns the power off
after powering up and no operation for
20 minutes ± 30 seconds.
Disable Auto Power off Function
Press > 2Sec
OFF
Press
Press “Power” button, and during the LCD
shows battery capacity, press “backlight”
button until the LCD shows “AoFF” to disable
Auto Power Off Function.