Overload protection : 600Vrms 
Open circuit Voltage : -1.3V approx. 
* 1 < 100 dgt rolling. 
* 2 < 10 dgt rolling. 
 
Diode Check and Continuity 
Resolution Accuracy 
10 mV ±(1.5%+5 dgt)* 
*For 0.4V ~ 0.8V 
 
Max.Test Current : 1.5mA 
 
Max. Open Circuit Voltage : 3V 
 
Overload Protection : 600Vrms 
Continuity : Built-in buzzer sounds when 
resistance is less than approximately  
100Ω. Response time is approximately 
100 msec. 
 
DC µA
 
Range Accuracy 
4.000nF ±(3.0% + 20 dgt) 
40.00nF 
400.0nF 
4.000µF 
40.00µF 
400.0µF 
4.000mF * 
±(2.0% + 8 dgt) 
Function Range Accuracy 
A
(50~60Hz) 0.0 ~399.9A 
400~600A ±(1.9% + 5 dgt) 
Overload protection : 600 Arms 
AC Conversion Type : 
Average sensing rms indication. 
Position Error : ±1.5% of reading. 
 
Temperature 
Function Range Accuracy 
-40.0 °C ~ 0.0°C 1% ±4 °C 
0.0 °C ~ 400.0 °C 1% ±3 °C 
°F 
-40 °F ~ 32 °F 1% ±8 °F 
32 °F ~ 750 °F 1% ±6 °F 
°C 
750 °F ~ 1000 °F 2% ±8 °F 
Auto Power Off (APO)
 
 
The meter will automatically shut itself 
off after approximately 10 minutes after 
power on. 
Maintenance 
 
Do not attempt to repair this Meter. 
It contains no user-serviceable parts. 
Repair or serving should only be 
performed by qualified personal. 
 
 
Cleaning 
 
Periodically wipe the case with a dry 
cloth and detergent do not use abrasives 
or solvents. 
AC Current 
Range Accuracy 
400.0 µA 
4000 µA ±(1%+2 dgt)* 
Voltage barden : < 5mV / µA 
Overload Protection : 600Vrms 
 
Capacitance 
Overload Protection : 600Vrms 
* < 50 dgt of reading rolling. 
Overload Protection : 600Vrms 
Specifications 
1-1 General Specifications 
 
LCD display digits : 
3 3/4 digit large scale LCD readout. 
Display count : 4000 counts. 
Measuring rate : 1.5 times / sec. 
Overrange display : 
“OL” is displayed for “Ω” functions, 
shows the real value for “A” and “V” and 
“ °C/°F “function. 
Automatic power off time : 
Approximately 10 minutes after power on. 
Low battery indicator : 
is displayed. 
Power requirement : 9V battery 
Battery life : ALKALINE 9V 200 hours 
 
1-2 Environmental Conditions 
 
Indoor Use. 
Calibration : 
One year calibration cycle. 
Operating temperature : 
0°C ~ 30°C (≦80% RH) 
30°C ~ 40°C (≦75% RH) 
40°C ~ 50°C (≦45%RH) 
Storage temperature : 
-20 to +60°C, 0 to 80% RH (batteries not 
fitted). 
Overvoltage category : 
IEC 61010-1 600V CAT.Ⅲ. 
 
CAT Application field 
ⅠThe circuits not connected to 
mains. 
ⅡThe circuits directly connected to 
Low-voltage installation. 
ⅢThe building installation. 
ⅣThe source of the Low-voltage 
installation. 
Operating altitude : 2000m (6562 ft) 
Conductor Size : 32mm diameter. 
Pollution degree : 2 
EMC : EN 61326-1 
Shock vibration : 
Sinusoidal vibration per MIL-T-28800E 
(5 ~ 55 Hz, 3g maximum). 
1-3 Electrical Specifications 
Accuracy is ±(% reading + number 
of digits) at 23°C ±5°C < 80%RH. 
Temperature coefficient : 
Add 0.2 x (Specified accuracy) / °C, 
< 18°C, > 28°C . 
 
DC / AC Volts
 
Range DC Accuracy AC Accuracy 
4.000V 
±(0.9% + 2dgt) 
±(1.5% + 5dgt) 
50Hz ~ 500Hz 
 
40.00V 
400.0V 
DC1000V 
AC750V  
Over voltage protection : 
DC 1000 V or AC 750 V 
 
Input Impedance : 
10MΩ// less than 100pF. 
 
CMRR / NMRR : 
(Common Mode Rejection Ratio) 
(Normal Mode Rejection Ratio) 
 
VAC : CMRR > 60dB at DC, 50Hz / 60Hz 
VDC : CMRR > 100dB at DC, 
 50Hz / 60Hz 
 NMRR > 50dB at DC, 50Hz / 60Hz 
 
AC Conversion Type :  
Average sensing rms indication. 
 
Resistance 
Range Accuracy 
400.0 Ω*2  ±(1.0% + 2 dgt)  
4.000 KΩ 
40.00 KΩ 
400.0 KΩ 
4.000MΩ  ±(1.0% + 2 dgt)  
40.00 MΩ*1  ±(1.5% + 2 dgt)  
±(0.7% + 2 dgt)  
Droop Protection :  
4 feet droop to hardwood on concrete 
floor.