Off / On
Auto Power Off
Auto Power Off disable :
Press backlight buttons than switch the
rotary knob to power on the Meter.
Warning
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . .
Safety sheet
. . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . .
. . . . . . . . .

Read First
-Avoid working alone so assistance can
be rendered.
- Do not use test leads or the Meter if
they look damaged.
- Do not use the Meter if the Meter is
not operating properly or if it is wet.
- Use the Meter only as specified in the
Instruction card or the protection by
the Meter might be impaired.
- Use extreme caution when working
around bare conductors or bus bar.
Contact with the conductor could
result in electric shock.
- Use caution with voltages above 30 V
ac rms, or 60 V dc. These voltages
pose a shock hazard.
Symbols as marked on the Meter
and Instruction card
Risk of electric shock
See instruction card
DC measurement
Equipment protected by double or
reinforced insulation
Battery
Earth
AC measurement
Conforms to EU directives

Safety Information
To ensure safe operation and service
of the Meter, follow these instructions.
Failure to observe warnings can result
in severe injury or death.
AC V / DC V
Resistance

Caution
If the meter is used in the vicinity of
equipment which generates electromag-
netic interference, the display may
become unstable or the measurements
show may be subject to large errors.
10min
Clamp Meter Card
A10N Instruction
Diode / Continuity
<
DC µA
Capacitance
Temperature ACA
In order to ensure the best accuracy in
measurement of low capacitance, note
and subtract the stray capacitance of
the open circuited test leads from
subsequent measurements.
Data Hold
392A
0A
392A
CAT.Ⅲ.600V
with respect to
earth for the jaw.
Tactile Barrier for
hand guard.
Do not hold
the meter across
the Tactile Barrier.
I
I
I
I
OK
OK
OK I + (-I) = 0
Back Light
Overload protection : 600Vrms
Open circuit Voltage : -1.3V approx.
* 1 < 100 dgt rolling.
* 2 < 10 dgt rolling.
Diode Check and Continuity
Resolution Accuracy
10 mV ±(1.5%+5 dgt)*
*For 0.4V ~ 0.8V
Max.Test Current : 1.5mA
Max. Open Circuit Voltage : 3V
Overload Protection : 600Vrms
Continuity : Built-in buzzer sounds when
resistance is less than approximately
100Ω. Response time is approximately
100 msec.
DC µA
Range Accuracy
4.000nF ±(3.0% + 20 dgt)
40.00nF
400.0nF
4.000µF
40.00µF
400.0µF
4.000mF *
±(2.0% + 8 dgt)
Function Range Accuracy
A
(50~60Hz) 0.0 ~399.9A
400~600A ±(1.9% + 5 dgt)
Overload protection : 600 Arms
AC Conversion Type :
Average sensing rms indication.
Position Error : ±1.5% of reading.
Temperature
Function Range Accuracy
-40.0 °C ~ 0.0°C 1% ±4 °C
0.0 °C ~ 400.0 °C 1% ±3 °C
°F
-40 °F ~ 32 °F 1% ±8 °F
32 °F ~ 750 °F 1% ±6 °F
°C
750 °F ~ 1000 °F 2% ±8 °F
Auto Power Off (APO)
The meter will automatically shut itself
off after approximately 10 minutes after
power on.
Maintenance
Do not attempt to repair this Meter.
It contains no user-serviceable parts.
Repair or serving should only be
performed by qualified personal.
Cleaning
Periodically wipe the case with a dry
cloth and detergent do not use abrasives
or solvents.
AC Current
Range Accuracy
400.0 µA
4000 µA ±(1%+2 dgt)*
Voltage barden : < 5mV / µA
Overload Protection : 600Vrms
Capacitance
Overload Protection : 600Vrms
* < 50 dgt of reading rolling.
Overload Protection : 600Vrms
Specifications
1-1 General Specifications
LCD display digits :
3 3/4 digit large scale LCD readout.
Display count : 4000 counts.
Measuring rate : 1.5 times / sec.
Overrange display :
“OL” is displayed for “Ω” functions,
shows the real value for “A” and “V” and
“ °C/°F “function.
Automatic power off time :
Approximately 10 minutes after power on.
Low battery indicator :
is displayed.
Power requirement : 9V battery
Battery life : ALKALINE 9V 200 hours
1-2 Environmental Conditions
Indoor Use.
Calibration :
One year calibration cycle.
Operating temperature :
0°C ~ 30°C (≦80% RH)
30°C ~ 40°C (≦75% RH)
40°C ~ 50°C (≦45%RH)
Storage temperature :
-20 to +60°C, 0 to 80% RH (batteries not
fitted).
Overvoltage category :
IEC 61010-1 600V CAT.Ⅲ.
CAT Application field
ⅠThe circuits not connected to
mains.
ⅡThe circuits directly connected to
Low-voltage installation.
ⅢThe building installation.
ⅣThe source of the Low-voltage
installation.
Operating altitude : 2000m (6562 ft)
Conductor Size : 32mm diameter.
Pollution degree : 2
EMC : EN 61326-1
Shock vibration :
Sinusoidal vibration per MIL-T-28800E
(5 ~ 55 Hz, 3g maximum).
1-3 Electrical Specifications
Accuracy is ±(% reading + number
of digits) at 23°C ±5°C < 80%RH.
Temperature coefficient :
Add 0.2 x (Specified accuracy) / °C,
< 18°C, > 28°C .
DC / AC Volts
Range DC Accuracy AC Accuracy
4.000V
±(0.9% + 2dgt)
±(1.5% + 5dgt)
50Hz ~ 500Hz
40.00V
400.0V
DC1000V
AC750V
Over voltage protection :
DC 1000 V or AC 750 V
Input Impedance :
10MΩ// less than 100pF.
CMRR / NMRR :
(Common Mode Rejection Ratio)
(Normal Mode Rejection Ratio)
VAC : CMRR > 60dB at DC, 50Hz / 60Hz
VDC : CMRR > 100dB at DC,
50Hz / 60Hz
NMRR > 50dB at DC, 50Hz / 60Hz
AC Conversion Type :
Average sensing rms indication.
Resistance
Range Accuracy
400.0 Ω*2 ±(1.0% + 2 dgt)
4.000 KΩ
40.00 KΩ
400.0 KΩ
4.000MΩ ±(1.0% + 2 dgt)
40.00 MΩ*1 ±(1.5% + 2 dgt)
±(0.7% + 2 dgt)
Droop Protection :
4 feet droop to hardwood on concrete
floor.
Limited Warranty
This Meter is warranted to the original
purchaser against defects in material
and workmanship for 1 year from the
date of purchase. During this warranty
period, manufacturer will, at its option,
replace or repair the defective unit,
subject to verification of the defect or
malfunction.
This warranty does not cover fuses,
disposable batteries, or damage from
abuse, neglect, accident, unauthoriz-
ed repair, alteration, contamination,
or abnormal conditions of operation
or handling.
Any implied warranties arising out of
the sale of this product, including but
not limited to implied warranties of
merchantability and fitness for a
particular purpose, are limited to the
above. The manufacturer shall not be
liable for loss of use of the instrument
or other incidental or consequential
damages, expenses, or economic
loss, or for any claim or claims for
such damage, expense or economic
loss. Some states or countries laws
vary, so the above limitations or
exclusions may not apply to you.
APPA TECHNOLOGY CORP.
9F, 119-1 Pao-Zong Rd.,
Shin-Tien, Taipei, 23115, Taiwan.
P.O.Box. 12-24 Shin-Tien, Taiwan.
Tel: +886-2-29178820
Fax: +886-2-29170848
e-mail : info@appatech.com
http: //www.appatech.com

Other APPA Measuring Instrument manuals

APPA A3D User manual

APPA

APPA A3D User manual

APPA A1 User manual

APPA

APPA A1 User manual

APPA A3 User manual

APPA

APPA A3 User manual

APPA M2 User manual

APPA

APPA M2 User manual

APPA 76 User manual

APPA

APPA 76 User manual

APPA 606 User manual

APPA

APPA 606 User manual

APPA sFLEX-D User manual

APPA

APPA sFLEX-D User manual

APPA ATL-PV User manual

APPA

APPA ATL-PV User manual

APPA 31 User manual

APPA

APPA 31 User manual

APPA 93N User manual

APPA

APPA 93N User manual

APPA 703 User manual

APPA

APPA 703 User manual

APPA 103N User manual

APPA

APPA 103N User manual

APPA A17 Technical manual

APPA

APPA A17 Technical manual

Popular Measuring Instrument manuals by other brands

Emerson PACSystems RX3i user manual

Emerson

Emerson PACSystems RX3i user manual

ALCO BASELINE AL-63329 instruction manual

ALCO

ALCO BASELINE AL-63329 instruction manual

Bender ISOMETER isoPV1685DP manual

Bender

Bender ISOMETER isoPV1685DP manual

Hioki 3390 instruction manual

Hioki

Hioki 3390 instruction manual

Ponpe 8351 Operation manual

Ponpe

Ponpe 8351 Operation manual

XS Instruments OXY 7 Vio instruction manual

XS Instruments

XS Instruments OXY 7 Vio instruction manual

MATSYS Fluidized Filling System user guide

MATSYS

MATSYS Fluidized Filling System user guide

ETS HPDA-3 instructions

ETS

ETS HPDA-3 instructions

Besantek BST-HVD21 instruction manual

Besantek

Besantek BST-HVD21 instruction manual

Bartec TECH450PRO user guide

Bartec

Bartec TECH450PRO user guide

Dostmann Electronic 5020-0322 instruction manual

Dostmann Electronic

Dostmann Electronic 5020-0322 instruction manual

Sealey VS0563 instructions

Sealey

Sealey VS0563 instructions

Quantel Ultra quick start guide

Quantel

Quantel Ultra quick start guide

janitza UMG 512-Pro User manual and technical data

janitza

janitza UMG 512-Pro User manual and technical data

MSA ALTAIR 5X operating manual

MSA

MSA ALTAIR 5X operating manual

Black Box SW718A Specifications

Black Box

Black Box SW718A Specifications

AutomationDirect OP-406 manual

AutomationDirect

AutomationDirect OP-406 manual

Ecotech Serinus 40 user manual

Ecotech

Ecotech Serinus 40 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.