Arbiter Systems 1040C Installation guide

1
Training Manual
Arbiter Systems
Model 1040C
Panel Meter Calibrator
ARBITER SYSTEMS, INC.
PASO ROBLES, CALIFORNIA
U.S.A.


1
Overview

2
AUX.VOLTAGE
REMOTE
VOLTAGE
CURRENT
1040CPANELMETERCALIBRATOR
OVERLOAD
HIGH
VOLTAGE
POWER
GP-IB
OSRQORMT
OLSNORLK

3
SYNCHRO
SCOPE
OPERATING
SEQUENCE
FIGURE
2.1
D.C.
VOLTAGECURRENT
OPERATE
LOOP
RESET
60HZ400
HZ
POWER/
RECALLOPERATE
LOOP
VOLTAGECURRENTFREQUENCYPOWER/
OPERATE
LOOP
CURRENTFREQUENCY
OPERATE
LOOP
VOLTAGEFREQUENCY
OPERATE
LOOP
OPERATE
LOOP
ENTER
VALUE
AND
UNITS
MODIFYVALUEOPERATESTANDBYSTORENORMAL
%
DEVIATION
VOLTAGEFREQUENCY
OPERATE
LOOP
OPERATE
LOOP
VARS
VARS
POWER
FACTOR
PHASE
POWER
FACTOR
PHASE


5
1040C Lab Session
♦DC Voltage / Current
♦AC Voltage / Current
♦Power / VARS / PF
♦% Deviation
♦Store Recall
♦Calibration

6
DC Voltage Operating Procedure
CAUTION: Prior to making any connections, place the PMC in Standby Mode.
Connect the Model 918B to the PMC as shown below and perform the following steps:
1. Press DC.
2. Press VOLTAGE.
3. Enter the desired voltage using the keypad and the units key.
4. Press OPER.
5. MODIFY, MEMORY, or % DEV may be used.
6. To remove the output signal, press STBY when the measurement is complete.
+
—
—
+
1000 Vrms
MAX
1000 Vdc
MAX
45 - 500
H
CT OUTPUT ONLY
25 mArms MAX
25V
MAX
POWE
R
AT POW E R ON
USE "HOLD" KEY TO CHANGE CT
RATIO
USE "PHASE" KEY TO CHANGE PHASE
MODE
Vdc VA
VA
R
RANG
E
V
rm
A
rm W/Wh HOL
D
SHIF
T
PHAS
E
PF STAR
T
PHAS
E
FRE
Q
MIN/M
A
X
∆
CURRENT TRANSFORMER TERMINALS ARE AT
SAME
POTENTIAL AS (–) VOLTAGE
INPUT
USE PROPER
PRECAUTIONS
FUSED
AUX.
VOLTAGE
REMOT
E
V
OLTAG
E
CURRE
NT
1040C PANEL METER
CALIBRATOR
OVERLO
A
D
HIG
H
VOLTAG
E
POWE
R
GP-
IB
O SRQ ORMT
OLSN O RLK

7
DC Current Operating Procedure
CAUTION: Prior to making any connections, place the PMC in Standby Mode.
Connect the Model 918B to the PMC as shown below and perform the following steps:
1. Press DC.
2. Press CURRENT.
3. Enter the desired current using the keypad and the units key.
4. Press OPER.
5. MODIFY, MEMORY, or % DEV may be used.
6. To remove the output signal, press STBY when the measurement is complete.

8
AC Voltage Operating Procedure
CAUTION: Prior to making any connections, place the PMC in Standby Mode.
Connect the Model 918B to the PMC as shown below and perform the following steps:
1. Press 60 or 400 HZ.
2. Press VOLTAGE.
3. Enter the desired voltage using the keypad and the units key.
4. Press OPER.
5. MODIFY, MEMORY, or % DEV may be used. The FREQUENCY or VOLTAGE may
be modified.
6. To remove the output signal, press STBY when the measurement is complete.
+——
+
1000 Vrms
MAX
1000 Vdc
MAX
45 - 500
H
CT OUTPUT ONLY
25 mArms MAX
25V
MAX
POWE
R
AT POW E R ON
USE "HOLD" KEY TO CHANGE CT
RATIO
USE "PHASE" KEY TO CHANGE PHASE
MODE
Vdc VA
VA
R
RANG
E
V
rm Arm W/Wh HOL
D
SHIF
T
PHAS
E
PF STAR
T
PHAS
E
FRE
Q
MIN/MA
X
∆
CURRENT TRANSFORMER TERMINALS ARE AT
SAME
POTENTIAL AS (–) VOLTAGE
INPUT
USE PROPER
PRECAUTIONS
FUSED
AUX.
VOLTAGE
REMOT
E
V
OLTAG
E
CURRE
NT
1040C PANEL METER
CALIBRATOR
OVERLO
AD
HIG
H
VOLTAG
E
POWE
R
GP-
IB
O SRQ ORMT
OLSN O RLK

9
AC Current Operating Procedure
CAUTION: Prior to making any connections, place the PMC in Standby Mode.
Connect the Model 918B to the PMC as shown below and perform the following steps:
1. Press 60 or 400 HZ.
2. Press CURRENT.
3. Enter the desired current using the keypad and the units key.
4. Press OPER.
5. MODIFY, MEMORY, or % DEV may be used. The FREQUENCY or VOLTAGE may
be modified.
6. To remove the output signal, press STBY when the measurement is complete.
+
—
—
+
1000 Vrms
MAX
1000 Vdc
MAX
45 - 500
H
CT OUTPUT ONLY
25 mArms MAX
25V
MAX
POWE
R
AT POW E R ON
USE "HOLD" KEY TO CHANGE CT
RATIO
USE "PHASE" KEY TO CHANGE PHASE
MODE
Vdc VA
VA
R
RANG
E
V
rm
A
rm W/Wh HOL
D
SHIF
T
PHAS
E
PF STAR
T
PHAS
E
FRE
Q
MIN/M
A
X
∆
CURRENT TRANSFORMER TERMINALS ARE AT
SAME
POTENTIAL AS (–) VOLTAGE
INPUT
USE PROPER
PRECAUTIONS
FUSED
AUX.
VOLTAGE
REMOT
E
V
OLTAG
E
CURRE
NT
1040C PANEL METER
CALIBRATOR
OVERLO
A
D
HIG
H
VOLTAG
E
POWE
R
GP-
IB
O SRQ ORMT
OLSN O RLK
Arb
iter
Sy
ste
ms
10
00:
1
CT

10
AC Power / VARS Operating Procedure
CAUTION: Prior to making any connections, place the PMC in Standby Mode.
Connect the Model 918B to the PMC as shown below and perform the following steps:
1. Press 60 or 400 HZ.
2. Press POWER.
3. Press VOLTAGE.
4. Enter the desired voltage using the keypad and the units key.
5. Press CURRENT.
6. Enter the desired current using the keypad and the units key.
7. Press POWER. The power calculated from the entered voltage, current, and phase
values will be displayed. Press POWER again and VARS will be displayed. Note that
modifying Power or VARS changes the current output while leaving the voltage and
phase fixed.
8. Press OPER.
9. MODIFY, MEMORY, or % DEV may be used. POWER, VARS, CURRENT,
VOLTAGE or FREQUENCY may be modified. Note that modifying Power changes
the current output while leaving the voltage fixed.
10. To remove the output signal, press STBY when the measurement is complete.
+
—
—
+
1000 Vrms
MAX
1000 Vdc
MAX
45 - 500
H
CT OUTPUT ONLY
25 mAr ms MA X
25V
MAX
POWE
R
AT POW ER ON
USE "HOLD" KEY TO CHANGE CT
O
USE "PHASE" KEY TO CHANGE PHASE
MODE
Vdc VA
VA
R
RANG
E
V
rm
A
rm W/Wh HOL
D
SHIF
T
PHAS
E
PF STAR
T
PHAS
E
FRE
Q
MIN/MA
X
∆
CURRENT TRANSFORMER TERMINALS ARE AT
SAME
POTENTIAL AS (–) VOLTAGE
INPUT
USE PROPER
PRECAUTIONS
FUSED
AUX .
VOLTAGE
REMOT
E
V
OLTAG
E
CURRE
NT
1040C PANEL METER
CALIBRA TOR
OVERLO
A
D
HIG
H
VOLTAG
E
POWE
R
GP-
IB
O SRQ ORMT
OLSN O RLK
Arb
iter
Sy
ste
ms
10
00:
1
CT
+
—
—
+
1000 Vrms
MAX
1000 Vdc
MAX
45 - 500
H
CT OUTPUT ONLY
25 mAr ms M AX
25V
MAX
POWE
R
AT POW ER ON
USE "HOLD" KEY TO CHANGE CT
O
USE "PHASE" KEY TO CHANGE PHASE
MOD E
Vdc VA
VA
R
RANG
E
V
rm
A
rm W/ Wh HOL
D
SHIF
T
PHAS
E
PF STAR
T
PHAS
E
FRE
Q
MIN/M
A
X
∆
CURRENT TRANSFORMER TERMINALS ARE AT
SAME
POTENTIAL AS (–) VOLTAGE
INPUT
USE PR OPE R
PRECAUTIONS
FUSED

11
Power Factor / Phase Operating Procedure
CAUTION: Prior to making any connections, place the PMC in Standby Mode. The
following procedure employs initial output values of 120 volts and 5 amps at 60 Hz and
0°, 30° and 60° phase values. A full meter calibration would include other values.
The Phase Convention in the PMC is determined by the following equations:
Voltage: V(t) = Vo sinωt
Current: (t) = Vo sin (ωt + Θ)
Power: Vrms Irms cos Θ
VARS: Vrms Irms sin ( -Θ)
Power Factor: Cos Θ
Where: Θ= phase angle in degrees between voltage and current.
Phase Angle Sign of Power Sign of VARS Power Factor
0 < Θ≤90 + - Lead +
90 < Θ≤180 - - Lead -
-180 < Θ≤-90 - + Lag -
-90 < Θ≤0 + + Lag +
Turn-OFF the PMC and connect to the Model 931A System Analyzer as shown below and
perform steps of Table 1:
AUX.
VOLTAGE
REMOT
E
V
OLTAG
E
CURRE
NT
1040C PANEL METER
CALIBRATOR
OVERLO
A
D
HIG
H
VOLTAG
E
POWE
R
GP-
IB
O SRQ ORMT
OLSN O RLK
™

12
Table 1. Power Factor/Phase Operating Procedure
STEP OPERATOR ACTION RESULTS
1. Apply power to PMC After successful completion of Function Tests the PMC
will momentarily display PASSED FUNCTION TEST.
The PMC will return to RESET mode and display DC, 60
Hz, or 400 Hz? The RESET, STBY, and NORMAL
LED's (and the STORE LED if user settings are presently
stored in the user memory) will be illuminated.
2. Press 60 HZ RESET LED will extinguish, the 60 HZ and
FREQUENCY LED's will illuminate. The PMC now
displays FREQUENCY 60.00 Hz. (The Frequency may
now be modified or a new value entered).
3. Press POWER FACTOR FREQUENCY LED will extinguish, the POWER
FACTOR LED will illuminate and the PMC now
displays POWER FACTOR 1.000 (Power Factor may
now be modified or a new value entered).
4. Press VOLTAGE VOLTAGE LED is illuminated and the PMC now
displays VOLTAGE 15.00V (15 Volts rms is the lowest
range available for Power Factor/Phase output mode).
5. Press 1 The PMC displays ENTRY = 1
6. Press 2 The PMC displays ENTRY = 12
Press 0 The PMC displays ENTRY = 120. (The
MINUS/BACKSPACE key may be used to erase
incorrect entries).
7. Press V/A Pressing the units key V/A completes the entry of the
voltage setting. The PMC now displays VOLTAGE
120.0 V (The voltage may now be modified using the
rotary Knob or the Up/Down modify keys.).
9. Press CURRENT The VOLTAGE LED will extinguish, the CURRENT
LED will illuminate, the PMC displays CURRENT
.1000A (.1 amps rms is the lowest current available for
Power Factor/Phase output mode).
10. Press 5 PMC displays ENTRY = 5

13
Table 1. Power Factor/Phase Operating Procedure - cont’d
STEP OPERATOR ACTION RESULTS
11. Press V/A Pressing the units key V/A completes the entry of the
current setting. The PMC displays CURRENT 5.000A
(The current may now be modified).
12. Press OPER The OPER LED will illuminate, the HIGH VOLTAGE
indicator will flash On and Off, and a <, =, or > will
a
pp
ear on the dis
p
la
y
indicatin
g
the status of the out
p
ut.
Within the s
p
ecified settlin
g
time the dis
p
la
y
shoul
d
show = and the meter should read a
p
ower factor of
1.000.
13. Press PHASE The POWER FACTOR and CURRENT LED's will
extin
g
uish, the PHASE LED will illuminate, and the
PMC displays PHASE 00.00ò.
14. Adjust Phase to 30ò The PMC will dis
p
la
y
PHASE 30.00ò
(
ad
j
ustments of
the output value is accomplished by either entering a new
value from the ke
yp
ad, Sec. 2.4.2, or usin
g
the modif
y
controls, Sec. 2.4.5
)
. The meter shows a la
gg
in
g
p
owe
r
factor of .8661.
15. Press POWER FACTOR The PHASE LED will extinguish, the POWER FACTOR
LED will illuminate, and the PMC dis
p
la
y
s POWE
R
FACTOR .8661.
16. Ad
j
ust Power Factor to
.4999
The PMC dis
p
la
y
POWER FACTOR .4999
(
Due to the
minimum
p
hase increments some Power Factor values
such as .5000 are not allowed
)
. The meter should show
.5 lag.
17. Press PHASE The POWER FACTOR LED will extin
g
uish, PHASE
LED will illuminate, and the PMC dis
p
la
y
s PHASE
60.00ò.
18. Press POWER
FACTOR, PHASE,
POWER/VARS,
FREQUENCY,
VOLTAGE or
CURRENT as desired.
New values for any of these functions may be entered
using the keypad or the old settings may be modified.
The PMC will remain in Operate mode unless the new
Voltage or Current settings entered are outside of the
present operating range.
19. Press STBY The STBY LED will illuminate, and OPER LED will
extinguish. No output signals will be present at the
output terminals.

14
Normal/Deviation
In the % deviation mode the PMC computes and displays the output value according to the
formula:
PMC Display Value = (METER C.P. - PMC output value) X 100 %
METER F.S.
The PMC must have both the meter full scale value (Meter F.S.) and the meter calibration
point (Meter C.P.) to perform the necessary calculation.
To enable this function, press %DEV and set the PMC output value to the meter's full scale
value (it is not necessary to press OPER) . The display will read METER F.S. = XXXXX
YY where XXXXX YY represents the meter's full scale value and units.
Press %DEV again and the display will change to METER C.P. = XXXXX YY. Enter
the meter calibration point by using the keypad or RECALLing a stored setting. Do not
use the Modify Controls.
Press the %DEV key again to display METER = 00.00%. The PMC now displays the
deviation from the selected calibration point in percent of meter full scale. The previous
displays can be reviewed and/or changed by pressing the %DEV key.
Repeatedly press the % DEV key until the METER = XX.XX% is displayed again. Press
OPER to intitiate the output. Use the rotary knob or the modify keys to change the output
level and the % deviation displayed. Do not enter a value from the keypad or use the
RECALL feature. A new full scale or calibration point must be entered using the keypad
or RECALL feature. Changing the full scale or calibration point resets the percent
deviation to zero.
Memory
The Memory feature is useful for storing frequently used output settings and quickly
recalling them later. An output setting can be entered into the memory by simply pressing
the STORE key. Illumination of the STORE LED indicates that stored values are present.
Up to 99 settings can be stored. The display will flash memory store # after each entry and
will continue to flash on and off if all of the memory locations have been filled.
Output settings in memory are recalled by pressing the RECALL key. The memory
operates on a first in, first out (FIFO) basis. This sequence will continue whereby the first
setting stored will also follow the last when recalling through all of the stored settings.
Pressing the CLEAR key clears all memory settings.
The keypad may be used to help move through the stored settings. Pressing 0 then
RECALL recalls the first stored setting. Entering a number other than zero and pressing
RECALL recalls the output setting at that location.
Calibration
External Adjustments
This section consists of test procedures that check the PMC’s output amplitude and phase.
If a discrepancy with the specified tolerances is encountered, modifications are made to the
stored calibration constants.

15
Calibration Mode
The stored calibration constants can only be modified while in the calibrate mode. The
calibrate mode is identified by a circle with a cross character appearing in the right-hand
position of the display, usually after the units designator. While in the calibrate mode any
or all of the stored constants may be modified. There are two methods of exiting the
calibrate mode. If STORED is pressed, all of the modified calibration constants will be
stored in non-volatile memory for use during regular operation. If RESET is pressed, the
modified calibration constants will be discarded and the PMC will revert to normal
operation using the previously stored constants.
PMC operation in the Calibrate Mode differs from normal operation. Only five output
amplitude levels are available for each range. These are 1/10, 1/4, 1/2, 3/4 and top of
range. Once in the desired range, the CLEAR key cycles the output through these five
values. Also, only 60 and 400 HZ frequencies and 30° phases are available.
Calibration Constants
There is one calibration constant for the bottom and one for the top of each range. When
the output is at 1/10 of range, the bottom constant may be modified. The top of range
constant may be modified only while the output is set to top of range. No adjustments can
be made at the 1/4, 1/2, or 3/4 settings.
Make modifications by entering a number on the keypad and pressing ENTER. During
amplitude alignment the output will change approximately .01% times the number entered.
For phase the factor is 0.1°. A long settling time should be allowed (10 seconds or more)
after each adjustment is made.
NOTE: the two calibration constants for each range affect each other. A +0.1%
change at the top of range will produce a 0.1% change at the bottom of range . A
+1.0% change at the bottom of range will produce a +0.1% change at the top of
range.
After aligning the top and bottom of ranges, recheck both outputs to find any interaction in
the alignments.

16
Entering Calibration Mode
To access the calibrate mode, the operator must know the Key Code for the particular
instrument. The Key Code is set to the instrument’s serial number when shipped from the
factory. Access the calibrate mode in the following manner:
1. With the PMC in the reset mode, press and hold the MODIFY UP arrow key. Press
RESET. The display will show: ENTER KEY #0000.
2. Use the numeric keypad to enter the Key Code and press ENTER.
NOTE: If an incorrect Key Code is entered the PMC will revert to Reset.
3. Once the correct Key Code is entered, the display will change to CHANGE KEY #
XXXX. The key Code may now be modified.
CAUTION: Ensure to properly record the Key Code if modified and restored.
4. Press ENTER for calibrate mode operation signified by a circle with a cross character
in the right-hand portion of the display.
Typical Calibrate Routine
1. Enter Calibrate Routine as previously described.
2. Set the PMC for operation in desired mode and range using the top of the range value.
As an example, use the following sequential steps to test the top of the 100-volt DC
range:
a. Press DC.
b. Press VOLTAGE.
c. On the numeric keypad, enter 100.
d. Press V.
e. Press OPER.
f. Allow the output to settle and perform measurement.
3. The top of range calibration constant and the output level can now be modified by
entering a number on the numeric keypad and pressing ENTER. During amplitude
alignment the calibration constant and the output level will change approximately
0.01% times the number entered. For phase corrections the factor is 0.01°. Adjustments
to calibration constants may be made at the top of range and 1/10 settings only.
4. The 1/10, 1/4, 1/2, 3/4, and top of range outputs are programmed and available by
pressing the Clear Key. The PMC will not revert to Standby when CLEAR is pressed,
but the output will change to the newly displayed value.
5. To adjust Voltage or Current during Power, setup Power using the normal sequence,
e.g., 60 Hz, POWER, VOLTAGE, 150V, CURRENT, and 1A. Then select either
VOLTAGE or CURRENT to monitor. The Clear Key will cycle the setting selected.
6. While most functions operate in their normal fashion in calibrate mode, several
characteristics are worth noting: Frequency and phase are not adjustable; only 60 / 400
Hz are available and the phase is set to 30°.

17
Meter Interconnect
Diagram

18
Red Black Red Black
Voltage Current
+-
Ammeter
Power Factor Meter
12
34
12
34
12
34
VAR Meter
Wattmeter
Interconnect Diagram Demonstration Meters
Other manuals for 1040C
1
Table of contents
Other Arbiter Systems Measuring Instrument manuals

Arbiter Systems
Arbiter Systems 940A User manual

Arbiter Systems
Arbiter Systems 921A Guide

Arbiter Systems
Arbiter Systems 940A User manual

Arbiter Systems
Arbiter Systems Portable Power Sentinel 933A User manual

Arbiter Systems
Arbiter Systems 1092A User manual

Arbiter Systems
Arbiter Systems 1133A User manual

Arbiter Systems
Arbiter Systems 931A User manual

Arbiter Systems
Arbiter Systems 931A Installation and operating instructions