
14.148.- / A414D en / Software 1.30 14.10.2003 7
1.5 List of the interrogation- and configuration functions:
CAUTION! Before you make a new configuration or replace an EPROM or the mainboard make a note of
the programmed values. Afterwards you can re-enter the old not writeprotected values.
Address Write Default Function
(#) prot. Value
00= No PIN code active
11= Yes Run indication by running decimal point
20= No Key "ON/OFF" only at STOP valid
30= No Rate change only at STOP valid
40= No Key "STOP" delayed (time at #361)
50= No 2nd entry of rate (#3=1), rate calculation disabled.
60= No Static alarm (staff alerting system)
70= No Display elapsed time in run mode (#8=0) ****
80= No Select remaining time (#7=1)
91= Yes Alternative time input (10h, 1h, 1/4h)
10 0 = No Operation without drop detector
11 1 = Yes Recall "ml/h" (rate) at next power on
12 0 = No Recall "ml total" (end volume) at next power on
13 0 = No Recall "ml inf." (volume inf.) at next power on
14 0 = No SBS Step By Step function
15 0 = No Display VTBI (Volume To Be Infused)
16 0 = No Display "SEt -X-"’ if only 1 set enabled
17 1 = Yes KVO (KOR), mode see #60
18 0 = No Drop alarm only if bottle is empty (#10=0)
19 1 = Yes Buzzer at start
20 0 = No Menu "CLr" (clear "ml inf.") (#15=0) ****
21 0 = No Menu "uOP" (Micro operation) ****
22 0 = No Menu "trA" (transport) (#10=0 and #18=0) ****
23 1 = Yes Menu "PrL" (pressure alarm limit) ****
24 1 = Yes Menu "CAP" (battery capacity) ****
25 0 = No Menu "SEt Fill" ****
26 0 = No Menu "InF" ("ml inf." since last power on) ****
27 0 = No Menu "dLo" (data-lock) ****
28 0 = No Menu "Stb" (stand-by) ****
29 0 = No Menu "MEd" (medication number) ****
30 0 = No Menu "tM " (timer alarm) ****
31 0 = No
32 0 = No Menu "boL" (release bolus) ****
33 0 = No Menu "boLr" (bolus rate) (#32=1) ****
34 0 = No Menu "tot" (bolus total) (#32=1) ****
35 0 = No
36 0 = No
37 0 = No
38 0 = No Bolus application automatic (#32=1 and #34=1)
39 0 = No
40 0 = No Demo mode (all menus enabled) ****
41 0 = No Clear "ml/h" after infusion completed
42 0 = No Clear "ml total" after infusion completed (#41=1)
43 0 = No Air volume accumulated (1ml over 0.5 hrs)
44 1 = Yes Automatic pressure release after occlusion
45 1 = Yes Pressure display 20/40/60/80/100% (Bargraph ON)
46 0 = No Bargraph with indicator (25% steps, #45=1)
47 0 = No Standby- and battery prealarm low volume
48 1 = Yes Flashing numeric display at alarm
49 0 = No Alarm acknowledge only with key "MODE"