Arkray AUTION ELEVEN AE-4022 User manual

SEMI-AUTOMATED URINE ANALYZER
AUTION ELEVEN
™
AE-4022
Quick Reference Guide

Table of Contents
System Overview........................ 1
Sample Preparation .................... 2
Sample Measurement.............. 3-5
Control Measurement............... 6
Check Measurement................. 7
Maintenance ......................... 8-12
Cleaning the Feeder .............8-10
Cleaning the Waste Box ..........11
Replacing Thermal Paper ....... 12
Warnings .................................. 13
Errors ....................................14 -15
This Quick Reference Guide is a brief
summary of the operation and routine
maintenance of the AUTION ELEVEN
Semi-Automated Urine Analyzer.
Read the Operating Manual before
using the instrument.
Wear protective gloves to prevent
exposure to pathogenic microbes.
SEMI-AUTOMATED URINE ANALYZER
AUTION ELEVEN
™
AE-4022

System Overview
System Overview

SEMI-AUTOMATED URINE ANALYZER
AUTION ELEVEN
™
AE-4022 System Overview
1
Measurement
Name of each part
Display
Operator panel
Test strip tray
Carrying arm Cover open button
(on both sides)
Waste box
Maintenance cover
Built-in printer
Rails (white)
Test strip detection
window
Preparation of measurement
Setup (test strip, measurement No., patient ID)
Measurement
Check that there is no used test strips in the waste box.
with buttonsSelect
[ Select test strip type ] [ Menu screen ]
Select the type of test strip
Dip the entire test strip pad area in the sample for two seconds.
IMPORTANT
Do NOT wet the black mark.
After dipping the test strip in the sample,
remove excess urine by letting the strip
touch the edge of the collection cup.
Set the measurement number
Place a test strip over the detection area of the test
strip detection window, crossing the two white rails
of the test strip tray.
The carrying arm sends the test strip into the
analyzer, and the measurement starts automatically.
Set the patient ID
When measuring continuously, repeat steps 1 to 3 .
Set measurement numbers and patient IDs as necessary.
When the buzzer setting is ON, the beeper sounds
for two seconds to signal the timing for dipping the
next test strip.
Dip the test strip while you hear the beep.
Turn off the instrument after completing all the measurement procedures.
Turn on the power switch at the back of the instrument.
Prepare a sufficient volume of sample in urine collection cup so
that the entire test strip pad area can be dipped in a single
motion. Stir well.
Waste box
Test strip
Test strip
pad area
Do NOT wet the
black mark.
Test strip
Rails (white)
Test strip detection window
Detection
area
(up to 13 digits)
,
(up to 4 digits)
System Overview

Sample Preparation
Sample Preparation

2
SEMI-AUTOMATED URINE ANALYZER
AUTION ELEVEN
™
AE-4022 Sample Preparation
1. Turn on power switch at back of instrument.
2. After a 20 second warm up the standby screen will appear.
3. Prepare suffi cient volume of sample in urine collection cup so entire test strip pad
area can be dipped in a single movement.
4. Mix well.
Daily Maintenance
Power switch
Turn off the instrument.
Waste box
Discard used test strips.
140820YH
Wear protective gloves to prevent
exposure to pathogenic microbes.
Discard used test strips and
protective gloves separately from
general waste according to local
regulations on biohazardous waste.
Detach the carrying arm.
Open the maintenance cover.
Feed lever
Lever
Detach the feed lever by sliding the white lever
in the center towards you.
Wash the detached carrying arm, waste box,
test strip tray, suction port, test strip feed tray
and feed lever with water to remove any dirt.
Carrying arm
Suction port
Waste box
Test strip feed tray
Feed lever
Test strip tray
Wipe the parts with cloth and fully dry them before
attaching them back on.
This Quick Reference Guide is a brief
summary of the measurement operation
and daily maintenance. Before using the
instrument, make sure to read the Operat-
ing Manual.
Clean the area circled in red below.
Test strip feed tray
Detach the test strip feed tray.
Suction port
Test strip tray
Detach the suction port.
Test strip tray
Detach the test strip tray.
Measurement
Name of each part
Display
Operator panel
Test strip tray
Carrying arm Cover open button
(on both sides)
Waste box
Maintenance cover
Built-in printer
Rails (white)
Test strip detection
window
Preparation of measurement
Setup (test strip, measurement No., patient ID)
Measurement
Check that there is no used test strips in the waste box.
with buttonsSelect
[ Select test strip type ] [ Menu screen ]
Select the type of test strip
Dip the entire test strip pad area in the sample for two seconds.
IMPORTANT
Do NOT wet the black mark.
After dipping the test strip in the sample,
remove excess urine by letting the strip
touch the edge of the collection cup.
Set the measurement number
Place a test strip over the detection area of the test
strip detection window, crossing the two white rails
of the test strip tray.
The carrying arm sends the test strip into the
analyzer, and the measurement starts automatically.
Set the patient ID
When measuring continuously, repeat steps 1 to 3 .
Set measurement numbers and patient IDs as necessary.
When the buzzer setting is ON, the beeper sounds
for two seconds to signal the timing for dipping the
next test strip.
Dip the test strip while you hear the beep.
Turn off the instrument after completing all the measurement procedures.
Turn on the power switch at the back of the instrument.
Prepare a sufficient volume of sample in urine collection cup so
that the entire test strip pad area can be dipped in a single
motion. Stir well.
Waste box
Test strip
Test strip
pad area
Do NOT wet the
black mark.
Test strip
Rails (white)
Test strip detection window
Detection
area
(up to 13 digits)
,
(up to 4 digits)
2.3 Preparation for measurement
AE-4022 2-15
bPress .
• The recorded patient ID number is stored and the display returns
to the [Standby screen].
• Once the patient ID number is set, the “measurement number +
test strip type” and “patient ID number” are displayed alternately
on the [Standby screen].
To enter a patient ID number using a barcode
1 Call up the screen
aWith the [Standby screen] displayed, press .
• The [Patient ID number input screen] appears.
2 Enter patient ID number by reading its barcode
aUse a hand-held barcode reader to scan the barcode on the label,
which will enter the ID number when the [Patient ID number
input screen] is displayed.
Displayed alternately
IMPORTANT:
Do not centrifuge samples.
Centrifugal force may cause undesirable sedimentation of blood cells and/or other
sample contents, thus causing several measurement items to have incorrect results.

Sample Measurement
Sample Measurement

3
SEMI-AUTOMATED URINE ANALYZER
AUTION ELEVEN
™
AE-4022 Sample Measurement
1. Enter operator ID number (if function is ON).
• From the Standby Screen, press and hold
Measurement
Name of each part
Display
Operator panel
Test strip tray
Carrying arm Cover open button
(on both sides)
Waste box
Maintenance cover
Built-in printer
Rails (white)
Test strip detection
window
Preparation of measurement
Setup (test strip, measurement No., patient ID)
Measurement
Check that there is no used test strips in the waste box.
with buttonsSelect
[ Select test strip type ] [ Menu screen ]
Select the type of test strip
Dip the entire test strip pad area in the sample for two seconds.
IMPORTANT
Do NOT wet the black mark.
After dipping the test strip in the sample,
remove excess urine by letting the strip
touch the edge of the collection cup.
Set the measurement number
Place a test strip over the detection area of the test
strip detection window, crossing the two white rails
of the test strip tray.
The carrying arm sends the test strip into the
analyzer, and the measurement starts automatically.
Set the patient ID
When measuring continuously, repeat steps 1 to 3 .
Set measurement numbers and patient IDs as necessary.
When the buzzer setting is ON, the beeper sounds
for two seconds to signal the timing for dipping the
next test strip.
Dip the test strip while you hear the beep.
Turn off the instrument after completing all the measurement procedures.
Turn on the power switch at the back of the instrument.
Prepare a sufficient volume of sample in urine collection cup so
that the entire test strip pad area can be dipped in a single
motion. Stir well.
Waste box
Test strip
Test strip
pad area
Do NOT wet the
black mark.
Test strip
Rails (white)
Test strip detection window
Detection
area
(up to 13 digits)
,
(up to 4 digits)
for 3 seconds.
• Enter operator ID number using numeric and [-] keys or using barcode reader.
• Press
Measurement
Name of each part
Display
Operator panel
Test strip tray
Carrying arm Cover open button
(on both sides)
Waste box
Maintenance cover
Built-in printer
Rails (white)
Test strip detection
window
Preparation of measurement
Setup (test strip, measurement No., patient ID)
Measurement
Check that there is no used test strips in the waste box.
with buttonsSelect
[ Select test strip type ] [ Menu screen ]
Select the type of test strip
Dip the entire test strip pad area in the sample for two seconds.
IMPORTANT
Do NOT wet the black mark.
After dipping the test strip in the sample,
remove excess urine by letting the strip
touch the edge of the collection cup.
Set the measurement number
Place a test strip over the detection area of the test
strip detection window, crossing the two white rails
of the test strip tray.
The carrying arm sends the test strip into the
analyzer, and the measurement starts automatically.
Set the patient ID
When measuring continuously, repeat steps 1 to 3 .
Set measurement numbers and patient IDs as necessary.
When the buzzer setting is ON, the beeper sounds
for two seconds to signal the timing for dipping the
next test strip.
Dip the test strip while you hear the beep.
Turn off the instrument after completing all the measurement procedures.
Turn on the power switch at the back of the instrument.
Prepare a sufficient volume of sample in urine collection cup so
that the entire test strip pad area can be dipped in a single
motion. Stir well.
Waste box
Test strip
Test strip
pad area
Do NOT wet the
black mark.
Test strip
Rails (white)
Test strip detection window
Detection
area
(up to 13 digits)
,
(up to 4 digits)
to confi rm entry.
2. Confi rm display shows 10EA as test strip type.
3. Set measurement number (optional).
• If using 0001 for fi rst measurement number, go to step 4.
• Enter new number using keypad
Measurement
Name of each part
Display
Operator panel
Test strip tray
Carrying arm Cover open button
(on both sides)
Waste box
Maintenance cover
Built-in printer
Rails (white)
Test strip detection
window
Preparation of measurement
Setup (test strip, measurement No., patient ID)
Measurement
Check that there is no used test strips in the waste box.
with buttonsSelect
[ Select test strip type ] [ Menu screen ]
Select the type of test strip
Dip the entire test strip pad area in the sample for two seconds.
IMPORTANT
Do NOT wet the black mark.
After dipping the test strip in the sample,
remove excess urine by letting the strip
touch the edge of the collection cup.
Set the measurement number
Place a test strip over the detection area of the test
strip detection window, crossing the two white rails
of the test strip tray.
The carrying arm sends the test strip into the
analyzer, and the measurement starts automatically.
Set the patient ID
When measuring continuously, repeat steps 1 to 3 .
Set measurement numbers and patient IDs as necessary.
When the buzzer setting is ON, the beeper sounds
for two seconds to signal the timing for dipping the
next test strip.
Dip the test strip while you hear the beep.
Turn off the instrument after completing all the measurement procedures.
Turn on the power switch at the back of the instrument.
Prepare a sufficient volume of sample in urine collection cup so
that the entire test strip pad area can be dipped in a single
motion. Stir well.
Waste box
Test strip
Test strip
pad area
Do NOT wet the
black mark.
Test strip
Rails (white)
Test strip detection window
Detection
area
(up to 13 digits)
,
(up to 4 digits)
(up to 4 digits).
• Press
Measurement
Name of each part
Display
Operator panel
Test strip tray
Carrying arm Cover open button
(on both sides)
Waste box
Maintenance cover
Built-in printer
Rails (white)
Test strip detection
window
Preparation of measurement
Setup (test strip, measurement No., patient ID)
Measurement
Check that there is no used test strips in the waste box.
with buttonsSelect
[ Select test strip type ] [ Menu screen ]
Select the type of test strip
Dip the entire test strip pad area in the sample for two seconds.
IMPORTANT
Do NOT wet the black mark.
After dipping the test strip in the sample,
remove excess urine by letting the strip
touch the edge of the collection cup.
Set the measurement number
Place a test strip over the detection area of the test
strip detection window, crossing the two white rails
of the test strip tray.
The carrying arm sends the test strip into the
analyzer, and the measurement starts automatically.
Set the patient ID
When measuring continuously, repeat steps 1 to 3 .
Set measurement numbers and patient IDs as necessary.
When the buzzer setting is ON, the beeper sounds
for two seconds to signal the timing for dipping the
next test strip.
Dip the test strip while you hear the beep.
Turn off the instrument after completing all the measurement procedures.
Turn on the power switch at the back of the instrument.
Prepare a sufficient volume of sample in urine collection cup so
that the entire test strip pad area can be dipped in a single
motion. Stir well.
Waste box
Test strip
Test strip
pad area
Do NOT wet the
black mark.
Test strip
Rails (white)
Test strip detection window
Detection
area
(up to 13 digits)
,
(up to 4 digits)
to confi rm entry.
4. Set patient ID.
• Press
Measurement
Name of each part
Display
Operator panel
Test strip tray
Carrying arm Cover open button
(on both sides)
Waste box
Maintenance cover
Built-in printer
Rails (white)
Test strip detection
window
Preparation of measurement
Setup (test strip, measurement No., patient ID)
Measurement
Check that there is no used test strips in the waste box.
with buttonsSelect
[ Select test strip type ] [ Menu screen ]
Select the type of test strip
Dip the entire test strip pad area in the sample for two seconds.
IMPORTANT
Do NOT wet the black mark.
After dipping the test strip in the sample,
remove excess urine by letting the strip
touch the edge of the collection cup.
Set the measurement number
Place a test strip over the detection area of the test
strip detection window, crossing the two white rails
of the test strip tray.
The carrying arm sends the test strip into the
analyzer, and the measurement starts automatically.
Set the patient ID
When measuring continuously, repeat steps 1 to 3 .
Set measurement numbers and patient IDs as necessary.
When the buzzer setting is ON, the beeper sounds
for two seconds to signal the timing for dipping the
next test strip.
Dip the test strip while you hear the beep.
Turn off the instrument after completing all the measurement procedures.
Turn on the power switch at the back of the instrument.
Prepare a sufficient volume of sample in urine collection cup so
that the entire test strip pad area can be dipped in a single
motion. Stir well.
Waste box
Test strip
Test strip
pad area
Do NOT wet the
black mark.
Test strip
Rails (white)
Test strip detection window
Detection
area
(up to 13 digits)
,
(up to 4 digits)
• Enter up to 13 digits using numeric and [-] keys or scan barcoded sample
with barcode reader.
• Press
Measurement
Name of each part
Display
Operator panel
Test strip tray
Carrying arm Cover open button
(on both sides)
Waste box
Maintenance cover
Built-in printer
Rails (white)
Test strip detection
window
Preparation of measurement
Setup (test strip, measurement No., patient ID)
Measurement
Check that there is no used test strips in the waste box.
with buttonsSelect
[ Select test strip type ] [ Menu screen ]
Select the type of test strip
Dip the entire test strip pad area in the sample for two seconds.
IMPORTANT
Do NOT wet the black mark.
After dipping the test strip in the sample,
remove excess urine by letting the strip
touch the edge of the collection cup.
Set the measurement number
Place a test strip over the detection area of the test
strip detection window, crossing the two white rails
of the test strip tray.
The carrying arm sends the test strip into the
analyzer, and the measurement starts automatically.
Set the patient ID
When measuring continuously, repeat steps 1 to 3 .
Set measurement numbers and patient IDs as necessary.
When the buzzer setting is ON, the beeper sounds
for two seconds to signal the timing for dipping the
next test strip.
Dip the test strip while you hear the beep.
Turn off the instrument after completing all the measurement procedures.
Turn on the power switch at the back of the instrument.
Prepare a sufficient volume of sample in urine collection cup so
that the entire test strip pad area can be dipped in a single
motion. Stir well.
Waste box
Test strip
Test strip
pad area
Do NOT wet the
black mark.
Test strip
Rails (white)
Test strip detection window
Detection
area
(up to 13 digits)
,
(up to 4 digits)
to confi rm entry.
• The measurement number and patient ID number will display alternately
on screen.
Chapter 2 Measurement Operation
AE-4022 2-12
2.3.4 Entering your operator ID number
When the operator ID function is ON, you need to enter your operator ID number before performing a test.
1 Call up the screen
aWith the [Standby screen] displayed, press and hold for 3
seconds.
• The [Operator ID number input screen] appears.
2 Enter your operator ID number
aEnter your operator ID number correctly using the numeric and
keys.
bPress .
• If your entry is valid, the display returns to the [Standby screen].
You are then permitted to perform measurement.
REFERENCE:
You can also enter the operator ID number by reading its
barcode.
REFERENCE:
Elapse of Operator ID Effective Time
When the Operator ID effective time elapses, the keys will not respond and/or testing will not start when a strip is
placed until you enter your operator ID.
Chapter 2 Measurement Operation
AE-4022 2-14
2.3.6 Entering patient ID numbers
Each patient ID number can have up to 13 digits of numeric figures and “–” characters entered. The patient ID number
entered is uniquely allocated to the measurement made just after this entry is made. After a group of measurements is
completed, the patient ID numbers are automatically deleted.
To manually input a patient ID number
1 Call up the screen
aWith the [Standby screen] displayed, press .
• The [Patient ID number input screen] appears.
2 Enter a patient ID number
aEnter a patient ID number.
• You can enter up to 13 digits using the numeric and keys.
REFERENCE:
In addition to the patient ID number, a measurement number is also allocated to each sample. The
measurement number is automatically incremented by one for each successive measurement from the start of
measurement until the instrument is turned OFF.
When using a hand-held barcode reader (optional), you need not manually enter the patient ID number. Follow
the procedures in “To enter a patient ID number using a barcode” on page 2-15.
NOTE:
If the input value exceeds 13 digits, the first digit (the
leftmost digit) will be deleted.
REFERENCE:
If a patient ID number has already been set, the highest-order digit (leftmost digit) flashes. If you then input a
number or “–”, the next digit to the right flashes. The flashing digit accepts key input.
If you press during input, the currently input patient ID number is cancelled and the instrument returns to
the [Standby screen].
2.3 Preparation for measurement
AE-4022 2-15
bPress .
• The recorded patient ID number is stored and the display returns
to the [Standby screen].
• Once the patient ID number is set, the “measurement number +
test strip type” and “patient ID number” are displayed alternately
on the [Standby screen].
To enter a patient ID number using a barcode
1 Call up the screen
aWith the [Standby screen] displayed, press .
• The [Patient ID number input screen] appears.
2 Enter patient ID number by reading its barcode
aUse a hand-held barcode reader to scan the barcode on the label,
which will enter the ID number when the [Patient ID number
input screen] is displayed.
Displayed alternately
2.3 Preparation for measurement
AE-4022 2-15
bPress .
• The recorded patient ID number is stored and the display returns
to the [Standby screen].
• Once the patient ID number is set, the “measurement number +
test strip type” and “patient ID number” are displayed alternately
on the [Standby screen].
To enter a patient ID number using a barcode
1 Call up the screen
aWith the [Standby screen] displayed, press .
• The [Patient ID number input screen] appears.
2 Enter patient ID number by reading its barcode
aUse a hand-held barcode reader to scan the barcode on the label,
which will enter the ID number when the [Patient ID number
input screen] is displayed.
Displayed alternately
Continued on next page…

4
SEMI-AUTOMATED URINE ANALYZER
AUTION ELEVEN
™
AE-4022 Sample Measurement
5. Confi rm normal measurement icon
AE-4020
2-16
Chapter 2 Measurement Operation
Take out the required number of test strips from
the bottle.
Close the test strip bottle cap securely.
IMPORTANT: After taking the test strips out
of the bottle, immediately close
the cap or the test strips in the
bottle will absorb moisture or dirt
in the air, which may deteriorate
reagents and make the test
strips useless.
5
Start normal measurement
Conrm that the icon ( ) for normal measurement
appears on the [Standby screen].
IMPORTANT: Different procedures at this
stage are required, depending
on the setting of the “opera-
tional mode when turning ON”
parameter. (See “Operational
mode” on page 2-3.)
The standard setting is “Auto start” mode. To change the “operational
mode when turning ON” parameter, see “3-7-8. No.005: Operational
mode when turning ON” on page 3-18.
REFERENCE: When the buzzer setting is ON, the buzzer sounds in the following
intervals to signal the dipping timing for the test strips.
Slow beeping
Fast beeping
Preparation for dipping:
3 seconds
Dipping period:
2 seconds
Test strip setting:
1.4 seconds
●When using the “Auto start” mode, go to step 6.
●When using the “Cycle start” mode, press and go to step 6.
6
Call up the [Clarity input screen]
Press and hold for about 2 seconds.
The [Clarity input screen] appears.
appears on standby screen.
6. Enter Clarity Value (If feature is ON).
• Press and hold
Measurement
Name of each part
Display
Operator panel
Test strip tray
Carrying arm Cover open button
(on both sides)
Waste box
Maintenance cover
Built-in printer
Rails (white)
Test strip detection
window
Preparation of measurement
Setup (test strip, measurement No., patient ID)
Measurement
Check that there is no used test strips in the waste box.
with buttonsSelect
[ Select test strip type ] [ Menu screen ]
Select the type of test strip
Dip the entire test strip pad area in the sample for two seconds.
IMPORTANT
Do NOT wet the black mark.
After dipping the test strip in the sample,
remove excess urine by letting the strip
touch the edge of the collection cup.
Set the measurement number
Place a test strip over the detection area of the test
strip detection window, crossing the two white rails
of the test strip tray.
The carrying arm sends the test strip into the
analyzer, and the measurement starts automatically.
Set the patient ID
When measuring continuously, repeat steps 1 to 3 .
Set measurement numbers and patient IDs as necessary.
When the buzzer setting is ON, the beeper sounds
for two seconds to signal the timing for dipping the
next test strip.
Dip the test strip while you hear the beep.
Turn off the instrument after completing all the measurement procedures.
Turn on the power switch at the back of the instrument.
Prepare a sufficient volume of sample in urine collection cup so
that the entire test strip pad area can be dipped in a single
motion. Stir well.
Waste box
Test strip
Test strip
pad area
Do NOT wet the
black mark.
Test strip
Rails (white)
Test strip detection window
Detection
area
(up to 13 digits)
,
(up to 4 digits)
for about 2 seconds.
• Enter the clarity value of sample using numeric keys
• Enter a number between 1 and 5 using table below:
• Press
Measurement
Name of each part
Display
Operator panel
Test strip tray
Carrying arm Cover open button
(on both sides)
Waste box
Maintenance cover
Built-in printer
Rails (white)
Test strip detection
window
Preparation of measurement
Setup (test strip, measurement No., patient ID)
Measurement
Check that there is no used test strips in the waste box.
with buttonsSelect
[ Select test strip type ] [ Menu screen ]
Select the type of test strip
Dip the entire test strip pad area in the sample for two seconds.
IMPORTANT
Do NOT wet the black mark.
After dipping the test strip in the sample,
remove excess urine by letting the strip
touch the edge of the collection cup.
Set the measurement number
Place a test strip over the detection area of the test
strip detection window, crossing the two white rails
of the test strip tray.
The carrying arm sends the test strip into the
analyzer, and the measurement starts automatically.
Set the patient ID
When measuring continuously, repeat steps 1 to 3 .
Set measurement numbers and patient IDs as necessary.
When the buzzer setting is ON, the beeper sounds
for two seconds to signal the timing for dipping the
next test strip.
Dip the test strip while you hear the beep.
Turn off the instrument after completing all the measurement procedures.
Turn on the power switch at the back of the instrument.
Prepare a sufficient volume of sample in urine collection cup so
that the entire test strip pad area can be dipped in a single
motion. Stir well.
Waste box
Test strip
Test strip
pad area
Do NOT wet the
black mark.
Test strip
Rails (white)
Test strip detection window
Detection
area
(up to 13 digits)
,
(up to 4 digits)
to confi rm entry.
7. Dip entire test strip pad area in sample for two (2) seconds.
• If buzzer setting is ON follow the timing chart below:
Measurement
Name of each part
Display
Operator panel
Test strip tray
Carrying arm Cover open button
(on both sides)
Waste box
Maintenance cover
Built-in printer
Rails (white)
Test strip detection
window
Preparation of measurement
Setup (test strip, measurement No., patient ID)
Measurement
Check that there is no used test strips in the waste box.
with buttonsSelect
[ Select test strip type ] [ Menu screen ]
Select the type of test strip
Dip the entire test strip pad area in the sample for two seconds.
IMPORTANT
Do NOT wet the black mark.
After dipping the test strip in the sample,
remove excess urine by letting the strip
touch the edge of the collection cup.
Set the measurement number
Place a test strip over the detection area of the test
strip detection window, crossing the two white rails
of the test strip tray.
The carrying arm sends the test strip into the
analyzer, and the measurement starts automatically.
Set the patient ID
When measuring continuously, repeat steps 1 to 3 .
Set measurement numbers and patient IDs as necessary.
When the buzzer setting is ON, the beeper sounds
for two seconds to signal the timing for dipping the
next test strip.
Dip the test strip while you hear the beep.
Turn off the instrument after completing all the measurement procedures.
Turn on the power switch at the back of the instrument.
Prepare a sufficient volume of sample in urine collection cup so
that the entire test strip pad area can be dipped in a single
motion. Stir well.
Waste box
Test strip
Test strip
pad area
Do NOT wet the
black mark.
Test strip
Rails (white)
Test strip detection window
Detection
area
(up to 13 digits)
,
(up to 4 digits)
IMPORTANT:
• Do NOT wet the black
mark on test strip.
• After dipping test strip in
sample, gently remove
excess urine by drawing
the entire length of strip
against the rim of the
collection cup.
AE-4020
2-16
Chapter 2 Measurement Operation
Take out the required number of test strips from
the bottle.
Close the test strip bottle cap securely.
IMPORTANT: After taking the test strips out
of the bottle, immediately close
the cap or the test strips in the
bottle will absorb moisture or dirt
in the air, which may deteriorate
reagents and make the test
strips useless.
5
Start normal measurement
Conrm that the icon ( ) for normal measurement
appears on the [Standby screen].
IMPORTANT: Different procedures at this
stage are required, depending
on the setting of the “opera-
tional mode when turning ON”
parameter. (See “Operational
mode” on page 2-3.)
The standard setting is “Auto start” mode. To change the “operational
mode when turning ON” parameter, see “3-7-8. No.005: Operational
mode when turning ON” on page 3-18.
REFERENCE: When the buzzer setting is ON, the buzzer sounds in the following
intervals to signal the dipping timing for the test strips.
Slow beeping
Fast beeping
Preparation for dipping:
3 seconds
Dipping period:
2 seconds
Test strip setting:
1.4 seconds
●When using the “Auto start” mode, go to step 6.
●When using the “Cycle start” mode, press and go to step 6.
6
Call up the [Clarity input screen]
Press and hold for about 2 seconds.
The [Clarity input screen] appears.
1 2 3 4 5
Clear Slightly
Cloudy
Cloudy Turbid Other
2.4 Measurement operation
AE-4022 2-19
6 Call up the [Clarity input screen]
aPress and hold for about 2 seconds.
• The [Clarity input screen] appears.
7 Enter the clarity value
aEnter the clarity value of the sample using the numeric keys.
• Enter a number between 1 and 5 using the table below.
bPress .
• The display returns to the [Standby screen].
8 Dip a test strip in the sample for 2 seconds
aDip the test strip in the sample for 2 seconds, and then withdraw it.
InputNo.12345
Clarity Clear Slightly
cloudy Cloudy Turbid Other
REFERENCE:
When the clarity value of the sample is “1”, just press without pressing any numeric keys.
IMPORTANT:
Dip the entire test strip pad area in the sample at once. Do not allow the
black mark shown in the figure to become wet. If the test strip is dipped too
deeply and the black mark is moistened, correct measurement may not be
obtained.
After dipping the test strip in the sample, gently remove excess urine by
drawing the entire length of the test strip against the rim of the collection cup.
In the “Auto start” mode, the buzzer does not sound for the first measurement. Always dip the test strip for 2
seconds. Insufficient dipping of test strips may cause insufficient color change, while dipping too long may
cause reagents to drain out of the test strip, either of which will prevent correct measurement.
Do not dip the
black mark.
Continued on next page…

SEMI-AUTOMATED URINE ANALYZER
AUTION ELEVEN
™
AE-4022 Sample Measurement
5
8. Place test strip on test strip tray.
• Place strip in the detection area of the test strip detection window
with the black marker on right or left.
• The carrying arm sends the test strip into the analyzer and
measurement starts automatically.
9. When buzzer setting is ON, the beeper sounds for two seconds to
signal the timing for dipping the next test strip. Dip the test strip
while you hear the beep.
When measuring continuously, repeat steps 2 to 4.
Set measurement numbers and patient IDs as necessary.
Measurement
Name of each part
Display
Operator panel
Test strip tray
Carrying arm Cover open button
(on both sides)
Waste box
Maintenance cover
Built-in printer
Rails (white)
Test strip detection
window
Preparation of measurement
Setup (test strip, measurement No., patient ID)
Measurement
Check that there is no used test strips in the waste box.
with buttonsSelect
[ Select test strip type ] [ Menu screen ]
Select the type of test strip
Dip the entire test strip pad area in the sample for two seconds.
IMPORTANT
Do NOT wet the black mark.
After dipping the test strip in the sample,
remove excess urine by letting the strip
touch the edge of the collection cup.
Set the measurement number
Place a test strip over the detection area of the test
strip detection window, crossing the two white rails
of the test strip tray.
The carrying arm sends the test strip into the
analyzer, and the measurement starts automatically.
Set the patient ID
When measuring continuously, repeat steps 1 to 3 .
Set measurement numbers and patient IDs as necessary.
When the buzzer setting is ON, the beeper sounds
for two seconds to signal the timing for dipping the
next test strip.
Dip the test strip while you hear the beep.
Turn off the instrument after completing all the measurement procedures.
Turn on the power switch at the back of the instrument.
Prepare a sufficient volume of sample in urine collection cup so
that the entire test strip pad area can be dipped in a single
motion. Stir well.
Waste box
Test strip
Test strip
pad area
Do NOT wet the
black mark.
Test strip
Rails (white)
Test strip detection window
Detection
area
(up to 13 digits)
,
(up to 4 digits)
10. Read test results from printout.
11. Turn off instrument after completing all measurement procedures.

Control Measurement
Control Measurement

6
SEMI-AUTOMATED URINE ANALYZER
AUTION ELEVEN
™
AE-4022 Control Measurement
1. Change measurement mode.
• From the [standby screen] press
Measurement
Name of each part
Display
Operator panel
Test strip tray
Carrying arm Cover open button
(on both sides)
Waste box
Maintenance cover
Built-in printer
Rails (white)
Test strip detection
window
Preparation of measurement
Setup (test strip, measurement No., patient ID)
Measurement
Check that there is no used test strips in the waste box.
with buttonsSelect
[ Select test strip type ] [ Menu screen ]
Select the type of test strip
Dip the entire test strip pad area in the sample for two seconds.
IMPORTANT
Do NOT wet the black mark.
After dipping the test strip in the sample,
remove excess urine by letting the strip
touch the edge of the collection cup.
Set the measurement number
Place a test strip over the detection area of the test
strip detection window, crossing the two white rails
of the test strip tray.
The carrying arm sends the test strip into the
analyzer, and the measurement starts automatically.
Set the patient ID
When measuring continuously, repeat steps 1 to 3 .
Set measurement numbers and patient IDs as necessary.
When the buzzer setting is ON, the beeper sounds
for two seconds to signal the timing for dipping the
next test strip.
Dip the test strip while you hear the beep.
Turn off the instrument after completing all the measurement procedures.
Turn on the power switch at the back of the instrument.
Prepare a sufficient volume of sample in urine collection cup so
that the entire test strip pad area can be dipped in a single
motion. Stir well.
Waste box
Test strip
Test strip
pad area
Do NOT wet the
black mark.
Test strip
Rails (white)
Test strip detection window
Detection
area
(up to 13 digits)
,
(up to 4 digits)
or
Measurement
Name of each part
Display
Operator panel
Test strip tray
Carrying arm Cover open button
(on both sides)
Waste box
Maintenance cover
Built-in printer
Rails (white)
Test strip detection
window
Preparation of measurement
Setup (test strip, measurement No., patient ID)
Measurement
Check that there is no used test strips in the waste box.
with buttonsSelect
[ Select test strip type ] [ Menu screen ]
Select the type of test strip
Dip the entire test strip pad area in the sample for two seconds.
IMPORTANT
Do NOT wet the black mark.
After dipping the test strip in the sample,
remove excess urine by letting the strip
touch the edge of the collection cup.
Set the measurement number
Place a test strip over the detection area of the test
strip detection window, crossing the two white rails
of the test strip tray.
The carrying arm sends the test strip into the
analyzer, and the measurement starts automatically.
Set the patient ID
When measuring continuously, repeat steps 1 to 3 .
Set measurement numbers and patient IDs as necessary.
When the buzzer setting is ON, the beeper sounds
for two seconds to signal the timing for dipping the
next test strip.
Dip the test strip while you hear the beep.
Turn off the instrument after completing all the measurement procedures.
Turn on the power switch at the back of the instrument.
Prepare a sufficient volume of sample in urine collection cup so
that the entire test strip pad area can be dipped in a single
motion. Stir well.
Waste box
Test strip
Test strip
pad area
Do NOT wet the
black mark.
Test strip
Rails (white)
Test strip detection window
Detection
area
(up to 13 digits)
,
(up to 4 digits)
until the
AE-4020
2-24
Chapter 2 Measurement Operation
5
Start the Control measurement
IMPORTANT: Different procedures at this
stage are required, depending
on the setting of the “opera-
tional mode when turning ON”
parameter. (See “Operational
mode” on page 2-3.)
The standard setting is “Auto
start” mode. To change the
“operational mode when turn-
ing ON”, see “3-7-8. No.005:
Operational mode when turn-
ing ON” on page 3-18.
REFERENCE: When the buzzer setting is ON, the buzzer sounds in the following
intervals to signal the dipping timing for the test strips.
Slow beeping Fast beeping
Preparation for
dipping: 3 seconds
Dipping period:
2 seconds
Test strip setting:
1.4 seconds
●When using the “Auto start” mode, go to step 6.
●When using the “Cycle start” mode, press and go to step 6.
6
Dip a test strip in the sample for 2 seconds
Do not dip the
black mark.
Dip the test strip in the sample for 2 seconds, and then withdraw it.
IMPORTANT: ●Dip the entire test strip pad
area in the sample at once.
Do not allow the black mark
shown in the gure to become
wet. If the test strip is dipped
too deeply and the black
mark is moistened, correct
measurement may not be
obtained.
●After dipping the test strip in the sample, remove excess urine
using the edge of the collection cup.
●In the “Auto start” mode, the buzzer does not sound for the rst
measurement. Always dip the test strip for 2 seconds. Insufcient
dipping of test strips may cause insufcient color change, while
dipping too long may cause reagents to drain out of the test strip,
either of which will prevent correct measurement.
REFERENCE: When the buzzer setting is ON, the buzzer sounds for 2 seconds to
signal the dipping timing.
icon is displayed.
2. Enter measurement number to identify control level.
3. Dip entire test strip pad area in sample for two (2) seconds.
• If buzzer setting is ON follow the timing chart below:
AE-4020
2-16
Chapter 2 Measurement Operation
Take out the required number of test strips from
the bottle.
Close the test strip bottle cap securely.
IMPORTANT: After taking the test strips out
of the bottle, immediately close
the cap or the test strips in the
bottle will absorb moisture or dirt
in the air, which may deteriorate
reagents and make the test
strips useless.
5
Start normal measurement
Conrm that the icon ( ) for normal measurement
appears on the [Standby screen].
IMPORTANT: Different procedures at this
stage are required, depending
on the setting of the “opera-
tional mode when turning ON”
parameter. (See “Operational
mode” on page 2-3.)
The standard setting is “Auto start” mode. To change the “operational
mode when turning ON” parameter, see “3-7-8. No.005: Operational
mode when turning ON” on page 3-18.
REFERENCE: When the buzzer setting is ON, the buzzer sounds in the following
intervals to signal the dipping timing for the test strips.
Slow beeping
Fast beeping
Preparation for dipping:
3 seconds
Dipping period:
2 seconds
Test strip setting:
1.4 seconds
●When using the “Auto start” mode, go to step 6.
●When using the “Cycle start” mode, press and go to step 6.
6
Call up the [Clarity input screen]
Press and hold for about 2 seconds.
The [Clarity input screen] appears.
4. Place test strip on test strip tray.
• Place strip in the detection area of the test strip detection window with the black marker on right or left.
• The carrying arm sends the test strip into the analyzer and measurement starts automatically.
5. When buzzer setting is ON, the beeper sounds for two seconds to signal the timing for dipping the next test strip.
Dip the test strip while you hear the beep.
When measuring continuously, repeat steps 2 to 4.
Set measurement numbers as necessary.
6. Read test results from printout.
7. Turn off instrument after completing all measurement procedures.
2.4 Measurement operation
AE-4022 2-27
2.4.3 Control measurement
Controls should be used daily, whenever a new bottle of strips is opened, before reporting unusual or unexpected
results, and when training new operators. Refer to the control solution package insert for expected values. Each
laboratory should establish its own goals for adequate standards of performance by following local, state and federal
requirements when establishing their QC program.
1 Prepare the Control
aPrepare the Control.
2 Change the measurement mode
aCheck that the [Standby screen] is displayed.
bPress or to switch to the Control measurement mode.
• On the [Standby screen], theicon is displayed and the
mode changes to the Control measurement mode. When the
measurement mode is switched, the measurement number
restarts from “0001”.
Wear protective gloves to prevent exposure to pathogenic microorganisms.
Pad area
IMPORTANT:
Prepare a sufficient volume of Control in a vessel so that the
entire test strip pad area can be dipped at once.
NOTE:
Before using a Control, read carefully its package insert.

Check Measurement
Check Measurement

7
SEMI-AUTOMATED URINE ANALYZER
AUTION ELEVEN
™
AE-4022 Check Measurement
1. Clean the feeder and waste box to prevent check strips from becoming soiled.
2. Change measurement mode.
• From the [standby screen] press
Measurement
Name of each part
Display
Operator panel
Test strip tray
Carrying arm Cover open button
(on both sides)
Waste box
Maintenance cover
Built-in printer
Rails (white)
Test strip detection
window
Preparation of measurement
Setup (test strip, measurement No., patient ID)
Measurement
Check that there is no used test strips in the waste box.
with buttonsSelect
[ Select test strip type ] [ Menu screen ]
Select the type of test strip
Dip the entire test strip pad area in the sample for two seconds.
IMPORTANT
Do NOT wet the black mark.
After dipping the test strip in the sample,
remove excess urine by letting the strip
touch the edge of the collection cup.
Set the measurement number
Place a test strip over the detection area of the test
strip detection window, crossing the two white rails
of the test strip tray.
The carrying arm sends the test strip into the
analyzer, and the measurement starts automatically.
Set the patient ID
When measuring continuously, repeat steps 1 to 3 .
Set measurement numbers and patient IDs as necessary.
When the buzzer setting is ON, the beeper sounds
for two seconds to signal the timing for dipping the
next test strip.
Dip the test strip while you hear the beep.
Turn off the instrument after completing all the measurement procedures.
Turn on the power switch at the back of the instrument.
Prepare a sufficient volume of sample in urine collection cup so
that the entire test strip pad area can be dipped in a single
motion. Stir well.
Waste box
Test strip
Test strip
pad area
Do NOT wet the
black mark.
Test strip
Rails (white)
Test strip detection window
Detection
area
(up to 13 digits)
,
(up to 4 digits)
or
Measurement
Name of each part
Display
Operator panel
Test strip tray
Carrying arm Cover open button
(on both sides)
Waste box
Maintenance cover
Built-in printer
Rails (white)
Test strip detection
window
Preparation of measurement
Setup (test strip, measurement No., patient ID)
Measurement
Check that there is no used test strips in the waste box.
with buttonsSelect
[ Select test strip type ] [ Menu screen ]
Select the type of test strip
Dip the entire test strip pad area in the sample for two seconds.
IMPORTANT
Do NOT wet the black mark.
After dipping the test strip in the sample,
remove excess urine by letting the strip
touch the edge of the collection cup.
Set the measurement number
Place a test strip over the detection area of the test
strip detection window, crossing the two white rails
of the test strip tray.
The carrying arm sends the test strip into the
analyzer, and the measurement starts automatically.
Set the patient ID
When measuring continuously, repeat steps 1 to 3 .
Set measurement numbers and patient IDs as necessary.
When the buzzer setting is ON, the beeper sounds
for two seconds to signal the timing for dipping the
next test strip.
Dip the test strip while you hear the beep.
Turn off the instrument after completing all the measurement procedures.
Turn on the power switch at the back of the instrument.
Prepare a sufficient volume of sample in urine collection cup so
that the entire test strip pad area can be dipped in a single
motion. Stir well.
Waste box
Test strip
Test strip
pad area
Do NOT wet the
black mark.
Test strip
Rails (white)
Test strip detection window
Detection
area
(up to 13 digits)
,
(up to 4 digits)
until the
2-27
AE-4020
Chapter 2 Measurement Operation
3
Prepare the check strips
Check strip (white)
Check strip (gray)
Reectance label
Prepare the check strips. The check strip bottle has
a reectance label on it, which must be used when
evaluating reectance after the check measurement.
Keep this labeled bottle in an easily accessible
place.
4
Start the check measurement
Place a white check strip on the test strip tray.
REFERENCE: Place the check strip on the test strip tray with the black mark facing up.
NOTE: Place a test strip in the detection
area of the test strip detection
window as shown in the right gure.
If the system does not detect the test
strip in the auto start mode, remove
and replace it in the center of the test
strip detection widow. At this time,
the test strip must cross both of the
white rails in the detection area. If the
test strip is not placed in the correct
position, it may not be transported
properly, leading to a strip jam or
inaccurate measurement result.
Rails (white)
Test strip
detection window
Detection
area
Pay attention to the carrying arm when it moves, so as not to
have your ngers caught or pinched.
IMPORTANT: Different procedures at this stage are required, depending on the
setting of the “operational mode when turning ON” parameter. (See
“Operational mode” on page 2-3.)
The standard setting is “Auto start” mode. To change the operational
mode when turning ON, see “3-7-8. No.005: Operational mode when
turning ON” on page 3-18.
●When using the “Cycle start” mode, press to
start the check measurement of the rst strip.
●In the “Auto start” mode, the system automatically
starts the check measurement of the rst strip
when placement of the check strip is detected.
icon is displayed.
3. Place white check strip on test strip tray and perform measurement.
4. Place gray check strip on test strip tray and perform measurement.
5. Remove check strips from waste box and return to check strip bottle.
6. Evaluate results
• Compare the refl ectance values on the test result printout to the ranges on the refl ectance label on the check strip bottle.
• If any of the measurement results are outside of the specfi ed ranges, repeat check measurement using the other set of
check strips.
• If any of the measurement results are still outside the ranges, contact ARKRAY Technical Support at 1.855.646.3108.
2.4 Measurement operation
AE-4022 2-33
bPress or to switch to check measurement mode.
• On the [Standby screen], the icon is displayed, and the
measurement mode is switched to the Check measurement mode.
3 Prepare the check strips
aPrepare the check strips. The check strip bottle has a reflectance
label on it, which must be used when evaluating reflectance after
the check measurement. Keep this labeled bottle in an easily
accessible place.
4 Start the check measurement
aPlace a white check strip on the test strip tray.
Reflectance label
Check strip (gray)
Check strip (white)
REFERENCE:
Place the check strip on the test strip tray with the black mark facing up.
Detection
area
Test strip
detection window
Rails (white)
NOTE:
Place a test strip in the detection area of the test strip
detection window as shown in the right figure. If the system
does not detect the test strip in the auto start mode, remove
and replace it in the center of the test strip detection
window. At this time, the test strip must cross both of the
white rails in the detection area. If the test strip is not placed
in the correct position, it may not be transported properly,
leading to a strip jam or inaccurate measurement result.
Pay attention to the carrying arm when it moves, so as not to have your fingers caught or
pinched.
IMPORTANT:
Different procedures at this stage are required, depending on the setting of the “operational mode when
turning ON” parameter. (See “Operational mode” on page 2-3.)
The standard setting is “Auto start” mode. To change the operational mode when turning ON, see “3.7.8.
No.005: Operational mode when turning ON” on page 3-27.

Maintenance
Maintenance

8
SEMI-AUTOMATED URINE ANALYZER
AUTION ELEVEN
™
AE-4022 Maintenance – Daily
Daily Maintenance
Power switch
Turn off the instrument.
Waste box
Discard used test strips.
140820YH
Wear protective gloves to prevent
exposure to pathogenic microbes.
Discard used test strips and
protective gloves separately from
general waste according to local
regulations on biohazardous waste.
Detach the carrying arm.
Open the maintenance cover.
Feed lever
Lever
Detach the feed lever by sliding the white lever
in the center towards you.
Wash the detached carrying arm, waste box,
test strip tray, suction port, test strip feed tray
and feed lever with water to remove any dirt.
Carrying arm
Suction port
Waste box
Test strip feed tray
Feed lever
Test strip tray
Wipe the parts with cloth and fully dry them before
attaching them back on.
This Quick Reference Guide is a brief
summary of the measurement operation
and daily maintenance. Before using the
instrument, make sure to read the Operat-
ing Manual.
Clean the area circled in red below.
Test strip feed tray
Detach the test strip feed tray.
Suction port
Test strip tray
Detach the suction port.
Test strip tray
Detach the test strip tray.
Daily Maintenance
Power switch
Turn off the instrument.
Waste box
Discard used test strips.
140820YH
Wear protective gloves to prevent
exposure to pathogenic microbes.
Discard used test strips and
protective gloves separately from
general waste according to local
regulations on biohazardous waste.
Detach the carrying arm.
Open the maintenance cover.
Feed lever
Lever
Detach the feed lever by sliding the white lever
in the center towards you.
Wash the detached carrying arm, waste box,
test strip tray, suction port, test strip feed tray
and feed lever with water to remove any dirt.
Carrying arm
Suction port
Waste box
Test strip feed tray
Feed lever
Test strip tray
Wipe the parts with cloth and fully dry them before
attaching them back on.
This Quick Reference Guide is a brief
summary of the measurement operation
and daily maintenance. Before using the
instrument, make sure to read the Operat-
ing Manual.
Clean the area circled in red below.
Test strip feed tray
Detach the test strip feed tray.
Suction port
Test strip tray
Detach the suction port.
Test strip tray
Detach the test strip tray.
Daily Maintenance
Power switch
Turn off the instrument.
Waste box
Discard used test strips.
140820YH
Wear protective gloves to prevent
exposure to pathogenic microbes.
Discard used test strips and
protective gloves separately from
general waste according to local
regulations on biohazardous waste.
Detach the carrying arm.
Open the maintenance cover.
Feed lever
Lever
Detach the feed lever by sliding the white lever
in the center towards you.
Wash the detached carrying arm, waste box,
test strip tray, suction port, test strip feed tray
and feed lever with water to remove any dirt.
Carrying arm
Suction port
Waste box
Test strip feed tray
Feed lever
Test strip tray
Wipe the parts with cloth and fully dry them before
attaching them back on.
This Quick Reference Guide is a brief
summary of the measurement operation
and daily maintenance. Before using the
instrument, make sure to read the Operat-
ing Manual.
Clean the area circled in red below.
Test strip feed tray
Detach the test strip feed tray.
Suction port
Test strip tray
Detach the suction port.
Test strip tray
Detach the test strip tray.
Daily Maintenance
Power switch
Turn off the instrument.
Waste box
Discard used test strips.
140820YH
Wear protective gloves to prevent
exposure to pathogenic microbes.
Discard used test strips and
protective gloves separately from
general waste according to local
regulations on biohazardous waste.
Detach the carrying arm.
Open the maintenance cover.
Feed lever
Lever
Detach the feed lever by sliding the white lever
in the center towards you.
Wash the detached carrying arm, waste box,
test strip tray, suction port, test strip feed tray
and feed lever with water to remove any dirt.
Carrying arm
Suction port
Waste box
Test strip feed tray
Feed lever
Test strip tray
Wipe the parts with cloth and fully dry them before
attaching them back on.
This Quick Reference Guide is a brief
summary of the measurement operation
and daily maintenance. Before using the
instrument, make sure to read the Operat-
ing Manual.
Clean the area circled in red below.
Test strip feed tray
Detach the test strip feed tray.
Suction port
Test strip tray
Detach the suction port.
Test strip tray
Detach the test strip tray.
1. Turn off instrument. 2. Detach carrying arm.
• Pull towards you and
lift up to remove.
3 Open maintenance cover.
• Press and hold the two open
buttons on either side of
instrument and lift up to open.
4. Detach test strip tray.
• Slide the test strip tray to the
front and lift up to detach.
5. Detach suction port.
• Pull suction port straight
up to detach.
Clean the Feeder
Daily Maintenance
Power switch
Turn off the instrument.
Waste box
Discard used test strips.
140820YH
Wear protective gloves to prevent
exposure to pathogenic microbes.
Discard used test strips and
protective gloves separately from
general waste according to local
regulations on biohazardous waste.
Detach the carrying arm.
Open the maintenance cover.
Feed lever
Lever
Detach the feed lever by sliding the white lever
in the center towards you.
Wash the detached carrying arm, waste box,
test strip tray, suction port, test strip feed tray
and feed lever with water to remove any dirt.
Carrying arm
Suction port
Waste box
Test strip feed tray
Feed lever
Test strip tray
Wipe the parts with cloth and fully dry them before
attaching them back on.
This Quick Reference Guide is a brief
summary of the measurement operation
and daily maintenance. Before using the
instrument, make sure to read the Operat-
ing Manual.
Clean the area circled in red below.
Test strip feed tray
Detach the test strip feed tray.
Suction port
Test strip tray
Detach the suction port.
Test strip tray
Detach the test strip tray.
Daily Maintenance
Power switch
Turn off the instrument.
Waste box
Discard used test strips.
140820YH
Wear protective gloves to prevent
exposure to pathogenic microbes.
Discard used test strips and
protective gloves separately from
general waste according to local
regulations on biohazardous waste.
Detach the carrying arm.
Open the maintenance cover.
Feed lever
Lever
Detach the feed lever by sliding the white lever
in the center towards you.
Wash the detached carrying arm, waste box,
test strip tray, suction port, test strip feed tray
and feed lever with water to remove any dirt.
Carrying arm
Suction port
Waste box
Test strip feed tray
Feed lever
Test strip tray
Wipe the parts with cloth and fully dry them before
attaching them back on.
This Quick Reference Guide is a brief
summary of the measurement operation
and daily maintenance. Before using the
instrument, make sure to read the Operat-
ing Manual.
Clean the area circled in red below.
Test strip feed tray
Detach the test strip feed tray.
Suction port
Test strip tray
Detach the suction port.
Test strip tray
Detach the test strip tray.
6. Detach test strip feed tray.
• Slide to front and lift up
to detach.
Continued on next page…

9
SEMI-AUTOMATED URINE ANALYZER
AUTION ELEVEN
™
AE-4022 Maintenance – Daily
Daily Maintenance
Power switch
Turn off the instrument.
Waste box
Discard used test strips.
140820YH
Wear protective gloves to prevent
exposure to pathogenic microbes.
Discard used test strips and
protective gloves separately from
general waste according to local
regulations on biohazardous waste.
Detach the carrying arm.
Open the maintenance cover.
Feed lever
Lever
Detach the feed lever by sliding the white lever
in the center towards you.
Wash the detached carrying arm, waste box,
test strip tray, suction port, test strip feed tray
and feed lever with water to remove any dirt.
Carrying arm
Suction port
Waste box
Test strip feed tray
Feed lever
Test strip tray
Wipe the parts with cloth and fully dry them before
attaching them back on.
This Quick Reference Guide is a brief
summary of the measurement operation
and daily maintenance. Before using the
instrument, make sure to read the Operat-
ing Manual.
Clean the area circled in red below.
Test strip feed tray
Detach the test strip feed tray.
Suction port
Test strip tray
Detach the suction port.
Test strip tray
Detach the test strip tray.
Daily Maintenance
Power switch
Turn off the instrument.
Waste box
Discard used test strips.
140820YH
Wear protective gloves to prevent
exposure to pathogenic microbes.
Discard used test strips and
protective gloves separately from
general waste according to local
regulations on biohazardous waste.
Detach the carrying arm.
Open the maintenance cover.
Feed lever
Lever
Detach the feed lever by sliding the white lever
in the center towards you.
Wash the detached carrying arm, waste box,
test strip tray, suction port, test strip feed tray
and feed lever with water to remove any dirt.
Carrying arm
Suction port
Waste box
Test strip feed tray
Feed lever
Test strip tray
Wipe the parts with cloth and fully dry them before
attaching them back on.
This Quick Reference Guide is a brief
summary of the measurement operation
and daily maintenance. Before using the
instrument, make sure to read the Operat-
ing Manual.
Clean the area circled in red below.
Test strip feed tray
Detach the test strip feed tray.
Suction port
Test strip tray
Detach the suction port.
Test strip tray
Detach the test strip tray.
Daily Maintenance
Power switch
Turn off the instrument.
Waste box
Discard used test strips.
140820YH
Wear protective gloves to prevent
exposure to pathogenic microbes.
Discard used test strips and
protective gloves separately from
general waste according to local
regulations on biohazardous waste.
Detach the carrying arm.
Open the maintenance cover.
Feed lever
Lever
Detach the feed lever by sliding the white lever
in the center towards you.
Wash the detached carrying arm, waste box,
test strip tray, suction port, test strip feed tray
and feed lever with water to remove any dirt.
Carrying arm
Suction port
Waste box
Test strip feed tray
Feed lever
Test strip tray
Wipe the parts with cloth and fully dry them before
attaching them back on.
This Quick Reference Guide is a brief
summary of the measurement operation
and daily maintenance. Before using the
instrument, make sure to read the Operat-
ing Manual.
Clean the area circled in red below.
Test strip feed tray
Detach the test strip feed tray.
Suction port
Test strip tray
Detach the suction port.
Test strip tray
Detach the test strip tray.
7. Detach feed lever by sliding
the white lever in the center
towards you.
8. Clean the table.
• Clean the areas circled in red below with water.
Clean the Feeder
Continued on next page…

10
SEMI-AUTOMATED URINE ANALYZER
AUTION ELEVEN
™
AE-4022 Maintenance – Daily
9. Wash detached test strip tray, suction port, test strip feed tray and feed lever with water.
10. Sterilize the carrying arm using alcohol. Then wash with water.
11. Wipe the parts with a cloth and dry completely before reattaching to instrument.
Daily Maintenance
Power switch
Turn off the instrument.
Waste box
Discard used test strips.
140820YH
Wear protective gloves to prevent
exposure to pathogenic microbes.
Discard used test strips and
protective gloves separately from
general waste according to local
regulations on biohazardous waste.
Detach the carrying arm.
Open the maintenance cover.
Feed lever
Lever
Detach the feed lever by sliding the white lever
in the center towards you.
Wash the detached carrying arm, waste box,
test strip tray, suction port, test strip feed tray
and feed lever with water to remove any dirt.
Carrying arm
Suction port
Waste box
Test strip feed tray
Feed lever
Test strip tray
Wipe the parts with cloth and fully dry them before
attaching them back on.
This Quick Reference Guide is a brief
summary of the measurement operation
and daily maintenance. Before using the
instrument, make sure to read the Operat-
ing Manual.
Clean the area circled in red below.
Test strip feed tray
Detach the test strip feed tray.
Suction port
Test strip tray
Detach the suction port.
Test strip tray
Detach the test strip tray.
Daily Maintenance
Power switch
Turn off the instrument.
Waste box
Discard used test strips.
140820YH
Wear protective gloves to prevent
exposure to pathogenic microbes.
Discard used test strips and
protective gloves separately from
general waste according to local
regulations on biohazardous waste.
Detach the carrying arm.
Open the maintenance cover.
Feed lever
Lever
Detach the feed lever by sliding the white lever
in the center towards you.
Wash the detached carrying arm, waste box,
test strip tray, suction port, test strip feed tray
and feed lever with water to remove any dirt.
Carrying arm
Suction port
Waste box
Test strip feed tray
Feed lever
Test strip tray
Wipe the parts with cloth and fully dry them before
attaching them back on.
This Quick Reference Guide is a brief
summary of the measurement operation
and daily maintenance. Before using the
instrument, make sure to read the Operat-
ing Manual.
Clean the area circled in red below.
Test strip feed tray
Detach the test strip feed tray.
Suction port
Test strip tray
Detach the suction port.
Test strip tray
Detach the test strip tray.
Daily Maintenance
Power switch
Turn off the instrument.
Waste box
Discard used test strips.
140820YH
Wear protective gloves to prevent
exposure to pathogenic microbes.
Discard used test strips and
protective gloves separately from
general waste according to local
regulations on biohazardous waste.
Detach the carrying arm.
Open the maintenance cover.
Feed lever
Lever
Detach the feed lever by sliding the white lever
in the center towards you.
Wash the detached carrying arm, waste box,
test strip tray, suction port, test strip feed tray
and feed lever with water to remove any dirt.
Carrying arm
Suction port
Waste box
Test strip feed tray
Feed lever
Test strip tray
Wipe the parts with cloth and fully dry them before
attaching them back on.
This Quick Reference Guide is a brief
summary of the measurement operation
and daily maintenance. Before using the
instrument, make sure to read the Operat-
ing Manual.
Clean the area circled in red below.
Test strip feed tray
Detach the test strip feed tray.
Suction port
Test strip tray
Detach the suction port.
Test strip tray
Detach the test strip tray.
Daily Maintenance
Power switch
Turn off the instrument.
Waste box
Discard used test strips.
140820YH
Wear protective gloves to prevent
exposure to pathogenic microbes.
Discard used test strips and
protective gloves separately from
general waste according to local
regulations on biohazardous waste.
Detach the carrying arm.
Open the maintenance cover.
Feed lever
Lever
Detach the feed lever by sliding the white lever
in the center towards you.
Wash the detached carrying arm, waste box,
test strip tray, suction port, test strip feed tray
and feed lever with water to remove any dirt.
Carrying arm
Suction port
Waste box
Test strip feed tray
Feed lever
Test strip tray
Wipe the parts with cloth and fully dry them before
attaching them back on.
This Quick Reference Guide is a brief
summary of the measurement operation
and daily maintenance. Before using the
instrument, make sure to read the Operat-
ing Manual.
Clean the area circled in red below.
Test strip feed tray
Detach the test strip feed tray.
Suction port
Test strip tray
Detach the suction port.
Test strip tray
Detach the test strip tray.
Daily Maintenance
Power switch
Turn off the instrument.
Waste box
Discard used test strips.
140820YH
Wear protective gloves to prevent
exposure to pathogenic microbes.
Discard used test strips and
protective gloves separately from
general waste according to local
regulations on biohazardous waste.
Detach the carrying arm.
Open the maintenance cover.
Feed lever
Lever
Detach the feed lever by sliding the white lever
in the center towards you.
Wash the detached carrying arm, waste box,
test strip tray, suction port, test strip feed tray
and feed lever with water to remove any dirt.
Carrying arm
Suction port
Waste box
Test strip feed tray
Feed lever
Test strip tray
Wipe the parts with cloth and fully dry them before
attaching them back on.
This Quick Reference Guide is a brief
summary of the measurement operation
and daily maintenance. Before using the
instrument, make sure to read the Operat-
ing Manual.
Clean the area circled in red below.
Test strip feed tray
Detach the test strip feed tray.
Suction port
Test strip tray
Detach the suction port.
Test strip tray
Detach the test strip tray.
Clean the Feeder

11
SEMI-AUTOMATED URINE ANALYZER
AUTION ELEVEN
™
AE-4022 Maintenance – Daily
Clean the Waste Box
Chapter 4 Maintenance
AE-4022 4-10
4.1.2 Cleaning the waste box
The waste box gets full after approximately 100 measurements. Discard the used test strips, and sterilize and clean the
waste box.
Items required: Alcohol, cloth, and protective gloves.
1 Discard used test strips
aCheck that the [Standby screen] is displayed.
bPull out the waste box and discard the used test strips.
2 Sterilize and clean the waste box
aSterilize the waste box using alcohol and then wash with water to
remove all traces of dirt.
bWipe all moisture off with a cloth to dry the part.
Wear protective gloves to prevent exposure to pathogenic microorganisms.
Discard used test strips in accordance with local regulations for biohazardous waste.
NOTE:
When sterilizing and cleaning the waste box, avoid the use of thinner, other organic solvents, or ultrasonic
cleaning. These sterilizing/cleaning methods may deform or discolor the test strip tray and thus make it
unserviceable for further testing.
Waste box
1. Remove waste box and
discard used test strips.
2. Sterilize and clean waste box.
• Sterilize with alcohol then
wash with water and wipe dry.
3. Attach waste box.
Chapter 4 Maintenance
AE-4022 4-10
4.1.2 Cleaning the waste box
The waste box gets full after approximately 100 measurements. Discard the used test strips, and sterilize and clean the
waste box.
Items required: Alcohol, cloth, and protective gloves.
1 Discard used test strips
aCheck that the [Standby screen] is displayed.
bPull out the waste box and discard the used test strips.
2 Sterilize and clean the waste box
aSterilize the waste box using alcohol and then wash with water to
remove all traces of dirt.
bWipe all moisture off with a cloth to dry the part.
Wear protective gloves to prevent exposure to pathogenic microorganisms.
Discard used test strips in accordance with local regulations for biohazardous waste.
NOTE:
When sterilizing and cleaning the waste box, avoid the use of thinner, other organic solvents, or ultrasonic
cleaning. These sterilizing/cleaning methods may deform or discolor the test strip tray and thus make it
unserviceable for further testing.
Waste box
4.1 Daily maintenance
AE-4022 4-11
3 Attach the waste box.
aAttach the waste box to the unit.
NOTE:
When attaching the waste box to the unit, do not spread
tissue paper (Kleenex) or anything else in the bottom of the
box. Doing so may cause a problem later when disposing of
waste.

12
SEMI-AUTOMATED URINE ANALYZER
AUTION ELEVEN
™
AE-4022 Maintenance – As Needed
1. Check that the [Standby screen] is displayed,
then open the printer cover.
2. Press
Chapter 4 Maintenance
AE-4022 4-12
4.2 Replacing the thermal recording paper
When a red line appears on each side of the thermal recording paper, the paper will soon run out. Replace the depleted
paper roll with a new one before it runs out. A new roll of thermal recording paper can be used for approximately 450
measurements.
Items required: New roll of thermal recording paper, scissors
1 Cut the thermal recording paper
aCheck that the [Standby screen] is displayed.
bOpen the printer cover.
cIf some thermal recording paper still remains in the printer, cut
the remaining paper strip using scissors and remove the depleted
roll’s paper core.
If the paper has completely run out, remove the paper core from
the paper holder and go to step 3.
2 Remove the remaining thermal recording paper
aPress . The thermal recording paper remaining in the printer
is fed out. Remove the paper by holding it with your fingertips.
Printer cover
Take out.
Cut using scissors.
to remove any remaining paper in
the printer.
3. Cut the fi rst (outermost) layer of a new paper roll and trim
the end to make it straight. A skewed or odd-shaped end
may cause the paper to jam.
4. Place new roll of thermal recording paper in paper holder so
that paper feeds from the bottom. Guide the trimmed end of the
thermal recording paper into and through the slot that receives
the recording paper. The thermal recording paper is automatically
wound forward and fed into the unit.
5. Close printer cover.
Replacing thermal recording paper
NOTE:
When paper is removed from printer,
error message E004 appears and a
beep sounds for 1 minute. The alarm
can be stopped by pressing
4.2 Replacing the thermal recording paper
AE-4022 4-13
3 Prepare a new roll of thermal recording paper
aCut the first (outermost) layer of a new thermal recording paper
roll, and trim the end to make it straight. A skewed or odd-shaped
end may cause the thermal recording paper to jam.
4 Set the new thermal recording paper roll
aPlace the new roll of thermal recording paper in the paper holder,
orienting it correctly so that the paper feeds from the bottom.
Guide the trimmed end of the thermal recording paper into and
through the slot that receives the recording paper.
The thermal recording paper is automatically wound forward and
fed into the unit.
NOTE:
When the thermal recording paper is completely removed
from the printer, the screen message shown in the right
figure appears and a continuous beep sounds for
approximately 1 minute. The alarm sound can be stopped
by pressing .
Paper holder Set the new roll of thermal
recording paper in the
paper holder in the proper
orientation so that the end
feeds from the bottom.
NOTE:
Always press the button after replacing the thermal
recording paper.If you do not fully feed the thermal
recording paper through the unit by pressing the button,
the printer will fail to print the measurement results.
REFERENCE:
If the printer fails to properly wind and feed the thermal recording paper, press and verify that the paper feeds
correctly.
.
4.2 Replacing the thermal recording paper
AE-4022 4-13
3 Prepare a new roll of thermal recording paper
aCut the first (outermost) layer of a new thermal recording paper
roll, and trim the end to make it straight. A skewed or odd-shaped
end may cause the thermal recording paper to jam.
4 Set the new thermal recording paper roll
aPlace the new roll of thermal recording paper in the paper holder,
orienting it correctly so that the paper feeds from the bottom.
Guide the trimmed end of the thermal recording paper into and
through the slot that receives the recording paper.
The thermal recording paper is automatically wound forward and
fed into the unit.
NOTE:
When the thermal recording paper is completely removed
from the printer, the screen message shown in the right
figure appears and a continuous beep sounds for
approximately 1 minute. The alarm sound can be stopped
by pressing .
Paper holder Set the new roll of thermal
recording paper in the
paper holder in the proper
orientation so that the end
feeds from the bottom.
NOTE:
Always press the button after replacing the thermal
recording paper.If you do not fully feed the thermal
recording paper through the unit by pressing the button,
the printer will fail to print the measurement results.
REFERENCE:
If the printer fails to properly wind and feed the thermal recording paper, press and verify that the paper feeds
correctly.
NOTE:
Always press the
Chapter 4 Maintenance
AE-4022 4-12
4.2 Replacing the thermal recording paper
When a red line appears on each side of the thermal recording paper, the paper will soon run out. Replace the depleted
paper roll with a new one before it runs out. A new roll of thermal recording paper can be used for approximately 450
measurements.
Items required: New roll of thermal recording paper, scissors
1 Cut the thermal recording paper
aCheck that the [Standby screen] is displayed.
bOpen the printer cover.
cIf some thermal recording paper still remains in the printer, cut
the remaining paper strip using scissors and remove the depleted
roll’s paper core.
If the paper has completely run out, remove the paper core from
the paper holder and go to step 3.
2 Remove the remaining thermal recording paper
aPress . The thermal recording paper remaining in the printer
is fed out. Remove the paper by holding it with your fingertips.
Printer cover
Take out.
Cut using scissors.
button after replacing the thermal recording
paper. If you do not fully feed the thermal recording paper through
the unit by pressing the
Chapter 4 Maintenance
AE-4022 4-12
4.2 Replacing the thermal recording paper
When a red line appears on each side of the thermal recording paper, the paper will soon run out. Replace the depleted
paper roll with a new one before it runs out. A new roll of thermal recording paper can be used for approximately 450
measurements.
Items required: New roll of thermal recording paper, scissors
1 Cut the thermal recording paper
aCheck that the [Standby screen] is displayed.
bOpen the printer cover.
cIf some thermal recording paper still remains in the printer, cut
the remaining paper strip using scissors and remove the depleted
roll’s paper core.
If the paper has completely run out, remove the paper core from
the paper holder and go to step 3.
2 Remove the remaining thermal recording paper
aPress . The thermal recording paper remaining in the printer
is fed out. Remove the paper by holding it with your fingertips.
Printer cover
Take out.
Cut using scissors.
button, the printer will fail to print the
measurement results.
4.2 Replacing the thermal recording paper
AE-4022 4-13
3 Prepare a new roll of thermal recording paper
aCut the first (outermost) layer of a new thermal recording paper
roll, and trim the end to make it straight. A skewed or odd-shaped
end may cause the thermal recording paper to jam.
4 Set the new thermal recording paper roll
aPlace the new roll of thermal recording paper in the paper holder,
orienting it correctly so that the paper feeds from the bottom.
Guide the trimmed end of the thermal recording paper into and
through the slot that receives the recording paper.
The thermal recording paper is automatically wound forward and
fed into the unit.
NOTE:
When the thermal recording paper is completely removed
from the printer, the screen message shown in the right
figure appears and a continuous beep sounds for
approximately 1 minute. The alarm sound can be stopped
by pressing .
Paper holder Set the new roll of thermal
recording paper in the
paper holder in the proper
orientation so that the end
feeds from the bottom.
NOTE:
Always press the button after replacing the thermal
recording paper.If you do not fully feed the thermal
recording paper through the unit by pressing the button,
the printer will fail to print the measurement results.
REFERENCE:
If the printer fails to properly wind and feed the thermal recording paper, press and verify that the paper feeds
correctly.
Correct Incorrect Incorrect
Paper Holder
4.2 Replacing the thermal recording paper
AE-4022 4-13
3 Prepare a new roll of thermal recording paper
aCut the first (outermost) layer of a new thermal recording paper
roll, and trim the end to make it straight. A skewed or odd-shaped
end may cause the thermal recording paper to jam.
4 Set the new thermal recording paper roll
aPlace the new roll of thermal recording paper in the paper holder,
orienting it correctly so that the paper feeds from the bottom.
Guide the trimmed end of the thermal recording paper into and
through the slot that receives the recording paper.
The thermal recording paper is automatically wound forward and
fed into the unit.
NOTE:
When the thermal recording paper is completely removed
from the printer, the screen message shown in the right
figure appears and a continuous beep sounds for
approximately 1 minute. The alarm sound can be stopped
by pressing .
Paper holder Set the new roll of thermal
recording paper in the
paper holder in the proper
orientation so that the end
feeds from the bottom.
NOTE:
Always press the button after replacing the thermal
recording paper.If you do not fully feed the thermal
recording paper through the unit by pressing the button,
the printer will fail to print the measurement results.
REFERENCE:
If the printer fails to properly wind and feed the thermal recording paper, press and verify that the paper feeds
correctly.
Set the new roll of thermal recording
paper in the paper holder in the
proper orientation so that the end
feeds from the bottom.
Other manuals for AUTION ELEVEN AE-4022
1
Table of contents
Other Arkray Measuring Instrument manuals

Arkray
Arkray Lab 004 User manual

Arkray
Arkray SPOTCHEM EZ Mounting instructions

Arkray
Arkray Lab 004 User manual

Arkray
Arkray AUTION ELEVEN AE-4022 User manual

Arkray
Arkray AUTION MAX AX-4030 User manual

Arkray
Arkray SPOTCHEM EZ SP-4430 User manual

Arkray
Arkray SPOTCHEM EL User manual

Arkray
Arkray PocketChem UA User manual
Popular Measuring Instrument manuals by other brands

Agilent Technologies
Agilent Technologies E8356A Installation note

Seres OL
Seres OL TOPAZ Total Hardness Operator's manual

Toptica
Toptica FemtoFiber pro Series manual

Autometers Systems
Autometers Systems Q-180 quick start guide

Extech Instruments
Extech Instruments EX845 user guide

Savio
Savio AE-01 user manual