Arkray SPOTCHEM EL User manual


d01 ■ SE-1520
Premise
This manual contains important information on the functions of the
SPOTCHEMELSE-1520
Thismanualisissuedby:ARKRAY,Inc.
Readcarefullypriortostartinguptheunit.
Itisrecommendedtoretainthismanualforfutureuse.
ThisproductconformstotheEMCStandardEN61326-2-6:2006.
Classofemission:CISPR11ClassA
ThisinstrumentisanIVDmedicalinstrument.
ThisproductconformstoEuropeanDirective98/79/EC
NOTE: This instrument has been tested and found to comply with the
limitsforaClassAdigitaldevice,pursuanttopart15oftheFCCRules.
These limits are designed to provide reasonable protection against
harmful interference when the instrument is operated in a commercial
environment. This instrument generates, uses, and can radiate radio
frequencyenergyand,ifnotinstalledandusedinaccordancewiththe
referencemanual,maycauseharmfulinterferencetoradio
communications.
Operation of this instrument in a residential area is likely to cause
harmfulinterferenceinwhichcasetheuserwillberequiredtocorrect
theinterferenceathisownexpense.
Theelectromagneticenvironmentshouldbeevaluatedpriorto
operation of the device. Do not use this device in close proximity to
sources of strong electromagnetic radiation, as these may interfere
withtheproperoperation.

eSE-1520 ■ 02
Introduction
TheSPOTCHEMELSE-1520isanelectrolytemeasuringsystem.
BeforeoperatingtheSE-1520,pleasereadthismanualcarefully.
This manual contains outlines, instructions for the operation, maintenance
andtroubleshootingfortheSPOTCHEMELSE-1520.
Followtheinstructionsinthismanualinordernottodefeatthepurposeof
protectivefeaturesoftheinstrument.
Pleaseretainthismanualforfuturereference.
Alwaysbecarefulwhenhandlingbloodsamplesorurine
samples.Incorrector impreciseproceduresmayresult in
exposuretopathogenicmicrobes.
This system must be operated only by those trained in proper
proceduresforclinicaltestingandhandlingofhazardouswaste.
Trainedpersonnelmustassistanyonewhooperatesthissystem
forthefirsttime.
If blood sample is spilt, the user has responsibility for carrying
outappropriatedecontamination.
Never touch the E-plate, pipetting tip or other places where
sample residue may adhere with bare hands. When performing
maintenance,alwayswearprotectiveglovestoprevent
exposuretopathogenicmicrobes.
Separateused samples, E-Plate,pipettingtips and cleaning
equipmentfromgeneralwasteanddisposeofthemaccordingto
localregulationsonbiohazardouswaste.
Thissystemmaybecomeinfectiousinthecourseofuse.
Discard the product in accordance with local regulations for
biohazardouswaste.
●
●
●
●
●
●
Allrightsreserved.Reproductionofthismanualisprohibited.
Thecontentsofthismanualaresubjecttochangewithoutpriornotice.
Althoughwetake all possible measurestoensurethe contents of thismanual,pleasenotify your
distributorwhenyouhavequestions,orfinderrorsoromission.
Beforeusing any cleaningor decontamination methodsexceptthose
recommendedbythemanufacturer,usersshouldcheckwiththe
manufacturerthattheproposedmethodwillnotdamagetheinstrument.

d03 ■ SE-1520
Notation
Thefollowingsymbolsareusedinthismanualandlabelsonthisinstrument
tocallyourattentiontospecificitems.
■Regardingdamageandperformanceofproducts
Toobtainaccurateresults,tofollowtheinstructions
givenherein.
Topreventdamagetotheinstrumentandassistyou
inmakingbestuseofthecapabilitiesofthe
instrument, additional explanations and notes are
providedherein.
IMPORTANT
NOTE
▲
Referenceinformationonoperation,additional
explanationsand relatedfunctions areprovided
herein.
■Regardingaccidentsresultingininjuryordeath
Topreventexposuretopathogenicmicrobes,followthe
instructionsgivenherein.
To prevent injury and property damage, follow the instructions
givenherein.

d
CautionLabels
1
Plate setting position
Do not touch the Plate setting position with bare hands. When placing
reagents or cleaning the area, wear protective gloves to prevent exposure
to pathogenic microbes.
Plate tray
The Plate tray holds reagent to which samples are adhered. When
discarding reagents or cleaning the case, wear protective gloves to
prevent exposure to pathogenic microbes.
2
Probe pin
Do not touch the probe pin with bare hands. When cleaning the probe pin,
wear protective gloves and use swabs to prevent exposure to pathogenic
microbes.
1
3
2
3
SE-1520 ■ 04

d05 ■ SE-1520
Chapter1
Introduction
Chapter2
Measurement
Electrolytemeasuringsystem,SE-1520OperatingManual
Contents
Premise…………………………………………………………………………………
Introduction…………………………………………………………………………
Notation………………………………………………………………………………
Contents…………………………………………………………………………………
1-1 Outline…………………………………………………………………………
1-1-1 Features……………………………………………………………
1-1-2 Measurement Types………………………………………
1-1- 3 Measurement Principles……………………………………
1-1-4 Specifications……………………………………………………
1-2 Shipping Carton………………………………………………………
1-2-1 Shipping Carton (Analyzer and Accessories)
1-3 Parts Description and Function………………………
1-3-1 Front of the Analyzer………………………………………
1-3- 2 OperatorPanel……………………………………………………
1-3- 3 BackoftheAnalyzer…………………………………………
1-3-4 TwinPipette……………………………………………………
1-3-5 Twin Pipette II……………………………………………………
1-4 Installation……………………………………………………………
1-4-1 Cautions…………………………………………………………
1-4- 2 Setting Up…………………………………………………………
1-4-3 First Operation after Installation…………………
1-4-4 Precautions in Instrument Relocation………………
2-1 Outline…………………………………………………………………………
2-1-1 OperatingProcedures……………………………………
2-1-2 Exclusive E-Plate……………………………………………
2-1-3 Measurement……………………………………………………
2-1-4 Calibration…………………………………………………………
2-2 Cautions………………………………………………………………………
2-2-1 Cautions on Operation………………………………………
2-2-2 Handling Samples……………………………………………
2-2-3 HandlingExclusiveE-Plates………………………………
2-2-4 HandlingMagneticCard………………………………
2-2- 5
Handling Twin Pipette and Twin Pipette II
……………
2-2-6 Handling Reference Solution……………………………
2-3 Preparation………………………………………………………………
2-3-1 Preparation……………………………………………………
2-3-2 Startup………………………………………………………………
2-3-3 Checks before Measurement……………………………
2-3-4 Preparation of Samples……………………………………
2-3-5 Suction of Reference Solution and Samples……
2-4 Measurement……………………………………………………………
01
02
03
04
1-02
1-02
1-03
1-03
1-04
1-05
1-05
1-06
1-06
1-07
1-08
1-09
1-10
1-11
1-11
1-13
1-15
1-16
2-02
2-02
2-03
2-03
2-04
2-05
2-05
2-06
2-06
2-07
2-07
2-09
2-10
2-10
2-11
2-12
2-14
2-15
2-18

d
Chapter3
SubMenu
Chapter4
Maintenance
Chapter5
Troubleshooting
Chapter6
Appendix
2-4-1 NormalMeasurement…………………………………………
2-5 Calibration…………………………………………………………………
2-5-1 Overview……………………………………………………………
2-5-2 MagneticCard Calibration……………………………
2-6 Measurement Result………………………………………………
2-6-1 PrintingtheNormalMeasurementResults………
3-1 Overview………………………………………………………………………
3-1-1 Contents of Each Menu……………………………………
3-2 MeasurementResultsMenu…………………………………
3-2-1 Printing Measurement Results…………………………
3-2-2 Transmitting Measurement Results…………………
3-2-3 DeletingMeasurementResults…………………………
3-2-4 Wildcards…………………………………………………………
3-3 Parameter Menu…………………………………………………………
3-3-1 PrintingParameters………………………………………
3-3-2 Entering Parameters………………………………………
3-3-3 Initializing Parameters……………………………………
3-4 Maintenance Menu…………………………………………………
3-4-1 Cleaning the Probe……………………………………………
3-4-2 CleaningtheTable………………………………………………
3-4-3 Check Measurement…………………………………………
3-5 Mode Menu…………………………………………………………………
3-5-1 Survey Mode………………………………………………………
3-6 Built-in Clock Adjustment …………………………………
4-1 Outline…………………………………………………………………………
4-1-1 FrequencyofMaintenance…………………………………
4-2 DailyMaintenance……………………………………………………
4-2-1 Cleaning the Plate Tray……………………………………
4-2-2 CleaningthePlateTransferPart………………………
4-3 PeriodicalMaintenance…………………………………………
4-3-1 Replacement of Thermal Printer Paper…………
4-3-2 Cleaning the Probe……………………………………………
4-3-3 Cleaning the Twin PipetteⅡ ……………………………
4-3-4 ReplacementofNozzleO-ring……………………………
5-1 Error Messages…………………………………………………………
5-2 TroubleMessages………………………………………………………
6-1 TransmissionSpecifications……………………………………
6-1-1 External Output Format………………………………………
6-1-2 Block Structure……………………………………………………
6-1-3 FormatofMeasurement Results…………………
6-2 After-sales Service……………………………………………………
2-18
2-23
2-23
2-23
2-25
2-25
3-02
3-02
3-04
3-04
3-06
3-08
3-09
3-10
3-10
3-12
3-14
3-16
3-16
3-16
3-16
3-17
3-17
3-18
4-02
4-02
4-03
4-03
4-03
4-04
4-04
4-06
4-08
4-09
5-02
5-05
6-02
6-02
6-03
6-04
6-05
SE-1520 ■ 06

d07 ■ SE-1520

dSE-1520 ■ 1-01
The SPOTCHEM EL SE-1520 is an electrolyte measurement
systemusingaexclusiveE-Plate.
Chapter1containsfunctionsandmeasurementprinciplesofSE-
1520.
Chapter1
Introduction
1-1 Outline
1-1-1 Features
1-1-2 MeasurementTypes
1-1-3 MeasurementPrinciples
1-1-4 Specifications
1-2 ShippingCarton
1-2-1 ShippingCarton(AnalyzerandAccessories)
1-3 PartsDescriptionandFunction
1-3-1 FrontoftheAnalyzer
1-3-2 OperatorPanel
1-3-3 BackoftheAnalyzer
1-3-4 TwinPipette
1-3-5 TwinPipetteII
1-4 Installation
1-4-1 Cautions
1-4-2 SettingUp
1-4-3 FirstOperationafterInstallation
1-4-4 PrecautionsinInstrumentRelocation

Simpleand fast settings in
anemergency
Compactdesign
Upto 50 measurement
resultscanbestored.
Hand-held barcode reader
canbeused.
Theion selectivityelectrodeenables
simultaneousmeasurementofbasic
threeitems(Na+,K+,andCl−).
By adopting a pipette support structure
for the analyzer, pipetting to the
disposable electrode (exclusive plate) is
easy.
Auto-start function starts measurement
justafter exclusive platesetting and
pipettingwiththetwinpipette.
ByusingthetwinpipetteII(tradename),
whichcontinuously distributes reference
solution,samplingoperationisnot
neededeachtime,someasurementsare
donemoreeasily.
The size is as small as that of an A5
sizedpaper.Thesmallunitcontains
various components such as a display,
printer, measurement devices and flash
memory.
A maximum of 50 measurement results
canbestored. Dataisautomatically
deletedfromtheoldestdatawhen
storeddataexceeds50samples.
Ahand-heldbarcodereadercanbeused
(optional).Each readbarcode datais
allocatedIDasameasurementresult.
1-02 ■ SE-1520
The SPOTCHEM EL SE-1520 is an electrolyte measurement
systemthatusesadisposableionselectivityelectrode.
Itissuitableforimmediatetestsperformedinclinicsandsmall
tolargehospitals.
●
●
●
●
●
●
●
1-1 Outline Chapter1 Introduction
1-1-1 Features

SE-1520 ■ 1-03
TheSPOTCHEM EL SE-1520adopts a potentiometrymethod ofion selective
membrane electrode as its principle, which provides measurement for various ion
concentrationsinbodyfluid.Theconstructionoftheionselectiveelectrodeusedfor
electrolytemeasurementasfollows.
1-1-3 MeasurementPrinciples
■Normal
Measurement
■Check
Measurement
A (normal) sample is measured. The measurement result
is identified by the measurement number indicated as
No.XXXX,whichisautomaticallyupdated.Whenpower
isturnedonmeasurementnumberissetatNo.0001
anditincreasessequentiallyuntilthepoweristurnedoff.
Exclusivecheckplate isusedforCheck Measurement,
itbe able to control theprecision and the resultis
output.Maintenanceofthedeviceisrequired,
dependingontheresult.
1-1 SP-4430の概要 Chapter1 Introduction
1-1-2 MeasurementTypes
Internal
electrode
Internal
electrode
Ion-selective
membrane
Ion-selective
membrane
Reference
solution Sample
Solution
junction
part
Referenceelectrode Bridge Measurementelectrode
Er Ej Es
The identical structure of ion selective electrode is used for both reference electrode and
measurement electrode. When reference solution is measured with the reference electrode,
electricpotentialofreferenceelectrodeEris obtained.Also,whenthedesiredionin
measurement solution is measured with the measurement electrode, electric potential of
measurementEsisobtained.
By setting up a solution junction part between the reference and measurement solutions,
measurementelectricpotentialEisobtainedbetweenthereferenceelectrodeand
measurement electrode. There is a relation of Nernst equation between generated electric
potential and ion activity (ion concentration). Ion concentration is obtained by measuring
potentialdifferenceE.
E=Es−Er+Ej
E= log(as)−log(ar)+Ej
E= log(as)+Eo
ZF
2.303RT
ZF
2.303RT
ZF
2.303RT
[note]
Es:Electricpotentialbyionactivityinsample
Er:Electricpotentialbyionactivityinreferencesolutionconstant)
Ej:Electricpotentialofsolutionjunction(constant)
Eo:Ej− log(ar)(constant)
as:Ionactivityinsample
ar:Ionactivityinreferencesolutionconstant)
( )

e1-04 ■ SE-1520
Sample
Measurementitem
Measurementprinciple
Measurementtime
Consumedsample
Samplesupply
Incubationtemperature
Display
Built-inprinter
Externaloutput
Transmissionmethod
Transmissionspeed
Memory
Measurementconditions
Environment during transpor
Storage environment
Powersupply
Powerconsumption
Dimensionsandweight
Locationofuse
Altitude
Pollutiondegree
Overvoltagecategory
Expectedlife
Wholeblood,Serum,Plasma,Urine
Na+,K+,Cl−
Potentiometricmethodwithionselectiveelectrodes
Approx.1minute
22μL
Manualpipettingbytwinpipette
32℃
LCD(20charactersx2lines)
36-characterthermalprinter(Paperwidth:58mm)
RS-232Cinterface(serial)
One-way/two-waytransmission
9600bps
50measurements
Temperature:10-30℃,Humidity:20-80%RH(Non-condensing)
Temperature:-10-60°C
Humidity:20-80%RH(Non-condensing)
Temperature1-30°C
Humidity:20-80%RH(Non-condensing)
AC100-240V(Mainpowersupplyvoltagevariationmust
bewithin±10%),50/60Hz
40VA
Dimensions:135(W)x225(D)x138(H)
Weight:Approx1.5kg(Analyzeronly)
Indooruseonly
2,000m
2
Ⅱ
5years(accordingtocompanydata)
1-1-4 Specifications
Chapter1 Introduction

b
1-2 ShippingCarton Chapter1 Introduction
1-2-1 ShippingCarton(AnalyzerandAccessories)
1・Analyzer(SE-1520)
4・Accessorybox 5・Operatingmanual 6・Pipettingtip
2・ACadapter 3・Powercord
Openthepackageandconfirmthatallitemsareincluded.
Description
SE-1520(SPOTCHEMEL)
Thisbooklet
100pcs.
No.
1
2
3
4
5
6
Item
Analyzer
ACadapter
Powercord
Accessorybox
Operatingmanual
Pipettingtip
Amount
1unit
1unit
1unit
1box
1copy
1box
SE-1520 ■ 1-05
Description
1sheet
5pcs.
No.
1
2
3
4
Item
Checkplate
Cottonbuds
Platetray
Thermalprinterpaper
Amount
1unit
1set
1unit
1roll
3・Platetray1・Checkplate 2・Cottonbuds 4・Thermalprinterpaper
Analyzer
Accessorybox

b1-06 ■ SE-1520
3・Built-inprinter
1・Display
2・Operatorpanel
5・Platetray
4・Platesettingposition
6・Pipetteinsertionpart
7・Magnetic-cardreader
Platetransferpart
Functions
Displaysinformation,suchasoperatingstateoftheunitanderrormessages.
StartorStopmeasurement,andentervariouskindsofvalues.
Thermal line printer for printing out measurement results and parameter
settings.
SettheexclusiveE-plate.
Usedplateisdiscarded.
Insertpositionforpipettingsamples
Insertmagneticcard,ifanewitemisaddedorlotcardisloaded
No.
1
2
3
4
5
6
7
Item
Display
Operatorpanel
Built-inprinter
Platesettingposition
Platetray
Pipetteinsertionpart
Magnetic-cardreader
1-3
PartsDescriptionandFunction
Chapter1 Introduction
1-3-1 FrontoftheAnalyzer

1-3 各部の名称とはたらき Chapter1 Introduction
1-3-2 OperatorPanel
Functions
Startmeasurement.SelectYes fromtheYes/Nooption.
StopmeasurementorCanceltheentry.SelectNo
fromtheYes/Nooption.
Feedpaperthebuilt-inprinterintothebuilt-inprinter
whilebeingpressed.
Switchthepageoneachmenudisplay.
Enteralphanumericcharactersandsymbols.
Selectitems,move thecursor,switch thepageson
thedisplay,andenterminussignanddecimalpoint.
Confirmtheentry.Checkmessagetomoveontothe
nextoperation.
Item
START
STOP
FEED
MENU
0〜9(numerickeys)
ー/.(hyphen/period)
ENTER
1-2 梱包品の確認
SE-1520 ■ 1-07

1-08 ■ SE-1520
1-4 装置の設置
1・Powerswitch
2・Terminal
3・RS-232C
4・CONT.
5・B.C.R.
6・Papercover
Functions
Turnon/offthepower.
ConnectwiththeattachedACadapter.
Connectwiththecableofexternaldevice.
Adjustthecontrastonthedisplay.Toenhancethecontrast,
turntheknobclockwise;toattenuateit,counterclockwise.
Connectwithhandybarcodereader.(optional)
Openthisparttoexchangethermalpaper.
No.
1
2
3
4
5
6
Item
Powerswitch
Terminal
RS-232C
CONT.
B.C.R.
Papercover
Chapter1 Introduction
1-3-3 BackoftheAnalyzer

1-3 各部の名称とはたらき Chapter1 Introduction
1-3-4 TwinPipette
1-2 梱包品の確認
SE-1520 ■ 1-09
1
2
3
4
Referencesolutionbutton
Samplebutton
Ejectbutton
Reflectionboard
Suctionanddischargeofreferencesolution
Suctionanddischargeofsample
Detachmentofpipettingtip
Informsthesystemoftheconditionofpipetteoperationintheeventofsampling
Item FunctionsNo.
1・Referencesolutionbutton 2・Samplebutton
3・Ejectbutton
4・Reflectionboard

1-10 ■ SE-1520
1-4 装置の設置 Chapter1 Introduction
1-3-5 TwinPipetteII
1
2
3
4
5
6
7
8
9
10
11
Referencesolutionbutton
Samplebutton
Ejectionbutton
Reflectionboard
Nozzlecaphook
Referencesolutioncylinder
Referencesolution
replacementmark
Nozzlecap
Referencesolutiondischargelever
Referencesolutionsuctionlever
NozzleO-ring
Suctionanddischargeofreferencesolution
Suctionanddischargeofsample
Detachmentofpipettetip
Informsthesystemoftheconditionofpipetteoperationintheeventofsampling
Removedcapscanbeinsertedandfixedhere.
Drawnupreferencesolutionentershere.
Marktoseetheresidualquantityofreference
solutioninthecylinder.Ifthesolationamount
doesntreachthemark,replacereferencesolation.
Protectionofreferencesolationnozzletipandpreventionofevaporationwhennotused.
Pushdowntheleverwhenreferencesolutioninthecylinderisdischarged.
Pushdowntheleverwhenreferencesolutioninthecylinderisdrawnup.
Fixtheattachedpipettingtip
No. Item Functions
1・Reference
solutionbutton
2・Sample
button
3・Ejection
button
5・Nozzlecap
hook
8・Nozzlecap
9・Reference
solution
discharge
lever
10・Reference
solution
suction
lever
6・Reference
solution
cylinder 7・Reference
solution
replacementmark
11・NozzleO-ring
4・Reflection
board

dSE-1520 ■ 1-11
Allowaspaceof
10cmormore
Allowa space of 10 cm or more
betweenthe rear of the analyzer
and the wall. Failure to do so may
cause overheating. Excessive load
onthecableconnectionmaycause
fire. Correct measurement results
maynotbeobtained.
Also,youwillhavetroubletryingto
turnoffthepowerswitchand
disconnectconnectorsintheevent
oferrorsortrouble.
Beforesettingupthe
analyzer,readthe
followingnotesand
alwaystakeproper
safetyprecautions.
●
AC100-120V
AC220-240V
50 / 60Hz
Operatethe analyzer with power of
thecorrectvoltageandfrequency.
Otherwiseitmaycausefireor give
somedamagetotheanalyzer.
●
DoNOTunnecessarilydisassemble
or modify the analyzer. Such
actions may cause danger of
exposure to pathogenic microbes,
fireordamage.
●Placethe analyzer on a stableand
levelsurfacefreefromvibration.
Failure to do so may damage the
analyzer,correctmeasurement
resultsmaynotbe obtained,or
causeinjury.
●
To avoid electric shock and/or fire,
use the attached power cord only
toconnectwithapoweroutlet.
●Connect the power plug directly to
anoutlet,notviaanextensioncord
or power tap. The power supply for
theanalyzeris40VA.
●
1-4 Installation
1-4-1 Cautions
Chapter1 Introduction

b1-12 ■ SE-1520
Do NOT set the analyzer where
chemicals are stored nearby, or where
corrosive gases or electrical noise are
generated.Theymaydamagethe
analyzer,andmalfunctionsand/or
injurymayoccur.Correctmeasurement
resultsmaynotbeobtained.
●Avoidexposingtheanalyzerto
humidity,aircontainingsulfur,direct
sunlightorwind,etc.
Otherwisecorrectmeasurement
resultsmaynotbeobtained,and
itmaycausedeformationor
malfunctionoftheanalyzer.
●
Toconnecttheanalyzertoexternal
devices,besuretouseproper
cablestoavoidelectricshock
and/orfire.Fordetails,contactyour
distributor.
●
Temperatures
10 〜 30℃
Humidity
20 〜 80%
Placetheanalyzerin aroomat
temperaturesbetween10℃ and
30 ℃ with humidity between 20%
and80%.Otherwise,correct
measurement results may not be
obtained.
●
1-3 各部の名称とはたらき
Beforesettingupthe
analyzer,readthe
followingnotesand
alwaystakeproper
safetyprecautions.
Chapter1 Introduction
Becarefulnottodroptheunitfrom
thetable.
●
This manual suits for next models
1
Table of contents
Other Arkray Measuring Instrument manuals

Arkray
Arkray PocketChem UA User manual

Arkray
Arkray AUTION ELEVEN AE-4022 User manual

Arkray
Arkray Lab 004 User manual

Arkray
Arkray Lab 004 User manual

Arkray
Arkray SPOTCHEM EZ SP-4430 User manual

Arkray
Arkray AUTION MAX AX-4030 User manual

Arkray
Arkray AUTION ELEVEN AE-4022 User manual

Arkray
Arkray SPOTCHEM EZ Mounting instructions