manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Ashcroft
  6. •
  7. Measuring Instrument
  8. •
  9. Ashcroft 1131 User manual

Ashcroft 1131 User manual

Installat on Instruct ons for
ASHCROFT®Model 1131
D fferent al Pressure Gauge/Sw tch
© Ashcroft Inc., 250 E. Main St., Stratford, CT 06614-5145, USA, 203-378-8281, 203-385-0357 Fax, www.ashcroft.com
All sales subject to standard terms and conditions of sale. I&M008-10088-12/01 07/13
For efficient working of your gauge, please read all instructions
carefully before attempting to install.
CAUTION: Do not exceed maximum operating pressure given
on the gauge label. Check fluid compatibility with wetted parts
before use.
OPERATING PRINCIPLE
High and ow pressures are separated by a sensor assembly
consisting of a magnet, diaphragm and a range spring. The
difference in pressure causes the sensor assembly to move in
proportion to the change against a range spring.
A rotary magnet, located in a separate body compartment and
isolated from the acting pressures is rotated by a magnetic
coupling following the linear movement of the sensor assembly.
A pointer attached to the rotary magnet indicates differential
pressure on the dial.
SWITCH: Reed switches (optional) are located adjacent to the
pressure chamber and are activated by the magnetic field of
the sensor assembly.
Note: The gauges are calibrated to give ±2% full scale accura-
cy on ascending readings.
INSTALLATION
For better performance the gauge should be mounted horizon-
tally with the dial vertical. Depressurize the system and con-
nect the high and low pressure lines of your system to the
“High” and “ ow”ports of the gauge, respectively.
It is recommended to use “O” rings with male connectors to
avoid excessive tightening and to prevent leakage (for parallel
threads). The gauge is now ready for operation. Apply “High”
and “ ow”pressures simultaneously, to avoid damage to the
internal parts.
Model 1131 consists of a rolling diaphragm to sense the pres-
sure difference. This model cannot see reverse pressure. It can
withstand maximum operating pressure up to 3000 psi for all
ranges. If pressure exceeds the rated maximum pressure, “O”
rings and diaphragm inside the pressure chamber, will be
damaged. If maximum operating pressure is within the allow-
able limit of 3000 psi, but the differential pressure exceeds
gauge range, there will be no damage to the gauge. The pointer
will only go to the extreme right end of the scale.
REED SWITCH
0
ROTARYMAGNET
POINTER
MAGNET
HIGH
20
10
30
ROLLING DIAPHRAGM
RANGESPRING
LOW
40
50
(OPTIONAL)
Model 1131
Roll ng D aphragm
PRECAUTIONS
Do not connect “High” and “ ow” ports to the wrong pipe ends.
Do not pressurize low side beyond 25 bar. Do not subject the
gauge to excessive vibration. The gauge is never to be used in
an area where a magnetic field is present. It may show the
wrong readings.
As the gauge works on a magnetic coupling principle, use only
non-magnetic fittings, parts, etc. in areas closer than 2˝ on all
sides, otherwise calibration will be affected. Panel mounted
gauges should be installed in non-ferrous panel material.
However, gauges with 2.5˝ dials, mounted in steel panels, may
require pointer adjustment. This is done by the manufacturer,
upon request.
Do not try to open any part of the gauge for any reason, be-
cause if not reassembled properly calibration will be affected.
SWITCH SETTING
Please follow these instructions when your differential pressure
gauges are supplied with a switch. The switches are normally
factory set to save the customer time. However, they are field
adjustable.
CAUTION: Supply voltage should not exceed switch rating.
For higher supply, use of relay circuit is recommended.
SWITCH ADJUSTMENT
The following procedure must be done by putting the gauge on
test bench or while in actual service. The switch adjustment
screw is located on the plastic cover. Rotate the screw counter
clockwise to increase the setpoint and clockwise to decrease
the setpoint. One or two trials may be necessary to attain the
exact setpoint.
Increase setpoint
counter clockwise
View from high pressure side
Load
L
(+)
N
(-)
N
(-)
R
RB
B
Switch #1Switch #2
Load
L
(+)
N
(-)
RB
Switch #1
6
6
5
5
4
4
3
3
2
2
1
1
6
6
5
5
4
4
3
3
2
2
1
1
Lor(+) Lor(+)
Load
Nor(-)Nor(-)
Nor(-)
1
3
2
E
B
R
E
1
3
2SPST
Switch
Load for
Load for
Switch #1
3
2
E
1
3
1
E
R
B
2
B
Switch #2
Switch #1Switch #2
View of plug after removing the socket View of plug after removing the socket
Installat on Instruct ons for
ASHCROFT®Model 1131
D fferent al Pressure Gauge/Sw tch
© Ashcroft Inc., 250 E. Main St., Stratford, CT 06614-5145, USA, 203-378-8281, 203-385-0357 Fax, www.ashcroft.com
All sales subject to standard terms and conditions of sale. I&M008-10088-12/01 07/13
Load
NC NO
1
3
2
E
N
(-)
L
(+)
N
(-)
Load
Load
NC
NO
B
RY
6
6
5
5
4
4
3
3
2
2
1
1
N
(-)
L
(+)
N
(-)
N
(-)
Load
Load
Load
Load
N
(-)
NC
NC
NO
NO
B
RYY
B
6
6
5
5
4
4
3
3
2
2
1
1
Lor(+)
Nor(-)
Nor(-)
Y
R
3
12
EB
SPDT
Switch
Switch #1Switch #1Switch #2
SPST SWITCH
Spec f cat ons
Contact Rating : 10 VA ac (rms) or dc (max)
Switching Current : 0.5 Amp ac (rms) or dc (max)
Switch Voltage : 100 Vac/Vdc (max)
SPDT SWITCH
Spec f cat ons
Contact Rating : 3 VA ac (rms) or dc (max)
Switching Current : 0.3 Amp ac (rms) or dc (max)
Switch Voltage : 30 Vac/Vdc (max)
ONE SPST SWITCH
Reed Sw tches and
Term nal Str p Connect on
TWO SPST SWITCHES
Reed Sw tches and
Term nal Str p Connect on
Reed Sw tches and
DIN Plug Connect on
View of socket for supply connections
Reed Sw tches and
DIN Plug Connect on
View of socket for supply connections
Reed Sw tches and
DIN Plug Connect on
View of socket for supply connections
ONE SPDT SWITCH
Reed Sw tches and
Term nal Str p Connect on
TWO SPDT SWITCHES
Reed Sw tches and
Term nal Str p Connect on
View of plug after removing the socket
R = Red; B = Black; Y = Yellow; = ive or + Supply; N = Neutral or – Supply. *Body to be suitably grounded while using gauge + switch and only switch.

Other Ashcroft Measuring Instrument manuals

Ashcroft a4a Manual

Ashcroft

Ashcroft a4a Manual

Ashcroft T5500 Technical specifications

Ashcroft

Ashcroft T5500 Technical specifications

Ashcroft F5509 User manual

Ashcroft

Ashcroft F5509 User manual

Ashcroft N5500 Instruction Manual

Ashcroft

Ashcroft N5500 Instruction Manual

Ashcroft t5500e Instruction Manual

Ashcroft

Ashcroft t5500e Instruction Manual

Ashcroft F5503 Instruction Manual

Ashcroft

Ashcroft F5503 Instruction Manual

Ashcroft DURALIFE 1009 Installation and operating instructions

Ashcroft

Ashcroft DURALIFE 1009 Installation and operating instructions

Ashcroft T5500 User manual

Ashcroft

Ashcroft T5500 User manual

Ashcroft TC User manual

Ashcroft

Ashcroft TC User manual

Ashcroft F5509 Instruction Manual

Ashcroft

Ashcroft F5509 Instruction Manual

Ashcroft P6500 Instruction Manual

Ashcroft

Ashcroft P6500 Instruction Manual

Ashcroft F5504 User manual

Ashcroft

Ashcroft F5504 User manual

Ashcroft A5400 Instruction Manual

Ashcroft

Ashcroft A5400 Instruction Manual

Ashcroft T5500 Instructions for use

Ashcroft

Ashcroft T5500 Instructions for use

Ashcroft F5510 User manual

Ashcroft

Ashcroft F5510 User manual

Ashcroft Digital Test Gauge User manual

Ashcroft

Ashcroft Digital Test Gauge User manual

Ashcroft 1132 User manual

Ashcroft

Ashcroft 1132 User manual

Ashcroft 28HPX User manual

Ashcroft

Ashcroft 28HPX User manual

Ashcroft 2074 User manual

Ashcroft

Ashcroft 2074 User manual

Ashcroft 50 HPS User manual

Ashcroft

Ashcroft 50 HPS User manual

Ashcroft 2032 User manual

Ashcroft

Ashcroft 2032 User manual

Ashcroft 8008 User manual

Ashcroft

Ashcroft 8008 User manual

Ashcroft A Instruction Manual

Ashcroft

Ashcroft A Instruction Manual

Ashcroft 2074 User manual

Ashcroft

Ashcroft 2074 User manual

Popular Measuring Instrument manuals by other brands

Young 86004 instructions

Young

Young 86004 instructions

Sealey TA203.V2 instructions

Sealey

Sealey TA203.V2 instructions

techem Compact IV c operating instructions

techem

techem Compact IV c operating instructions

ATAGO PAL-57S instruction manual

ATAGO

ATAGO PAL-57S instruction manual

Mohr CT100B Series Operator's manual

Mohr

Mohr CT100B Series Operator's manual

Honeywell AD1200 product manual

Honeywell

Honeywell AD1200 product manual

Medem AGDS-4 installation instructions

Medem

Medem AGDS-4 installation instructions

Sfere LNF96ZY user manual

Sfere

Sfere LNF96ZY user manual

Ingersoll-Rand ETA2 Operator's manual

Ingersoll-Rand

Ingersoll-Rand ETA2 Operator's manual

Refco Refmate quick guide

Refco

Refco Refmate quick guide

AISWEI AiCom Quick installation guide

AISWEI

AISWEI AiCom Quick installation guide

Rigaku R-XAS instruction manual

Rigaku

Rigaku R-XAS instruction manual

IFM SI0550 operating instructions

IFM

IFM SI0550 operating instructions

VEE GEE BTX-1 Operation manual

VEE GEE

VEE GEE BTX-1 Operation manual

Tenma 72-10465 operating manual

Tenma

Tenma 72-10465 operating manual

Fuji Electric NRF51 user manual

Fuji Electric

Fuji Electric NRF51 user manual

LOVATO ELECTRIC DMED100T1MID instruction manual

LOVATO ELECTRIC

LOVATO ELECTRIC DMED100T1MID instruction manual

PCB Piezotronics 354M56 Installation and operating manual

PCB Piezotronics

PCB Piezotronics 354M56 Installation and operating manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.