ATAGO 4553 User manual

Error Messages
The following messages alert the user when an operation has failed.
The battery is low.
The prism temperature is below the temperature range.
The ZERO button was pressed with nothing or something
other than water on the prism.
The prism temperature is above the temperature range.
The START button was pressed with nothing or an
insufficient amount of sample on the prism.
The instrument is faulty. (Replace the batteries. Contact
ATAGO if this error persists.)
The sample measured outside the measurement range.
Too much light is entering the prism, and the instrument
cannot measure accurately. (Shade the sample stage with
your hand and take a measurement again.)
Automatic Temperature Compensation
The readings are corrected, based on the temperature of the prism, within the automatic temperature compensation range.
【Caution】
◇Measurements may fluctuate with hot or cold samples. Wait for approximately 20 seconds to press the START button. Measurements will stabilize
once the instrument acclimates to the sample temperature.
Storage and Maintenance
Store the instrument in a dry place away from
direct sunlight. Exposure to humidity and heat
may damage the instrument.
Clean and dry the sample stage thoroughly, following the
“Cleaning”instructions.
Store the unit away from direct sunlight at a stable temperature
with as little fluctuation as possible.
Repair and Warranty
The instrument is warranted for one year from the date of purchase. This warranty is void if the instrument shows evidence of the following. Send the
included batteries as well if they are still in use.
Having been disassembled by unauthorized personnel
Damages to the prism and/or sample stage
Water damage or having been dropped
Having been misused and/or operated outside the environmental specifications
Leakage from batteries other than those included with the unit
Repair services are available for a fee after the warranty expires. Contact an ATAGO authorized service center for service and support.
Please have the serial number information ready when contacting a service center.
Specifications
Measurement range
Ethyl alcohol 0.0 to 53.0%(mL/100mL)
The 1:1 dilution : 0-100%
Maximum number of data
history
100
5.0 to 30.0℃
Output
NFC Forum Type 4 Tag
Resolution
0.1% (The 1:1 dilution 0.2%)/0.1℃
ISO/IEC 14443 Type A
Accuracy
±1.0%(The 1:1 dilution ±3%)/ ±1℃
Output category : Date Time,
Automatic temperature
5 to 30℃
Ethyl alcohol [mL/100mL], Temp [degC]
compensation range
(e.g.) 2017/08/17 09:30:45, 79.2, 20.4
Ambient temperature
10 to 30℃
Power supply
Two (2) AAA alkaline batteries
range
Battery life
Approx. 11,000 measurements
Sample volume
At least 0.3mL
(when using alkaline batteries)
Measurement time
Approx. 3 seconds
International
IP65
Backlight
The backlight stays on for 30 seconds
Protection class
after any button is pressed.
Dimensions and weight
55 (W) x 31 (D) x 109 (H)mm, 100g (main unit only)
Patent Granted in Japan, United States, Germany, China and Taiwan.
Headquarters: The Front Tower Shiba Koen,
23rd Floor 2-6-3 Shiba-koen, Minato-ku,
Tokyo 105-0011, Japan
http://www.atago.net/
TEL: 1-425-637-2107
TEL: 91-22-2854-4915 / 4071-3232
TEL: 66-21948727-9 ,66-21171549
TEL: 55 16 3913-8400
TEL: 39 02 36557267
TEL: 86-20-38108256
TEL: 7-812-777-96-96
TEL: 234-707-558-1552
TEL: 7-727-257-08-95
2004K Printed in Japan
4553-E01
Digital Hand-held “Pocket”Ethyl alcohol Refractometer
Instruction Manual
PAL-COVID-19 Sanitizer % Checker Cat.No.4553
Parts
Contents
Main unit ........ 1
Instruction Manual (this book)..... 1
Calibration Report...1
AAA batteries........... 2
Beaker 100mL...........1
Digital scale .. 1
AAA alkaline batteries are included. Remove the tape from the battery compartment before inserting the batteries.
ATAGO instruments are rigorously inspected to ensure each unit meets the highest standards of quality assurance.
Introduction
Thank you for purchasing the instrument. Carefully read and follow all instructions. Keep this manual for future reference.
Safety Instructions
Read and follow all safety instructions before operating the instrument. Failure to comply with the following instructions may result in personal injury or
property damage.
◇Ensure safety when handling hazardous materials. Observe precautionary measures and use protective equipment, Be aware of the hazards of such
chemicals and emergency response guidelines.
◇ATAGO may not be held liable for any injury or damage arising in connection with handling of hazardous materials during the use of the instrument.
◇Do not drop the instrument or subject it to strong physical shock.
◇Do not attempt to repair, modify, or disassemble the instrument.
◇Carefully read this manual to have basic knowledge of the function of each component.
◇ATAGO is not liable for any loss and damage caused by the measurement and use of this instrument.
◇Some acids may corrode the glass prism and/or metal sample stage, which may cause erroneous measurements.
◇Do not use metal tools, such as a spoon, as they may scratch the prism, resulting in erroneous measurements.
◇Do not use water above 50℃to rinse the instrument.
◇Only use the specified battery type. Observe proper polarities, properly aligning the anodes and cathodes.
◇Store the instrument away from direct sunlight/heat sources and excessive amounts of dust/debris.
◇Do not expose the instrument to a rapid change in ambient temperature.
◇Do not subject the instrument to strong vibration.
◇Do not subject the instrument to extreme cold temperature.
◇Do not place the instrument under anything heavy.
◇Loosen the battery compartment cover for air transportation.
<International Protection Classification IP65>
◇The instrument is water-resistant, not waterproof, and should not be submerged.
<Chemical Resistance of Body Case>
◇The body case is made of PBT resin. Do not expose it to water vapor. Some solvents may compromise the structural integrity of the instrument.
Press to take
measurements and
hold down to turn off
the display.
START button
(Power botton)
Place and remove
batteries from here.
Battery
compartment
ZERO button
Press to perform
zero-setting.
START button &
ZERO button
Press to set date, time,
and delete data history.
Lanyard hole
Sample stage
Apply water and
samples on the glass
prism located in the
center of the sample
stage.
Image is for explanation purposes only. It may be
different than the actual product purchased.
LCD
Measurement results,
prism temperature,
remaining battery
charge, etc., are
displayed.
The displayed value
is an example.

Zero-setting and Measurement
Zero-setting
【Caution】
◇Zero-set the instrument at the beginning of each day before use as well as after replacing the batteries.
◇Let water on the prism acclimate to the temperature of the instrument before zero-setting.
◇When “AAA”is displayed, wipe the prism clean, apply water, and press the ZERO button again.
1. Wipe the prism clean and
apply about 0.3mL of tap water.
2. Press the START
button, and after “---”
appears, the
measurement and the
prism temperature are
displayed.
3.
4. Zero-setting is
complete when “000”blinks
twice and then stops.
Measurement : not 0.0
Press the ZERO
button while the
water is on the
prism.
Measurement : 0.0
Proceed to measurement.
Wipe the prism clean with
tissue paper.
The instrument is zero-set
correctly.
Dilution
<Reference>
Ethyl alcohol 76.9-81.4vol%
The 1:1 dilution
Apply some diluted
sample
←Water、100mL
}99%Ethyl alcohol 80mL
Water 1mLEthyl alcohol 1mL
Gently stir.
←Water、100mL
}95%Ethyl alcohol 83mL
Measurement
1. Wipe the prism clean and
apply about 0.3mL of the sample.
2. Press the START
button
3. After “---”appears, the
measurement and the prism
temperature are displayed.
Displayed
=measured value×2
(dilution)
Offset
y=ax+b
Default:a=2 ,b=0
【Caution】
◇This instrument displays the concentration of ethanol in water as a percentage of total volume.
◇When a solution contains components other than ethanol and water, the readings may not correspond to alcohol by volume.
◇Do not use metal tools to apply samples on the prism as they may scratch the prism.
◇Initial measurements may fluctuate with hot or cold samples. Wait for the instrument to acclimate to the sample temperature,
approximately 20 seconds, to press the START button. Alternatively, press the START button multiple times until
measurements become stable.
◇Do not splash water above 50℃. The plastic may warp, which may compromise the water resistance.
◇The displayed temperature is that of the prism and may not necessarily match the temperature of the sample.
<LCD Auto Shut-off>
The instrument will turn itself off after 2 minutes of inactivity. To manually turn it off, hold down the START
button for more than 2 seconds.
<Option>
An anti-volatile cover, MAGIC, is available (sold separately) to prevent volatilization of the sample.
Part No. RE-39446 Apply some sample on the prism, then place the MAGIC so it fits tightly over the sample.
Press the START key whilst leaving the MAGIC on the sample.
Cleaning
【Caution】
◇Do not scratch the prism.
◇The instrument is water-resistant, not waterproof, and should not be submerged.
1. Wipe off the sample.
2. Clean the prism and sample stage using a mild soap and thoroughly rinse with water.
3. Dry the area with tissues thoroughly.
Offset Function
1.Hold down
ZERO while it
is turned on.
“b”wil appear.
2.Press ZERO
to select either
addition (b) or
subtraction (-b).
3.Press START
to confirm.
4.Enter the addition/subtraction number.
ZERO to change the number:
0, 1, 2, …, 8, 9, 10, 0, 1, 2…
START to confirm and move to the next
decimal place.
When the 1st place is confirmed with “10,”
the number selections for the decimal
places are skipped.
5.Press START
to confirm the
addition/subtracti
on number.
Next is to program
a coefficient.
(Factory default
2.00)
6.Enter the
coefficient.
Refer to the
step 4.
7.Press
START to
confirm the
coefficient.
※To disable the Offset feature, set the Offset value to factory default value
(a:1.00, b:0.00).
※The measurement range is shifted according to the offset settings.
※Screen images when offset is on
y = ax + b y:Actual concentration
x:PAL-102S readings
a:Coefficient (multiplication)
Range: 0.01 to 10.00(Factory default 2.00)
b:Addition/subtraction number
Range: -10.00 to 10.00(Factory default 0.00)
Replacing the Batteries
【Caution】
◇Fasten the battery compartment cover tightly to prevent water ingress or poor connection, which will cause erroneous
measurements. Push the cover in firmly and turn.
◇When the O-ring on the cover is dirty or damaged, the water resistance may be compromised. (Figure 1)
◇When the battery icon indicates the low power level ( ), replace both batteries with a brand new set of AAA alkaline
batteries (1.5V).
◇Static images may occasionally appear on LCD. Such retained pixel charges do not indicate a faulty display, consume the
battery power, or affect the instrument’s performance in any way.
◇Check the expiration dates on batteries before purchase.
◇Zero-set the instrument after the batteries are replaced.
1. Insert a coin in the
groove on the battery
compartment cover. Turn
the coin counterclockwise
to remove the cover.
2. Insert batteries,
observing the correct
polarity.
3. Align the cover and
push it down.
4. Close the battery compartment
cover by pushing the cover in with
a coin in the groove and turning it
clockwise until it stops.
*Turning excessively may cause
malfunction.
1 2 3
Figure1
1
O-ring
Groove
Approx.
2mm
Approx. 0.3mL of sample
Approx. 0.3mL of water
Approx.
2mm
For addition, only
“b”will appear.
1 2 3
Water after
zero-setting
Offset “b”=
addition of 0.30
During
measurement
For coefficient (a)
For addition/subtraction (b)
No plus sign will
be displayed.
Mild
soap
MAGI
C
20.4℃
79.2
Other ATAGO Measuring Instrument manuals

ATAGO
ATAGO PAL-36S User manual

ATAGO
ATAGO PRM-2000 A User manual

ATAGO
ATAGO RX-5000i Manual

ATAGO
ATAGO PAL-BXISALT User manual

ATAGO
ATAGO NAR-2T User manual

ATAGO
ATAGO RX-5000i User manual

ATAGO
ATAGO MASTER Series User manual

ATAGO
ATAGO PAL-98S User manual

ATAGO
ATAGO PAL-121S User manual

ATAGO
ATAGO PAL-04S User manual

ATAGO
ATAGO PAL-3 User manual

ATAGO
ATAGO 4430 User manual

ATAGO
ATAGO PAL-55S User manual

ATAGO
ATAGO PAL-Easy ACID14 User manual

ATAGO
ATAGO PAL-HIKARi 8 User manual

ATAGO
ATAGO 5468 User manual

ATAGO
ATAGO 4437 User manual

ATAGO
ATAGO PAL-14S User manual

ATAGO
ATAGO PAL-49S User manual

ATAGO
ATAGO Pocket PAL-1 User manual