6
• The ATMOS®Endoscopy and Stroboscopy System is
produced according to IEC 60601/EN 60601 and listed in
the following classes:
- VDE Class of protection I
- Class I Medical Devices Act (German law for medical
products)
• The device may only be connected to a properly installed
safety socket.
• The ATMOS®Endoscopy and Stroboscopy System is
designed in line with approved quality standards and
safety-related regulations. It should only be used in
accordance with regulations and under consideration of
the information stated in the operating instructions.
• Please do not perform any modications of the ATMOS®
Endoscopy and Stroboscopy System without approval
by the manufacturer. Please observe general rules for
accident avoidance as well as for environment protection.
In order to guarantee correct function, please only use
original spare parts ordered from the manufacturer.
• The ATMOS®Endoscopy and Stroboscopy System may
only be used by skilled staff who have been authorised
by ATMOS and trained in its operation (IEC 60601-1 / EN
60601-1).
• The mains voltage indicated on the type plate must
correspond to the values of the supply network.
• Make sure prior to every application of the equipment that
it is technically safe and in proper condition. Damaged
cables must be replaced immediately!
• Correct conguration in assembly of country-specic
connections:
- green/yellow: protective conductor (PE)
- blue: neutral conductor (N)
- black or brown: phase (L)
• Please observe that the mobile parts could cause injury
and contusion!
• The ambient conditions specied in the "Technical data"
(chapter 7.0) must be strictly observed!
• Switch off all function modules when you nish your daily
business.
• The ATMOS®Endoscopy and Stroboscopy System may
be operated only in rooms used for medical purposes, but
not in areas subject to explosion hazards.
• The ATMOS®Endoscopy and Stroboscopy System meets
the immunity to interference requirements of IEC 60601-
1-2 / EN 60601-1-2 „Electromagnetic Compatibility –
Medical Electrical Devices“.
• Please note that the insulation transformer has to be
installed in the base of the trolley if more than one
camera respectively stroboscopy module as well as
further consumer loads are connected.
• The ATMOS®Endoscopy and Stroboscopy System may
not be operated with devices not complying with the
requirements of standard EN 60601-1 „Medical Electrical
Equipment“ and EN 60601-1-2 „Electromagnetic
Compatibility“ (Medical Electrical Equipment).
• There are no warranty claims whatsoever on defects
which arise from the use of third party accessories or
consumables.
• ATMOS is not liable for personal injury and damage to
property if
- no original ATMOS parts are being used,
- the advice for use in these operating instructions is
not being observed,
- assembly, new settings, alterations, extensions
and repairs have been carried out by personnel not
authorised by ATMOS.
• Also pay attention to the safety information in following
chapters.
2.0 For your safety
!
For your safety