Atom IncuArch User manual

TO THE OPERATOR AND THE PERSON IN CHARGE OF
MAINTENANCE AND CARE OF THE UNIT:
zRead this Manual carefully before operating the unit.
zKeep this Manual where it is readily accessible for reference when needed.
zThis Manual describes operations of this unit and instructions on how to
operate the pulse oximeter and oxygen monitor. You may skip any section(s)
describing functions that are not included in your specific unit.
Operation Manual
Equipment for neonatal and premature infants: Incubator
0123
ORIGINAL
Lithium-ion battery type

Via Libia, 54 - 20081 Abbiategrasso (MI) - Italy
Tel: +39 02 9466329 Fax: +39 02 9081428
EU OFFICE

1
INTRODUCTION
Thank you for purchasing IncuArch.
This Operation Manual deals with the specifications, operation and maintenance of the IncuArch. Atom is by no
means responsible for any malfunction arising from a user ignoring the instructions for operation and mainte-
nance described in this Manual as well as any accident attributable to repair by someone other than technical
personnel belonging to or authorized by Atom.
Read this Manual carefully and familiarize yourself with its contents before operating the unit. Keep this Manual
where it is readily accessible for reference when needed. If any technical problems should arise, please contact
your local Atom representative.
The unit is shipped without being disinfected. Be sure to clean and disinfect the unit before using
it for the first time after purchase.
The product and its parts that are past their service lives should be disinfected before being dis-
posed of in accordance with the applicable waste management laws.
This product maintains the ambient temperature around an infant by drawing in the air outside
and warming it. It does not have the capability to cool the air drawn from outside. Be sure to set
the incubator air temperature at least 3°C higher than the ambient temperature. In particular, if a
phototherapy unit, a heated humidifier, or some other heat-generating device is used with the in-
cubator, set the incubator air temperature at least 5°C higher than the ambient temperature. If a
lower setting is selected, the incubator may not be controlled correctly.
This product is categorized as a portable medical device. It is intended for installation in an ambu-
lance or rotorcraft. Make sure to perform test operation when installing it to confirm that there is
no problem with the operation of the product or the vehicle where it is installed. For the device
characteristics, see "[15] Device Characteristics - EMC Level and Classification."
INTENDED USE
The IncuArch is a closed-care transport incubator for neonatal and premature infants and is intended to transport
low birth weight infants and seriously ill neonates.

2
SAFETY INFORMATION
Instructions to ensure the safe operation of the unit are found throughout this Manual.
Please read the Manual carefully before operating the unit. Please follow the instructions when operating the
unit.
Basic Instructions
1. Follow the instructions for the safe use of the unit.
Please follow the operating instructions described in this Manual to help ensure safe use of the unit.
2. Inspect the unit on a periodical basis.
It is necessary to carry out appropriate periodical inspections in order to maintain the unit in optimum
condition.
3. Never use the unit if it is found to be defective.
If any damage or malfunction of the unit should be noticed, stop using it immediately and contact your local
Atom representative.
4. Follow the EMC information given in this Manual.
Electrical equipment for medical use needs special precautions regarding EMC. It needs to be installed and
put into service according to the EMC information provided in this Manual.
Definitions of Warning Indication
Three levels of warning indication are used throughout this Manual and on the unit. They are defined as
follows.
A DANGER notice indicates an immediate hazardous situation which, if
not avoided, will result in death or serious injury, serious damage to property
such as total loss of use of equipment or fire.
A WARNING notice indicates an indirect (potentially) hazardous situation
which, if not avoided, will result in death or serious injury, serious damage to
property such as total loss of use of equipment or fire.
A CAUTION notice indicates a hazardous situation which, if not avoided, can
result in minor or moderate injuries, partial damage to equipment, and
loss of data stored in computers.

3
Definition of Symbols
1. Symbols to indicate danger, warning or caution
Symbol Title and indication
Caution:Withstand load
Indicates that the use exceeding the indicated maximum load capacity (withstand load) may
cause falling or damage of the unit.
General attention
Indicates unspecified general danger, warning or caution.
Caution: Hot surface
Indicates that the surface can be dangerously hot under certain conditions.
2. Symbols to prohibit action
Symbol Title and indication
General prohibition
Indicates unspecified general prohibition.
Prohibition of disassembly
Indicates prohibition of disassembly of the unit where it may cause an electric shock or other
hazards.
Prohibition of use of fire
Indicates prohibition of use of fire where an external use of fire may cause the unit to ignite un-
der certain conditions.
Prohibition of contact
Indicates that touching a certain part of the unit where it may cause injury is prohibited under
certain conditions.
3. Symbols to give instructions for action
Symbol Title and indication
General instruction
Indicates unspecified general action on the part of the user.
Connect a ground wire
Instructs the user to connect the ground wire without fail where the unit is provided with a
ground terminal.
Remove the power plug from the power outlet
Instructs the user to remove the power plug from the power outlet in the case of malfunction or
when there is a threat of lightning.

4
4. Symbols of international standards (IEC)
Symbols Title and indication
Type B applied part
Indicates that the device is classified as Type B in terms of the degree of protection against elec-
tric shock.
See Operation Manual
Follow Operation Manual.
Type BF applied part
Indicates that the device is classified as Type BF in terms of the degree of protection against
electric shock.
See Operation Manual
Follow operation manual.
Date of manufacture
Indicates the date when the unit was manufactured in the factory.
Lighting
Indicates a switch that controls lighting.
Power
Indicates a switch to turn the power on or off.
Electrostatic discharge
Indicates a caution regarding electrostatic discharges.
Recycle mark
Indicates that recycling is recommended.
WEEE symbol
In the EC area, an electrical and electronic product falling in one of the categories specified by
“DIRECTIVE 2002/96/EC OF THE EUROPEAN PARLIAMENT AND OF THE COUNCIL of 27
January 2003 on Waste Electrical and Electronic Equipment (WEEE)” should be disposed of in a
manner consistent with relevant laws and regulations.
This symbol indicates that the above-mentioned requirement applies to this product.
Manufacturer
This symbol indicates the name and the address shown adjacent to the symbol is of the manufacturer.
Authorised representative in the european community
This symbol indicates the name and the address shown adjacent to the symbol is of the authorised
representative in the European Community.

5
5. Other symbols
Symbol Title and indication
Setting
Indicates that a setting is increased.
Setting
Indicates that a setting is decreased.
AC Power
Indicates that the AC power is being supplied if an adjacent indicator is on.
Indicates that the AC power is not being supplied if an adjacent indicator is off.
Alarm
Indicates that an audible alarm is silenced.
Alarm
Indicates that a system of the unit is in an abnormal condition.
Alarm
Indicates that the rotation of the fan is abnormal or the fan has not been attached.
Alarm
Indicates an alarm related to temperature.
Alarm
Indicates an alarm related to a skin temperature probe.
Battery
Indicates the power level of battery 1 and 2.
24V12V
DC power connecting port
Indicates the DC power connecting port of the main body.
Alarm
Indicates excessive temperature condition.
Alarm
Indicates a low battery level.
Alarm
Indicates an alarm related to the SpO2patient cable. (For Masimo pulse oximeter)
Alarm
Indicates an alarm related to a failure of the SpO2patient cable. (For Masimo pulse oximeter)
Alarm
Indicates an alarm related to the SpO2sensor.
Alarm
Indicates an alarm related to a failure of the SpO2sensor.
Alarm
Indicates an alarm related to a failure of the SpO2sensor function.

6
Symbol Title and indication
(Red marking)
Alarm
Indicates the inability to detect pulse during measurement. (For Nellcor pulse oximeter)
Alarm
Indicates an alarm related to an oxygen sensor.
Sensitivity setting indicator
Indicates that the sensitivity mode is set to “Max”. (For Masimo pulse oximeter)
Sensitivity setting indicator
Indicates that the sensitivity mode is set to “APOD”. (For Masimo pulse oximeter)
Pulse search indicator
Indicates that a pulse is being detected. (For Masimo pulse oximeter)
Interference indicator
Indicates that interference is detected. (For Masimo pulse oximeter)
Fast response mode indicator
Indicates that the response mode of the pulse oximeter is set to “Fast.” (For Nellcor pulse oxim-
eter)
Pulse search indicator
Indicates that a pulse is being detected. (For Nellcor pulse oximeter)
Interference indicator
Indicates that interference is detected. (For Nellcor pulse oximeter)
10kg MAX. Load capacity
Indicates the maximum load capacity.
Gross weight
Indicates that the gross weight when peripheral devices of maximum loading capacity are at-
tached to the product is as inscripted.
Precautions on Jamming
The unit is intended for use in hospitals.
Portable and mobile communication equipment and other devices which generate high frequency
should be used at least 30cm away from the unit, including cables specified by the manufacturer.
Failure to do so may cause malfunction.
Note: The emission characteristics of this unit make it suitable for use in industrial areas or hospi-
tals (CISPR11 Class A). Take care when using the unit in residential environments (normally CIS-
PR11 Class B), as it may not provide adequate protection from wireless frequency communication
services. The user may also need to relocate or realign the unit to alleviate the problem.

7
Responsibility for Care of Equipment
The user (a hospital, a doctor’s office, a clinic) is responsible for the operation, maintenance and
care of the electrical equipment for medical use.
The equipment should be used only by medical personnel.
Prohibition of Modification
Do not disassemble or modify the unit.
Otherwise, a fire, electric shock or injury may result.
Periodical Inspection
Proper periodical inspection is needed to use the unit in the optimum condition.
In Case of Trouble
If any abnormal condition or trouble should occur to the unit, indicate on the unit that it is out of
order and contact your local Atom representative or service engineer immediately.
If any abnormal condition or trouble should occur, do not use the unit until it has been repaired
completely by a service engineer so as to prevent possible danger.

Table of Contents
INTRODUCTION........................................................... 1
PLEASE READ WITHOUT FAIL
[1] Safety Instructions.........................................10
1- 1. DANGER ................................................. 10
1-2. WARNING ............................................... 11
1-3. CAUTION ................................................ 14
[2] Parts Identification.........................................16
2-1. Main body ...................................................... 16
2-2. Operation panel ............................................ 18
2-3. Stands ............................................................ 19
PREPARATION
[3] Preparation before Use................................20
3-1. Assembly........................................................ 20
3-1-1. For HL Stand (Option) ........................ 20
3-1-2. For Cabinet Stand (Option)................ 20
3-2. Adjusting the HL Stand ................................ 21
3-3. Where to Install the Incubator..................... 21
3-4. Locking the Casters...................................... 22
3-4-1. For HL Stand (Option) ........................ 22
3-4-2. For Cabinet Stand (Option)................ 22
3-5. Securing the Unit to an Ambulance ........... 23
3-6. Inspections Before Start .............................. 23
[4] Power requirements ......................................24
4-1. Power Outlet and Grounding....................... 24
4-2. Using an AC Power Source
(AC100-240V)................................................ 25
4-3. Using the Built-in Battery Power Supply.... 27
4-4. Using an External DC Power Source
(DC12V/24V) ................................................. 31
4-4-1. Requirements for External
DC Power Source................................ 31
4-4-2. Connecting External DC Power
Source and Connection Cable .......... 31
OPERATION
[5] Display Screens ..............................................34
5-1. Overview of Display Screens...................... 34
5-2. Main screen ................................................... 35
5-3. Menu Screen ................................................. 39
[6] Setting the Incubator Temperature and
Placing the Infant in the Incubator.........41
6-1. Setting the Incubator Temperature............. 41
6-2. Placing the Infant in the Incubator ............. 43
6-3. Drawing out the Mattress Platform............. 46
6-4. Skin Temperature Monitoring ...................... 46
[7] Oxygen Supply ................................................49
7-1. Supplying Oxygen......................................... 50
7-2. Using the Oxygen Monitor........................... 51
7-2-1. Calibrating the Oxygen Monitor......... 51
7-2-2. Setting the Upper and Lower Alarm
Limits for Oxygen Level ...................... 53
[8] Pulse Oximeter ................................................55
8-1. Attaching the SpO2Sensor ......................... 55
8-2. Connecting the Patient Cable ..................... 56
8-3. Setting Alarm Limits...................................... 56
8-3-1. Setting SpO2Upper/Lower Alarm
Limits..................................................... 57
8-3-2. Setting Pulse Rate Upper/Lower
Alarm Limits ......................................... 58
[9] Other Operation Procedures .....................59
9-1. Working with the Menu Screen................... 59
9-1-1. Setting the Alarm Volume................... 60
9-1-2. Setting the LCD Screen Brightness.. 61
9-1-3. Setting the Display Size...................... 62
9-1-4. Setting the Language.......................... 63
9-1-5. Setting the Synchronizing Pulse
Beep Volume........................................ 64
9-1-6. Setting the Sensitivity Mode ............. 65
9-1-7. Setting the Averaging Time ................ 66
9-1-8. Setting the Pulse Oximeter Power
Frequency............................................. 67
9-1-9. Setting the Response Mode .............. 68
9-2. Feeding Cords and Tubes into or out of
the Incubator.................................................. 69
9-3. Lighting lamp ................................................. 69
9-4. Multi-purpose pole (option) ......................... 70
MAINTENANCE
[10] Cleaning and Disinfection..........................71
10-1. Hood Assembly ............................................. 72
10-2. Mattress Platform and Inside of
Incubator ........................................................ 74
10-3. Others............................................................. 76
[11] Maintenance and Inspection.....................77
11-1. Inspection Checklist ..................................... 77
11-1-1. Inspection Checklist - Before Use..... 78
11-1-2. Inspection Checklist - During Use .... 79
11-1-3. Inspection Checklist - After Use ........ 80
11-2. Periodic Replacement of Parts ................... 81
11-3. Replacing the Filter....................................... 82
11-4. Replacing the Oxygen Sensor.................... 83
11-5. Replacing the Batteries................................ 85
11-6. Expected Life................................................. 85
11-7. Disposal.......................................................... 85

PREPARATION OPERATION MAINTENANCE APPENDIX
PLEASE READ WITHOUT FAIL
[12] Alarms .................................................................86
[13] Troubleshooting..............................................89
APPENDIX
[14] Technical Data .................................................91
[15] EMC Level and Classification...................97

Safety Instructions
10
[1] Safety Instructions
Please follow the operating instructions described in this Manual to help ensure safe use of the unit. The unit
should be operated only by those who have received relevant training and instruction regarding its opera-
tion. The unit should only be operated according to its intended use.
1- 1. DANGER
Death or serious injury, damage to equipment or a fire will result if the instructions given below are not fol-
lowed.
Monitor the infant’s skin temperature when operating the unit.
Be sure to keep the admittance panels and snap-open access ports closed while the unit
is in use.
Using the unit with the admittance panel or snap-open access port left open may cause the infant to fall out of
the baby compartment. Be sure to close the admittance panels and the snap-open access ports when adminis-
tering phototherapy to an infant.
Do not leave the unit unattended when any of the admittance panels or snap-open access
ports are open.
The infant may be in danger of falling.
Stop using the unit immediately and request repairs if you find any abnormalities, such as
a misaligned admittance panel or snap-open access port, or a loose press bar.
The admittance panel or snap-open access port may unexpectedly open, causing the infant to fall out of the
baby compartment.
When providing phototherapy to an infant in the incubator, watch for any increase in the in-
cubator air temperature.
Since a mature infant emits a lot of heat, the incubator air temperature may rise if you place the infant in the in-
cubator and perform phototherapy. The incubator air temperature may also rise if you use multiple photothera-
py units at the same time, or if the room temperature is high. In such a case, proceed with the phototherapy af-
ter placing the infant in a cot or an open-type incubator, as directed by the doctor.
Do not overtighten the disposable safety bands.
Doing so may harm the infant.
Any type of ignition source, such as a hand warmer, must be kept away from the incubator.
Hand warmers or other items that may catch fire, or devices that generate sparks may cause an explosion or a
fire if used near the unit when oxygen is being supplied.
Do not use the unit in the presence of a flammable anesthetic gas.
Otherwise, a fire or an explosion may result.

PLEASE READ WITHOUT FAIL
Safety Instructions
11
Do not use ether, alcohol or any other flammable substances.
Even a small amount of ether, alcohol or other flammable substances may cause a fire when mixed with the oxy-
gen in the incubator.
Analyze arterial gas levels repeatedly when a high oxygen environment is required.
It has been reported that when setting the concentration of oxygen in the incubator to a high level, it is ex-
tremely important and essential to repeatedly measure the oxygen concentrations inside the incubator and ana-
lyze arterial blood gas in order to correctly maintain oxygen concentrations. Follow the doctor’s instructions
for measuring the oxygen concentration because ignoring essential requirements may increase the risk of reti-
nopathy of prematurity and other adverse effects.
1-2. WARNING
Death or serious injury, damage to equipment or a fire will result if the instructions given below are not fol-
lowed.
Be sure to follow the doctor’s instructions in setting the incubator air temperature.
The doctor is responsible for making a decision on transporting a premature infant. Watch
the infant continuously during transport.
Transporting a premature infant is associated with a risk of intracerebral hemorrhage due to vibration during
transport.
Smoking is prohibited in the room where the unit is installed. Do not place any potential ig-
nition sources in the room.
Follow the doctor’s instructions when selecting an appropriate method of oxygen adminis-
tration, an appropriate oxygen concentration, and an appropriate duration of administra-
tion.
Incorrect use of supplementary oxygen may cause serious side effects including loss of sight, damage to the
brain, and death. The risks of side effects vary with infants. The concentration of oxygen delivered to an incu-
bator must be followed by the doctor’s instructions.
Use oxygen for medical use only.
Be sure to bear in mind the following precautions when supplying oxygen.
zA hand warmer, flashlight, oils and fats, or combustible vaporized gases should not be placed in or near the
incubator.
zUse pure cotton for the infant’s clothing, bed sheets, etc. Do not use any material that is easily charged with
static electricity.
zBe sure that the clothing of doctors, nurses and ambulance attendants who handle this unit is made from
pure cotton or fire-proof materials.
Observe the following precautions when oxygen supply equipment is in use.
zA spontaneous and violent ignition may occur if oil, grease, or greasy substances come in contact with pres-
surized oxygen. These substances must be kept away from oxygen regulators, cylinder valves, tubing and
connections, and all other oxygen supply equipment.
zOn a high-pressure oxygen cylinder, only use tested pressure reducing or regulating valves marked for oxy-
gen service. Do not use these valves for air or gases other than oxygen. It is dangerous to use a valve to sup-
ply a gas other than air or oxygen and then to supply oxygen again.

Safety Instructions
12
Do not use any devices that generate high frequencies near the unit.
Using electric surgical knives, mobile phones, or other devices that generate high frequencies near the unit
may cause malfunctions due to jamming.
Ensure the unit is securely grounded when connecting the unit to a power source.
This can prevent electric shock due to current leakage. In order to complete the ground connection, be sure to
connect the power plug to a three-prong, correctly grounded outlet. Do not operate the unit if you have any
doubts about its ground connection.
Ensure the peripheral electrical equipment is securely grounded when connecting the unit
to a power source.
Take care of the usage environment.
Using the unit in a strong electromagnetic environment may disrupt accurate control over the incubator air
temperature or prevent accurate skin temperature from being displayed.
Do not use the unit next to other devices.
Using the unit next to or stacked with other devices may cause malfunction.
When it is unavoidable to do so, it is necessary to check that the unit and the other devices are operating cor-
rectly.
Avoid damaging the power cord.
A damaged power cord may cause a fire or electric shock. If a power cord is damaged, it should be immediately
replaced with a new one.
zDo not allow the power cord to get caught between the unit and wall, shelf or floor.
zDo not place the power cord near a heating apparatus or heat it.
zDo not place any heavy objects on the power cord.
zAlways grasp the power plug with your hand to remove the power cord from the power outlet.
Use only the power cord supplied with the unit.
Otherwise, a fire or an electric shock may result. A power cord or an AC adapter other than the one supplied
with the unit can generate high frequency noise and cause jamming to other electric devices for medical use
and thus result in malfunction, or make the unit more susceptible to high frequency noise emitted by other
electrical devices and thus result in malfunction.
Do not put multiple loads on one power outlet.
The power outlet should be located near the unit in order to prevent accidental contact with a trailing power
cord. Use a separate power outlet for each unit.
Avoid connecting the unit to a power source other than a appropriately rated power source.
The electrical rating of this unit is as follows:
Power voltage - AC100-240V; frequency - 50/60Hz; power consumption - 220VA; operating voltage range
-AC100-240V±10%. Do not connect the unit to any other power source.
Do not touch the power plug with wet hands.
Touching the power plug with wet hands may cause electric shock.
Do not use a battery that has reached the end of its service life.
The unit uses a battery that has a limited life span. If the battery is used past its life span of two years, not only
will it no longer function, but it may also short the electronic circuits inside or cause other damages to the bat-
tery, which may result in smoke or fire.
If the battery is subjected to a shock, it may also cause smoke or fire. If it happens, stop using or recharging
the battery, and contact your local Atom representative immediately.

PLEASE READ WITHOUT FAIL
Safety Instructions
13
If a fire should occur, extinguish it using a fire extinguisher for an electrical fire (Class C
fire), notify the fire department, and contact our sales representative immediately.
zRemove the power cord to stop the power supply.
zKeep the battery away from objects that may cause a secondary fire, such as cylinders and flammables.
Stop using the battery if problems such as damage or deterioration are found, and discard
the battery as described below:
zIn the case of medical institutions: Immediately request a specialized contractor to dispose of the batteries as
medical waste.
zIn the case of a company or public institution: Dispose of it as industrial waste.
zIn other cases such as where the unit belongs to an individual: Dispose of the battery according to the in-
struction of your local government.
If you are not sure how to dispose of the battery, contact your local Atom representative.
Charge the battery as instructed.
Failure to do so may cause the battery to leak, generate heat or smoke, rupture, or ignite.
If the battery is not fully charged after the specified charging time period has elapsed, stop
charging it.
Failure to do so may cause the battery to leak, generate heat or smoke, rupture, or ignite.
Do not charge the battery in locations such as gas stations where flammable gases (pro-
pane gas, gasoline, etc.) or fine particulates are produced.
Otherwise, gases may ignite.
Doing so may cause an explosion or fire.
If a leakage is suspected or you detect a strange odor, keep the battery away from fire.
Failure to do so may cause the battery to leak, generate heat or smoke, rupture, or ignite.
Do not directly touch electrolyte that has leaked from the battery cover.
If any electrolyte comes in contact with your eyes, you may lose your sight. If it comes in contact with your skin
or clothing, it may cause irritation. Flush it immediately with plenty of water and consult a physician.
Install the unit on a stable surface.
Installing it on an unstable surface or a slope may cause it to topple, fall, or move, resulting in injury or damage.
Therefore, before installing the unit, make sure that the place where the unit is to be installed is stable and
strong enough to support the weight of the unit.
Do not subject the oxygen sensor to impact.
If the oxygen sensor is damaged, the electrolyte inside may leak out of the sensor. If contact with the skin or
clothing occurs, rinse the area immediately with running water.
Do not touch the heaters during use or shortly after use.
The heaters are very hot during use and shortly after use. Allow the heaters to cool down sufficiently before
cleaning and disinfecting them in order to avoid injury.
Do not attempt to disassemble or modify the unit.
A fire, electric shock or injury may result.

Safety Instructions
14
The unit should only be serviced by qualified personnel in accordance with the appropriate
service manual.
Be sure to inspect the unit at the start of each day.
Operating the unit without inspecting it at the start of each day may result in a defect passing unnoticed and
cause a potentially unfavorable outcome.
If the unit fails to operate correctly, stop using it immediately.
Indicate on the incubator that it is out of order and contact your local Atom representative.
Do not reuse the disposable safety bands.
Change the disposable safety bands for every infant. Reusing them could lead to infection.
Do not stare into the lighting lamp or expose your eyes to the light.
Doing so may cause retinal damage.
1-3. CAUTION
Injury or damage to surrounding objects may result if the instructions given below are not followed.
Place the infant in the incubator only after the incubator air temperature has stabilized.
Prior to placing the infant in the incubator, pre-warm the incubator to keep the incubator air temperature stable.
Do not forcefully pull on or twist cords.
When feeding the cords and tubes into the unit, be very careful not to let them wind or
tighten around the infant.
Do not press strongly or rub on the LCD screen.
Doing so may result in damage or malfunction.
Do not allow any chemicals to adhere to the display.
If any chemicals do adhere to the display, wipe them off immediately.
Check that the peripheral devices operate correctly.
A device that transmits or receives faint signals may be affected by the electromagnetic waves generated by the
unit. If such a device is used in the vicinity of the unit, check the operation of that device for any detrimental ef-
fects before using the unit in clinical environments. Stop using the unit immediately if any problems occur.
Unplug the power plug before moving the unit to another location or when the unit is not
used for an extended period of time.
Moving the unit to another location with the power plug connected to a power outlet will damage the power cord
and may cause a fire or electric shock.
Do not apply impact to the unit or hit it on something.
Doing so may result in damage or malfunction.

PLEASE READ WITHOUT FAIL
Safety Instructions
15
Do not install the unit in excessively hot or humid places, excessively dusty environments,
or where it will be exposed to steam.
Installing the unit in such a place may cause a fire or electric shock.
Do not install the unit in direct sunlight or near a heating apparatus.
Keep the unit out of reach of small children.
Do not place any heavy objects on the unit.
Do not pour liquids directly onto the unit.
A fire or electric shock may result, or the unit may malfunction. If a large amount of liquid has spilled on the
unit, immediately wipe it up and call your local Atom representative to request checking of the internal elec-
tronic circuit.
Be sure to clean and disinfect the unit before using it for the first time after purchase.
The unit is shipped without being disinfected.
Remove the power plug from the power outlet before cleaning and disinfecting the unit.
Cleaning and disinfecting the unit with the power plug connected to the power outlet may cause electric shock.
Do not connect any equipment items other than those we specify.
The user must operate the unit at a position facing the operation panel.

Parts Identification
16
2-1. Main body
J Left side J Right side J Rear of the main body
No. Name
Admittance panel
Hood
Lighting lamp
Admittance panel latch
Tube introduction slit
Oxygen sensor module
Handle
Operation panel
Multi-purpose pole (option)
Oxygen supply port
No. Name
Snap-open access port
Oxygen sensor connection port
Skin temperature probe connection
port
Connector for SpO2sensor
AC inlet
DC power breaker (for 24V)
DC power breaker (for 12V)
External DC power inlet
Filter cover
[2] Parts Identification

PLEASE READ WITHOUT FAIL
Parts Identification
17
J Mattress Platform and Inside of Incubator
No. Name
Mattress
Disposable safety band
Mattress platform
Packing for main body
Middle board
No. Name
Fan (common to both V-707 and
V-808)
Heater
Battery 1
Battery 2 (option)
Battery cover

Parts Identification
18
2-2. Operation panel
No. Name Description
Display area Displays skin temperature, set temperature, SpO2, pulse rate, oxygen
concentration, etc.
Control switches Used to switch display screens or to specify various settings.
Light switch Used to turn the light ON or OFF or to change brightness.
Alarm silence switch Used to silence alarms.
Power switch Used to turn the power ON or OFF.
AC power indicator Lights up when the power cord is connected to an AC outlet.
Table of contents
Other Atom Medical Equipment manuals