BUFFALO FILTER® Smoke Evacuation Systems – Rev G
2.1 System Controls
The electronic system controls on the PlumeSafe
Smoke Evacuation System(s) are easy to
understand and simple to use. The membrane
control panel contains the power on/off switch,
suction power adjustment, remote activation
indicator, filter life reset button, filter life
indicator, and service indicator light. See Figure 1.
Note: Please be sure to read all instructions
before installing accessories or operating this
equipment. Failure to do so may result in
damage to the unit and/or personal injury.
•POWER ON/OFF & Standby
There are two ON/OFF switches on the PlumeSafe
Smoke Evacuation System(s). The ON/OFF
switch that completely disables the electrical
power to the unit is normally located on the back
or side of the unit. The ON/OFF switch on the
electronic membrane control panel is located in
the upper right hand corner of the membrane
panel. To power up the machine, place the power
switch on the rear or side of the machine in the
ON position and press the ON/OFF switch on the
membrane panel until the suction control lights
illuminate. Place the unit in “standby” by
pressing the ON/OFF membrane switch to the
OFF position. Turn the system main power off by
placing the power switch on the back or side of the
system in the OFF position.
•SUCTION CONTROL
(Membrane Control Panel)
The amount of suction may be adjusted by
pressing the UP and DOWN Arrow keys.
Pressing the UP arrow button increases suction
and pressing the DOWN arrow button decreases
suction. The suction control should be set at the
lowest practical setting to completely remove the
surgical smoke from the operative site. Each time
either arrow button is pressed, the suction will
increase or decrease by 10%. When two lights are
lit on the flow suction indicator, the actual air flow
is the value between the two illuminated
percentages.
•WHISPER TURBO BUTTON
The “TURBO” button is located just below the
lower left corner of the membrane keypad on the
front of the PlumeSafe Whisper TURBO system.
During standard operation, the amount of suction
may be adjusted by pressing the UP and DOWN
Arrow keys. Pressing the UP arrow button
increases suction and pressing the DOWN arrow
button decreases suction. The suction control
should be set at the lowest practical setting to
completely remove the surgical smoke from the
operative site. Each time either arrow button is
pressed, the suction will increase or decrease by
10%. When two lights are lit on the flow suction
indicator, the actual air flow is the value between
the two illuminated percentages.
When additional smoke removal is necessary
during any surgical procedure, the TURBO button
may be pressed at anytime. The TURBO button
will illuminate once pressed. TURBO mode
effectively doubles the airflow to quickly remove
smoke plume from the surgical site.
•FOOTSWITCH / REMOTE SWITCH
ACTIVATOR (Membrane Control Panel)
A footswitch or a Remote Switch activator may be
added to any system by simply plugging in a
BUFFALO FILTER activation accessory into the
appropriate jack on the back or side of the unit.
When the footswitch is plugged in, the unit may be
turned on or off by depressing the footswitch pedal
once for each operation. For footswitching
operation, the power switch on the membrane
control panel should be in the ON position. When
the electronic footswitch is plugged in, the light on
the membrane control panel entitled REMOTE
SWITCH will light to indicate that the electronic
footswitch is active. For directions on using the
Remote Switch Activator, please see instructions
that accompany that product.
The Whisper TURBO Smoke Evacuation System
also comes equipped with a pneumatic footswitch
which can easily connected by inserting the
pneumatic tubing that accompanies the footpedal
into the connector port. Upon insertion of the
Operating Instructions