ELECTRICAL SPECIFICATIONS/EQUIPMENT CLASSIFICATION
The Model IC-BAP-DL Bio Arterial Plus interior components are “double insulated” and do
not require a “protecve ground.” The system is equipped with an 18 gauge, 2-wire, 10.
Power cord, secured through the pump casing with a Heyco strain relief brushing as well as
an addional “hold-down” clamp for added safety.
1. Class of protecon against electrical shock: CLASS II EQUIPMENT
2. The degree of protecon against electric shock: APPLIED PART-TYPE B
3. Mode: CONTINUOUS OPERATION WITH INTERMITTENT LOADING
4. According degree of protecon against ingress of water: IXP0
EMC Manufacturer’s Declaraon
Model IC-BAP-DL electromagnec emissions—manufacturer’s declaraon
The model IC-BAP-DL is intended for use in the electromagnec environment specied below. The customer or the user of the model IC-BAP-DL should assure that
it is used in such an environment.
Emissions test Compliance Electromagnec environment— guidance
RF emissions
CISPR 11
Group 1 The model IC-BAP-DL uses RF energy only for its internal funcons. Therefore, its RF emissions are
very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The model IC-BAP-DL is suitable for use in all establishments, including domesc establishments and
those directly connected to the public low voltage power supply network that supplies building used
for domesc purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage uctuaons / icker
emissions
IEC 61000-3-3
Not applicable
Guidance and manufacturer’s declaraon—electromagnec immunity
The model IC-BAP-DL is intended for use in the electromagnec environment specied below. The customer or the user of the model IC-BAP-DL should assure that
it is used in such an environment.
Immunity test IEC 60601
test level Compliance level Electromagnec environment—guidance
Electrostac
discharge (ESD)
IEC 61000-4-2
± 2, 4, and6 kV contact
± 2, 4 and 8 kV air
± 2, 4, and6 kV contact
± 2, 4 and 8 kV air
Floors should be wood, concrete or ceramic
le. If oors are covered with synthec ma-
terial, the relave humidity should be at least
30%.
Electrostac
fast transient / burst
IEC 61000-4-4
± 2 kV for power supply lines
± 1 kV for input / output lines
± 2 kV
Not applicable
Mains power quality should be that of a
typical home use locaon.
Surge
IEC 61000-4-5
± 0.5 and 1 kV line(s) to line(s)
± 2 kV line(s) to earth
± 0.5 and 1 kV dierenal mode
Not applicable, no ground wire
Mains power quality should be that of a
typical home use locaon.
Voltage dips, short
interrupons and voltage
variaons on power supply
input lines
IEC 61000-4-11
<5 % UT (>95% dip in UT) for 0.5 cycles
40 % UT (60% dip in UT) for 6 cycles
70 % UT (30% dip in UT) for 30 cycles
<5 % UT (>95% dip in UT) for 5 s
<5 % UT (>95% dip in UT) for 0.5 cycles
40 % UT (60% dip in UT) for 6 cycles
70 % UT (30% dip in UT) for 30 cycles
<5 % UT (>95% dip in UT) for 5 s
Mains power quality should be that of a
typical home use locaon. If the user of the
model IC-BAP-DL requires connued opera-
on during mains power interrupons, it is
recommended that the model IC-BAP-DL be
powered from an uninterrupble power
supply or a baery.
Power frequency (50/60
Hz) magnec eld
IEC 61000-4-8
3 A/m 3 A/m The power frequency magnec elds should
be at the levels found in a typical home use
locaon.
NOTE: UT is the AC mains voltage prior to applicaon of the test level.