ATP Electronics APD Series User manual

APD: Bench scale with weight indicator which can be mo nted
attached to the platform or can be mobile.
APD_01_11.05_EN_U
User Manual
ATPMesstechnik GmbH
J. B. von Weiss Strasse 1
D- 77955 Ettenheim
Internet: www.atp-messtechnik.de
Tel: 0 7822-8624 0 - FAX: 0 7822-8624 40

APD series bench scale APD_01_11.05_EN_U
2
INDEX
1. INTRODUCTION.................................................................................................................................................................3
2. MAIN TECHNICAL SPECIFICATIONS...............................................................................................................................4
3. SYMBOLS...........................................................................................................................................................................4
4. INSTALLATION ..................................................................................................................................................................5
5. POWER SUPPLY AND START UP....................................................................................................................................7
5.1 POWER SUPPLY.........................................................................................................................................................7
5.2 TURN NG ON / OFF THE NSTRUMENT....................................................................................................................7
6. FRONT PANEL KEYS AND INDICATORS ........................................................................................................................8
7. SYMBOLS ON THE DISPLAY............................................................................................................................................9
8. BASIC FUNCTIONS .........................................................................................................................................................11
8.1 ZERO SCALE.............................................................................................................................................................11
8.2 TARE OPERAT ONS .................................................................................................................................................11
8.3 L M TAT ON OF THE TARE FUNCT ONS FOR D RECT SALE................................................................................12
8.4 AUTO POWER OFF FUNCT ON ...............................................................................................................................12
8.5 LOW BATTERY WARN NG .......................................................................................................................................12
8.6 MULT RANGE FUNCT ON NG (for legal for trade approved instruments) ...............................................................13
8.7 CONVERS ON OF THE UN T OF MEASURE ...........................................................................................................13
8.8 D SPLAY OF METR C DATA (inFO)..........................................................................................................................13
9. TOTALISATION................................................................................................................................................................14
9.1 OPERAT ONS OF TOTAL SAT ON ...........................................................................................................................14
9.2 REACT VAT ON OF THE TOTAL SAT ON................................................................................................................15
9.3 V SUAL ZAT ON OF THE TOTAL..............................................................................................................................15
9.4 ZERO NG OF THE TOTAL ........................................................................................................................................15
10. INSTRUMENT MESSAGES WHILE IN USE ..................................................................................................................16
DECLARATION OF CONFORMITY .....................................................................................................................................17
WARRANTY .........................................................................................................................................................................17

APD series bench scale APD_01_11.05_EN_U
3
1. INTRODUCTION
The purpose of this manual is to help the user get to know the scale’s various functioning modes, the keys’ functions and
the display indications. t is possible that one may incur into the phrase “TECH.MAN.REF.”: this means that an advanced
function is being described (therefore, for the technical personnel) and which is further explained in the corresponding
technical manual.
We advise to caref lly follow the instr ctions for programming the scale; by taking actions not indicated in this
man al, one co ld ca se the scale to not work properly.
n addition to having all the characteristics of a high precision scale, the instrument has the unit of measure / pounds
conversion function and the weighs totaliser.
The scale adapts itself to normal weighing applications in either industrial settings, such as during factory production
processes, or that of commerce, such as legal for trade applications.
This man al has been made as caref lly and exactly as possible; in any case, yo r s ggestions are always
welcome.
WARNING
Any attempt to repair or alter the unit can expose the user to the danger of electric shock and it will void our warranty. This
instrument is covered under warranty provided that IT HAS NOT BEEN OPENED BY THE USER for any reason. f any
problem with the unit or system has been experienced please notify the manufacturer or the dealer from which the
instrument was acquired.
The 6V rechargeable battery has to be completely recharged (12 ho rs) in the first installation of the instr ment;
we RECOMMEND disconnecting the battery if the instr ment is not going to be sed for more than 30 days.
In order to avoid the deterioration of the rechargeable battery:
In standard conditions:
- never leave the battery partially or completely ncharged; at least once a week recharge it completely.
In case the instr ment is not sed for a long period, one needs to:
- completely recharge the battery before the system is switched off for the last time;
- recharge completely every 3 months.
Do not po r liq ids on the indicator!
!!
!
Do not se solvents to clean the indicator!
!!
!
Do not expose instr ment to either direct s n light or any heat so rces!
!!
!
Always mo nt the indicator and platform in a vibration free setting!
!!
!
Read caref lly & apply what described in the POWER SUPPLY AND START-UP section!
Do not install in an environment with any risk of explosion!
!!
!
All the connections of the instr ment have to be made respecting the r les applicable in the zone
and in the installing environment.
Everything not expressly described in this man al has to be considered as improper se of the
eq ipment.
The crossed-out wheeled bin on the product means that at the product end of life, it must be taken to
separate collection or to the reseller when a new equivalent type of equipment is purchased. The
adequate differentiated refuse collection in having the product recycled, helps to avoid possible
negative effects on the environment and health and supports the recycling of the materials of which the
equipment is made. The unlawful disposal of the product by the user will entail fines foreseen by the
current regulations.

APD series bench scale APD_01_11.05_EN_U
4
2. MAIN TECHNICAL SPECIFICATIONS
POWER SUPPLY 6Vdc trough 110-240 Vac 50-60Hz adapter or rechargeable battery (6V – 4,5 Ah).
NOTE: recharge the battery using only the adapter provided.
MAX MUM POWER 5 VA
OPERAT NG TEMPERATURE From -10 to +40 °C (14 to 104 °F) (with even temperature).
D SPLAYED D V S ONS 10000e, 3X3000e for legal for trade use expandable to 800.000 for internal use
(with minim m signal coming from the 1,6mV/V cell).
MAX MUM NPUT S GNAL 6 mV/V.
M N MUM VOLTAGE PER D V S ON 0.3 µV (approved instrument); 0.03 µV (non approved instrument).
RESOLUT ON N CALCULAT ON 1'500'000 points (with signal in input equal to 3mV/V).
KEYBOARD Water resistant polycarbonate mechanic keys with tactile and acoustic feedback.
PROTECT VE CASE OF THE ABS console ( P 65 protection)
ND CATOR
BODY OF THE SCALE AND Stainless steel.
WE GH NG PLATE
TARE FUNCT ON Available on the entire capacity.
AUTO POWER OFF Programmable from 1 to 255 minutes, or disinserted.
LOW BATTERY WARN NG “Low Batt “ will appear on the display.
BATTERY RECHARGE T ME 12 hours.
LOAD CELL POWER SUPPLY 5Vdc ± 5%, 120mA (max 8 cells of 350 Ohms)
LOAD CELL CONNECT ONS 6 wires (CELL1) with Remote Sense, 4 wires (CELL 2, 3, 4) without Remote
Sense.
THE PARTS OF THE INSTRUMENT CONTAINING DANGEROUS ELECTRICAL TENSION ARE ISOLATED AND
INACCESSIBLE TO THE USER UNLESS IT HAS BEEN DAMAGED, OPENED, OR ALTERED.
3. SYMBOLS
To call the attention of the user, the following symbols are used both in the manual and on the instrument itself:
WARN NG! This operation must be performed only by qualified personal.
Conforms to the standards of the European Union.
dentifies the Class Of Precision defined by the O ML to represent 3000 divisions.
“TECH.MAN.REF.”
means that an advanced function is being described (therefore for the technical personnel) which
will be further explained in the corresponding technical manual.

APD series bench scale APD_01_11.05_EN_U
5
4. INSTALLATION
1) nput for the external power supply.
2) nput for the cable of the cell and battery power supply.
1
2
2

APD series bench scale APD_01_11.05_EN_U
6
The digital weight indicator has an ABS case whose external dimensions are represented in figures.
The indicator can be mounted attached to the platform or it can be installed in any desired position.
n the following figures, the dimensions of the scale and the plate are shown:

APD series bench scale APD_01_11.05_EN_U
7
5. POWER SUPPLY AND START UP
5.1 POWER SUPPLY
The instrument can be powered with a 6 Vdc tension supplied from an AC/DC external adapter which should be connected
to the 110-240 Vac mains voltage. Safety norms must be respected for the connection to the mains voltage including the
use of a line which has to be free from noise generated by other electronic equipment.
t is also possible to power the scale through the rechargeable built-in battery.
6V RECHARGEABLE BATTERY FEATURES
Material Lead
Power 4,5 Ah
Output 6 V
THE BATTERY MUST ONLY BE REPLACED WITH AN ORIGINAL FROM THE MANUFACTURER.
n order TO POWER the instrument through the 110-240Vac mains or TO RECHARGE the 6V battery, insert the plug end
of the AC/DC power adapter into the socket in the lower part of the indicator (chapter “ NSTALLAT ON”) and the adapter in
your main current source.
Do not connect other eq ipment to the same socket as the one that the adapter is in.
Do not step on or cr sh the power s pply cable.
5.2 TURNING ON / OFF THE INSTRUMENT
TO TURN ON the scale press the MODE/C key until the indicator turns on; then release.
The display shows:
XX.YY is the installed software version.
The instrument turns on all the display segments and symbols.
MAX XXX.XXX capacity of channel 1
bt XXX in which XXX is a number from 0 to 100 which indicates the battery level (if present).
The scale has an “auto zero at start-up” function: in other words it means that if at start-up a weight within +/- 10% of the
capacity is detected, it will be zeroed; if the weight is not within this tolerance, with a non approved instrument the display
shows the present weight after a few instants, while with an approved instrument “ZEro” is shown continuously on the
display, until the weight does not re-enter within this tolerance; the auto zero function at start-up may be disabled in the set-
up environment (only with non approved instrument); see SEt P >> ConFiG >> PArAM >> A to-0 parameter
(TECH.MAN.REF.).
By pressing the ZERO
key for an instant while the version is shown in the display, the indicator will show the following in
this order:
XX.YY in which XX indicates the instrument type, YY indicates the metrological software version.
XX.YY.ZZ is the installed software version.
APd is the name of the installed software.
bt XXX in which XXX is a number from 0 to 100 which indicates the battery level (if present).
-K- X.YY in which K identifies the type of keyboard: K=0 5-key keyboard.
X.YY is the installed software version.
After this, it shows the programmed capacity and the minimum division, “hi rES” (in case of NOT approved instrument) or
“LEGAL” (in case of approved instrument), the g gravity value, and finally it executes a countdown (self-check).
TO RESTART the scale keep the MODE/C key pressed for 8 seconds long: the indicator will turn off, then release the key
and the indicator will automatically turn on.
TO TURN OFF the scale keep the MODE/C key pressed until the “- OFF–“ message appears on the display; then release
the key.

APD series bench scale APD_01_11.05_EN_U
8
6. FRONT PANEL KEYS AND INDICATORS
The front panel of the indicator is designed for quick but simple weighing applications. t consists of a display with 6 digits,
25 mm in height and a 5 keys water-proof film keyboard.
While weighing various multifunction symbols indicating the functioning status will turn on (see the section “SYMBOLS ON
THE D SPLAY”).
SCALE KEY FUNCTION
ZERO
- Zeroes the displayed gross weight, if it is within +/- 2% of the total capacity.
- Cancels the value of the tare if the scale is unloaded
.
- When entering numbers it decreases the digit to be modified.
TARE
- f pressed for an instant it carries out the semiautomatic tare.
- f pressed at length it allows entering the manual tare from keyboard.
- Cancels the value of the tare if the scale is unloaded.
- n the numeric input phase it increases the digit to be modified.
M+
- t carries out the totalization if the conditions for this operation are satisfied.
- n the numeric input phase it selects the digit to be modified, from left to right.
MR
- f pressed for an instant, it carries out the display of the number of totalizations and the
present total.
- n the numeric input phase, it confirms the entry made.
- n the SET-UP, it allows to enter a step or to confirm a parameter within a step.
MODE/C
- t turns the scale on and off.
- f pressed during the display of the number of totalizations and present total, it clears the
total.
-
n the numeric input phase, it quickly zeros the present value.
- n the SET-UP, it allows to exit a step without confirming the change made.
- Allows viewing the scale’s metric information: capacity, division, minimum weigh for each
configured range.
Mc - Memory clear function: by pressing the MR and MODE/C keys in rapid succession the total is
cleared.

APD series bench scale APD_01_11.05_EN_U
9
7. SYMBOLS ON THE DISPLAY
The display has symbols which show the scale’s functioning status; you will find the description for each symbol below.
D SPLAY
NUMBER SYMBOL FUNCTION
(1)
0
The weight detected on the weighing system is near zero, within the interval of –1/4
÷ +1/4 of the division.
(2)
~ The weight is unstable.
(3) Not managed.
(4)
NET The displayed weight is a net weight.
(5)
G The displayed value is a gross weight, if the talian or English language is selected.
(6) B The displayed value is a gross weight, if the German, French or Spanish language is
selected.
(7) ndicates the battery charge level: see the section “LOW BATTERY WARN NG”.
(8) MAX=
MIN=
e=
When viewing the metric information, it identifies the indicated capacity range.
When viewing the metric information, it identifies the indicated minimum weigh range.
When viewing the metric information, it identifies the indicated division range.
(9) LT The locked tare is enabled.
(10) PT The manual tare is active.
(11) W1
W2
W3
The instrument is in the first weighing range.
The instrument is in the second weighing range.
The instrument is in the third weighing range.
(12)
Scale nr. 1 is always shown.

APD series bench scale APD_01_11.05_EN_U
10
(13) PCS Not used.
(14) kg ndicates the unit of measure in use (“kg” for kilogram, “g” gram).
(15)
%
Not used.
(16) t ndicates the unit of measure in use (tons).
(17) LB ndicates the unit of measure in use (pounds).
(18)
Not managed.
(19) * ndicates that a key has been pressed.
(20) PEAK Not used.
(21) HOLD Not used.
(22) SP1
SP2
SP3
SP4
Not used.
Not used.
Not used.
Not used.

APD series bench scale APD_01_11.05_EN_U
11
8. BASIC FUNCTIONS
8.1 ZERO SCALE
By pressing the ZERO key, it is possible to zero a gross weight value which is within +/- 2% of the capacity; after the
zeroing, the display shows 0 weight and the relative pilot lights are turned on.
8.2 TARE OPERATIONS
AUTOMATIC TARE
By loading any weight on the scale, the corresponding value is acquired as tare.
NOTE: The automatic tare will be acquired only if the weight is AT LEAST A D V S ON, STABLE (instability ~ led off) and
VAL D (in other words, the OVERLOAD condition must not be created).
SEMI-AUTOMATIC TARE
By pressing the TARE key any weight value present on the display is tared: the display shows “tArE” for an instant and then
0 (net weight); the pilot lights turn on.
NOTE: The semiautomatic tare will be acquired only if the weight is AT LEAST A D V S ON, STABLE (instability ~
led off)
and VAL D (in other words, the OVERLOAD condition must not be created).
ENTERING THE MANUAL TARE FROM KEYBOARD
Press TARE for a few seconds: the display shows “– tM –“ and then "000000". Enter the desired value using the following
keys:
ZERO decreases the blinking digit.
TARE increases the blinking digit.
M+ selects the digit to be modified (blinking); the scrolling of the digits takes place from left to right.
MODE/C if pressed for an instant it quickly zeros the present value; if pressed at length it allows to return to weighing
without saving the changes made.
Confirm with the MR key; the value will be subtracted from the weight present on the plate and the relative pilot lights will
turn on.
NOTE: f the entered value is not a multiple of the scale’s minimum division, it will be rounded off.
CANCELLING A TARE
One can manually cancel the tare value in different ways:
- unload the scale and press the TARE or ZERO key.
- carry out the tares in deduction, partially unloading the scale and pressing TARE to zero the display.
- enter a manual tare equal to 0.
NOTE: it is possible to a tomatically cancel the tare val e; see the following section.
LOCKED/UNLOCKED/AUTO/DISABLED TARE SELECTION
Normally, when a tare value is entered (semi-automatic or manual), by unloading the scale plate, the display shows the tare
value with a negative sign (LOCKED TARE). For one’s convenience it is also possible to choose that the tare value cancels
itself automatically each time that the scale is unloaded (UNLOCKED TARE or AUTO TARE), or disable the tare functions.
With the UNLOCKED tare:
In case of SEMIAUTOMATIC TARE the net weight, before nloading the scale, may also be 0.
In case of MANUAL TARE the net weight before nloading the scale m st be greater than 2 divisions and stable.
To set the type of tare:
-Turn on the scale, press the TARE while the firmware version is displayed (the display shows the “F.ModE” menu).
-Press MR to enter the menu.
-Press ZERO many times (to scroll ahead through the parameters) or TARE (to scroll backwards) until one finds the
“tArE” parameter.
-Press MR to enter the parameter.
-With the ZERO or TARE keys select the possible options: “LoCK” (locked tare), “unLoCK” (unlocked tare), “Auto”
(automatic tare), “diSAb” (disabled tare).
-Confirm with MR.

APD series bench scale APD_01_11.05_EN_U
12
-Press the MODE/C key many times until the display shows the message “SAVE?”.
-Press MR to confirm the changes made or another key for not saving.
The instrument stores the last selection made, also after it is turned off.
8.3 LIMITATION OF THE TARE FUNCTIONS FOR DIRECT SALE
With approved instrument, it is possible to limit the tare functions, by setting “yES” in the step SEt P >> d.SALE
(TECH.MAN.REF.). n this case, all the tare operations will be disabled.
With approved instrument, the d.SALE step is read-only.
8.4 AUTO POWER OFF FUNCTION
t is possible to automatically turn off the scale (from 1 to 255 minutes), or disable it; the auto power off takes place when,
with nloaded scale, the weight has not been moved or a key has not been pressed for the time set: the display shows the
“- OFF – “ blinking message and an acoustic signal is emitted; after this the scale turns off.
For the setting, follow the procedures below:
- Turn on the scale, press TARE while the firmware version is displayed (the display shows the “F.ModE” menu).
-Press MR to enter the menu.
-Press ZERO many times (to scroll ahead through the parameters) or TARE (to scroll backwards) until one finds the
“En.SAVE” parameter.
-Press MR to enter the menu
-Press ZERO many times (to scroll ahead through the parameters) or TARE (to scroll backwards) until one finds the
“AutoFF” parameter.
-Press MR to enter the parameter.
-With the ZERO or TARE keys select the possible options: “diSAb” (auto switch-off disabled), “EnAb” (auto switch-off
enabled).
-Confirm with MR; if “EnAb” has been selected, one will be asked to enter the number of minutes after which the scale
should turn off: enter a number between 1 and 255 (using the M+ key to select the digit to be modified and ZERO/TARE
to decrease/increase it) and confirm with MR.
-Press many times the MODE/C key until the display shows “SAVE?”.
-Press MR to confirm the changes made or another key for not saving.
8.5 LOW BATTERY WARNING
One should set the supply mode by setting the step SEt P >> ALiM.
The charge level is shown in the weighing phase by the battery symbol:
-
:
battery is charged.
- : battery is partially charged.
- : battery is discharged: connect the scale to the mains for recharge the battery or replace it.
Furthermore, for a few seconds the "Low.bat " message appears on the display (minimum level voltage).
When recharging the battery, the display shows the recharging phase below:
RECHARGING PHASE: …
RECHARGE IS COMPLETED:
NOTES:
- While recharging, the instrument can be used as usual.
- The instrument automatically turns off when the voltage goes below the minimum level.
- t’s possible to view the recharge percentile of the battery by pressing the ZERO key upon start-up (see the section
“POWER SUPPLY AND START-UP”).

APD series bench scale APD_01_11.05_EN_U
13
8.6 MULTI RANGE FUNCTIONING (for legal for trade approved instr ments)
The multi range functioning allows to subdivide the scale capacity in two or three ranges, each which is up to 3000
divisions, improving in this way the first range division in the dual range and the first two ranges in the triple range.
For example, with a 30 kg cell platform it is possible to approve the weighing system with:
-A single range: 6 kg capacity and 2 g division (3000 div.).
-Dual range: 6 / 3 kg capacity and 2/1 g division (3000 + 3000 div.).
-Triple range: 15 / 6 / 3 kg capacity and 5/2/1 g division (3000 + 3000 + 3000 div.).
NOTES:
- For the approval of the weighing system in dual and triple range the cell must have better technical features in comparison
to the cell used for the approval in a single range.
The multirange functioning is shown by the turning on of the relative symbol on the display which identifies the range in
which one is operating; by passing to the second range, the second range division is enabled; by passing to the third range,
the third range division is enabled. At this point the first range division is restored only by passing by the gross zero of
the scale.
- The selection of the range number with multirange functioning is made during the scale’s calibration (TECH.MAN.REF.).
8.7 CONVERSION OF THE UNIT OF MEASURE
t is possible to execute the weight conversion from the unit of measure of the scale to pounds and vice versa, by pressing
the MODE/C key.
To enable or disable this function, follow the procedure below:
-Turn on the scale, press the TARE key while the firmware version is displayed (the display shows the “F.ModE” step).
-Press MR to enter the step.
-Press ZERO many times (to scroll forwards through the parameters) or TARE (to scroll backwards) until one finds the
“un.ConV” step.
-Press MR to enter the parameter.
-With the ZERO or TARE keys select the possible options: “diSAbL” (disables the unit conversion), “EnAbLE” (enables
the unit conversion).
-Confirm with MR.
-Press the MODE/C key many times until the message “SAVE?” is shown on the display.
-Press MR to confirm the changes made or another key to not save.
NOTES:
-The present total will be cleared when the conversion takes place.
-The conversion causes the state of instability of the weight.
-The conversion can take place for any unit of measure set during the calibration.
-With approved instrument the weight in pounds is displayed for 5 seconds, after which the display returns to the scale
unit of measure.
8.8 DISPLAY OF METRIC DATA (inFO)
The scale is fitted with a function named “ NFO”, thanks to which it is possible to view the configuration metric data:
-Keep the MODE/C key pressed until the display shows “inFo”, and release.
-The capacity value of the first range will appear.
-Press the ZERO key to scroll the following data, in this order:
Capacity 1° range Minimum weigh 1° range Division 1° range
Capacity 2° range Minimum weigh 2° range Division 2° range
Capacity 3° range Minimum weigh 3° range Division 3° range
Capacity 1° range ………
-Press the TARE key to scroll backwards the metric data.
-Press the MODE/C key to return to weighing.
NOTES:
-The minimum weigh corresponds to 20 net weight divisions.
-The data of the second and third range appear only if actually configured.

APD series bench scale APD_01_11.05_EN_U
14
9. TOTALISATION
nside the F.ModE >> totAL step of the setup environment it is possible to set the parameters for the function of
totalisation:
tot.Mod: TYPE OF TOTAL ZAT ON
t is possible to select the type of totalisation: manual (MAnuAL) or automatic (Auto).
n.WEiGh: NUMBER OF WE GHS
f one sets this parameter as “yES”, for each operation of totalisation the number of weighs and the total weight are
displayed. f the total weight has more than 6 digits, its visualisation takes place in two stages.
f one sets the parameter as “no”, for each operation of totalisation only the "-tot-" message is displayed.
WAit.St: WA T STAB L TY
Once one has set the wait stability time with a value different from 0, each operation of totalisation will be executed only
if the weight is in stability. When the weight is not stable and M+ key is pressed, the unstable weight is displayed blinking
and one is not allowed to press any key; the totalisation will remain pending. f the wait stability time is expired and the
weight is not stable, the present accumulation will be aborted with the display of the “unStAb” message and error beeping.
f one sets this parameter with the value 255, the waiting stability time is infinite on each accumulation.
Once one has set the wait stability time at 0, if one press the M+ key and the weight is not in stability, the operation of
totalisation is instantly aborted with the display of the “unStAb” message and error beeping.
NOTE: This parameter takes effect only if the manual totalisation has been set.
MAx.tot: NUMBER OF CONSECUT VE TOTAL SAT ONS AFTER WH CH THE TOTAL S AUTOMAT CALLY CLEARED
After having carried out the number of totalisations set in this step, the accumulated total is shown on the display and
cleared; one can set a value between 0 and 255.
NOTE: the value 0 disables the function.
9.1 OPERATIONS OF TOTALISATION
This scale allows to totalize the weight in load and unload.
Totalisation in load:
f the manual totalisation has been set, in order to carry out the totalisation it is necessary to load the weight on the scale
and press the M+ key; then the weight is accumulated in the total.
f the automatic totalisation has been set, there is the automatic acquisition of the stable weight loaded on the scale and
then the accumulation of the weight in the total.
To totalize, the net weight must be:
- at least 1 division with not approved instrument and manual totalisation;
- at least 10 divisions with not approved instrument and automatic totalisation;
- at least 20 divisions with approved instrument.
Totalisation in nload:
f a tare is active, one can totalize also negative values of weight: after having executed the tare, by removing part of the
weight from the scale, the negative value of the unloaded weight is displayed and it is possible to totalize this value. The
totalisation is executed by pressing the M+ key if the manual totalisation has been set or automatically when the weight is
stable if the automatic totalisation has been selected.
To totalize, the negative net weight must be:
- at least 1 division with not approved instrument and manual totalisation;
- at least 10 divisions with not approved instrument and automatic totalisation;
- at least 1 division with approved instrument and manual totalisation;
- at least 20 divisions with approved instrument and automatic totalisation.
Note: to totalize the negative net weight, with not approved instrument the gross weight must be greater than or equal to
zero; with approved instrument the gross weight must be at least 20 divisions.

APD series bench scale APD_01_11.05_EN_U
15
9.2 REACTIVATION OF THE TOTALISATION
To avoid undesired totalisations, the M+ key is active just once; its reactivation depends on the setting of the “rEACt”
parameter (TECH.MAN.REF.) in the setup environment. t can be reactivated after the passing of the weight by the zero of
the scale, by instability of the weight or always.
f the reactivation on instability has been set, the totalisation is reactivated by instability of the weight with manual
totalisation or by instability and change in weight of 10 or 20 divisions (respectively with not approved or approved
instrument) with automatic totalisation.
By pressing the M+ key again, witho t having reenabled the totalisation, the "no.0.unS" error message is displayed and
an acoustic signal is emitted.
NOTES:
-f the gross or net weight in absolute value is less than the requested minimum weight, by pressing the M+ key the
display shows the “LoW” error message.
-f the scale is in the under load or over load status, by pressing the M+ key the display shows the “un.oVEr” error
message.
9.3 VISUALIZATION OF THE TOTAL
By pressing the MR key one can temporarily view on the display the number of weighs and the net weight total. f the total
weight has more than 6 digits, the visualisation of this value takes place in two stages.
9.4 ZEROING OF THE TOTAL
The instrument increases the total weight upon each totalisation; it can be cleared in the following modes:
- by executing the Mc (memory clear) function: press in rapid succession the MR and MODE keys (the “tot.Clr” message is
displayed);
- by executing the conversion of the unit of measure;
- by turning off the instrument;
- automatically after having carried out the number of totalisations set in the <<MAX.tot>> step.

APD series bench scale APD_01_11.05_EN_U
16
10. INSTRUMENT MESSAGES WHILE IN USE
MESSAGE
DESCRIPTION
ZEro The scale is zeroing the weight.
nStAb One is trying to totalize with an unstable weight.
n.oVEr One is trying to totalize with the weight in underload or in overload, in other words, with a weight of 9 divisions
greater than the capacity or of 100 divisions below the gross zero.
LoW Weight less than the minimum weight provided for the totalisation or the transmission of the string, standard
or extended, upon pressing of the print key.
no.0. nS Weight not passed by net 0 or by instability.
ErMot Unstable weight during the calibration.
PrEC. t is displayed if one tries to calibrate the zero point without first having confirmed the number of calibration
points.
ErPnt During the acquisition of a calibration point a null value has been read by the converter.
Er – 11 Calibration error: a too small sample weight has been used; it is advisable to use a weight equal to at least
half of the scale capacity.
Er – 12 Calibration error: the acquired calibration point (tP1 o tP2 o tP3) is equal to the zero point (tP0).
Er – 37 The number of converter points per scale division is less than two. Carry out again the calibration with special
attention to the capacity and the division.
Er – 39 t is displayed when the instrument has not yet been calibrated and initialized.
Press the TARE key when the instrument displays “Er – 39” to enter the technical set-up environment. Carry
out the initialization of the indicator (“dEFAu” parameter) and finally the programming of all the parameters of
the set-up environment and the calibration.
Er – 85 t is displayed when the instrument has been initialized but not yet been calibrated.
Press the TARE key when the instrument displays “Er – 85” to enter the technical set-up environment and
carry out the calibration.
C.Er. – 36 During the calibration some internal negative points have been calculated:
- the calibration point is less than the zero point.
- the signal is negative (check the connections)
ndEr
(blinking)
The weight in underload (i.e. a weight of 100 divisions below the gross zero, if the instrument is approved) is
notified through this message and by an acoustic signal.
oVEr
(blinking)
The weight in overload (i.e. a weight of 9 divisions greater than the capacity) is notified through this message
and by an acoustic signal.

APD series bench scale APD_01_11.05_EN_U
17
DECLARATION OF CONFORMITY
This device conforms to the essential standards and norms relative to the
applicable European regulations. The Declaration of Conformity is available
in the web site www.diniargeo.com
WARRANTY
The TWO YEARS warranty period begins on the day the instrument is
delivered. t includes spare parts and labour repair at no charge if the
NSTRUMENT S RETURNED prepaid to the DEALER’S PLACE OF
BUS NESS. Warranty covers all defects NOT attributable to the Customer
(such as improper use) and NOT caused during transport.
f on site service is requested (or necessary), for any reason, where the
instrument is used, the Customer will pay for all of the service technician’s
costs: travel time and expenses plus room and board (if any).
The Customer pays for the transport costs (both ways), if the instrument is
shipped to DEALER or manufacturer for repair.
The WARRANTY is VO DED if any of the following occurs: repairs or
attempted repairs are made by unauthorised personnel, connected to
equipment installed by others, or is incorrectly connected to the power
supply, or instrument has defects or damage due to carelessness or failure
to follow the guidelines in this instruction manual.
This warranty DOES NOT provide for any compensation for losses or
damages incurred by the Customer due to complete or partial failure of
instruments, even during the warranty period.
AUTHORIZED SERVICE CENTRE STAMP
Table of contents
Other ATP Electronics Scale manuals
Popular Scale manuals by other brands

Total
Total TP2000 user guide