AVICHE ES10 User manual

Automatic Digital Blood Pressure Monitor
User Manual
ES10
Issued time: January, 2019
Version 2.0
Attention: The user manual is in the packaging, please read all
instructions before using and keep them properly for later use.


1
Introduction
Product Features
Thank you for purchasing our product. It is used to measure
human blood pressure and pulse, help people to know their health
status. With automatic control measurement technique, This digital
blood pressure monitor will inflate intelligently according to people's
systolic pressure and arm circumference, reducing the uncomfortable
feeling caused by excessive pressure.
· LCD backlit display
· Automatical intelligent inflation
· Upper arm blood pressure measurement
· Voice broadcast
· Alarm clock
· Two memories, each with 60 measured values
· Hypertension alarm
· Date/Time display
· Indicating average of the latest 3 readings
· Adapter is optional

2

3
Table of Contents
Unit Description
Icons in Display
Inserting and Replacing Batteries
Setting Date, Time, Voice, Unit, and Alarm Clock
Setting Memory Space
Measuring Position
Fitting the Cuff
Measuring Blood Pressure
Reading Memory
Deleting Memory
Blood Pressure Knowledge
Safety Cautions
Important Information
Troubleshooting
Maintenance
Specification
EMC Declaration
Blood Pressure Diary
After-safes Service and Warranty Card
………………………………………………………
………………………………………………………
………………………………
…………
……………………………………………
…………………………………………………
…………………………………………………………
………………………………………
……………………………………………………
……………………………………………………
………………………………………
………………………………………………………
………………………………………………
………………………………………………………
……………………………………………………………
……………………………………………………………
……………………………………………………
………………………………………………
………………………
4
5
6
7
9
9
10
11
12
12
13
14
14
15
16
17
18
23
27

4
Unit Description
Voice Broadcast
Adapter Socket
(USB Type-C)
Cuff Socket
Memory Button
(MEM)
Power Button
(START/STOP)
LCD Display
Setting Button
(SET)
Cuff
Cuff Connector
1. If user needs to purchase the adapter, please refer
to the specifications shown on the right figure.
2. Please use the power source in 100-240V.
3. Do not pull adapter with wet hands.
4. The standard packaging is without adapter,
you can buy it separately if you need.
Attention if use AC adapter
Input: AC100-240V
50/60Hz
Output: DC5V 1A
AC adaptor

5
Display Symbols
Storing or reading memory symbol
Low battery symbol, replacing battery
Alarm clock
Hypertension classification
Voice function
Pulse symbol
Memory group 1
Memory group 2
Average of latest 3 readings
Inflating error
the pressure is more than 300mmHg when inflate manually
measuring error
Icons in Display
Memory group
Voice broadcast
Blood pressure
classification
Memory number
Date/Time
Systolic pressure
Diastolic pressure
Pulse

6
Inserting and Replacing Batteries
1. Remove the battery compartment lid according to the arrow direction.
2. Insert 4pcs AA batteries with correct polarity, then put on the battery
cover.
Attention:
1. Replace the batteries when indicating low battery symbol or no
display after switching on machine.
2. Please make sure adapter unconnected when replacing the batteries
and battery polarity is correct. Never use new and old batteries at the
same time.
3. Remove batteries from device when not in operation for more than 3
months.
4. Attached batteries is for testing only, we advise good alkaline
batteries.
5. Battery life will change with the ambient temperature, shorter in cold
situation.
6. Please dispose batteries according to local environmental protection
law.
7. Batteries are dangerous article, don't dump them into regular trash.
8. Use only the approved arm cuff for this device. Use of other
arm cuffs may result in incorrect measurement results.

7
Setting Date, Time, Voice, Unit, and Alarm Clock
Right time setting can help you to record every measurement time
correctly.
1. Press button for 3 seconds, then enter the setting interface, the
year in display starts to flash (Figure 1).
2. Press button to adjust year, the figure will be added for every
press, circulate between 00 to 99.
3. Press to confirm when appear the right year, then enter to
month and date setting (Figure 2).
4. Repeat above three procedure to set month, date and time (Figure
3).
5. Set voice broadcast function. After setting the time, it will automatical
ly enter the voice broadcast settings, press button to open or
close the voice broadcast function.
Figure 1 Figure 2 Figure 3

8
6. mmHg and kPa unit conversion setting. After setting the voice, it will
automatically enter the unit setting and user can press button
to adjust mmHg to kPa.
7. 3 groups of alarm clock remind you to measure blood pressure on
time, after unit setting it automatically enter A1 alarm setting, press
to adjust ON or OFF, press button to confirm, after finish
setting, enter A2 alarm time setting, repeat the above steps, can set
A2, A3 alarm time setting.

9
Setting Memory Space
Measuring Position
1. Press button to choose memory or .
Take a seat and two feet flat on
the floor, place your arm on the table
comfortably, palms up, ensure that
the arm cuff is at heart height. Keep
still and do not talk during the
measurement.

10
Fitting the Cuff
1. Open arm cuff, slide the open end of cuff through the metal bracket
so that the velcro fastener is on outside, insert the cuff tube connec-
tor solidly into blood pressure monitor body.
2. Slide the cuff over the left upper arm, wrapped the arm cuff tightly,
artery indicator close to artery of upper arm (the inner arm).
3. Left palm up, the cuff should be 2-3cm above the elbow crease.
2-3 cm
0.8-1.2 inches
Aorta

11
Measuring Blood Pressure
1. Wrap the arm cuff properly on the arm (refer to cuff using).
2. Press button display full screen, ready to measure.
3. Press button to start measuring, the device will inflate
until the pressure is suitable for you, and then deflate, keep still and
quiet during measurement. It will display systolic pressure, diastolic
pressure and pulse after measuring finished.
Reference table for blood pressure warning
Range of blood
pressure value
Normal
Moderate hypertention
Mild hypertention
High-normal
≤120
>140
131-140
121-130
≤80
>90
86-90
81-85
Systolic (mmHg) Diastolic (mmHg) Indicator

12
Reading Memory
Deleting Memory
1. When you press button, the first display is the average of the
latest 3 measured value.
2. Press button again, it will display the record of the previous time.
1. Press button for 3 seconds, all records will be deleted.
Attention:
1. You can not delete only some of the values, all stored results will be
cleared if you press deletion. (You can make record for future
reference before deleting).
2. The measuring values in the monitor will not get lost if batteries are
taken out.

13
Blood Pressure Knowledge
Blood pressure is the pressure exerted by circulating blood upon the
walls of blood vessels, which refers to systolic pressure and diastolic
pressure. Systolic pressure is the highest pressure output by blood
during systole. Diastolic pressure is the lowest pressure that blood flow
back to heart during diastole.
The World Health Organization (WHO) and International Society of
Hypertension (ISH) made a set of standard for blood pressure. It is a
general rule for people of different age. It is important to consult your
doctors frequently, they will tell you the critical point which need
attention.
Blood pressure is not constant, it is influenced by season, time,
emotion, sport, alcohol, tobacco, coffee, medicine etc, therefore, the
result will be more accurate when you measure blood pressure in a
quite environment and physical comfort.
What is the blood pressure ?
Blood pressure classification standard
Change of blood pressure
Optimal
Normal
High Normal
Stage 1 Hypertension
Stage 3 Hypertension
Stage 2 Hypertension
Systolic
Pressure
Diastolic Pressure

14
Safety Cautions
Important Information
1. Battery leakage may damage the machine body, remove the
batteries if the product will not be used for a long time.
2. The product is not allowed to use on infants and young children.
3. Do not inflate before wrapping the cuff.
4. Patients can not judge the results and care themselves, please follow
doctor's instruction and ask professional person to explain the
measured value.
5. For patients with cardiovascular and cerebrovascular diseases,
please use it under the guidance of doctors.
6. Be sure to use an adapter with specifications as shown on page 4.
7. Please be sure to use AC100-240V power source.
8. Please don't use wet hand to pull out the power adapter in case of
electric shock.
9. Avoid using mobile phone near the device as it may cause malfunc-
tion.
10. Do not use this device in electromagnetic fields. It may result in
incorrect measurement results.
1. Take the measurement one hour after meal.
2. Do not smoke, drink alcohol, tea or coffee one hour before measure-
ment.
3. Do not measure when you are over fatigue or emotional.
4. keep quiet for 10-20 minutes before measurement.
5. As far as possible, make everyday measurement in the same time,
same condition and same position, wait 5 minutes between each
measurement.
6. If the device stored in low temperature environment, place it in
normal temperature environment for at least one hour before using.
7. Do not use it in moving vehicles (cars, planes), otherwise it can't
measure correctly.
8. Please use it in the required temperature and humidity environment,
or it may not be able to measure correctly.

Press the
button, no display
15
Troubleshooting
9. There may be risk if the air bag is over inflated for a long time.
10. When common arrhythmia (such as atrial premature, premature
ventricular, atrial fibrillation) appears, measured values may not be
accurate or blood pressure can't be measured.
The accuracy of this monitor has been rigorously tested. It is
recommended to check and calibrate once a year to ensure the monitor
is functioning properly and accurately. The monitor has a static
pressure detection mode for the relevant technical department to
perform the test. For more details, please call the customer service
hotline for consultation.
Abnormality Reason Checkout
Batteries low Replace new batteries
LCD shows E1,
abnormal inflating
LCD shows
measuring error
“EE” or abnormal
high (low) blood
pressure value
Check if the Cuff tube
connector is firmly fixed
Reinsert the tube
connector
Check if the cuff fixed
Well
Speak or move arm
during measurement
Refix the cuff
Keep quiet and take a
measurement again
If abnormal phenomenon still cannot be solved, please contact the
manufacturer or consult the distributor. Please do not disassemble
or repair if not knowing the reasons well.
If there are following abnormal phenomenon, you may refer to the
corresponding solutions:
Check if the batteries
are installed with the
correct polarity
Re-install the batteries

16
Maintenance
1. Please keep the blood pressure cuff and accessories in the saving
bag when not using.
2. In order to avoid the damage of the device, do not expose the blood
pressure monitor and cuff to high temperature, humidity, full of water
vapor or direct sunlight. Avoid strike and fall of the device.
3. To keep the machine clean, please use soft dry cloth to clean it. Do
not use diluent, alcohol, petrol and other chemical reagent to clean
the device body and cuff.
Note: Please clean the components of arm cuff after one year applica-
tion.
4. Water and other liquid are kept out of this product.
5. Do not fold the cuff or air pipe tightly.
6. The calibration interval is 1 year. Users can do it in advance if
necessary. The device must be tested and calibrated after repair.
7. Do not attempt to open, repair any part, if any problems, contact with
distributor of this product.

About 279g (without batteries and adapter)
22-36cm or 22-42cm (optional)
129(L) × 110 (W) × 77.5(H) mm
17
Specification
Product name Automatic Digital Blood Pressure Monitor
Model number ES10
Power supply 4 “AA” batteries 1.5V or USB power line (Type-C)
Measuring method Oscillometric method
Measure range Pressure: 0 to 280 mmHg (0-37.3kPa)
Pressure sensor Semiconductor
Inflation Automatic
Deflation Automatic
Memory groups 2 group, each group has 60 values
Operating environment Temperature 5~40℃
relative humidity ≤80%
Air pressure 80.0kPa-105.0kPa
Temperature -20~55℃
relative humidity 10%~90%
Air pressure 80.0kPa-105.0kPa
Transportation and storing environment
Device weight
Cuff circumference
Overall dimension
Automatic off time 1 minute
Operation method Continuous operation
Equipment category Class II, internal electric source
Class II equipment
Limited users Adults above 18 years old
BF type product
Read the user manual
Symbol for warning
Dispose the used product to the recycling collection
point according to local regulation
The product conforms to the requirement of EC
Directive MDD(93/42/EEC)
Disinfection, sterilization method According to the method provided by manufacturer
Safety sort
Cannot be used in the condition of flammable
anesthesia gas mixed with air or with oxygen or
nitric oxide
Electromagnetic compatibility Group 1, class B equipment
Ingress protection grade
Prevent solid objects into greater than 12 mm,
15 degrees tilt when still can prevent water
intrusion
IP22
Pulse: 40 to 199 beats/min
Accuracy Pressure: ±3 mmHg (±0.4kPa)
Pulse: ±5% of display reading

18
EMC Declaration
The “blood pressure monitor” is intended for use in the electromagnetic
environment specified below. The customer or the user of the “blood
pressure monitor” should ensure that it is used in such an environment.
Guidance and manufacturer's declaration - electromagnetic immunity
Electromagnetic environment
— guidance
Immunity test Compliance
level
Electrostatic
discharge (ESD)
IEC 61000-4-2
IEC 60601
test level
±6kV contact
±8kV air
±6kV contact
±8kV air
Floor should be wood, concrete, or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2kV for power
supply lines
±1kV for
input/output lines
±2kV for power
supply lines
±1kV for
input/output lines
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±1kV differential
mode
±2kV common
mode
±1kV differential
mode
±2kV common
mode
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5%UT
(>95% dip in UT)
for 0.5 cycle
40%UT
(60% dip in UT)
for 5 cycles
70%UT
(30% dip in UT)
for 25 cycles
<5%UT
(>95% dip in UT)
for 5 sec
<5%UT
(>95% dip in UT)
for 0.5 cycle
40%UT
(60% dip in UT)
for 5 cycles
70%UT
(30% dip in UT)
for 25 cycles
<5%UT
(>95% dip in UT)
for 5 sec
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
“blood pressure monitor” requires
continued operation during power
mains interruptions, it is recomme-
nded that the “blood pressure
monitor” be powered from an unin-
terruptible power supply or a
battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level.
3A/m 3A/m
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical com-
mercial or hospital environment.