Components of the UltraSoundGate 116Hb
47
Introduction
Thank you for purchasing Avisoft UltraSoundGate 116Hb This bus-
powered USB device supports single-channel high-speed data
acquisition at sampling rates of up to 300 kHz
The accompanying recording software Avisoft-RECORDER USGH
provides either continuous or triggered direct-to-disk recording with
real-time spectrogram displays
Getting started
The supplied RECORDER USGH software can be launched from
Start / All Programs / Avisoft ioacoustics / RECORDER USGH
On the first program start, the configuration dialog box will be
launched automatically (otherwise it is available from Options /
Configuration). Select the desired Sampling rate from the Input
Device Settings section and click at Ok. Then click at the Pause
button (Monitoring/Pause) and the Start button (Monitoring/Start)
You will then see the real-time spectrogram displaying the incoming
signals
8
1 XLR input connector
The 5-pole XLR input connector represents the analog inputs of the
recording device and provide power supply voltages for external
amplifiers and microphones The connector scheme is as follows:
1
2
3-
+
differential source
45
+200V
+5V
1
2
3-
+
single-ended source
45
+200V
+5V
2 US 2.0 interface
3 PHONES
This stereo 3 5mm mini jack connector allows to connect
headphones or a small speaker for acoustically monitoring the
incoming ultrasounds A undersampling technique is being used
to convert the ultrasonic sounds into audible signals The
undersampling ratio can be adjusted from the Advanced USGH
Settings dialog box of the RECORDER USGH software (see
section RECORDER USGH Settings for details)
4 TRG button
This button can control the wav file recording process in the
RECORDER software To enable this mode of operation, one of
the following Trigger source options must be selected from the
configuration dialog box:
USG TRG button auto hold : Pressing the button for more than 2 seconds will
activate an auto hold mechanism (the recording continues after releasing the button and
will stop once the button is pressed again) If the button is pressed for less than two
seconds, it will only record as long as the button is being pressed
USG TRG button : The software will record as long as the button is pressed
USG TRG button inversed : The software will record as long as the button is not
pressed (or as long as the external TRG signal is not active (logic high))
USG TRG button toggled : The software will start recording once the button is
pressed and continues until the button is pressed again
5 REC indicator
This amber colored LED will flash once the device is connected to
the PC It will be switched off once the RECORDER USGH
software is running the in the monitoring mode In this mode, the
REC LED indicates whether the RECORDER software is recording
the incoming data onto disk
6 OVERload indicator
The red OVER LED indicates clipping (over-modulation) If this
happens, the gain should be reduced by turning the gain control
knob to the left
7 POWER indicator
This green LED indicates that the unit is connected to the USB
power supply
8 GAIN control knob
This control knob adjusts the analog input recording level
9 VOLUME control knob
This knob adjusts the volume of the phones monitor output (5)
5
6
3
1
2
9
1 Ground
2 Positive input
3 Negative input
4 +5V supply voltage (max current 20 mA)
5 +200V polarization voltage
Installation procedure
First download and install the RECORDER USGH software from the
Avisoft Bioacoustics website (www avisoft com/downloads htm or
directly www avisoft com/RECORDER USGH exe) This installation
program will install both the RECORDER USGH application
(rec_usgh exe) and the required device drivers (usgh_xx16h inf,
usgh sys) for the UltraSoundGate xx16H devices When the
installation procedure has completed, the UltraSoundGate unit can
be connected to the computer The device should then be detected
as “Avisoft-UltraSoundGate 116H” and the pre-installed driver
should be finally activated
Under some circumstances it might happen that the silent
installation of the device driver fails If that happens, navigate to the
Windows Control Panel > Hardware and Sound > Device Manager
and right-click at the entry Other devices > Aviso t Aviso t-
UltraSoundGate 116H and select the Update Driver So tware..
option Then click at Browse my computer or device driver
so tware, click at Browse and navigate to the folder C:\Program
Files (x86)\Aviso t Bioacoustics\RECORDER USGH\Drivers and
finally click at Next The completed device driver installation will
then look like this:
To become familiar with the RECORDER USGH
software use the online help system that can be
accessed through the drop-down menu Help > Help...,
the Help... buttons on the individual dialog boxes or
the website at
http://www avisoft com/Help/RECORDER/content htm
and the section RECORDER USGH Software
Settings in this guide