SPECIFICATIONS
Model: AVR-E3-B2
Maximum amplitude3: (50Ω load8) 100V
Rise time (20%-80%): ≤ 0 5 ns
Fall time (80%-20%): ≤ 1 0 ns
Pulse width (FWHM): standard units: 10 ns to 200 ns
with -W1 option: 1 ns to 200 ns
with -W2 option: 50 ns to 5 us
with -W3 option: 1 ns to 5 us
Maximum PRF:
Standard units, or with -W1 option:
Units with -W2 or -W3 options:
100 kHz
20 kHz
Maximum duty cycle: 5%
Polarity: Positive or negative or both (specify4)
DC offset or bias insertion: Option available5 Apply required DC offset or bias in the range of ± 25 Volts,
(250 mA max) to back panel solder terminals
Trigger modes: Internal trigger, external trigger (TTL level pulse, > 10 ns, 1 kΩ input impedance),
front-panel “Single Pulse” pushbutton, or single pulse trigger via computer command
Variable delay: 0 to 1 0 seconds, for all trigger modes (including external trigger)
Propagation delay: ≤ 160 ns (Ext trig in to pulse out)
Jitter: (Ext trig in to pulse out) ± 35ps ± 0 015% of sync delay
Sync output: > +3 Volts, > 50 ns, will drive 50 Ohm loads
Gate input: Synchronous or asynchronous, active high or low, switchable Suppresses triggering when active
Connectors: Out: SMA, Other: BNC
Power requirements: 100 - 240 Volts, 50 - 60 Hz
GPIB and RS-232 control2: Standard on -B units
LabView drivers: Check http://www avtechpulse com/labview for availability and downloads
Ethernet port, for remote control
using VXI-11 3, ssh, telnet, & web:
Optional6 Recommended as a modern alternative to GPIB / RS-232
See http://www avtechpulse com/options/vxi for details
Settings resolution: The resolution of the timing parameters (pulse width, delay, period) varies,
but is always better than 0 15% of (|set value| + 20 ns)
The amplitude resolution is < 0 1% of the maximum amplitude
Settings accuracy: Typically ± 3% (±2 ns or ± 2% of max amplitude) after 10 minute warmup For high-accuracy
applications requiring traceable calibration, verify the output with a calibrated oscilloscope
Dimensions (H x W x D): 100 mm x 430 mm x 375 mm (3 9” x 17” x 14 8”)
Chassis material: cast aluminum frame and handles, blue vinyl on aluminum cover plates
Temperature range: +5°C to +40°C
1) -C suffix indicates stand-alone lab instrument with internal clock and line powering (See http://www avtechpulse com/formats for additional details of
the four basic instrument formats)
2) -B suffix indicates IEEE-488 2 GPIB and RS-232 control of amplitude, pulse width, PRF and delay (see http://www avtechpulse com/gpib)
3) For operation at amplitudes of less than 10% of full-scale, best results will
be obtained by setting the amplitude near full-scale and using external attenuators on the output
4) Indicate desired polarity by suffixing model number with -P or -N (i e positive or negative) or -PN for dual polarity option Polarity reversal achieved
by means of a two-position switch that controls the polarity of the signal output port on -C units, and via keypad control on -B units -PN option not
available on -W3 units
5) For DC offset option suffix model number with -OS
6) Add the suffix -VXI to the model number to specify the Ethernet port
7) Fall time increases to < 3 ns for pulse widths less than 15 ns
8) A 50Ω load is required Other loads may damage the instrument Consult Avtech (info@avtechpulse com) if you need to drive other
load impedances
8