SPECIFICATIONS
Model1: AVR-E2-B-P-TEK1C
Amplitude2: +10V to +50V
Rise time (10%-90%): Between 0 6 and 1 0 ns
Fall time (90%-10%): Between 0 6 and 2 0 ns
Pulse width (FWHM): 50 ns - 200 ns
Maximum PRF: 100 kHz
Pulse top aberrations at
maximum amplitude (+50V):
< +/-5% within the first 20 ns and last 12 ns of the pulse,
and < +/-2% for the remaining middle portion of the pulse
Required load3: 50Ω (±10%)
Trigger modes: Internal trigger, external trigger (TTL level pulse, > 10 ns, 1 kΩ input impedance),
front-panel “Single Pulse” pushbutton, or single pulse trigger via computer command
Variable delay:
(Sync to main out) 0 to 1 0 seconds, for all trigger modes (including external trigger)
Propagation delay: ≤ 200 ns (Ext trig in to pulse out)
Jitter: ± 35ps ± 0 015% of sync delay (Ext trig in to pulse out)
Sync output: > +3 Volts, > 50 ns, will drive 50 Ohm loads
Gate input: Synchronous or asynchronous, active high or low, switchable Suppresses triggering when active
Connectors: Out: SMA, Other: BNC
Power requirements: 100 - 240 Volts, 50 - 60 Hz
GPIB and RS-232 control1: Standard on -B units
LabView drivers: Check http://www avtechpulse com/labview for availability and downloads
Ethernet port, for remote
control using VXI-11 3, ssh,
telnet, & web:
Included Recommended as a modern alternative to GPIB / RS-232
See http://www avtechpulse com/options/vxi for details
Settings resolution: The resolution of the timing parameters (pulse width, delay, period) varies,
but is always better than 0 15% of (|set value| + 20 ns)
The amplitude resolution is < 0 1% of the maximum amplitude
Settings accuracy, after 10
minute warm-up:
Amplitude: Typically ± (3% of setting) ± (2% of maximum)
Delay, Period: Typically ± (3% of setting ) ± (5 ns)
Pulse width: Typically ± (3% of setting ) ± (2 ns), at maximum amplitude As the
amplitude is reduced, the pulse width may shift ± 5 ns For high-accuracy applications
requiring traceable calibration, verify the output with a calibrated oscilloscope4
Dimensions (H x W x D): 100 mm x 430 mm x 375 mm (3 9” x 17” x 14 8”)
Chassis material: cast aluminum frame and handles, blue vinyl on aluminum cover plates
Temperature range: +5°C to +40°C
1) -B suffix indicates IEEE-488 2 GPIB and RS-232 control of amplitude, pulse width, PRF and delay (see http://www avtechpulse com/gpib)
2) For operation at amplitudes of less than 10% of full-scale, best results will be obtained by setting the amplitude near full-scale and using external
attenuators on the output
3) A 50Ω load is required Other loads may damage the instrument Consult Avtech (info@avtechpulse com) if you need to drive other load
impedances
4) These instruments are provided with a basic calibration checksheet, showing a selection of measured output parameters These measurements are
performed with equipment that is calibrated on a regular basis by a third-party ISO/IEC 17025:2005 accredited calibration laboratory However,
Avtech itself does not claim any accreditation For applications requiring traceable performance, use a calibrated measurement system rather than
relying on the accuracy of the pulse generator settings