Barco JAO15 User manual

BARCO LTD.
33F., No. 16 &No.16-1, Xizhan Rd., Banqiao Dist., New Taipei City 22 ,
Taiwan (R.O.C.)
Telephone: +886-2-7715- 299 Fax: +886-2-7715- 298
BARCO
Panel PC
JAO15/ JAO18
User Manual
Version: 1.2

Interactive Patient Care – Barco Healthcare Division
CONFIDENTIAL: The information contained herein is ro rietary and confidential information of Barco. No art of the information contained herein may
be disclosed in any form or by any means and/or stored in a data base or retrieval system without the rior written consent of Barco and/or used for any
ur ose other than authorized by Barco.
C
CC
Content
Preface ........................................................................................................... 3
Copyright ................................................................................................ 3
Disclaimer .............................................................................................. 3
Acknowledgements ............................................................................... 3
Regulatory Compliance Statements....................................................... 3
Declaration of Conformity ..................................................................... 3
RoHS Compliance ........................................................................................... 4
Safety Precautions ......................................................................................... 6
Technical Support and Assistance .................................................................. 7
Conventions Used in this Manual .................................................................. 7
Ordering Information ................................................................................... 1
Chapter 1: Product Introduction .................................................................. 11
JAO15 ................................................................................................... 11
JAO18 ................................................................................................... 14
IO introduction ............................................................................................. 17
Mechanical Dimensions ............................................................................... 19
JAO15 ................................................................................................... 19
JAO18 ................................................................................................... 19
Chapter2: Jumpers and Connectors ............................................................. 2
Before you Begin .................................................................................. 2
Jumper Setting ..................................................................................... 21
Locations of the Jumpers and Connectors ................................................... 22
Top View ............................................................................................... 22
Jumpers and DIP Switch Setting ................................................................... 22
CMOS Clear Select ................................................................................ 22
Connector Pin Definition .............................................................................. 23
External I/O Interface ........................................................................... 23
19V DC Power Input ..................................................................... 23
REAR IO USB PORT ....................................................................... 23
Service Call connector .................................................................. 24
Line-out Jack ................................................................................. 25
Internal Connectors .............................................................................. 26
SATA .............................................................................................. 26
Mini-PCIe Slot ............................................................................... 27
Block Diagram .......................................................... 錯誤
錯誤錯誤
錯誤! 尚未定義書籤
尚未定義書籤尚未定義書籤
尚未定義書籤。
。。
。
Chapter 3: System Setup ...................................................................... 28
Replace SATA Disk ................................................................................. 28
Replace RAM ........................................................................................ 28
Install Handset .................................................. 錯誤
錯誤錯誤
錯誤! 尚未定義書籤
尚未定義書籤尚未定義書籤
尚未定義書籤。
。。
。

Interactive Patient Care – Barco Healthcare Division
CONFIDENTIAL: The information contained herein is ro rietary and confidential information of Barco. No art of the information contained herein may
be disclosed in any form or by any means and/or stored in a data base or retrieval system without the rior written consent of Barco and/or used for any
ur ose other than authorized by Barco.
Preface
Copyright
This publication, including all photographs, illustrations and software, is
protected under international copyright laws, with all rights reserved. No
part of this manual may be reproduced, copied, translated or transmitted
in any form or by any means without the prior written consent from
BARCO.
Disclaimer
The information in this document is subject to change without prior notice
and does not represent commitment from BARCO International Co., Ltd.
However, users may update their knowledge of any product in use by
constantly checking its manual posted on our website:
http://www.barco.com. BARCO shall not be liable for direct, indirect,
special, incidental, or consequential damages arising out of the use of any
product, nor for any infringements upon the rights of third parties, which
may result from such use. Any implied warranties of merchantability or
fitness for any particular purpose is also disclaimed.
Acknowledgements
ST156B/ST185D is a trademark of BARCO. All other product names
mentioned herein are registered trademarks of their respective owners.
Regulatory Compliance Statements
This section provides the FCC compliance statement for Class A devices and
describes how to keep the system CE compliant.
Declaration of Conformity
FCC
This device complies with Part 15 of the FCC Rules. Operation is subject to
the following two conditions: (1) This device may not cause harmful
interference, and (2) this device must accept any interference received,
including interference that may cause undesired operation.
This equipment has been tested and found to comply with the limits for a
Class B digital device, pursuant to Part 15 of the FCC Rules. These limits
are designed to provide reasonable protection against harmful interference
in a residential installation. This equipment generates, uses and can radiate
radio frequency energy and, if not installed and used in accordance with
the instructions, may cause harmful interference to radio communications.

Interactive Patient Care – Barco Healthcare Division
CONFIDENTIAL: The information contained herein is ro rietary and confidential information of Barco. No art of the information contained herein may
be disclosed in any form or by any means and/or stored in a data base or retrieval system without the rior written consent of Barco and/or used for any
ur ose other than authorized by Barco.
However, there is no guarantee that interference will not occur in a
particular installation. If this equipment does cause harmful interference
to radio or television reception, which can be determined by turning the
equipment off and on, the user is encouraged to try to correct the
interference by one of the following measures:
Reorient or relocate the receiving antenna.
Increase the separation between the equipment and receiver.
Connect the equipment into an outlet on a circuit different from that
to which the receiver is connected.
Consult the dealer or an experienced radio/TV technician for help.
. FCC Caution:
Any changes or modifications not expressly approved by the party
responsible for compliance could void the user's authority to operate
this equipment.
This transmitter must not be co-located or operating in conjunction
with any other antenna or transmitter.
Radiation Expo ure Statement:
This equipment complies with FCC radiation exposure limits set forth for an
uncontrolled environment. This equipment should be installed and
operated with minimum distance 2 cm between the radiator & your body.
CE
The product(s) described in this manual complies with all applicable
European Union (CE) directives if it has a CE marking. For computer systems
to remain CE compliant, only CE-compliant parts may be used. Maintaining
CE compliance also requires proper cable and cabling techniques.
IEC/ISO
All components and wirings comply with IEC or ISO standard, except for
Plastic Enclosure is UL approved only
RoHS Compliance
BARCO RoHS Environmental Policy and Statu
Update
This publication, including all photographs, illustrations and software, is
protected under international copyright laws, with all rights reserved. No
part of this manual may be reproduced, copied, translated or transmitted
in any form or by any means without the prior written consent from
BARCO.
RoHS restricts the use of Lead (Pb) < .1% or 1, ppm, Mercury (Hg) <
.1% or 1, ppm, Cadmium (Cd) < . 1% or 1 ppm, Hexavalent
Chromium (Cr6+) < .1% or 1, ppm, Polybrominated biphenyls (PBB) <
.1% or 1, ppm, and Polybrominated diphenyl Ethers (PBDE) < .1% or
1, ppm.
In order to meet the RoHS compliant directives, BARCO has established an
engineering and manufacturing task force to implement the introduction of
green products. The task force will ensure that we follow the standard

Interactive Patient Care – Barco Healthcare Division
CONFIDENTIAL: The information contained herein is ro rietary and confidential information of Barco. No art of the information contained herein may
be disclosed in any form or by any means and/or stored in a data base or retrieval system without the rior written consent of Barco and/or used for any
ur ose other than authorized by Barco.
development procedure and that all the new RoHS components and new
manufacturing processes maintain the highest industry quality levels for
which BARCO are renowned.
The model selection criteria will be based on market demand. Vendors and
suppliers will ensure that all designed components will be RoHS compliant.

Interactive Patient Care – Barco Healthcare Division
CONFIDENTIAL: The information contained herein is ro rietary and confidential information of Barco. No art of the information contained herein may
be disclosed in any form or by any means and/or stored in a data base or retrieval system without the rior written consent of Barco and/or used for any
ur ose other than authorized by Barco.
Safety Precautions
1. Read these safety instructions carefully.
2. Keep this User Manual for later reference.
3. Disconnect this equipment from any AC outlet before cleaning. Use a damp
cloth. Do not use liquid or spray detergents for cleaning.
4. For plug-in equipment, the power outlet socket must be located near the
equipment and must be easily accessible.
5. Keep this equipment away from humidity.
6. Put this equipment on a stable surface during installation. Dropping it or
letting it fall may cause damage.
7. The openings on the enclosure are for air convection to protect the
equipment from overheating. DO NOT COVER THE OPENINGS.
8. Make sure the voltage of the power source is correct before connecting the
equipment to the power outlet.
9. Place the power cord in a way so that people will not step on it. Do not place
anything on top of the power cord. Use a power cord that has been
approved for use with the product and that it matches the voltage and
current marked on the product’s electrical range label. The voltage and
current rating of the cord must be greater than the voltage and current
rating marked on the product.
1 . All cautions and warnings on the equipment should be noted.
11. If the equipment is not used for a long time, disconnect it from the power
source to avoid damage by transient overvoltage.
12. Never pour any liquid into an opening. This may cause fire or electrical
shock.
13. Never open the equipment. For safety reasons, the equipment should be
opened only by qualified service personnel.
14. If one of the following situations arises, get the equipment checked by
service personnel:
i. The power cord or plug is damaged.
ii. Liquid has penetrated into the equipment.
iii. The equipment has been exposed to moisture.
iv. The equipment does not work well, or you cannot get it to work
according to the user’s manual.
v. The equipment has been dropped and damaged.
vi. The equipment has obvious signs of breakage.
15. Do not place heavy objects on the equipment.
16. The system follows the national requirement to dispose unit

Interactive Patient Care – Barco Healthcare Division
CONFIDENTIAL: The information contained herein is ro rietary and confidential information of Barco. No art of the information contained herein may
be disclosed in any form or by any means and/or stored in a data base or retrieval system without the rior written consent of Barco and/or used for any
ur ose other than authorized by Barco.
Technical Support and Assistance
1. For the most updated information of products, visit Barco website at
www.barco.com.
2. For technical issues that require contacting our technical support team
or sales representative, please have the following information ready
before calling:
– Product name and serial number
– Detailed information of the peripheral devices
– Detailed information of the installed software (operating system,
version, application software, etc.)
– A complete description of the problem
– The exact wordings of the error messages
Warning
1. Handling the unit: carry the unit with both hands and handle it with care.
2. Maintenance: to keep the unit clean, uses only approved cleaning
products or clean with a dry cloth.
3. To avoid risk of electric shock, this equipment must only be connected to
a supply with protective earth
4. Do not modify this equipment without authorization of the
manufacturer.
Warning
1. Manipulation de l'appareil : réaliser l'unité avec les deux mains et
manipuler avec soin .
2. Entretien: Pour garder l'unité propre, utilise des produits de nettoyage
approuvés ou nettoyer avec un chiffon sec .
3. Pour éviter tout risque de choc électrique , cet équipement ne doit être
connecté à une alimentation avec de la terre de protection
4. Ne pas modifier cet équipement sans l'autorisation du constructeur

Interactive Patient Care – Barco Healthcare Division
CONFIDENTIAL: The information contained herein is ro rietary and confidential information of Barco. No art of the information contained herein may
be disclosed in any form or by any means and/or stored in a data base or retrieval system without the rior written consent of Barco and/or used for any
ur ose other than authorized by Barco.
Conventions Used in this anual
Warning:
Information about certain situations, which if not observed, can cause
personal injury. This will prevent injury to you when performing a task.
Caution:
Information to avoid damaging components or losing data.
Note:
Provides additional information to complete a task easily

Interactive Patient Care – Barco Healthcare Division
CONFIDENTIAL: The information contained herein is ro rietary and confidential information of Barco. No art of the information contained herein may
be disclosed in any form or by any means and/or stored in a data base or retrieval system without the rior written consent of Barco and/or used for any
ur ose other than authorized by Barco.
Symbol Glossary
There are some symbols marks on user manual, label, carton and product,
each symbol has its purpose, please take notice the symbol identification
before use.
Book mark
Read the manual before start operate the system
Power key
Power key use to control system status such as reset or set
enable panel backlight.
RFID and SMC
This system work with contact and contactless reader.
Service call
The system has 4pin service call. It connects to back end
service system with 4 wire signal pin
USB
Usb port symbol. The system support USB2. IO.
Network
The system support 1 /1 /1 Mbps Ethernet connection.
Line out
3.5mm earphone jack supports line out signal.
HDMI
HDMI signal, the system support HDMI input on this port
GND
Ground pin for system.
(
Panel PC with external power supply: class II with functional
earth equipment)
AC power ymbol
Information for AC power input range of PSU
DC power ymbol
Information for DC power input range of system

Interactive Patient Care – Barco Healthcare Division
CONFIDENTIAL: The information contained herein is ro rietary and confidential information of Barco. No art of the information contained herein may
be disclosed in any form or by any means and/or stored in a data base or retrieval system without the rior written consent of Barco and/or used for any
ur ose other than authorized by Barco.
Ordering Information
The following information below provides ordering information.
JAO 15 / ST156BIPC000 (P/N: 10Q0ST15600X1)
15.6” bedside terminal, Intel Bay Trail J19 2. GHz, resistive touch screen, 2GB DDR3L, 32GB MLC SSD, 13.56MHz RFID and smart card reader
JAO18/ ST185DIPC000 (P/N: 10Q0ST18500X1)
18.5” bedside terminal, Intel Bay Trail J19 2. GHz, PCAP touch screen, 2GB DDR3L, 32GB MLC SSD, 13.56MHz RFID and smart card reader

Interactive Patient Care – Barco Healthcare Division
CONFIDENTIAL: The information contained herein is ro rietary and confidential information of Barco. No art of the information contained herein may
be disclosed in any form or by any means and/or stored in a data base or retrieval system without the rior written consent of Barco and/or used for any
ur ose other than authorized by Barco.
Product Introduction
JAO15
Overview
JAO 15 is designed to serve as a Point of Care (POC) and Point of
Information terminal (POI) within healthcare applications. Manufactured
using anti-bacterial plastics, the terminals provide a host of interactive
healthcare applications providing secure access to Electronic Medical
Records (EMR) and innovative triple-play solutions (web, IPTV, VoIP) for
patient entertainment.
Key Feature
16:9 15.6” 1366x768 Fanless LED Panel Computer
Intel® Celeron® J19 , Quad Core, Low Consumption CPU
Flat Panel by Resistive Touch Screen
With 1x GbE/ HDMI input/ 5x USB ports
Built in speaker/ microphone/ 2x line out
13.56MHz RFID/ Smart card reader
DDR3L 2GB/ 2.5” 32GB SSD
System Power Input 19V DC

Interactive Patient Care – Barco Healthcare Division
CONFIDENTIAL: The information contained herein is ro rietary and confidential information of Barco. No art of the information contained herein may
be disclosed in any form or by any means and/or stored in a data base or retrieval system without the rior written consent of Barco and/or used for any
ur ose other than authorized by Barco.
Specification
Panel
-LED size: 15.6”, 16:9
-Resolution: 1366x768
-Luminance: 3 cd/m2
-Contrast ratio: 5
-LED color: 16.7M
-Viewing Angle: 8 (U) 8 (D) 8 (L) 8 (R)
-Backlight: LED
Touch Screen
-5 wire Resistive touch
-Light transmission: 78%
-Interface: USB
Sy tem
-CPU: On-board Intel® Celeron® J19 , 2. GHz,2M L2 Cache
-BIOS: AMI BIOS
-IO chipset: ITE 8786
-System memory: 1x 2 4-pin DDR3L SO-DIMM socket, 2G
DDR3L (Default), support up to 8GB DDR3L-1333, Non-ECC
and un-buffered
-Storage Device: 1x 2.5” SATA HDD
-Watchdog timer: Watchdog timeout can be programmed by
software. from 1 second to 255 seconds and from 1 minute
to 255 minutes
(Tolerance 15% under room temperature 25°C)
-H/W status monitor: Monitoring system temperature, and
voltage
-Expansion: 1x mini-PCIe sockets
I/O
-Ethernet: 1x RJ45
-USB: 5 x USB 2.
-Audio port: 2x Line out/built in speaker and microphone
-HDMI input connector
-Front touch icon
-72 p camera
Audio
-HDA codec: Realtek ALC886-GR
-Audio interface: Line out Audio Jack
Ethernet
-LAN chip: Intel® I21 AT Gigabit LAN
-Ethernet interface: 1 /1 /1 Based-Tx Ethernet
compatible
Wifi & BT (optional)

Interactive Patient Care – Barco Healthcare Division
CONFIDENTIAL: The information contained herein is ro rietary and confidential information of Barco. No art of the information contained herein may
be disclosed in any form or by any means and/or stored in a data base or retrieval system without the rior written consent of Barco and/or used for any
ur ose other than authorized by Barco.
-Certified module BOINTE PE8 9-AA
-Support 8 2.11 b/g/n
-Support BT 4.
Mechanical & Environment
-Color: white, pantone RAL9 16
-Vibration:
IEC 68 2-64 (w/ HDD)
1Grms @ sine, 5~5 Hz, 1hr/axis (HDD Operating)
-Shock:
IEC 68 2-27
HDD: 2 G@panel mount, half sine, 11ms
-Operating temperature: °C to 4 °C
-Storage temperature: -2 °C to 6 °C
-Operating/Storage humidity: 2 %~8 % relative humidity,
non-condensing
-The humidity preconditioning treatment is performed in a
humidity cabinet containing air with a relative humidity of
93%RH @25°C 48hrs
-Storage Altitude: up to 3 M
-Dimension: 417.2x 266.6x5 mm
-Weight: 4.5 Kg
Power Supply
-Model: DELTA MDS- 9 AAS19 B
-Power input: 1 -24 V~ 1.5- .75A 5 -6 Hz
-Power Output: 19VDC, 4.74A
Certification
-CE
-FCC class B
-EN6 6 1-1
-UL6 6 1-1
-IEC 6 95 -1 CB
-ANSI/AAMI ES6 6 1-1*
-CAN/CSA-C22.2 No.6 6 1-1*
* This panel pc incorporates the ES60601-1 /IEC60601-1 certified power supply,
Model MDS-090AAS19 B is by DELTA. The power supply is used in the
conditions of overvoltage category II, Pollution Degree II.
*This panel pc incorporates the power supply is Class II with functional earth
AC/DC Adapter intended for supplying medical electrical apparatus with two
parts of plastic enclosure fixed together by screws

Interactive Patient Care – Barco Healthcare Division
CONFIDENTIAL: The information contained herein is ro rietary and confidential information of Barco. No art of the information contained herein may
be disclosed in any form or by any means and/or stored in a data base or retrieval system without the rior written consent of Barco and/or used for any
ur ose other than authorized by Barco.
JAO18
Overview
JAO 18 is designed to serve as a Point of Care (POC) and Point of
Information terminal (POI) within healthcare applications. Manufactured
using anti-bacterial plastics, the terminals provide a host of interactive
healthcare applications providing secure access to Electronic Medical
Records (EMR) and innovative triple-play solutions (web, IPTV, VoIP) for
patient entertainment.
Key Feature
16:9 18.5” 1366x768 Fanless LED Panel Computer
Intel® Celeron® J19 , Quad Core, Low Consumption CPU
Flat Panel by PCAP Touch Screen
With 1x GbE/ HDMI input/ 5x USB ports
Built in speaker/ microphone/ 2x line out
13.56MHz RFID/ Smart card reader
DDR3L 2GB/ 2.5” 32GB SSD
System Power Input 19V DC

Interactive Patient Care – Barco Healthcare Division
CONFIDENTIAL: The information contained herein is ro rietary and confidential information of Barco. No art of the information contained herein may
be disclosed in any form or by any means and/or stored in a data base or retrieval system without the rior written consent of Barco and/or used for any
ur ose other than authorized by Barco.
Specification
Panel
-LED size: 18.5”, 16:9
-Resolution: 1366X768
-Luminance: 3 cd/m2
-Contrast ratio: 1
-LED color: 16.7M
-Viewing Angle: 8 (U) 8 (D) 85(L) 85(R)
-Backlight: LED
Touch Screen
-PCAP touch
-Light transmission: 87%
-Interface: USB
Sy tem
-CPU: On-board Intel® Celeron® J19 , 2. GHz,2M L2 Cache
-BIOS: AMI BIOS
-IO chipset: ITE 8786
-System memory: 1x 2 4-pin DDR3L SO-DIMM socket, 2G
DDR3L (Default), support up to 8GB DDR3L-1333, Non-ECC
and un-buffered
-Storage Device: 1x 2.5” SATA HDD
-Watchdog timer: Watchdog timeout can be programmed by
software. from 1 second to 255 seconds and from 1 minute
to 255 minutes
(Tolerance 15% under room temperature 25°C)
-H/W status monitor: Monitoring system temperature, and
voltage
-Expansion: 1x mini-PCIe sockets
I/O
-Ethernet: 1x RJ45
-USB: 5 x USB 2.
-Audio port: 2x Line out/built in speaker and microphone
-HDMI input connector
-Front touch icon
-72 p camera
-Optional barcode scanner (IEC62471 certified)
Audio
-HDA codec: Realtek ALC886-GR
-Audio interface: Line out Audio Jack
Ethernet
-LAN chip: Intel® I21 AT Gigabit LAN
-Ethernet interface: 1 /1 /1 Based-Tx Ethernet
compatible
Wifi & BT (optional)

Interactive Patient Care – Barco Healthcare Division
CONFIDENTIAL: The information contained herein is ro rietary and confidential information of Barco. No art of the information contained herein may
be disclosed in any form or by any means and/or stored in a data base or retrieval system without the rior written consent of Barco and/or used for any
ur ose other than authorized by Barco.
-Certified module BOINTE PE8 9-AA
-Support 8 2.11 b/g/n
-Support BT 4.
Mechanical & Environment
-Color: white, pantone RAL9 16
-Vibration:
IEC 68 2-64 (w/ HDD)
1Grms @ sine, 5~5 Hz, 1hr/axis (HDD Operating)
-Shock:
IEC 68 2-27
HDD: 2 G@panel mount, half sine, 11ms
-Operating temperature: °C to 4 °C
-Storage temperature: -2 °C to 6 °C
-Operating/Storage humidity: 2 %~8 % relative humidity,
non-condensing
-The humidity preconditioning treatment is performed in a
humidity cabinet containing air with a relative humidity of
93%RH @25°C 48hrs
-Storage Altitude: up to 3 M
-Dimension: 481.6x3 8.2x5 mm
-Weight: 4.7 Kg
Power Supply
-Model: DELTA MDS- 9 AAS19 B
-Power input: 1 -24 V~ 1.5- .75A 5 -6 Hz
-Power Output: 19VDC, 4.74A
Certification
-CE
-FCC class B
-EN6 6 1-1
-UL6 6 1-1
-IEC 6 95 -1 CB
-ANSI/AAMI ES6 6 1-1*
-CAN/CSA-C22.2 No.6 6 1-1*
* This panel pc incorporates the ES60601-1 /IEC60601-1 certified power supply,
Model MDS-090AAS19 B is by DELTA. The power supply is used in the
conditions of overvoltage category II, Pollution Degree II.
*This panel pc incorporates the power supply is Class II with functional earth
AC/DC Adapter intended for supplying medical electrical apparatus with two
parts of plastic enclosure fixed together by screws

Interactive Patient Care – Barco Healthcare Division
CONFIDENTIAL: The information contained herein is ro rietary and confidential information of Barco. No art of the information contained herein may
be disclosed in any form or by any means and/or stored in a data base or retrieval system without the rior written consent of Barco and/or used for any
ur ose other than authorized by Barco.
IO introduction
Front ide
Camera
72 p camera with white LED
RFID and mart card reader
13.56MHz RFID and smart card reader. RFID tag also readable
while parking in smart card reader slot
Power Key
Touch icon input. Long press 5 seconds to turn on panel
backlight; long press 1 seconds to reset or power off system.

Interactive Patient Care – Barco Healthcare Division
CONFIDENTIAL: The information contained herein is ro rietary and confidential information of Barco. No art of the information contained herein may
be disclosed in any form or by any means and/or stored in a data base or retrieval system without the rior written consent of Barco and/or used for any
ur ose other than authorized by Barco.
Back ide
HDD ervice door
Lock with Torx screws, use for SSD replacement
RAM and miniPCIe ervice door
Lock with Torx screws, use for RAM replacement or install
extension module
Rear IO
19V DC input connect to PSU
Service call connect to back end service system (intra
network) with 4 wire signal pin
USB 3x USB 2. ports
LAN Gigabit Ethernet
Bottom IO
Line out 2x line out jack
USB 2x USB 2. ports with 1.5A V
HDMI use as HDMI input
Barcode Scanner (optional)
Optional install 1D/2D supported barcode scanner module
Microphone
Built in digital microphone
Speaker
Built in stereo speaker

Interactive Patient Care – Barco Healthcare Division
CONFIDENTIAL: The information contained herein is ro rietary and confidential information of Barco. No art of the information contained herein may
be disclosed in any form or by any means and/or stored in a data base or retrieval system without the rior written consent of Barco and/or used for any
ur ose other than authorized by Barco.
echanical Dimensions
JAO15
JAO18

Interactive Patient Care – Barco Healthcare Division
CONFIDENTIAL: The information contained herein is ro rietary and confidential information of Barco. No art of the information contained herein may
be disclosed in any form or by any means and/or stored in a data base or retrieval system without the rior written consent of Barco and/or used for any
ur ose other than authorized by Barco.
Jumpers and Connectors
This chapter describes how to set the jumpers and connectors on the
motherboard.
Before you Begin
Ensure you have a stable, clean working environment. Dust and dirt can get
into components and cause a malfunction. Use containers to keep small
components separated.
Adequate lighting and proper tools can prevent you from accidentally
damaging the internal components. Most of the procedures that follow
require only a few simple tools, including the following:
-A Philips screwdriver
-A flat-tipped screwdriver
-A set of jewelers screwdrivers
-A grounding strap
-An anti-static pad
Using your fingers can disconnect most of the connections. It is
recommended that you do not use needle-nosed pliers to disconnect
connections as these can damage the soft metal or plastic parts of the
connectors.
Before working on internal components, make sure that the power is off.
Ground yourself before touching any internal components, by touching a
metal object. Static electricity can damage many of the electronic
components. Humid environments tend to have less static electricity than
dry environments. A grounding strap is warranted whenever danger of static
electricity exists.
Precautions
Computer components and electronic circuit boards can be damaged by
discharges of static electricity. Working on computers that are still
connected to a power supply can be extremely dangerous.
Follow the guidelines below to avoid damage to your computer or yourself:
Always disconnect the unit from the power outlet whenever you are
working inside the case.
To fully disengage the power to the unit, please disconnect the power cord
from the AC inlet.
If possible, wear a grounded wrist strap when you are working inside the
computer case. Alternatively, discharge any static electricity by touching the
bare metal chassis of the unit case, or the bare metal body of any other
grounded appliance
Hold electronic circuit boards by the edges only. Do not touch the
components on the board unless it is necessary to do so. Don’t flex or stress
the circuit board.