BEINAT MME500 User manual

Belonging to
User'sGuide&
maintenance
Digital pressure gauge MME500
Version 01
MME500
DIFFERENTIALPRESSUREGAUGE
Thank you for choosing the BEINAT S.r.L. digital tool, model MME500.
This manual was designed to help you obtain the best results and efficiency
from the product
Important Note
Read these instructions carefully before use and always keep
them within hands reach while using the instrument.
The illustrations and texts shown on the screens in this
manual may differ from those actually displayed
500MBAR
TestsandMeasures
UNI 7129
Measurepositivepressures
Measurenegativepressures
Measuringscale500mbar
Checkthepresentpressure
Checkthemethane gaspressureonthe network
Checkthe LPG gas pressure
Checkthe pressureinthecombustion chamber
Checkthe efficiency of the paths

Page 2
The MME500 pressure gauge is a high-tech product that embodies the innovation and
elegance of BEINAT S.r.l.
Its best privilege is to assist the installer in his work both during the installation of new
plants and in the testing of existing plants.
Pressure gauge
It is a small portable differential pressure gauge, with easy use. Precision and autonomy
make this tool indispensable for testing and controlling heating systems.
It is calibrated with sophisticated testing machines, so its measurement is very precise. It is
calibrated for example for the measurement of:
Network gas pressure
Pressure in the combustion chamber
Fireplaces efficiency
For your safety
To prevent damage to the product or injury to you and third parties, read the following safety warnings carefully before using
the equipment.
Store so that anyone using the appliance can consult it in advance.
If you notice smoke escaping or an acrid or unusual odor coming from the instrument, and in case of malfunction, immediately
switch off the device, and send it to the nearest service center.
Do not attempt to disassemble the appliance.
Contact with internal parts of the detector can cause injury.
In case of fault, the product must be repaired only by qualified personnel.
Avoid contactwith liquid crystals.
In case of breakage of the display, be careful not to injure yourself with the glass fragments and avoid that the liquid crystals
come into contact with the skin, eyes or mouth.
To use your digital equipment for a long time and with satisfaction, use it and store it keeping in mind the following precautions.
MAKE SURE the integrity of the product after removing it from the packaging.
Any use other than that for which the detector was designed is to be considered improper, so that BEINAT S.r.l. declines all
responsibility for any damage caused to people, animals or things.
Do not wet it.
The detector is not waterproof, if immersed in water or exposed to high humidity it can be seriously damaged.
Don't let it fall.
Strong blows against hard surfaces and large vibrations can damage the appliance. Should the appliance break down after being
dropped or crushed, consult the Service Center for necessary repairs.
Avoid strongmagnetic fields.
This detector should not be used or stored in the presence of radiation or strong magnetic fields, static electricity or magnetic
fields produced by equipment such as radio transmitters may interfere during detection.
Avoidsuddenchangesintemperature.
Sudden changes in temperature can cause condensation and the batteries could supply less current.
Over (+ 45 ° C c / a) the monitor may turn black, to make it visible again it must be cooled by placing it in the refrigerator for
a few minutes.
Cleaning
Never clean the appliance with chemical products. If necessary, wash with a damp cloth.
Powered by 3 AAA alkaline batteries .......…........…………….............................................................. 4.5 V
Average consumption .................................................................................................8,30 mA Consum
Battery autonomy according to functions ..................................................................... 100 hours approx
Battery charge and discharge control ....................................................................................On Display
Pressure: from 0 bar to 500 mbar
Pressure detection probe ........................................................................................................ Built-in
Measuring range ............................................................................................. ............... +/- 500 mbar
Pressure overload .................................................................................................................. 3 bars.
Resolution ................................................. .................................................. .......................... 1 mbar
Accuracy ................................................. .................................................. ........................ 0.4% FS
Old button ............................................... ............................................................................. Built-in
Numeric alpha LCD display ........................................................................................................ 3 digits
Operating temperature ............................................................................................ -10 ° C ÷ + 50 ° C
Auto power off ................................................. ............................................................. after 60 minutes
Electromagnetic Compatibility "CE Reference standard .............................................................. UNI 7129
Dimensions and weight ............................................... ........................................ 67 * 135 * 35 mm 70gr
Readings and Measurements
These operations must be carried out in compliance with the methods described in this manual.
Remember measurements taken with batteries in need of recharge can give false measurements.
Rammentiamo che le misure eseguite con le batterie basse possono falsare le misurazioni
Technical specifications
500

Three macroscopic variables are necessary to define the state of static air, each subject to direct measurement.
These variables are pressure, temperature and volume.
V, pressure Pand temperature Tis defined by Boyles Law:
P*V =n*R*T. where n is the number of moles composing the gas and R is a universal constant.
The possible states of a gas with constant pressure and volume are expressed by the Gay-Lussac Law: with
constant pressure the volume of a gas increases as the temperature increases, or vice versa
Therefore, if the instrument is connected at variable temperatures, the pressure registered at the start will
definitely differ from final pressure.
This is normal and is not due to malfunctioning..
For each degree C. of temperature the pressure varies by c / a 10 mmH2O
InformationNotebetweenPressureandTemperature
Unit of measure Symbol Pa hPa bar mbar at mmH2O
Pascal mmH2O 1 0,01 0,00001 0,01 0,00001 0,1
Hect Pascal hPa 100 1 0,001 1 0,0001 10
bar bar 100.000 1.000 1 1.000 1 10.000
millibar mbar 0,01 1 0,001 1 0,001 10
Technical atm. at 100.000 1.000 1 1.000 1 10.000
millimetriH2OmmH2O 10 0,01 0,0001 0,1 0,0001 1
PSI PSI 0.000145038 0.0145038 1.45038 0.0145038 1.45038 0.00145038
Pressures Conversion Table
35
500
135
67
Page 3
Maximum pressure
Minimum pressure

Page 4
ButtonFunction
Power button ON/OFF
To switch the instrument on/off, keep the button pressed for 5 seconds.
Components and Controls
Viewfrom above
View from above
To connect the supplied pneumatic hoses, use the inputs we need:
1) Negative, to measure a negative pressure.
2) Positive, to measure positive pressure.
3) Both, to measure a differentiated pressure
500mbar
Maximum pressure
Minimum pressure

Page 5
WARNINGS !
These operations must be performed in accordance with the manner illustrated in this booklet.
We remind you that measurements made with low batteries can distort measurements
MME-500 GAUGE
To power the instrument on, keep the button pressed for 5 seconds.
The instrument automatically positions itself on the minimum scale "from 0 to 99 mbar"
Display reading:
max. 99.9 mbar.
N.B.The scale exchange
takes place automatically
Display reading:
max. 500 mbar.
N.B.The scale exchange
takes place automatically
Pressure Control
The pressure gauge has two measurement inputs, one of pressure (positive) and one of depression
(negative).
1) To measure the positive pressure, apply the supplied tube to the "RIGHT" input, making sure to leave
the "LEFT" input free. Minimum Input.
2) To measure the negative pressure (Depression), apply the supplied tube to the "LEFT" input, making
sure to leave the "RIGHT" input Maximum Input open.
3) To measure the difference of two pressures, apply the higher pressure to the positive inlet and the
other to the negative inlet.
Finita la lettura, sul display apparirà la differenza di pressione che lo strumento misurerà tra un ingresso e l’altro.
Turn on the instrument as described above
2) Insert the tube into the desired inlet
3) Connect the instrument, through the small tube, to the source of pressure or depression, which must be
measured.
How to use MME-500
99 mbar
500mbar
Display of the battery charge status
The battery figure appears on the display.
When it is loaded the figure is all black.
Downloading it becomes white
Flashing is empty. N.B. In this case do not take measurements.
SHUTDOWN. Press the OFF button for 5 seconds
500mbar
HOLD function
The power button also serves to store the measured pressure.
Press the button at the end of the measurement before removing the detection tube; The detected value will
be stored

Page 6
If the appliance does not switch on.
Check that the batteries are charged, if not replace them
If the appliance does not detect pressure
Check that the pneumatic tubes are well inserted and make contact.
In case of further problems, it is necessary to contact directly a specialized and / or authorized technician
or BEINAT S.r.l.
Problems - Solutions
Sostituzione Batterie
Technical data of non-rechargeable AAA Alkaline
batteries
Rated voltage of the 1.5 VDCV battery pack each
Operating time approximately 100 hours with charged
batteries
N.B. Do not replace the batteries in the ATEX area
Replacement of discharged batteries
When the low battery icon appears, proceed as follows.
1) Turn off the instrument
2) Unscrew the 4 screws located on the back of the instrument.
3) Raise the cough
4) Remove the batteries
5) Insert the new batteries, positioning them in the right
direction
(+ -)
6) Refit the cover and tighten the 4 screws
7) Turn on the instrument and continue with your work.
Use
The batteries must never be exposed to temperatures higher
than + 40 ° C
To ensure optimum durability, use them at room temperature.
If used both at low and high temperature the duration capacity
may decrease.
Calibration certificate on a sample instrument from the ASIT "ACCREDIA

Page 7
The MME-500 digital manometer with scale from 0 to 500 mbar leaves the factory accompanied by a test and
calibration certificate.
This corresponds to the calibration declared by the standard instrument, based on
international standards
regulations.
N.B. The sample instrument is kept at the production plant.
The assessment of measurement uncertainty is "category B"
NOTE!!
The ordinary and extraordinary maintenance of the digital pressure gauge and the CALIBRATION, whose
natural expiry is ONE YEAR, must be performed by authorized personnel, with equipment in accordance with
the law.
MME-500 calibration
NOTE
——————————————————————————————————————————————
All other technical features are present in the manual attached to the instrument.
The tests described above were carried out with the following references:
Temperature: 20°C ± 2°C
Atmospheric Pressure: 100 kPa ± 0,1 kPa
Relative humidity: 50% ± 15%
This certificate is valid for one year and may not be reproduced without our permission, in any case,
may not be reproduced in part.
first CERTIFICATE OF CALIBRATION
Instrument: Digital pressure gauge MME500
Version: V. 1.0
Digital pressure gauge +/- 0-500 mbar
——————————————————————————————————————————————
PRESSURE
Sample instrument Serial number Measurement range Uncertainty Resolution
DRUCK DPI 530 - 4bar 0745/99-09 0 ÷ 4 mbar ± 0.1% F.S. 1 mbar (100Pa)

IP40 Made in Italy
BEINATS.r.l.
Via Fatebenefratelli 122/C 10077, S. Maurizio C/se (TO) - ITALY
Tel. 011.921.04.84 - Fax 011.921.14.77
http:// www.beinat.com
Business - info@beinat.com
Help Desk - laboratorio@beinat.com
MME500 -Rev-1 18 20
Lo styling è della b & b design
Stamp of the Dealer
Purchase Date
Serial Number :
Beinat S.r.l. following the purpose of improving its products, it reserves the right to modify the technical, aesthetic and functional characteristics
at any time and without giving any notice.
Pressure gauge MME500
INSURANCE.This device is insured by the SOCIETÀ REALE MUTUA for the PRODUCT’S GENERAL LIABILITY
up to a maximum of 1.500.000,00 EURO against damages caused by the device in case of failures in functioning.
WARRANTY.The warranty term is 3 years from manufacturing date, in agreement with the following
conditions. The components acknowledged as faulty will be replaced free of charge, excluding the
replacement of plastic or aluminium cases, bags, packing, batteries and technical reports.
The device must arrive free of shipment charges to BEINAT S.r.L.. Defects caused by unauthorized personnel
tampering, incorrect installation and negligence resulting from phenomena outside normal functioning shall be
excluded from the warranty.
The BEINAT S.r.L. company is not liable for possible damage, direct or indirect, to people, animals, or things;
from product faults and from its enforced suspension of use.
DISPOSAL OF OLD ELECTRICAL & ELECTRONIC EQUIPMENT.
This symbol on the product or its packaging to indicates that this product shall not be treated as household waste. Instead, it shall be handed over to
the applicable collection point for the recycling of electrical and electronic equipment, such as for example:
- sales points, in case you buy a new and similar product
- local collection points (waste collection center, local recycling center, etc...).
By ensuring this product is disposed of correctly, you willhelp prevent potential negative consequence for the environment and human health, which could
otherwise be caused by inappropriate waste handing of this product. The recycling of materials will help to conserve natural resources. For more detailed
information about recycling of this product, please contact your local city office, your household waste disposal service or the shop where you purchased
theproduct.
Attention: in some countries of the European Union, the product is not included in the field of application of the National Law that applies the European
Directive 2002/96/EC and therefore these countries have no obligation to carry out a separate collection at the “end of life” of the product.
Table of contents
Other BEINAT Test Equipment manuals
Popular Test Equipment manuals by other brands

ZKETECH
ZKETECH EBC-A40L user manual

Imada
Imada DTX-15 Calibration manual

Kyoritsu Electrical Instruments Works, Ltd.
Kyoritsu Electrical Instruments Works, Ltd. KEW 4140 instruction manual

Tone
Tone TA419TQ instruction manual

Vanguard
Vanguard EZCT-2KA user manual

Aqua Medic
Aqua Medic pH controller Operation manual