6. PROGRAMMING & CALIBRATION
The BA324D is programmed and calibrated via
four push-buttons which are located behind the
instrument control cover. When frequent access
to the push-buttons is required, the indicator can
be supplied with external membrane push-buttons
mounted on the outside of the control cover.
All the programming functions are contained in an
easy to use menu which is shown diagramatically
in Fig 6. Each function is summarised in section
6.1 which includes references to more detailed
information. Although this simple menu driven
system enables most adjustments to be made
without repeated reference to this manual, we
recommend that at least the summary of the
programmable functions in section 6.1 is read
before attempting programming or recalibration.
When the indicator is fitted with alarms,
linearisation or a tare function, the basic menu is
expanded to include functions associated with
these optional features. All the optional functions
are described in section 9 of this manual.
Throughout this manual push-buttons are shown in
italics e.g. P or Up push-button, and legends
displayed by the indicator are shown within
inverted commas e.g. 'CAL' and ' ALr2'.
Access to the programme menu is obtained by
operating the Pand Epush-buttons
simultaneously. If the instrument is not protected
by a security code the first parameter 'root' will be
displayed. If a security code other than the default
code 0000 has already been entered, the indicator
will display 'COdE'. Press Pto clear this prompt
and enter the security code for the instrument
using the Up, Down and Ppush-buttons. If the
correct code has been entered pressing Ewill
cause the first parameter 'root' to be displayed. If
an incorrect code is entered the indicator will
return to the display mode.
Once within the menu the required parameter can
be reached by scrolling through the main menu
using the Up and Down push-buttons as shown in
Fig 6. When returning to the display mode
following recalibration or changes to any
parameters, the indicator will display circulating
decimal points for a few seconds while the new
information is stored in permanent memory.
All new BA324D indicators are supplied calibrated
as requested at the time of ordering. If calibration
is not requested, the indicator will be set to display
0.00 with 4mA input, and 100.00 with 20mA input,
but can easily be re-calibrated on-site.
6.1 Summary of programmable functions
This section summarises all of the programmable
functions and when read in conjunction with Fig 6
provides sufficient information to condition and
calibrate the indicator. A cross-reference to more
information is given for each function.
Display Description of function
'root' Square root extractor
Turns the square root extractor for
linearising the output from differential
flowmeters 'On' or 'OFF'. This function
is omitted when a lineariser is installed.
See section 6.2
'rESn' Display resolution
Selects the resolution of the least
significant display digit. May be set to
1, 2, 5 or 10 digits.
See section 6.3
'd.P.' Decimal point
Positions the dummy decimal point
between any of the digits or turns it off.
See section 6.4
'CAL' Calibration of display using external
current source.
Enables the zero and span of the
indicator to be adjusted using an
external current source such as a
calibrator. Also enables a complete
loop to be calibrated from primary
element to the indicator display.
When used with an accurate traceable
current source this is the preferred
method of calibration.
See section 6.5