Bin Master PROCAP I Series User manual

По вопросам продаж и поддержки обращайтесь:
Архангельск (8182)63‐90‐72
Астана (7172)727‐132
Белгород (4722)40‐23‐64
Брянск (4832)59‐03‐52
Владивосток (423)249‐28‐31
Волгоград (844)278‐03‐48
Вологда (8172)26‐41‐59
Воронеж (473)204‐51‐73
Екатеринбург (343)384‐55‐89
Иваново (4932)77‐34‐06
Ижевск (3412)26‐03‐58
Казань (843)206‐01‐48
Калининград (4012)72‐03‐81
Калуга (4842)92‐23‐67
Кемерово (3842)65‐04‐62
Киров (8332)68‐02‐04
Краснодар (861)203‐40‐90
Красноярск (391)204‐63‐61
Курск (4712)77‐13‐04
Липецк (4742)52‐20‐81
Магнитогорск (3519)55‐03‐13
Москва (495)268‐04‐70
Мурманск (8152)59‐64‐93
Набережные Челны (8552)20‐53‐41
Нижний Новгород (831)429‐08‐12
Новокузнецк (3843)20‐46‐81
Новосибирск (383)227‐86‐73
Орел (4862)44‐53‐42
Оренбург (3532)37‐68‐04
Пенза (8412)22‐31‐16
Пермь (342)205‐81‐47
Ростов‐на‐Дону (863)308‐18‐15
Рязань (4912)46‐61‐64
Самара (846)206‐03‐16
Санкт‐Петербург (812)309‐46‐40
Саратов (845)249‐38‐78
Смоленск (4812)29‐41‐54
Сочи (862)225‐72‐31
Ставрополь (8652)20‐65‐13
Тверь (4822)63‐31‐35
Томск (3822)98‐41‐53
Тула (4872)74‐02‐29
Тюмень (3452)66‐21‐18
Ульяновск (8422)24‐23‐59
Уфа (347)229‐48‐12
Челябинск (351)202‐03‐61
Череповец (8202)49‐02‐64
Ярославль (4852)69‐52‐93
Единый адрес: [email protected] Веб‐сайт: www.binmaster.nt-rt.ru
PROCAP I BinMaster

9250290 Rev A
PROCAP I Series
CAPACITANCE PROBE
OPERATING INSTRUCTIONS
PLEASE READ CAREFULLY
Price $5.00

925-0290 Rev A 2
TABLE OF CONTENTS
GENERAL SPECIFICATIONS .................................................................................................... 3
SAFETY SUMMARY................................................................................................................... 4
1.0 INTRODUCTION.................................................................................................................. 5
2.0 APPLICATIONS ................................................................................................................... 5
3.0 INSTALLATION .................................................................................................................... 5
3.1 Location and Mounting ............................................................................................ 5
3.2 Input Power and Field Wiring .................................................................................. 5
3.3 Grounding................................................................................................................ 6
4.0 FAIL-SAFE SELECTION ...................................................................................................... 7
4.1 Description .............................................................................................................. 7
4.2 Fail-Safe High.......................................................................................................... 7
4.3 Fail-Safe Low .......................................................................................................... 7
5.0 CALIBRATION...................................................................................................................... 8
5.1 Calibration Verification............................................................................................. 8
5.2 Calibration Procedure ............................................................................................. 8
5.3 Sensitivity Selection ............................................................................................... 8
6.0 EXTERNAL VIEW LED ....................................................................................................... 9
7.0 TIME DELAY ........................................................................................................................ 9
8.0 WARRANTY AND CUSTOMER SERVICE ........................................................................ 10
9.0 DISPOSAL ........................................................................................................................... 10
FIGURE 1.1 ................................................................................................................................ 11
FIGURE 1.2 ................................................................................................................................ 12
FIGURE 2.1 ................................................................................................................................ 13
DECLARATION OF CONFORMITY ........................................................................................... 14

925-0290 Rev A 3
PROCAP I
GENERAL SPECIFICATIONS
Supply: 24 TO 240 VAC or VDC
Load: 4VA
Ambient Temperature: (Electronics) -20°F to +145°F (-28°C to +62°C)
Sensitivity: 1 picofarad
Enclosure: PROCAP I Type 4X, 5, 9, & 12 (HAZ LOC CL II E, F, & G)
PROCAP I X Type 4X, 5, 7, 9, & 12 (HAZ LOC CL I C&D; CL II E,F, & G)
Fuse F1: 0.4A 250V Time Delay 5 X 20MM BinMaster P/N 328-0064
Relay Contact Rating: DPDT contacts; 10 Amps 250 VAC maximum
Fail-Safe: Switch selectable "High" or "Low" level modes.
Calibration: Set when probe is uncovered:
COARSE adjust; single turn potentiometer
FINE adjust; single turn potentiometer
Status Indicator: Internal LED indicates material in contact with probe.
PROCAP I has an external LED also
Time Delay: Independently adjustable up to 30 seconds
Probe Shield: Automatically compensates for material buildup on the
probe
Mounting: 1 1/4” or 3/4" NPT - When used in liquid applications the
stainless steel threads must be used to insure a proper
seal.
Conduit Entries: 3/4" NPT

9250290 Rev A
SAFETY SUMMARY
Review the following safety precautions to avoid injury and prevent damage to the equipment.
The product should be installed, commissioned, and maintained by qualified and authorized personnel only.
Install according to installation instructions and comply with all National and Local codes.
Use electrical wire that is sized and rated for the maximum voltage and current of the application.
Properly ground the enclosure to an adequate earth ground.
Observe all terminal and relay contact ratings as called out on the nameplate and in the installation manual.
Insure that the enclosure cover is in place and secured tightly during normal operation.
If this product is used in a manner not specified by the manufacturer the safety protection could be compromised.
Safety Terms and Symbols
WARNING: Warning statements identify conditions or practices that could result in injury or loss of life. Risk of
electrical shock.
CAUTION: Caution statements identify conditions or practices that could result in damage to this product or other
property.

925-0290 Rev A 5
1.0 INTRODUCTION
The Bin-Master PROCAP I is a point level control used to detect the presence or absence of solids or liq-
uids. The PROCAP I operates on the capacitance principle and incorporates a "Quick Set" feature to sim-
plify calibration. The PROCAP I includes switch selectable Fail-Safe output contacts and "Probe Shield"
technology for ignoring material build up.
Upon installation the PROCAP I is set up with the probe uncovered. The "Quick Set" feature allows simple
setup and calibration of the unit to achieve low, medium, or high sensitivity settings. The PRO CAP I has
an adjustable time delay for monitoring covered or uncovered conditions.
2.0 APPLICATIONS
For applications in pressurized vessels up to 500 PSI, the PROCAP I should be mounted with the 3/4"
or 1 1/4” stainless steel threads. When the PROCAP I is used in liquid applications the stainless steel
threads must be used to insure a proper seal.
Process temperatures for the various probes are:
1. Delrin Sleeved Stainless Steel: 250 Deg. F
2. Teflon Sleeved Stainless Steel: 500 Deg. F
3. Flush Mount (Polyethylene): 180 Deg. F
4. Flush Mount (Teflon): 450 Deg. F
5. Bare Stainless Steel: 250 Deg. F
6. Flexible Extension: 250 Deg. F
The Delrin or Teflon sleeved stainless steel probe should be used when mounting the PROCAP I In a
highly corrosive atmosphere. Confirm that the sleeve material is rated for use in the corrosive nature of the
application. Bare metal Stainless Steel probes are suitable for granular or liquid material that is not electri-
cally conductive. For electrically conductive material, in some applications, a coated (insulated) probe may
need be used for the unit to effectively sense the material. Contact BinMaster if there are any questions
about which probe to use for a specific application.
3.0 INSTALLATION
3.1 Location and Mounting
The probe should be located out of the direct flow of material. The PROCAP I is designed to mount utiliz-
ing either a 1 1/4" NPT coupling or 3/4" NPT coupling. When the PROCAP I is used with the flush mount
probe, the flush mount probe mounts directly to the vessel wall using a 7" diameter bolt circle. Refer to Fig-
ures 1.1 and 1.2 for illustrations on installation, mounting options, and different probe combinations.
3.2 Input Power and Field Wiring
The PROCAP I can be powered from either an AC or a DC power source with a voltage range from 24 up
to 240 volts. The power source can be either grounded or ungrounded. If powered from a grounded
source the grounded circuit conductor should be connected to the N terminal. In this way the ungrounded
circuit conductor is fused on the circuit board. A disconnecting means should be provided to disconnect
incoming power to this device. Field wiring should conform to all national and local electrical codes and
any other agency or authority having jurisdiction over the installation. For power input, use wire that is
sized and rated for the maximum voltage and current as per equipment ratings and has a temperature rat-
ing of at least 70 degrees C For relay output, use wire that is sized and rated for the maximum voltage and
current as per application, up to 250VAC 10A, and has a temperature rating of at least 70 degrees C .
Installation shall be done by qualified personnel.

925-0290 Rev A 6
3.3 Grounding
An equipment grounding connection (earth ground) must be supplied to the unit for safety and to insure
proper operation of the unit. A green ground screw is provided in the enclosure for connection of an equip-
ment grounding conductor as shown in the figure above. This unit uses earth ground as a reference for
calibration and detecting a covered or uncovered condition. Therefore, the vessel in which the PROCAP I
is mounted must be made of metal, or a suitable metal ground plate must be installed where the PROCAP
I is mounted.
1NC
1C
1NO
2NC
2NO
2C
N
L
COARSE FINE
INTERNAL
COVERED
LED
EXTERNAL
COVERED LED
Ground wire
DELAY
COVEREDUNCOVERED
FUSE

925-0290 Rev A 7
4.0 FAIL-SAFE SELECTION
4.1 Description
A Fail-Safe condition means that the relay contact positions are set up so that in the event of a power fail-
ure the relay will be de-energized and the contacts will indicate a condition that is deemed safe for the
application. (Refer to FIGURE 2.1 for the location of the Fail-Safe selection switch.)
4.2 Fail-Safe High
Fail-Safe High means that the relay will be energized when the probe is uncovered and will de-energize
when the probe is covered. In this mode, a power failure will cause the relay contacts to indicate that the
probe is covered whether it is or not.
4.3 Fail-Safe Low
Fail-Safe Low means that the relay will be de-energized when the probe is uncovered and will energize
when the probe is covered. In this mode, a power failure will cause the relay contacts to indicate that the
probe is uncovered whether it is or not.
NC C NO NO C NC
1 1 1 2 2 2
NC C NO NO C NC
111222
UNCOVERED COVERED
FAIL SAFE "HIGH" RELAY CONTACT POSITION
NC C NO NO C NC
1 1 1 2 2 2
NC C NO NO C NC
111222
UNCOVERED COVERED
FAIL SAFE "LOW" RELAY CONTACT POSITION

925-0290 Rev A 8
5.0 CALIBRATION
The PROCAP I Quick Set calibration uses two single turn potentiometers making calibration very simple.
One potentiometer labeled COARSE is used to compensate for the capacitance of the empty vessel. The
other potentiometer labeled FINE is used to set in the desired sensitivity. Refer to FIGURE 2.1 for the
location of these potentiometers on the printed circuit board.
NOTE: Accurate calibration requires that the appropriate probe is attached to the PROCAP I and
the unit installed in the vessel. The probe must be UNCOVERED and material well below it.
5.1 Calibration Verification
Following calibration adjustment, the sensitivity setting that you have selected should be checked by veri-
fying that the PRO CAP II senses a covered probe condition with your material.
5.2 Calibration Procedure
WARNING: During Calibration, this unit will need to be energized. Care should be taken to not
come in contact with any live voltage in the area around the terminal blocks and fuse!
CAUTION: The Calibration potentiometers are delicate electronic devices, do not use excessive
force when adjusting.
1. Turn both the COARSE and FINE potentiometers fully counter clockwise (CCW). The internal COV-
ERED indicator light should be OFF.
2. Turn the COARSE potentiometer slowly clockwise (CW) to the point where the COVERED indicator
light just turns ON and stays on.
3. Turn the FINE potentiometer slowly clockwise (CW) until the COVERED indicator light just turns OFF.
(If the COARSE potentiometer has been carefully adjusted, this should occur when the FINE potenti-
ometer is between the 8 and 10 O'clock position.) Now continue to turn the FINE potentiometer clock-
wise (CW) to the desired sensitivity setting.
HIGH sensitivity: 1/16 turn
MEDIUM sensitivity: 1/8 turn
LOW sensitivity: 1/4 to 1/2 turn
NOTE: It may be convenient to think of the FINE potentiometer as a clock face and envision the distance
between consecutive hour numbers. Turning the FINE potentiometer clockwise one hour position past the
point at which the COVERED indicator just turns OFF would provide a high sensitivity setting. Two-hour
positions past that point would provide a medium sensitivity setting..
5.3 Sensitivity Selection
Sensitivity Setting (typical) Dielectric Constant of Material
HIGH sensitivity: 4 or less
MEDIUM sensitivity: 4 to 10
LOW sensitivity: 10 or higher

925-0290 Rev A 9
6.0 EXTERNAL VIEW LED
PROCAP I models include an external view LED. PROCAP I X models do not have an external view LED.
This LED provides an externally visible status indication for the PROCAP I. When this LED is flashing, it
indicates a normal Uncovered condition. When this LED is ON steady, it indicates a Covered condition.
When this LED is OFF, it indicates a power failure or fault condition. This external LED is affected by any
time delay set for the relay contacts.
7.0 TIME DELAY
The PROCAP I has two single turn adjustable time delay potentiometers. The Delay potentiometer labeled
UNCVR is for applying a time delay on going from a covered indication to an uncovered indication.
The Delay potentiometer labeled CVR is for applying a time delay on going from an uncovered indication
to an covered indication. Each potentiometer can be set for a time delay of up to approximately 30 sec-
onds. This time delay affects the relay contacts and external LED only. The internal Covered LED
will immediately respond to a change in covered or uncovered condition regardless of the time delay set-
ting.
Minimum time delay is when the DELAY potentiometers are set fully counter-clockwise. (Refer to the Fig-
ure below for the location of the DELAY potentiometers.) Maximum time delay is with the DELAY potenti-
ometer set fully clockwise.
DELAY COARSE FINE
INTERNAL
COVERED
LED
EXTERNAL
COVERED LED
COVEREDUNCOVERED

925-0290 Rev A 10
8.0 WARRANTY AND CUSTOMER SERVICE
8.1 Limited Warranty
The manufacturer warrants this equipment for two (2) years according to the following terms:
1.) This warranty extends to the original purchaser only and commences on the date
of original purchase. The original purchaser must mail to the manufacturer the "Warranty
Registration" card to confirm the equipment purchase. Failure to do so may void the war-
ranty.
2.) The manufacturer will repair or replace any part of this equipment found to be
defective, provided such part is delivered prepaid, to the factory. Manufacturer's obligation
is limited to the cost of material and labor to repair or replace and does not include trans-
portation expenses.
3.) This warranty shall not apply to any product that has, in our judgment, been tam-
pered with, altered, subject to misuse, neglect or accident. In addition, the warranty does
not extend to repairs made necessary by normal wear.
4.) This warranty is in lieu of all other warranties, expressed or implied.
EXPLOSION HAZARD-SUBSTITUTION OF COMPONENTS MAY
IMPAIR INTRINSIC SAFETY
9.0
DISPOSAL
This product consists of materials that can be recycled by specialized recycling companies. It
uses recyclable materials and the electronics is designed to be easily separated. Consult local
authorities for proper disposal locations.

925-0290 Rev A 11
Figure 1.1
When the PROCAP I is used for liquid applications the stainless steel threads must be
used to insure a proper seal.

925-0290 Rev A 12
Figure 1.2

9250290 Rev A
FINE POT.
COARSE POT.
FAIL SAFE
SWITCH
TIME DELAY
POT.
UNCVR
CVR
FIGURE 2.1

9250290 Rev A
Declaration of Conformity
BinMaster declares that the PROCAP I and PROCAP IX level control devices as listed below
comply with the following directives and harmonized standards. This product if installed,
operated and maintained as described in this manual will provide a safe and reliable point
level control device for bulk solid materials.
Product: Capacitance point level control device.
Models: PROCAP I and PROCAP IX
All test reports and documentation are held and can be obtained from BinMaster.

9250290 Rev A
Mounting Instructions
Always Mount Unit With
Conduit opening down
When installing this level indicator in environments where it is possible for
moisture or moist air to enter the enclosure through the electrical conduit, the
conduit opening should be sealed with a duct seal compound or putty
appropriate for the purpose.
Conduit Seal

По вопросам продаж и поддержки обращайтесь:
Архангельск (8182)63‐90‐72
Астана (7172)727‐132
Белгород (4722)40‐23‐64
Брянск (4832)59‐03‐52
Владивосток (423)249‐28‐31
Волгоград (844)278‐03‐48
Вологда (8172)26‐41‐59
Воронеж (473)204‐51‐73
Екатеринбург (343)384‐55‐89
Иваново (4932)77‐34‐06
Ижевск (3412)26‐03‐58
Казань (843)206‐01‐48
Калининград (4012)72‐03‐81
Калуга (4842)92‐23‐67
Кемерово (3842)65‐04‐62
Киров (8332)68‐02‐04
Краснодар (861)203‐40‐90
Красноярск (391)204‐63‐61
Курск (4712)77‐13‐04
Липецк (4742)52‐20‐81
Магнитогорск (3519)55‐03‐13
Москва (495)268‐04‐70
Мурманск (8152)59‐64‐93
Набережные Челны (8552)20‐53‐41
Нижний Новгород (831)429‐08‐12
Новокузнецк (3843)20‐46‐81
Новосибирск (383)227‐86‐73
Орел (4862)44‐53‐42
Оренбург (3532)37‐68‐04
Пенза (8412)22‐31‐16
Пермь (342)205‐81‐47
Ростов‐на‐Дону (863)308‐18‐15
Рязань (4912)46‐61‐64
Самара (846)206‐03‐16
Санкт‐Петербург (812)309‐46‐40
Саратов (845)249‐38‐78
Смоленск (4812)29‐41‐54
Сочи (862)225‐72‐31
Ставрополь (8652)20‐65‐13
Тверь (4822)63‐31‐35
Томск (3822)98‐41‐53
Тула (4872)74‐02‐29
Тюмень (3452)66‐21‐18
Ульяновск (8422)24‐23‐59
Уфа (347)229‐48‐12
Челябинск (351)202‐03‐61
Череповец (8202)49‐02‐64
Ярославль (4852)69‐52‐93
Единый адрес: [email protected] Веб‐сайт: www.binmaster.nt-rt.ru
This manual suits for next models
2
Table of contents
Other Bin Master Measuring Instrument manuals

Bin Master
Bin Master MAXIMA+ Series User manual

Bin Master
Bin Master DPM-500 User manual

Bin Master
Bin Master 1000 User manual

Bin Master
Bin Master NCR-80 User manual

Bin Master
Bin Master DPM-500 User manual

Bin Master
Bin Master CNCR-210 User manual

Bin Master
Bin Master BMRX Series User manual

Bin Master
Bin Master SMARTBOB TS1 User manual

Bin Master
Bin Master BMRX-300 User manual

Bin Master
Bin Master GWR-2000 User manual