AST A150 User manual

AST A150
USER MANUAL
Accurate Sensors Technologies
Non-contact Infrared Pyrometers
We measure accurate temperature in extreme conditions
AST - Accurate Sensors Technologies
Misgav Industrial Park, Misgav 20173, Israel
Tel: +972-4-9990025, Fax: +972-4-9990031
Email: info@accuratesensors.com
Web: www.accuratesensors.com
(H)Relative energy(for two color pyrometer) The relative energy shows a signal weakening which can be caused by
contaminations of the optics or a viewing window or by dust in the field of view or a too small measuring object.
Relative energy shows the measured intensity compared to the intensity, a black body radiation source would have
at a determined ratio temperature of the pyrometer.
(I)Analog Range User can select the analog range from the option 4-20mA, 0-20 mA & 0-10V.
(J)Comm. Mode User can select the communication mode as per requirement [RS-232 / RS-485].
Note: For connection diagram from RS-232 to RS-485 & RS-485 to RS-232 Refer Page No:-9
To view parameters of multiple devices select the pyrometer name from the drop down list that appears at the top of
the screen.
Accurate Sensors Technologies
16
(lll) Reset Below Temperature(RBT) The user can set RBT within the limit of pyrometer sub range. This function sets
the temperature above which peak picker action starts. When the target temperature matches or is below the
selected value, the sensor indicates temperature without picking action.
Temp
Time
Without RBT
Temp
Time
RBT
RBT with Decay Rate

Index
1. Chapter - 1
General information
2. Chapter - 2 ....................................................................................................................... 2
Introduction
2.1 Application, range and working principle
2.2 Technical specification
2.3 Optics
3. Chapter - 3 ....................................................................................................................... 5
Installation of pyrometer
3.1 Installation
3.1.1 Location selection
3.1.2 Correct positioning of pyrometer
3.1.3 Mounting of pyrometer
3.1.4 Ambient temperature
3.1.5 Atmospheric condition
3.2 Mechanical accessories
3.3 Electrical installation
3.3.1 Connector pin assignment on the back of the pyrometer
3.3.2 Converter RS-232 ↔ RS-485
3.3.3 Display instrument
3.3.4 Display & parameterizer
4. Chapter - 4 ....................................................................................................................... 11
Software installation
4.1 Installation
4.2 Parameters in main screen.
6. Chapter - 6 ....................................................................................................................... 22
Calculate Spot size
7. Chapter - 7 ....................................................................................................................... 23
Serial communication protocol
....................................................................................................................... 1
5. Chapter - 5 ....................................................................................................................... 18
Android Application Software
5.1 Installation Process
5.2 Communication
5.3 Operation
5.4 Datalogging & Online graph
(H)Relative energy(for two color pyrometer) The relative energy shows a signal weakening which can be caused by
contaminations of the optics or a viewing window or by dust in the field of view or a too small measuring object.
Relative energy shows the measured intensity compared to the intensity, a black body radiation source would have
at a determined ratio temperature of the pyrometer.
(I)Analog Range User can select the analog range from the option 4-20mA, 0-20 mA & 0-10V.
(J)Comm. Mode User can select the communication mode as per requirement [RS-232 / RS-485].
Note: For connection diagram from RS-232 to RS-485 & RS-485 to RS-232 Refer Page No:-9
To view parameters of multiple devices select the pyrometer name from the drop down list that appears at the top of
the screen.
Accurate Sensors Technologies
16
(lll) Reset Below Temperature(RBT) The user can set RBT within the limit of pyrometer sub range. This function sets
the temperature above which peak picker action starts. When the target temperature matches or is below the
selected value, the sensor indicates temperature without picking action.
Temp
Time
Without RBT
Temp
Time
RBT
RBT with Decay Rate

We are pleased that you have chosen this high quality and highly efficient AST pyrometer for non-contact
temperature measurement.
Please read this manual carefully, step by step before performing any operation with the Pyrometer. It
contains all the necessary instructions for set up and operation of the pyrometer. When operating the instrument, it
is necessary to follow the general safety instructions.
1.1 Safety Measures
This section provides an overview about important safety regulations.
1.1.1 General
Each person working with the pyrometer must have read the user manual before operation. The Pyrometer
has only to be used for the purpose described in the manual.
1.1.2 Safety Precaution
The Pyrometer works only with a potential-free low voltage of range 24V DC. This voltage is not harmful for
the user.
1.1.3 Maintenance and use of Pyrometer
Pyrometer can be operated by the qualified person who has got instructions from the supervisor. It is strongly
prohibited to do technical modifications of the device without permission of the manufacturer.
1.1.4 Environmental Protection
The lens or its coating may contain harmful materials and hence it should not be disposed of with normal
waste.
1.1.5 Packaging and storage
Always use a shock-proof package for shipment of the pyrometer. It should be sealed to protect it against
humidity. Also protect the lens of the pyrometer with a cover. They should be stored at the temperature ranges from
-20° to +70° C.
1.1.6 Warranty
AST A150 instruments have a warranty of two years from the invoice date. AST will replace defective parts,
which arises from design errors or manufacturing faults. In case, if pyrometer is opened, disassembled or modified
then the guarantees will loss.
AST does not accept liability for any damage or losses which might occur, including consequential damages
and financial losses, as a result of use of the equipment.
1.1.7 Copyright ©
© All rights reserved. This document may contain proprietary information and shall be respected as a
proprietary document to AST with permission for review and usage given only to the rightful owner of the
equipment with which this document is associated.
Chapter - 1
General Information
1
(H)Relative energy(for two color pyrometer) The relative energy shows a signal weakening which can be caused by
contaminations of the optics or a viewing window or by dust in the field of view or a too small measuring object.
Relative energy shows the measured intensity compared to the intensity, a black body radiation source would have
at a determined ratio temperature of the pyrometer.
(I)Analog Range User can select the analog range from the option 4-20mA, 0-20 mA & 0-10V.
(J)Comm. Mode User can select the communication mode as per requirement [RS-232 / RS-485].
Note: For connection diagram from RS-232 to RS-485 & RS-485 to RS-232 Refer Page No:-9
To view parameters of multiple devices select the pyrometer name from the drop down list that appears at the top of
the screen.
Accurate Sensors Technologies
16
(lll) Reset Below Temperature(RBT) The user can set RBT within the limit of pyrometer sub range. This function sets
the temperature above which peak picker action starts. When the target temperature matches or is below the
selected value, the sensor indicates temperature without picking action.
Temp
Time
Without RBT
Temp
Time
RBT
RBT with Decay Rate

AST A150 model provides the advantage of non - contact temperature measurement of metallic surfaces,
graphite and ceramics. This instrument is designed for easy integration into standard 4 wire system. This format
combines the high accuracy of digital signal processing with the simple connection. The A150 pyrometer is used for
target temperature ranging from 75°C to 700°C.The electronic assembly is protected by IP65 rugged stainless steel
housing.
2.1 Application, Range and Working Principle
The applications in which AST pyrometers can be used are:
Standard Item supplied with AST A150 :-
1. AST pyrometer
2. Connection cable 12-core (length - 5 m)
3. Digital cable 3-core (length - 1.5 m)
4. USB Cable
5. AST software CD
6. Manual
7. Certificate of calibration
AST A150 is a specially designed highly accurate digital pyrometer operating at the near end of the infrared
spectrum and is therefore an excellent choice for measurement of ferrous and non ferrous metals above 100°C
because the emissivity of un-oxidized metal surfaces is higher at shorter wavelengths. The emissivity, analog output
sub-range, response time and peak picker can be preset ex-works or adjusted through software. This enables the
instrument to be adapted for various measuring tasks.
AST A150 is provided with USB 2.0 connector with cable. The pyrometer can be powered through USB port. In this
case no external power supply is required (+24V DC requires for operation of analog output, laser, RS- 232 / RS-485
output )
This pyrometer has solid body in stainless steel housing which provides high operation safety even in rough
industrial environment. Variety of working Distances with fixed focus available for different applications.
lInduction Heating
lCasting
lAnnealing
lWelding
lForging
lSintering
lMelting
lRolling mills
lHardening
2
Chapter-2
Introduction
118.00 mm
34.50 mm
Ø49.50 mm
Ø48.00 mm
Binder
Connector
AST A150
14.00 mm
(H)Relative energy(for two color pyrometer) The relative energy shows a signal weakening which can be caused by
contaminations of the optics or a viewing window or by dust in the field of view or a too small measuring object.
Relative energy shows the measured intensity compared to the intensity, a black body radiation source would have
at a determined ratio temperature of the pyrometer.
(I)Analog Range User can select the analog range from the option 4-20mA, 0-20 mA & 0-10V.
(J)Comm. Mode User can select the communication mode as per requirement [RS-232 / RS-485].
Note: For connection diagram from RS-232 to RS-485 & RS-485 to RS-232 Refer Page No:-9
To view parameters of multiple devices select the pyrometer name from the drop down list that appears at the top of
the screen.
Accurate Sensors Technologies
16
(lll) Reset Below Temperature(RBT) The user can set RBT within the limit of pyrometer sub range. This function sets
the temperature above which peak picker action starts. When the target temperature matches or is below the
selected value, the sensor indicates temperature without picking action.
Temp
Time
Without RBT
Temp
Time
RBT
RBT with Decay Rate

2.2 Technical Specifications
Note: - After power supply initialization, keep pyrometer under stable temperature condition for
25-30 minutes to get above stated accuracy.
Laser should be used only for targeting purpose. In normal measuring laser should be
turned off to get correct measurements.
3
Measured Temperature Ranges
(sub range adjustable)
Spectral Range, µm
Photodetector Type
Response Time
Emissivity
Distance to Spot size Ratio
Model AST A150
75°C - 700°C
2 to 2.6 µm (peak at 2.3 µm)
Extended InGaAs
2msec. Adjustable upto 10 sec
0.1….1 Adjustable
Digital Output
Analog Output
Bluetooth V2.0, RS-232 or RS-485 (Isolated), User Selectable
4-20mA or 0-20mA or 0-10 V, User selectable
Sighting Laser pilot light (PL)
Operating temperature range 0°C........+70°C, 0°C........+200°C (With cooling jacket)
Protection class IP 65
Isolation Power supply and digital output and analog output are galvanically
isolated against each other
Operating humidity
Dimensions/Weight
Adjustable Parameters via software
Unlimited
Dia. = Ø49.5 mm; Length = 118 mm / Weight = 0.6 kg
Response time, Peak picker, Analog output type, sub-range,
emissivity, interval of data storage etc.
50 : 1 (75°C - 700°C)
Power 24V DC ±1V, Ripple must be less than 50 mV
Laser Power <1mW
Storage temprature -20°........+70°C
Repeatability 0.1% reading in °C + 1°C(Ɛ = 1, t90 = 1s, T amb. = 30°C)
Uncertainty
Up to 400°C: 3°C
above 400°C: 0.5% of measured value in °C + 1°C
(Ɛ = 1, t 90 = 1 s, T amb. = 30°C; the pyrometer must be operate at least 30
min before these value are valid)
(H)Relative energy(for two color pyrometer) The relative energy shows a signal weakening which can be caused by
contaminations of the optics or a viewing window or by dust in the field of view or a too small measuring object.
Relative energy shows the measured intensity compared to the intensity, a black body radiation source would have
at a determined ratio temperature of the pyrometer.
(I)Analog Range User can select the analog range from the option 4-20mA, 0-20 mA & 0-10V.
(J)Comm. Mode User can select the communication mode as per requirement [RS-232 / RS-485].
Note: For connection diagram from RS-232 to RS-485 & RS-485 to RS-232 Refer Page No:-9
To view parameters of multiple devices select the pyrometer name from the drop down list that appears at the top of
the screen.
Accurate Sensors Technologies
16
(lll) Reset Below Temperature(RBT) The user can set RBT within the limit of pyrometer sub range. This function sets
the temperature above which peak picker action starts. When the target temperature matches or is below the
selected value, the sensor indicates temperature without picking action.
Temp
Time
Without RBT
Temp
Time
RBT
RBT with Decay Rate

2.3 Optics
The pyrometer measure temperature by receiving heat radiation from the object whose temperature has to be
measured. This heat radiation is passes through the lens to the sensor and is then converted to an electrical
signal. The farther the measured object is from the pyrometer, the larger the area that will be measured by the
pyrometer. Depending on customer need, the pyrometer is designed wit h fixed optics.
Some fixed optics focus is as below: -
4
distance (WD) mentioned on the
pyrometer.
If the pyrometer is not
installed at manufactured working
distance (WD) then spot size at
actual installed distance should be
calculated. For example, if factory
made working distance in 500mm
(FOV 50 : 1, Optics 500), then spot
size is 10mm (as given in figure). If
user installed this pyrometer at
800mm, user have to calculate by
the given method below.
* Manufactured working
Case-I: If installed working distance is greater than manufactured working distance
Installed Spot Size = Installed working distance (Case I)
WD x (S + A) - A
Case-II: If installed working distance is smaller than manufactured working distance
Installed Spot Size = Installed working distance (Case II)
WD x (S - A) + A
A
WD
(Case I)*
*Installed Working Distance
(Case II)*
Defocused Spot Ø
focused Spot Ø (S)
Pyrometer object
lens
Manufactured working
distance WD (mm) 350 500 700 Aperture(A) 800 1000 1500 2000 2500 3000 5000
7 10 14 16 20 30 40 50 60 100 18
Spot Size
(50:1)
(mm)
21
700
Spot Size (mm)
Distance (mm)
Optics 500
13 10 16 38
300 500 600 1000
(H)Relative energy(for two color pyrometer) The relative energy shows a signal weakening which can be caused by
contaminations of the optics or a viewing window or by dust in the field of view or a too small measuring object.
Relative energy shows the measured intensity compared to the intensity, a black body radiation source would have
at a determined ratio temperature of the pyrometer.
(I)Analog Range User can select the analog range from the option 4-20mA, 0-20 mA & 0-10V.
(J)Comm. Mode User can select the communication mode as per requirement [RS-232 / RS-485].
Note: For connection diagram from RS-232 to RS-485 & RS-485 to RS-232 Refer Page No:-9
To view parameters of multiple devices select the pyrometer name from the drop down list that appears at the top of
the screen.
Accurate Sensors Technologies
16
(lll) Reset Below Temperature(RBT) The user can set RBT within the limit of pyrometer sub range. This function sets
the temperature above which peak picker action starts. When the target temperature matches or is below the
selected value, the sensor indicates temperature without picking action.
Temp
Time
Without RBT
Temp
Time
RBT
RBT with Decay Rate

5
Chapter-3
Installation of the Pyrometer
3.1 Installation
3.1.1 Location selection
Qualified operating personnel should do the installation. Location should be good enough so that pyrometer
should get continuous infrared radiation.
Pyrometer distance from object is according to below points:-
1. Pyrometer spot size should be small than object size.
2. Know your pyrometer spot size according to point 2.3.
3. Pyrometer ambient temperature should be within 0° to 70°C (read 3.1.4).
3.1.2 Correct Positioning of the pyrometer
With pilot light (PL)
A laser targeting light will help to correct the position of the pyrometer. It is recommended that laser should
be switched off while measurement. It will increase the life of laser. To avoid measuring errors caused by a too big
spot size the pyrometer must be fixed in the correct measuring distance, so that the object under measurement fills
the spot size.
Note: The laser spot (PL) is only for indication of measuring spot, not exactly shows the measuring area.
3.1.3 Mounting of pyrometer
To install the sensor head at the place of measurement a mounting support is supplied as an accessory,
after losing the clamp screws, it can be fastened correctly.
3.1.4 Ambient temperature
The allowed operation temperature for the pyrometer is 0°C to +70°C. But the temperature at fiber optic &
optical head is 250°C max. Therefore if sensor head is to be used above 250°C water cooling jacket with built in
air purge unit is used otherwise it may damage the pyrometer. The ambient temperature is dependent on the
temperature and flow rate of cooling water. Details of air purge & water cooling jacket is given in point 3.2.
3.1.5 Atmospheric conditions
The pyrometer cannot receive the full infrared energy for the measurement if atmospheric
conditions like smoke, dust or steam are present and hence it result in measuring error. An air purge unit can be
helpful to avoid contamination such as dust and humidity on the lens. The air supplied should be at normal
temperature with oil & moisture free. The air purge generates an air stream shaped like a cone and blows particles
from the lens area.
(H)Relative energy(for two color pyrometer) The relative energy shows a signal weakening which can be caused by
contaminations of the optics or a viewing window or by dust in the field of view or a too small measuring object.
Relative energy shows the measured intensity compared to the intensity, a black body radiation source would have
at a determined ratio temperature of the pyrometer.
(I)Analog Range User can select the analog range from the option 4-20mA, 0-20 mA & 0-10V.
(J)Comm. Mode User can select the communication mode as per requirement [RS-232 / RS-485].
Note: For connection diagram from RS-232 to RS-485 & RS-485 to RS-232 Refer Page No:-9
To view parameters of multiple devices select the pyrometer name from the drop down list that appears at the top of
the screen.
Accurate Sensors Technologies
16
(lll) Reset Below Temperature(RBT) The user can set RBT within the limit of pyrometer sub range. This function sets
the temperature above which peak picker action starts. When the target temperature matches or is below the
selected value, the sensor indicates temperature without picking action.
Temp
Time
Without RBT
Temp
Time
RBT
RBT with Decay Rate

6
Air Purge Unit
(Reference no: 8000-04)
3.2 Mechanical accessories
Water pressure : < 10 bar
Air pressure : < 0.5 bar
Air consumption : 2...3 m³/h
Ambient temperature : < 200°C
Metal : Stainless steel
Weight : 2.75 Kg
Water Cooling Jacket with Adjustable Flange PL
(Reference no: 8000-02)
M40 x 1.5
60
48
156
80
287
54
75
18
24.50
Ø68
Ø63
Ø20
30.60 Grub Screw for
fixing Pyrometer
For 6 mm inside
Diameter hose
Adjustable Mounting Stand
(Reference no: 8000-07)
60°
R30
75
60
5
110
60
66
86
3 x M6 THRU AT 120°
Air pressure : < 0.5 bar
Air consumption : 2...3 m³/h
Metal : Aluminium
Weight : 0.15 Kg
Dry, clean air (Oil and dust free)
Metal : Stainless Steel
Weight : 0.9 Kg
(H)Relative energy(for two color pyrometer) The relative energy shows a signal weakening which can be caused by
contaminations of the optics or a viewing window or by dust in the field of view or a too small measuring object.
Relative energy shows the measured intensity compared to the intensity, a black body radiation source would have
at a determined ratio temperature of the pyrometer.
(I)Analog Range User can select the analog range from the option 4-20mA, 0-20 mA & 0-10V.
(J)Comm. Mode User can select the communication mode as per requirement [RS-232 / RS-485].
Note: For connection diagram from RS-232 to RS-485 & RS-485 to RS-232 Refer Page No:-9
To view parameters of multiple devices select the pyrometer name from the drop down list that appears at the top of
the screen.
Accurate Sensors Technologies
16
(lll) Reset Below Temperature(RBT) The user can set RBT within the limit of pyrometer sub range. This function sets
the temperature above which peak picker action starts. When the target temperature matches or is below the
selected value, the sensor indicates temperature without picking action.
Temp
Time
Without RBT
Temp
Time
RBT
RBT with Decay Rate

7
3.3. Electrical Accessories
Power Supply Unit
(Reference no: 9000-02)
AST A150 is powered by 24V DC (well stabilized ripple max
50mV).The input power supply is 110/230v AC check the polarity
before connecting the device
3.3.1 Connector pin assignment on the back of the pyrometer
RS-232 Connection
12-Core Cable
(Reference no: 7002-02)
Mounting Clamp
(Reference no: 8000-05)
110
70
50
70
3
Metal : Stainless Steel
Weight : 0.45 Kg
(H)Relative energy(for two color pyrometer) The relative energy shows a signal weakening which can be caused by
contaminations of the optics or a viewing window or by dust in the field of view or a too small measuring object.
Relative energy shows the measured intensity compared to the intensity, a black body radiation source would have
at a determined ratio temperature of the pyrometer.
(I)Analog Range User can select the analog range from the option 4-20mA, 0-20 mA & 0-10V.
(J)Comm. Mode User can select the communication mode as per requirement [RS-232 / RS-485].
Note: For connection diagram from RS-232 to RS-485 & RS-485 to RS-232 Refer Page No:-9
To view parameters of multiple devices select the pyrometer name from the drop down list that appears at the top of
the screen.
Accurate Sensors Technologies
16
(lll) Reset Below Temperature(RBT) The user can set RBT within the limit of pyrometer sub range. This function sets
the temperature above which peak picker action starts. When the target temperature matches or is below the
selected value, the sensor indicates temperature without picking action.
Temp
Time
Without RBT
Temp
Time
RBT
RBT with Decay Rate

8
12-pin Binder connector RS-232 communication connection
Male
Female
L
C
BA
J
K
G
H
M
F
E
D
LC
J
B
A
K
G
H
M
FE
D
D - Type 9 pin Female
Connector
LC
J
B
A
K
G
H
M
FE
D8
3
7
2
4
9
6
1
5
RxD
DGND
TxD
H
Pin
A
G
C
D
F
H
M
J
K
B
Colour
Red
White
Blue
Pink
Black
Yellow
Violet
Brown
Green
Yellow
Grey
Indication
+ 24 V DC In
Ground
TxD (RS-232) / D- (RS-485)
RxD (RS-232) / D+ (RS-485)
RS-232 communication
(-) 4-20mA / 0-20mA
(+) 4-20mA / 0-20mA
Short / Open both wires for laser
ON/OFF
- (0-10V)
+ (0-10V)
Used for
Power supply
RS-232/ communicationRS-485
Analog current output
Laser ON/OFF
Analog voltage output
E Blue-Yellow N/A
L Red-Grey N/A
Not used
3.3.2
USB 2.0 is provided for Parameter Setting and Data Logging. 2Mtr long USB cable is supplied with
pyrometer.
While USB 2.0 is connected RS-232 / RS-485 communication is not possible. LED light indicates USB 2.0
operation.
USB 2.0 Connection:
Open screw for USB 2.0
LED Light
(H)Relative energy(for two color pyrometer) The relative energy shows a signal weakening which can be caused by
contaminations of the optics or a viewing window or by dust in the field of view or a too small measuring object.
Relative energy shows the measured intensity compared to the intensity, a black body radiation source would have
at a determined ratio temperature of the pyrometer.
(I)Analog Range User can select the analog range from the option 4-20mA, 0-20 mA & 0-10V.
(J)Comm. Mode User can select the communication mode as per requirement [RS-232 / RS-485].
Note: For connection diagram from RS-232 to RS-485 & RS-485 to RS-232 Refer Page No:-9
To view parameters of multiple devices select the pyrometer name from the drop down list that appears at the top of
the screen.
Accurate Sensors Technologies
16
(lll) Reset Below Temperature(RBT) The user can set RBT within the limit of pyrometer sub range. This function sets
the temperature above which peak picker action starts. When the target temperature matches or is below the
selected value, the sensor indicates temperature without picking action.
Temp
Time
Without RBT
Temp
Time
RBT
RBT with Decay Rate

Power ON Pyrometer:
1. Connect the 12-core cable (Supplied with pyrometer) with pyrometer 12-pin binder connector given atthe back
side of pyrometer.
2. Connect the 12-core cable supply wire with +24V DC as given in table above.
3. Connect remaining wires as per your requirement (details given in table above).
4. Provide insulation for not used end points of 12-core cable.
5. Now, the pyrometer can be switch ON.
3.3.2 Converter RS-232 ↔ RS-485 :
The pyrometer can communicate with PC using RS-232 or RS-485.
RS-232 is used only for short distances.
RS-485 is well suited for long distance transmission. Standard on PC
is RS-232, so a converter is used which converts RS-485 to RS-232.
Converter RS-232 ↔ RS-485
9000-03)(Reference no:
RS-232 to RS-485 to RS-232 Connection
9
RS-485 to RS-232 Connection
(H)Relative energy(for two color pyrometer) The relative energy shows a signal weakening which can be caused by
contaminations of the optics or a viewing window or by dust in the field of view or a too small measuring object.
Relative energy shows the measured intensity compared to the intensity, a black body radiation source would have
at a determined ratio temperature of the pyrometer.
(I)Analog Range User can select the analog range from the option 4-20mA, 0-20 mA & 0-10V.
(J)Comm. Mode User can select the communication mode as per requirement [RS-232 / RS-485].
Note: For connection diagram from RS-232 to RS-485 & RS-485 to RS-232 Refer Page No:-9
To view parameters of multiple devices select the pyrometer name from the drop down list that appears at the top of
the screen.
Accurate Sensors Technologies
16
(lll) Reset Below Temperature(RBT) The user can set RBT within the limit of pyrometer sub range. This function sets
the temperature above which peak picker action starts. When the target temperature matches or is below the
selected value, the sensor indicates temperature without picking action.
Temp
Time
Without RBT
Temp
Time
RBT
RBT with Decay Rate

3.3.3 Display Instrument
Temperature Indicator
(Reference no: 9000-01)
Display & Parameterizer
P-120
(Reference no: 9001-01)
3.3.4 Display & Parameterizer
To display the measured temperature 7 segment digital
indicator is used. Device has following specifications:
Power supply I/P : 85 to 265 V, AC/DC
Analog I/P : 4….20 mA
Retransmission : 4….20 mA
Power supply O/P : 24V, DC
Display : 4 Digits
Alarm : 2
AST P-120 is a high precision LED indicator for non contact
temperature measurement. With indication of measured
temperature user can easily parameterize a connected AST digital
pyrometer without any PC.
Device has following specification:
Power supply : 100 – 240V, AC or 24V, DC
DC analog O/P : 0-20 mA or 4-20 mA
Digital interface : RS-232 or RS-485
Pilot light : Pilot light On/Off with keys
Multi-Pyrometer Connection
10
(H)Relative energy(for two color pyrometer) The relative energy shows a signal weakening which can be caused by
contaminations of the optics or a viewing window or by dust in the field of view or a too small measuring object.
Relative energy shows the measured intensity compared to the intensity, a black body radiation source would have
at a determined ratio temperature of the pyrometer.
(I)Analog Range User can select the analog range from the option 4-20mA, 0-20 mA & 0-10V.
(J)Comm. Mode User can select the communication mode as per requirement [RS-232 / RS-485].
Note: For connection diagram from RS-232 to RS-485 & RS-485 to RS-232 Refer Page No:-9
To view parameters of multiple devices select the pyrometer name from the drop down list that appears at the top of
the screen.
Accurate Sensors Technologies
16
(lll) Reset Below Temperature(RBT) The user can set RBT within the limit of pyrometer sub range. This function sets
the temperature above which peak picker action starts. When the target temperature matches or is below the
selected value, the sensor indicates temperature without picking action.
Temp
Time
Without RBT
Temp
Time
RBT
RBT with Decay Rate

Chapter - 4
Software Installation
The provided AST software “Infra 1.0.0” offers possibilities to connect three pyrometers simultaneously
for parameter setting, view real time graph, offline graph and to evaluate measuring data.
4.1 Installation
Install the pyrometer software using the installation guide file on CD ROM.After installation of the software;
Double click the application. It will open the screen of software.
Soft
4.2 Parameters in main screen
4.2.1 Communication
Communication between the AST pyrometer and the
software is implemented via a cable connected between
the pyrometer and the PC serial port. This enables the
acquisition and recording of data, as well as the transfer of
commands from the software application to the AST
pyrometer. Communication can be done by clicking on
“Communication panel" and select correct COM Port
address where pyrometer is connected. Also user has to
select address of the pyrometer(Example: Default 01/
printed on the pyrometer sticker). Then click on CONNECT
Button.
11
InfraSoft 1.0.0
(H)Relative energy(for two color pyrometer) The relative energy shows a signal weakening which can be caused by
contaminations of the optics or a viewing window or by dust in the field of view or a too small measuring object.
Relative energy shows the measured intensity compared to the intensity, a black body radiation source would have
at a determined ratio temperature of the pyrometer.
(I)Analog Range User can select the analog range from the option 4-20mA, 0-20 mA & 0-10V.
(J)Comm. Mode User can select the communication mode as per requirement [RS-232 / RS-485].
Note: For connection diagram from RS-232 to RS-485 & RS-485 to RS-232 Refer Page No:-9
To view parameters of multiple devices select the pyrometer name from the drop down list that appears at the top of
the screen.
Accurate Sensors Technologies
16
(lll) Reset Below Temperature(RBT) The user can set RBT within the limit of pyrometer sub range. This function sets
the temperature above which peak picker action starts. When the target temperature matches or is below the
selected value, the sensor indicates temperature without picking action.
Temp
Time
Without RBT
Temp
Time
RBT
RBT with Decay Rate

(A) Emissivity settings The emissivity can be set by clicking on
"Parameters” and select or type in the desired emissivity
directly in the description field. The emissivity value will be
transferred to pyrometer by hitting the "TAB " button.
(B) Match This parameter is used to match the known target
temperature value to be displayed by the pyrometer. The User
can enter the required temperature in window provided. This
value can be set by clicking on the “Match Temp.” parameter in
the software. The pyrometer automatically adjusts the
emissivity so that the temperature display indicates the same
temperature.
(C) Response time The desired response time can be chosen in
the panel Parameter by clicking the appropriate list box (as per
the values available in the drop box of response time). This
parameter is use to set the analog response time of pyrometer.
(D) Sub Range User can change the sub range of pyrometer in
the panel Parameter. Sub range must be within the basic range
of pyrometer, the minimum span between higher & lower range
is 51. Analog output will be automatically set according to the
sub-range by hitting "TAB" button.
(E) Sensor Type It shows pyrometer sensor type. User can
change sensor type from two color to single color and vice versa
(only applicable with two color pyrometer).
4.2.2 Temperature
It shows the temperature measured by the pyrometer
4.2.3 Parameter Setting
All user selectable device parameters can be set by using the software in the Panel "Parameter"
For communication of multiple pyrometers, select different com port and address.
Accurate Sensors Technologies
12
(H)Relative energy(for two color pyrometer) The relative energy shows a signal weakening which can be caused by
contaminations of the optics or a viewing window or by dust in the field of view or a too small measuring object.
Relative energy shows the measured intensity compared to the intensity, a black body radiation source would have
at a determined ratio temperature of the pyrometer.
(I)Analog Range User can select the analog range from the option 4-20mA, 0-20 mA & 0-10V.
(J)Comm. Mode User can select the communication mode as per requirement [RS-232 / RS-485].
Note: For connection diagram from RS-232 to RS-485 & RS-485 to RS-232 Refer Page No:-9
To view parameters of multiple devices select the pyrometer name from the drop down list that appears at the top of
the screen.
Accurate Sensors Technologies
16
(lll) Reset Below Temperature(RBT) The user can set RBT within the limit of pyrometer sub range. This function sets
the temperature above which peak picker action starts. When the target temperature matches or is below the
selected value, the sensor indicates temperature without picking action.
Temp
Time
Without RBT
Temp
Time
RBT
RBT with Decay Rate

(E)Switch off level% ( for two color pyrometer) The switch of level is the function that is used to avoid measurement
errors caused by signals, which are too low. Although factory default is set to 15 %, the switch off limit can be
adjusted between 2 and 50%.
(F)Unit User can change the measuring unit of temperature from “Centigrade" to “Fahrenheit" and vice versa.
(G)Peak Picker If the peak picker is switched on, the highest last temperature value will always be displayed and
stored. This feature is particularly useful when object temperature is not uniform across its dimension or the
pyrometer is not constantly viewing an object to be measured.
Peak Picker Functions :
Peak picker can be turn ON & OFF by using the software. When Peak picker is ON, the peak picker menu is enabled for
setting of the parameters like decay rate function, reset below temperature and peak picker delay.
Accurate Sensors Technologies
13
(I)Decay rate The Decay rate range is 0.00 to 166.66°C/sec. or 0.00
to 300°F/sec. depending upon °F/°C unit’s selection. The slowest
Decay rate is 0 degrees per sec. This feature helps to eliminate
erratic measurements and allows the peaked value to decay down
to lower process temperature values as they occur. Decay rate is set
to retain peak measured temperature value and ignore momentary
decreases in measured temperature.
(ll)Delay Time This function set the delay time in sec. before peak
picker function starts. The delay time is selectable in the range 0.02
to 10.00 sec. Zero (0) turns delay time OFF. This function is used to
delay the start of peaking action for upto 10 sec. following the
detection of leading edge of a new target.
Temp
Time
With Decay RateWithout Decay Rate
Temp
Time
Decay rate in
°C/sec.
Temp
Time
Delay Time (sec.)
(H)Relative energy(for two color pyrometer) The relative energy shows a signal weakening which can be caused by
contaminations of the optics or a viewing window or by dust in the field of view or a too small measuring object.
Relative energy shows the measured intensity compared to the intensity, a black body radiation source would have
at a determined ratio temperature of the pyrometer.
(I)Analog Range User can select the analog range from the option 4-20mA, 0-20 mA & 0-10V.
(J)Comm. Mode User can select the communication mode as per requirement [RS-232 / RS-485].
Note: For connection diagram from RS-232 to RS-485 & RS-485 to RS-232 Refer Page No:-9
To view parameters of multiple devices select the pyrometer name from the drop down list that appears at the top of
the screen.
Accurate Sensors Technologies
16
(lll) Reset Below Temperature(RBT) The user can set RBT within the limit of pyrometer sub range. This function sets
the temperature above which peak picker action starts. When the target temperature matches or is below the
selected value, the sensor indicates temperature without picking action.
Temp
Time
Without RBT
Temp
Time
RBT
RBT with Decay Rate

4.2.5 Record
Record is for continuous data logging. It records the measured temperature, emissivity with current date &
time.To start data logging click on start button. If user wants to record emissivity, click on record emissivity button.
After Clicking Start button window appears where user can specify the file name & location.
This screen shows the Model, basic range, serial number,
version, Head temperature, internal temperature, working
distance, spot size-apperture.
(H)Relative energy(for two color pyrometer) The relative energy shows a signal weakening which can be caused by
contaminations of the optics or a viewing window or by dust in the field of view or a too small measuring object.
Relative energy shows the measured intensity compared to the intensity, a black body radiation source would have
at a determined ratio temperature of the pyrometer.
(I)Analog Range User can select the analog range from the option 4-20mA, 0-20 mA & 0-10V.
(J)Comm. Mode User can select the communication mode as per requirement [RS-232 / RS-485].
Note: For connection diagram from RS-232 to RS-485 & RS-485 to RS-232 Refer Page No:-9
To view parameters of multiple devices select the pyrometer name from the drop down list that appears at the top of
the screen.
4.2.4 Device information
Pyrometer specific information will be displayed in the Info Panel
Accurate Sensors Technologies
14
(lll) Reset Below Temperature(RBT) The user can set RBT within the limit of pyrometer sub range. This function sets
the temperature above which peak picker action starts. When the target temperature matches or is below the
selected value, the sensor indicates temperature without picking action.
Temp
Time
Without RBT
Temp
Time
RBT
RBT with Decay Rate
(H)Relative energy(for two color pyrometer) The relative energy shows a signal weakening which can be caused by
contaminations of the optics or a viewing window or by dust in the field of view or a too small measuring object.
Relative energy shows the measured intensity compared to the intensity, a black body radiation source would have
at a determined ratio temperature of the pyrometer.
(I)Analog Range User can select the analog range from the option 4-20mA, 0-20 mA & 0-10V.
(J)Comm. Mode User can select the communication mode as per requirement [RS-232 / RS-485].
Note: For connection diagram from RS-232 to RS-485 & RS-485 to RS-232 Refer Page No:-9
To view parameters of multiple devices select the pyrometer name from the drop down list that appears at the top of
the screen.
Accurate Sensors Technologies
16
(lll) Reset Below Temperature(RBT) The user can set RBT within the limit of pyrometer sub range. This function sets
the temperature above which peak picker action starts. When the target temperature matches or is below the
selected value, the sensor indicates temperature without picking action.
Temp
Time
Without RBT
Temp
Time
RBT
RBT with Decay Rate

If user wants file in Spreadsheet format, user can export by choosing Excel Spreadsheet in file menu.
To record emissivity, click on Record Emissivity button.
Record will be saved as .txt format and the name of file will be user define.
Accurate Sensors Technologies
15
(H)Relative energy(for two color pyrometer) The relative energy shows a signal weakening which can be caused by
contaminations of the optics or a viewing window or by dust in the field of view or a too small measuring object.
Relative energy shows the measured intensity compared to the intensity, a black body radiation source would have
at a determined ratio temperature of the pyrometer.
(I)Analog Range User can select the analog range from the option 4-20mA, 0-20 mA & 0-10V.
(J)Comm. Mode User can select the communication mode as per requirement [RS-232 / RS-485].
Note: For connection diagram from RS-232 to RS-485 & RS-485 to RS-232 Refer Page No:-9
To view parameters of multiple devices select the pyrometer name from the drop down list that appears at the top of
the screen.
Accurate Sensors Technologies
16
(lll) Reset Below Temperature(RBT) The user can set RBT within the limit of pyrometer sub range. This function sets
the temperature above which peak picker action starts. When the target temperature matches or is below the
selected value, the sensor indicates temperature without picking action.
Temp
Time
Without RBT
Temp
Time
RBT
RBT with Decay Rate

To see previous record open the file by clicking on menu File open.
Screen containing historical trend & historical data will appear.
File will be stored in .xls format named as “export”. This “export.xls” file will be saved where the software is installed.
Accurate Sensors Technologies
16
(H)Relative energy(for two color pyrometer) The relative energy shows a signal weakening which can be caused by
contaminations of the optics or a viewing window or by dust in the field of view or a too small measuring object.
Relative energy shows the measured intensity compared to the intensity, a black body radiation source would have
at a determined ratio temperature of the pyrometer.
(I)Analog Range User can select the analog range from the option 4-20mA, 0-20 mA & 0-10V.
(J)Comm. Mode User can select the communication mode as per requirement [RS-232 / RS-485].
Note: For connection diagram from RS-232 to RS-485 & RS-485 to RS-232 Refer Page No:-9
To view parameters of multiple devices select the pyrometer name from the drop down list that appears at the top of
the screen.
Accurate Sensors Technologies
16
(lll) Reset Below Temperature(RBT) The user can set RBT within the limit of pyrometer sub range. This function sets
the temperature above which peak picker action starts. When the target temperature matches or is below the
selected value, the sensor indicates temperature without picking action.
Temp
Time
Without RBT
Temp
Time
RBT
RBT with Decay Rate

Note:
After connecting the pyrometer right click on the graph screen and choose "Set to default" option from pop-up
menu.
Temperature graph of connected pyrometer is displayed on right side of the screen. This shows the measured
temperature corresponding to the time.
Accurate Sensors Technologies
17
(H)Relative energy(for two color pyrometer) The relative energy shows a signal weakening which can be caused by
contaminations of the optics or a viewing window or by dust in the field of view or a too small measuring object.
Relative energy shows the measured intensity compared to the intensity, a black body radiation source would have
at a determined ratio temperature of the pyrometer.
(I)Analog Range User can select the analog range from the option 4-20mA, 0-20 mA & 0-10V.
(J)Comm. Mode User can select the communication mode as per requirement [RS-232 / RS-485].
Note: For connection diagram from RS-232 to RS-485 & RS-485 to RS-232 Refer Page No:-9
To view parameters of multiple devices select the pyrometer name from the drop down list that appears at the top of
the screen.
Accurate Sensors Technologies
16
(lll) Reset Below Temperature(RBT) The user can set RBT within the limit of pyrometer sub range. This function sets
the temperature above which peak picker action starts. When the target temperature matches or is below the
selected value, the sensor indicates temperature without picking action.
Temp
Time
Without RBT
Temp
Time
RBT
RBT with Decay Rate

Chapter - 5
Android Application Software
5.1 Installation Process
AST provides “infrasmart .apk” for android phone. This apk is in software CD.
Special Note: - This apk is compatible with android version 4.0 and above
Following are the predefined steps to install apk in a mobile.
(i) Copy apk file from CD in a mobile.
(ii) Click on “infrasmart .apk” (Fig. 5.1)
(iii) Next screen will be as shown below, click on settings (Fig. 5.2)
(IV) “On” unknown sources, Pop up window will appear as below. (Fig. 3)
Tick mark on “Allow this installation only “check box (indicated by arrow in below screen) and then click “OK”. (Fig.
5.4)
(V) Next click on Install tab (Fig. 5.5)
5.
(Fig. 5.1)
(Fig. 5.2)
(Fig. 5.4)(Fig. 5.3) (Fig. 5.5)
Accurate Sensors Technologies
18
(H)Relative energy(for two color pyrometer) The relative energy shows a signal weakening which can be caused by
contaminations of the optics or a viewing window or by dust in the field of view or a too small measuring object.
Relative energy shows the measured intensity compared to the intensity, a black body radiation source would have
at a determined ratio temperature of the pyrometer.
(I)Analog Range User can select the analog range from the option 4-20mA, 0-20 mA & 0-10V.
(J)Comm. Mode User can select the communication mode as per requirement [RS-232 / RS-485].
Note: For connection diagram from RS-232 to RS-485 & RS-485 to RS-232 Refer Page No:-9
To view parameters of multiple devices select the pyrometer name from the drop down list that appears at the top of
the screen.
Accurate Sensors Technologies
16
(lll) Reset Below Temperature(RBT) The user can set RBT within the limit of pyrometer sub range. This function sets
the temperature above which peak picker action starts. When the target temperature matches or is below the
selected value, the sensor indicates temperature without picking action.
Temp
Time
Without RBT
Temp
Time
RBT
RBT with Decay Rate
Table of contents
Other AST Measuring Instrument manuals

AST
AST TI-1800 User manual

AST
AST Embedded System Products TSA6000-USB User manual

AST
AST EL50 User manual

AST
AST Krystal Bond User manual

AST
AST AccuGauge AG200 User manual

AST
AST P390 User manual

AST
AST A450C+ PL/TL/Video User manual

AST
AST P-GM+ User manual

AST
AST AL390 User manual

AST
AST AL30 User manual
Popular Measuring Instrument manuals by other brands

AEMC instruments
AEMC instruments MN01 user manual

PCB Piezotronics
PCB Piezotronics IMI SENSORS 646B00 Assembly, installation and operating manual

Fluke
Fluke 1587 manual

TESTO
TESTO 816 instruction manual

Tektronix
Tektronix TLA 700 Series Technical reference manual

Honeywell
Honeywell W8600B installation instructions