Performing a Quality Control Test
9
PRECAUTION
- Your Control Solution results will not be calculated for average reading but still can be
recalled. The Control Solution Test result will be shown with " " symbol on the screen.
- Our suggestion environment for Control Solution Test is 15 - 40°C.
- Before " " and " " appears, please don't touch the Control Solution to the sample entry
on strip because the meter is still in an internal check. If you do so, the meter will show " "
and " ", and be accompanied by beeps (if volume is turned on)
- Don't drip the Control Solution to sample entry of the strip directly, doing
so may cause the contamination of the control solution or meter.
- Don't touch the tip of the control solution bottle. If you have touched it,
please clean up with water.
34
Maintain the products
9. Close the control solution cap immediately after use, and make sure to close it
tightly
10. The Control Solution result appears. Compare your Quality Control Test result to
the Control Solution Range printed on the test strip vial label.
Indirect transmission of Human Immunodeficiency Virus (HIV), Hepatitis B Virus (HBV) and Hepatitis
C Virus (HCV) during the delivery of healthcare services has been increasingly reported. Persons
using blood glucose monitoring systems have been identified as one risk group due to sharing of
lancets, lancing devices, and blood glucose meters.
The cleaning procedure is to remove dust, blood and body fluid from the surface and should be
performed whenever the meter or lancing device is visibly dirty. The disinfecting procedure is
necessary to kill pathogens such as HIV, HBV and HCV on the device.
NOTE: The cleaning procedure can only remove visible contaminates from surfaces. Only the
disinfecting procedure can eliminate non-visible pathogens.
If the meter is being operated by a second person who is providing testing assistance to the user, the
meter and lancing device should be decontaminated prior to use by the second person.
The following disinfecting wipe has been tested and may be used to clean and disinfect the meter
and lancing device.
CAVIWIPES DISINFECTING TOWELETTES, manufacturer: Metrex. It is with Isopropanol as the active
ingredient, have been shown to be safe for use with the meter and lancing device.