Block Optic CaraT User manual

Model „CaraT®“
Block Optic Design GmbH., Semerteichstr. 60, 44141 Dortmund, Tel.: +49 231 1087785-0, Fax: +49 231 1763065
Technical manual
Patient chair CaraT®
Sample image
Original manual since 2012
Update: October 2021
Subject to technical changes!

Model „CaraT®“
Block Optic Design GmbH., Semerteichstr. 60, 44141 Dortmund, Tel.: +49 231 1087785-0, Fax: +49 231 1763065
Foreword
Thank you for the confidence you have placed in us by
purchasing this patient chair. With the patient chair CaraT®, you
have chosen a modern, sophisticated product that has been
manufactured and tested according to strict quality criteria.
Continuous development may result in changes to the design
and scope of delivery. The illustrations in these operating
instructions may therefore differ from the delivered unit in
individual cases. The illustration shows the right-hand version.
For the left-hand version, the arrangement of the unit is
correspondingly mirror-inverted. If you have any questions or
require further information on your patient chair, please
contact us! Our service team will be happy to assist you.

Model „CaraT®“
Block Optic Design GmbH., Semerteichstr. 60, 44141 Dortmund, Tel.: +49 231 1087785-0, Fax: +49 231 1763065
Index
General ............................................................................................................................................ 1
1.1 Information about the user manual ........................................................................................ 1
1.2 General safety regulations ...................................................................................................... 1
1.3 Explanation of symbols............................................................................................................ 2
1.4 Copyrights and trademarks ..................................................................................................... 2
1.5 Limitation of liability ................................................................................................................ 3
1.6 Repair and spare parts ............................................................................................................ 3
1.7 Warranty conditions ................................................................................................................ 3
1.8 Customer service ..................................................................................................................... 3
Safety ............................................................................................................................................... 3
2.1 Manufacturer responsibility .................................................................................................... 3
2.2 Operator responsibility ........................................................................................................... 4
Technical specifications ................................................................................................................... 4
3.1 Space requirement .................................................................................................................. 4
3.2 Connection values in single installation or in combination with a Block Optic examination
unit without docking station ............................................................................................................... 4
3.3 Connection values in combination with a Block Optic examination unit and a docking
station 5
3.4 Operating condition ................................................................................................................ 5
3.5 Label ........................................................................................................................................ 5
Delivery ............................................................................................................................................ 6
4.1 Scope of delivery ..................................................................................................................... 6
4.2 Optional accessories ................................................................................................................ 6
4.3 Packaging ................................................................................................................................. 6
Mounting and electrical connection ............................................................................................... 7
5.1 Mounting ................................................................................................................................. 7
5.2 Electrical connection ............................................................................................................... 7
Basic concept ................................................................................................................................... 7
Handling........................................................................................................................................... 7
7.1 Functions via the keypad on a Block Optic examination unit ................................................. 8
Equipment features and optional accessories ................................................................................ 9
8.1 Function via foot switch (optional) ......................................................................................... 9
8.2 Footrest (optional) .................................................................................................................. 9
8.3 Manual seat adjustment (optional) ....................................................................................... 10
8.4 Rotation 360° of the patient chair (optional) ........................................................................ 10

Model „CaraT®“
Block Optic Design GmbH., Semerteichstr. 60, 44141 Dortmund, Tel.: +49 231 1087785-0, Fax: +49 231 1763065
8.5 Folding armrests .................................................................................................................... 11
8.6 Neck pad ................................................................................................................................ 11
8.7 Chair docking station (optional) ............................................................................................ 12
8.8 CaraT® chair without docking station with one examination unit or in single set-up ............. 13
Glass case (optional) ...................................................................................................................... 13
9.1 BriTa (optional) ...................................................................................................................... 14
Safety system............................................................................................................................. 14
Repairs ....................................................................................................................................... 15
11.1 Faults on the CaraT®patient chair ............................................................................................ 16
Maintenance.............................................................................................................................. 16
12.1 Service ................................................................................................................................... 16
12.2 Cleaning and care of the Block patient chairs ....................................................................... 17
12.3 Disinfection of Block patient chairs ....................................................................................... 17
Disposal! .................................................................................................................................... 18
EC Declaration of Conformity .................................................................................................... 19
CE Certificate ............................................................................................................................. 20

Model „CaraT®“
1
General
1.1 Information about the user manual
The instructions are an integral part of the patient chair CaraT® and must therefore be kept with the
patient chair CaraT®.
Before using the CaraT® patient chair, read these operating instructions carefully and familiarise
yourself with all functions.
If you have any questions about its use, please contact our customer service/field staff who will be
happy to help you.
Block Optic Design GmbH
Semerteichstr.60
D-44141 Dortmund
Germany
Tel: +49 231 108 77 850
Email: [email protected]om
Block Optic accepts no liability for damage resulting from improper operation of the CaraT® patient
chair and/or failure to observe the provisions of this operating manual.
Notice:
The illustrations in this manual may be shown differently from the actual delivery and are only for
understanding the function.
1.2 General safety regulations
The CaraT® patient chair is used exclusively in ophthalmic optics and ophthalmology. The patient chair
CaraT® fulfils the requirements of Annex 2 of the Medical Devices Act (MPBetriebV). In addition, the
local accident prevention regulations and general safety regulations for the area of application of the
patient chair CaraT® apply, as do the respective valid regulations of the VDE/EN standard.
WARN
I
NG!
Before putting the CaraT® patient chair into operation, read this manual carefully.
It contains important data and information on handling and safety for users and patients.
All electrical functions of the patient chair CaraT® are explained here
in connection with a Block Optic examination unit or as a stand-alone
chair with foot switch. In connection with examination units of other
manufacturers, not all electrical functions may be possible or the
symbols of the operation may be different. For more details, please
refer to the instructions of the corresponding manufacturer of
examination and patient chairs.

Model „CaraT®“
2
1.3 Explanation of symbols
Important text passages in these operating instructions are specially marked by highlighting and
keywords. The following highlighting is used in these operating instructions:
FORBIDDEN
!
Failure to observe these instructions may endanger the user and/or patient and may
damage the CaraT® patient chair.
CAUTION!
Indicates a potentially dangerous situation. Observe the precautions marked in this way to
avoid endangering persons or damaging the patient chair.
IMPORTANT!
Indicates important information. Read this information to maintain the high safety and
functional standard of the CaraT® patient chair.
NOTE
!
Indicates information on correct use. Please read this information to avoid incorrect
operation.
DANGER!
Indicates a potential danger to life due to electric shock.
SECURITY!
During electrical tests, the unit must be disconnected from the power supply and secured
against being switched on again.
SERVICE
!
Service should only be carried out by Block Optic or Block Optic authorised staff.
DANGER OF CROSSING!
Indicates a potential danger of crushing.
DISPOSAL!
Disposal information.
1.4 Copyrights and trademarks
All rights to this operating manual, in particular the right of reproduction and distribution as well as
translation, are reserved by Block Optic. In particular, these operating instructions may not be
reproduced or distributed without the agreement of Block Optic.
Contraventions are punishable and oblige to compensation. All rights to exercise industrial property
rights are reserved by Block Optic.

Model „CaraT®“
3
1.5 Limitation of liability
All information and notes in these operating instructions have been compiled taking into account the
applicable standards and regulations, the state of the art and our many years of knowledge and
experience.
The company Block Optic accepts no liability for damage due to
failure to follow the instructions
untrained personnel
unauthorised modifications
technical changes
use of spare parts not approved by Block Optic
1.6 Repair and spare parts
The CaraT® patient chair can only be repaired by Block Optic or by a specialist company authorised by
Block Optic.
1.7 Warranty conditions
The "General Terms and Conditions of Sale and Delivery" of the company Block Optic apply. These
can be viewed on our website www.block-optic.com.
1.8 Customer service
Our customer service is available for technical information.
In addition, our employees are constantly interested in new information and experiences resulting
from the application and which lead to the improvement of the patient chair CaraT®.
Safety
This section serves as an overview of all safety instructions that ensure the smooth and safe
operation of the patient chair CaraT®. Compliance with the handling instructions and the safety
instructions must be ensured.
Failure to comply may result in danger to operators and/or patients.
2.1 Manufacturer responsibility
As the manufacturer, we guarantee that the patient chair CaraT® has been manufactured according to
the latest state of the art and the recognised safety rules.
This applies in particular to the:
compliance with the 2014/30/EU
(electromagnetic compatibility)
compliance with the 2017/745/EU
(medical devices)
compliance with the EN 60601
-
1
-
2 2015
(
m
edical electrical devices)
compliance with the EN 55011
(
i
nterference emission
)
compliance with the IEC 801
(
i
nterference immunity)
CAUTION
!
Incorrect or non-original spare parts can impair safety and lead to damage, malfunction or
total failure.
Therefore: Only use original spare parts from the manufacturer.

Model „CaraT®“
4
2.2 Operator responsibility
The operator must clearly regulate and define the responsibilities for operation,
maintenance and cleaning.
The operator must observe the maintenance intervals as described in the operating
instructions.
The operator must check the safety devices at regular intervals.
The operator must check the CaraT® patient chair for visible damage at regular intervals.
In the event of damage, the operator must call in Block Optic or a specialist company
authorised by Block Optic to repair the damage.
Technical specifications
3.1 Space requirement
Depending on the design of the CaraT® patient chair, the dimensions may vary. An exact isometry can
be found in the appendix.
Example:
Version
:
m
ax.
dimensions
(L x
B
)
CaraT
®
with or without mecanical
ly
tiltable chair
600
mm x
560
mm
3.2 Connection values in single installation or in combination with a Block Optic
examination unit without docking station
Specification
Value
Line voltage
230 V AC
Frequen
cy
50 Hz
Permissible deviation from the
nominal line voltage
5,00
%
Max.
Watt
160
W
att
Max.
l
ifting capacity incl. chair top
250 kg
Stand
-
by power consumption when set up individually or
with a Block Optic examination unit without docking
station
2 Watt
Device fuse protection (on site)
Connection to a separate circuit with
fuse T16 AH, 250V
Device protection in combination with a Block
examination unit
Fixed committee to the
B
lock
electronics with fuse T2,5A, 250V
Protection class
I
Risk classification
IIa

Model „CaraT®“
5
3.3 Connection values in combination with a Block Optic examination unit and a
docking station
Specification
Value
Main voltage (from the examination unit)
36 V DC
Max. power consumption
160
W
att
Device fuse protection (SKF control box)
Direct connection to the circuit of the
Innovario® examination unit with fuse
T4, 250 V
Protection class
I
Risk classification
IIa
3.4 Operating condition
Specification
Value
Temperature range
+10°
C
to
+40°
C
Relative humidity
30
%
to
70
%
Air
pressure
700
hPa
to
1060 hPa
Mounting
in dry indoor rooms
3.5 Label
The label on the patient chair shows the following information:
Manufacturer with address
Model:
Type name
SN: Serial number
Year of manufacture
VAC Line voltage and frequency max.
VA max. power draw
CLASS Risk class
CE-label
Disposal note
Read manual carefully
Figure 1: Label patient chair

Model „CaraT®“
6
Delivery
4.1 Scope of delivery
The scope of delivery of the patient chair CaraT® varies depending on the equipment version.
The respective scope of delivery is listed in detail on the delivery note.
4.2 Optional accessories
A list of the current accessories for the patient chair CaraT® can be obtained from Block Optic or from
specialist dealers authorised by Block Optic. You will find a short excerpt in chapter Fehler!
Verweisquelle konnte nicht gefunden werden..
4.3 Packaging
If required, the patient chair CaraT® can be delivered in a special transport box; the dimensions and
weight of the packaging are variable depending on the equipment.
Upon delivery, please check the packaging boxes for external damage and observe the enclosed
freight instructions!
NOTE
!
The packaging weight is min. 100 Kg.
Please ensure that all the individual parts belonging to the CaraT® patient chair have been
removed.

Model „CaraT®“
7
Mounting and electrical connection
5.1 Mounting
The CT® patient chair is set up exclusively by Block Optic or by a company authorised by Block Optic.
5.2 Electrical connection
The CaraT® patient chair may only be installed in rooms that comply with the requirements of
VDE 0100-710.
Depending on the equipment of the CaraT® patient chair, it is integrated into the practice/clinic
building installation by means of a protective contact plug (individual installation) or connected to a
refraction unit.
Basic concept
The CaraT® refraction and patient chair has been specially developed for ophthalmologists and
opticians with high demands on equipment and operating comfort. The fully equipped version has a
wide range of characteristic features.
Handling
NOTE
!
Before each start-up, the patient chair and its external lines, if any, must be checked for
external damage.
If the chair is connected to a Block Optic examination unit, electrical functions on the CaraT® (up/down
and seat forward and back) can be controlled via the ergonomic membrane keyboard on the Block
Optic examination unit.
With an optional foot switch, you can operate the function of the patient chair CaraT® up/down
parallel to the keyboard.
CAUTION
!
The patient chair may only be set up by Block Optic or by a company authorised by Block
Optic.
FORBIDDEN
!
The electrical coupling with an examination unit may only be carried out by Block Optic or by a
company authorised by Block Optic.
NOTE
!
If the patient chair CaraT® is set up in combination with an examination unit that is NOT from
Block Optic, the corresponding safety devices of this third-party examination unit must be
used against a possible crushing hazard. To do this, contact the relevant supplier.

Model „CaraT®“
8
7.1 Functions via the keypad on a Block Optic examination unit
This symbolism only refers to Block Optic keyboards. If the patient chair CaraT® is connected to a
foreign examination unit, the symbols and also the function may be different.
Figure
Func
tion
Description
Chair UP The chair moves upwards by electric motor.
Chair DOWN The chair moves downwards by electric motor.
Chair AUTO DOWN The chair moves to the basic position by electric motor.
Chair FORWARD Function not available for the CaraT® patient chair.
Chair BACKWARD Function not available for the CaraT® patient chair.
Whenever you adjust the height or turn the chair, make sure that the patient is not exposed
to any danger. Especially when adjusting the height of the CaraT® patient chair, the seat shift
and footrest are a source of danger. Therefore, make sure that the patient always positions
his or her feet on the footrest.
CAUTION
!
If the patient's feet are not on the footrest when the chair is moved downwards, there is a
risk of crushing!
CAUTION
!
The motor of the patient chair CaraT® is not suitable for continuous up and down operation by
the electromotive height adjustment.
After continuous operation of the height adjustment for 1.5 minutes, a cooling time of at
least 8.5 minutes must be observed.
If the height adjustment is operated continuously for longer than 1.5 minutes, this can lead
to a defect in the height adjustment.

Model „CaraT®“
9
Equipment features and optional accessories
8.1 Function via foot switch (optional)
Using the optional foot switch in conjunction with a Block Optic examination unit or in stand-alone
set-up, you can move the CaraT® paten chair up and down with an electric motor.
8.2 Footrest (optional)
The foldable footrest (Figure 3) is fixed by a magnet so that it does not fold out by itself. If necessary,
it can be folded out by the user so that the patient can sit relaxed on the CaraT® patient chair.
A: Foldable footrest by 90°.
Figure 3: Footrest
NOTE
!
The footrest does not serve as a standing aid. Placing the feet on the footrest while standing
up from the CaraT® patient chair will damage the footrest.
The assignment of the pedals for
chair UP/DOWN can be done flexibly.
A Chair UP or DOWN.
B Chair UP or DOWN.
Figure 2: Foot switch

Model „CaraT®“
10
8.3 Manual seat adjustment (optional)
The patient chair CaraT® can be equipped with a manual seat adjustment with three different shifting
positions. Here, pull the lever upwards so that it disengages and push the seat in the corresponding
direction until the seat automatically engages again.
A: Release lever of the mechanical seat adjustment.
Figure 4: Release lever of the mechanical seat
adjustment
8.4 Rotation 360° of the patient chair (optional)
The upper part of the CaraT® patient chair can be swivelled, and the upper part locks into place every
90°. This possibility of rotating the upper part of the chair can be used advantageously when the
examination device is placed on a separately table.

Model „CaraT®“
11
8.5 Folding armrests
The armrests can be folded up 90°. Folding up the armrests allows obese patients to sit comfortably.
A: Folding armrests, can be folded up 90°.
Figure 5: Armrests
NOTE
!
The armrests do not serve as a standing aid. Resting the hands on the armrests while
standing up from the CaraT® patient chair can damage the armrests.
Tugging on the armrests to pull the chair out of its docking station can damage the armrests.
8.6 Neck pad
The neck pad of the CaraT® patient chair allows a stress-free posture of the patient’s head. Its height
can be continuously adjusted through a manual knurled screw placed at the reverse side of the
patient chair.
A: Neck pad
B: Manual knurled screw
setting the height of the neck
pad
Figure 6: Front view of the neck pad
Figure 7: Set screw of the neck pad

Model „CaraT®“
12
8.7 Chair docking station (optional)
The optional patient chair docking station, which is only possible in conjunction with Block Optic
examination units, is optimal for examining patients in wheelchairs, as it eliminates the need to
transfer the patient to the treatment chair. The chair is docked and undocked by means of a locking
pedal (Figure 8 and Figure 9).
Connection cables are not necessary with the patient chair docking station. After undocking, the
patient chair CaraT® can easily be moved in all directions via the castors.
A: Locking pedal
B: Patient chair with castors
Figure 8: Patient chair docking station locked
A: Locking pedal
B: Wireless contact surface chair and
docking station
C: Patient chair with castors
Figure 9: Patient chair docking station unlocked
FORBIDDEN
,
TIPPING HAZARD
!
It is prohibited to undock or dock and/or move the patient chair CaraT® when the patient is
sitting on the patient chair CaraT®.
NOTE
!
When undocked, no electric patient chair function is possible.

Model „CaraT®“
13
8.8 CaraT® chair without docking station with one examination unit or in single set-up
The CaraT® patient chair can also be used with a cable connection (power supply from the examination
unit) or in individual set-up with foot switch (up/down). Power supply via a protective contact socket
(Figure 10).
A: Motor column patient chair CaraT®
B: Status LED (voltage 230 V/AC) only in
function without docking station or individual
set-up
C: Cover motor electrics
D: Base plate patient chair CaraT® without
docking station and individual set-up
Glass case (optional)
The optional glass case (Figure 11) with telescopic
pull-out serves with the grooved frame as a
receptacle for trial glasses. It is ergonomically mounted on the backrest of the CaraT® patient chair.
A: Glass case with telescopic pull-out
Figure 11: Glass case
Figure 10: CaraT® without docking station

Model „CaraT®“
14
9.1 BriTa (optional)
The „BriTa“ (Figure 12) is the optimal solution for putting down a pair of glasses and hanging a bag.
A: Storage tray for one pair of glasses
(reading aid)
B: Holder for a bag up to max. 20 kg
C: Mounting point flexibly mounted for the „BriTa“
D: Armrest foldable 90°.
Figure 12: BriTa
NOTE
!
The bag must not exceed the maximum weight of 20 kg.
Safety system
A safety shut-off bar is fitted to each underside of a Block Optic examination unit to protect the
patient and prevent the patient's thighs from being pinched or crushed. Mechanical contact with the
safety bar will cause the patient chair drive to stop immediately. For more details, see the
instructions for a Block Optic examination unit.
NOTE
!
When using the Block Optic examination units, make sure that the patient rests his hands
comfortably on the armrests of the patient chair, as there is a risk of crushing at the headrest
holders!
When combining with products from other manufacturers, pay particular attention to the
safety requirements. The safety shut-off of the lifting function must be guaranteed.
Block Optic accepts no liability for improper use and any resulting damage.

Model „CaraT®“
15
Repairs
If a fault occurs that is not described in 11.1 or a repair needs to be carried out, contact Block Optic
or an authorised service company directly.
Have the serial number of the CaraT® patient chair and, if applicable, a customer number ready for
quick assistance by Block Optic. If possible, send us photos or a video of the problem with a short
description to support@block-optic.com.
NOTE
!
A comprehensive repair may only be carried out by Block Optic or by a specialist company
authorised by Block Optic.
A trained electrician can be consulted for an initial diagnosis.
This is required:
the appropriate circuit documents of the examination unit, which may be connected to the
CaraT® patient chair.
a measuring device with continuity tester for fuses
A visual inspection is not possible depending on the type of fuse.
Checking the fuses
Replacement of defective fuses
CAUTION DANGER
!
When testing fuses on a Block Optic examination unit, it must be de-energised and secured
against being switched on again.
IMPORTANT!
Only fuses with the same values may be used.

Model „CaraT®“
16
11.1 Faults on the CaraT®patient chair
Error
Possible cause
/
possible solution
Chair does not move
up
I
s
the
C
ara
T
®
pat
i
ent chair supplied with voltage
?
Switch on the examination unit or plug in the cold appliance plug of
the CaraT® patient chair in individual set-up.
In conjunction with a Block Optic Design examination unit:
Is the shut-off bar activated (obstacle)?
Remove obstruction under the shut-off bar, see the Block Optic
Design examination unit manual.
If the cut-off strip is hidden?
Check and straighten the cut-off strip plate see manual of the Block
Optic Design examination unit
Chair has reached its maximum height?
Chair does
not move
down
Is the
C
ara
T
®
patient
chair supplied with voltage?
Switch on the examination unit or plug in the cold appliance plug of
the patient chair CaraT® in individual set-up.
Maintenance
12.1 Service
If the patient chair is coupled in combination with Block Optic examination units, the CaraT® patient
chair is also automatically checked during maintenance.
NOTE!
In order to maintain the high safety standard and the high quality of the CaraT® patient chair,
we recommend regular maintenance and care.
Maintenance is divided into 2 categories. The maintenance interval type B is mandatory.
For the maintenance interval type A, this is max. one year for practical operation.
Depending on the frequency of use, the maintenance intervals may be shorter.
For operation in a clinic, the maintenance interval type A is one year.
Maintenance intervals type A:
Practice operation and clinic operation
max. 1 year
General check for correct
function
Checking the spring contacts in
the docking station.
Checking that the docking
station is correctly latched.
Maintenance intervals type B:
2 years
Safety inspection (STK incl. DGUV
A3 according to VDE 0751)
Check the condition of all
external cables (check for
insulation damage).
General check for correct
function.
Checking the spring contacts in
the docking station.
This manual suits for next models
1
Table of contents
Other Block Optic Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Olympus
Olympus ESG-410 Quick reference guide

Seers Medical
Seers Medical Signature CB0280 operating instructions

Ventec Life Systems
Ventec Life Systems V-Home Clinical and Technical Manual

Quest Engineering
Quest Engineering MPS 3 troubleshooting manual

Air Techniques
Air Techniques Mojave Mobile Installation and operating instructions

Fisher Wallace
Fisher Wallace CIRCADIA Series user guide