Boekel 145600 User manual

Boekel
Paraffin Wax Dispenser
Model Numbers
145600
145600-2
Operating Instructions
N2400317 Rev. 012/09

1
TABLE OF CONTENTS
Page
1. Safety............................................................................................................................2
2. Product Information..................................................................................................2
3. Assembly......................................................................................................................3
3.1 Unpacking ...............................................................................................................3
3.2 Installation................................................................................................................3
4. Operation....................................................................................................................3
4.1 Controls and Indicator Lamp (See Figure 1)..................................................3
4.2 Setting the Temperature......................................................................................4
5. Fault Diagnosis...........................................................................................................5
6. Technical Specifications.........................................................................................6
7. Maintenance and Service.....................................................................................6
7.1 Cleaning...................................................................................................................5
7.2 Replacement of Fuses (see Figure 2)...............................................................5
8. Warranty.......................................................................................................................7
9. Service..........................................................................................................................7

2
1. Safety
The following symbols marked on the equipment mean:
Caution: Read these operating instructions fully before use and pay particular
attention to sections containing this symbol.
Caution: Surfaces can become hot during use.
Always observe the following safety precautions:
•Use only as specified by the operating instructions or the intrinsic protection
may be impaired. After transport or storage in humid conditions, dry out the
unit before connecting it to the voltage source. During drying out, the
intrinsic protection may be impaired.
•Connect only to a voltage source with a voltage corresponding to that on
the serial number label.
•Connect only to a voltage source that provides a safety ground terminal.
•Before moving, disconnect the power cord from the unit.
•Do not touch surfaces that become hot during high temperature operation.
•Ensure that the power switch is easily accessible during use.
•It is the user’s responsibility to carry out appropriate decontamination if
hazardous material is spilled on or inside the equipment.
•This product must be used with an UL Listed / CSA Certified power cord set.
•The responsible body shall be made aware that, if the equipment is used in a
manner not specified by the manufacturer, the protection provided by the
equipment may be impaired.
•Unit is designed to melt Paraffin Wax only; no other matter should be placed
into the wax reservoir.
2. Product Information
The Boekel Paraffin Wax Dispenser is designed to help meet the needs of
today’s histology labs by having 5 liters of molten paraffin wax available
when required by user. The unit has a set temperature range of 40.0⁰C to
75.0⁰C. The unit has a heated dispenser valve to insure a consistent flow
and precise control of flowing wax. The unit also features a rapid warm up
and an advanced safety over-temperature cut-off.

3
3. Assembly
3.1 Unpacking
Remove packing materials carefully, and retain for future shipment or
storage of the unit. Inspect for damage. Report all shipping damage to
the carrier immediately. Shipping damage is covered by the carrier and
repair/replacement for shipping damages must be coordinated through
the carrier. Complete and return the Warranty Registration Card.
Contents include:
•
Wax Dispenser
•
Power Cord
•
Lid
•
Warranty Card
•
Operating Instructions
3.2 Installation
Place the Wax Dispenser on a flat and stable surface, preferably away
from drafts. Plug the power cord into a voltage source that matches the
voltage listed on the serial number label on the rear of the unit. Insure
that the surface on which the unit is placed will withstand the radiated
heat produced by typical laboratory equipment.
4. Operation
4.1 Controls and Indicator Lamp (see Figure 1)
1. Power Switch - Controls power to the unit.
2. Heater Control Dial – Used to set desired temperature.
3. Power on Lamp – When illuminated indicates unit is powered up.
4. Heater Indicator – When illuminated indicates heater is on.
5. Dispenser Valve –Used to dispense wax. The valve can be used to
dispense wax on demand, or rotate the handle 180° to lock the handle
into a constant flow position (See Figure 2).

4
FIGURE 1
FIGURE 2
Caution: Assure the reservoir is filled with a minimum of 2.5 cm of wax and
a maximum of 2.5 cm below the top of the reservoir and the dispenser
valve is positioned for “On Demand” dispensing before applying power.
Use caution when filling the reservoir with wax to avoid overfilling the
unit.

5
4.2 Setting the Temperature
1. When setting the temperature to melt the Paraffin Wax, refer to
Paraffin Wax manufacturers recommended melting temperature prior
to adjusting the heater control knob. Generally, it will take
approximately 2 hours to fully melt a reservoir full of Paraffin pellets.
2. Adjust the heater control dial to desired temperature.
3. The Heater Indicator Lamp will illuminate to indicate that the unit is
heating up. When the set temperature is reached the heater indicator
lamp will cycle on and off to maintain the set temperature.
5. Fault Diagnosis
Symptom
Possible Cause
Action Required
1. Power indicator will
not illuminate a. Unit is not turned on
b. Unit not plugged into
power supply
c. Fuses blown
d. Power supply failure
a. Turn power switch to the
on position.
b. Plug in, turn power switch
to the on position.
c. Replace fuses per Figure
7.2
d. Check that other electrical
appliances on the same
circuit are working
2. Wax does not melt a. Temperature control
knob not set high
enough
b. Temperature control
circuit fault
a. Check set temperature
b. Have unit checked by
competent person
3. Temperature
continues to rise
when not expected
a. Temperature control
knob set to high
b. Temperature control
circuit fault
a. Check set temperature
b. Have unit checked by
competent person
4. Heater Indicator
Lamp does not
illuminate
a. Temperature control
knob set to low
b. Safety over-temperature
cut out activated
a. Check set temperature
b. Have unit checked by
competent person
5. Wax will not flow
when dispenser valve
is engaged
a. Wax is not melted
b. Filter screen on bottom
of the wax reservoir is
clogged
a. See #2 above
b. Clear clog

6
6. Technical Specifications
Operating Temperature Range 20⁰C to 30⁰C
Settable Temperature Range
40.0
⁰
C to 75
⁰
C
Temperature Accuracy
±2.0
⁰
C
Reservoir Capacity
5.0 Liters
Electrical Model # 145600
Model # 145600-2
115VAC 50/60hz 250W
230VAC 50/60hz 250W
Approvals Model # 145600
Model # 145600-2
Certified to UL 61010-1 and CSA C22.2
CE Certified
7. Maintenance and Service
The reservoir screen in the bottom of the reservoir should be checked and
cleaned periodically.
7.1 Cleaning
Disengage power cord prior to cleaning. Paraffin build-up should be
avoided around the top area of the Wax Dispenser. The outer casing
may be cleaned with water and a damp cloth. Do not submerge or
immerse the Wax Dispenser in water. Before using any cleaning or
decontamination method, users should verify that the proposed method
would not damage the equipment.
7.2 Replacement of Fuses
There are two supply fuses located in the fuse drawer. To replace the
fuses:
•Disconnect the unit from the power supply.
•Remove the plug from the socket in the back of the unit.
•Pull back on the fuse drawer (see figure 2)
•Pull out the fuse drawer
•Check and replace with the correct fuses if necessary. The fuses
must be 5 mm x 20 mm fast acting, rated 4A and 250V.
•Push the fuse drawer back in. Reconnect unit to the power
supply.
Fuse
Drawer
Figure 2

7
8. Warranty
When used in laboratory conditions and according to these operating
instructions, Boekel warrants this product to be free of defective material
and workmanship for a period of two years from the date of manufacture.
The liability of Boekel for any defective equipment during the warranty
period shall be limited to the repair of such equipment or replacement
thereof without charge for parts or labor.
9. Service
It is required to obtain a Returned Material Authorization (RMA) number
before any Boekel products are returned for any reason. A
Decontamination Certificate must be completed, signed by the user, and
returned to Boekel Scientific prior to receiving the RMA number. Please be
sure to mark the outside of the returned goods package with this RMA
number to ensure prompt handling.