SECTION I
INTRODUCTION
Section I
Introduction
1-1. SCOPE OF MANUAL.
1-2.This instruction manual provides descriptive data,
operating instructions, theory of operation, maintenance
instructions, and aparts list for the Model 4210 R. F.
Microwattmeter. (See Figure 1-1.)
SAFETY NOTICE
Although this instrument has been designed in accordance
with international safety standards, general safety precau-
tions must be observed during all phases of operation and
maintenance of the instrument. Failure to comply with the
precautions listed in the Safety Summary located in the
front of this manual could result in serious injury or death.
Service and adjustment should be performed only by
qualified service personnel.
1-3. PURPOSE OF EQUIPMENT.
1-4.The instrument is amicroprocessor-controlled, solid-
state unit that features ease of operation, high sensitivity,
low input s.w.r., and low noise. It measures r.f. power levels
from 1nW (-60 dBm) to 100 mW (+20 dBm). The cali-
brated frequency range extends from 0.2 MHz to 18 GHz,
depending upon the accessory sensor used; useful response
for relative measurements is obtained from 20 kHz to
approximately 20 GHz. Representative uses of the instru-
ment include:
a. Low-power transmitter, signal generator, and oscilla-
tor measurements
b. S.W.R. and return-loss measurements with direc-
tional couplers and slotted lines
c. Gain and insertion-loss measurements
d. R.F. attenuation and s.w.r. measurements
e. Antenna measurements
1-5. DESCRIPTION OF EQUIPMENT.
1-6.The instrument is packaged as acompact bench unit.
When operated with Boonton Series 4210-4/ 5sensors, the
instrument displays r.f. power by measuring the voltage
across aprecision, non-inductive resistor in the sensor head
with specially selected diodes. Panel indications are cali-
brated in terms of power according to the relationship P=
E2/R. This detection system has important performance
advantages over power meters that use bolometer or
thermo-electric detection. The instrument sensitivity of 1
nanowatt (-60 dBm) is orders of magnitude better; temper-
ature stability of better than 0.001 dB/°C supports this
sensitivity; and burnout levels exceed 300 milliwatts.
average power for all types of waveforms. Above these
levels, the instrument displays, by means of internal shap-
ing, the true average power of c.w. signals. If the r.f.signal is
gated or amplitude modulated, the indicated power may
not be the true average power. The signal may be attenu-
ated to bring it within the r.m.s. region of measurement.
Alternatively, the instrument can be used with thermal
sensors Model 4210-7E/8E; in this case, the instrument mea-
sures and displays the true average power for all waveforms.
1-8 The outstanding design features are:
a. Wide Frequency Range. The calibrated frequency
range of the instrument is determined by the sensor used
with the instrument. The instrument is normally ordered
with one of the following sensors: /,
FREQUENCY RANGE”" 60/
+
to dDnn
Model (Impedance) Power Range
42 10-4A
4210-4B
42 1CMC
4210-4E
42 10-5 B
4210-5E
4210-7E
421 0-8 E
200 kHz to 7GHz (50 ohms)
200 kHz to 12.4 GHz (50 ohms)
200 kHz to IGHz (75 ohms)
200 kHz to 18 GHz (50 ohms)
200 kHz to 12.4 GHz (50 ohms)
200 kHz to 18 GHz (50 ohms)
10 MHz to 18 GHz (50 ohms)
10 MHz to 18 GHz (50 ohms)
1nW to 10 mW
1nW to 10 mW
1nW to 10 mW
1nW to 10 mW
10 nW to 100 mW
10 nW to 100 mW
1jaW to 10 mW
IOjuW to 100 mW
b. Wide Power Range. Depending on the selected sen-
sor, the instrument will measure r.f. power from 1nWup to
100 mW. Temporary overloads up to 300 mW with Series
42104 sensors, and up to 2Wwith Series 4210-5 sensors,
will do no permanent harm to the instrument or the sensor.
When measuring pulsed signals, the power indications are
accurate up to 20 microwatts peak power (up to 200 mi-
crowatts with Series 4210-5 sensors). External attenuators
may be used to extend the measurement range of the
instrument.
c. Low Noise. The instrument has been designed and
constructed to minimize noise from all sources. The sensor
cable is of aspecial low-noise design; vigorous flexing
causes only momentary minor excursions of the display on
the most sensitive range of the instrument. The sensors are
insensitive to shock and vibration; even sharp tapping on
the sensor barrel causes no visible excursions on any range.
Internal signal amplification occurs at approximately 94
Hz, thereby reducing susceptibility to 50- or 60-Hzfields. A
low-noise solid-state chopper is used.
1-7.Diode sensors are r.m.s. -responding for low-power
levels (below 20 microwatts for Series 42104 sensors and
200 microwatts for Series 4210-5 sensors). At these low-
power levels, the instrument measures and displays true
d. LED Display. Measured power levels are displayed
by a4-digit, LED-type readout with decimal points and
minus sign. Annunciators associated with the LED display
indicate the units of measurement. The result is aclear,
1-1