Borringia Rise III-130 Operator's manual

User’s Instruction Manual
Rise®IIIS-130 Hoisting Shower-, Commode- & Care Chair w. Tilting Seat
12C10S, English
January 2014

1
User’s Instruction Manual
Rise® IIIS-130 Hoisting Shower-, Commode- & Care Chair w. Tilting Seat
Table of Contents:
1. Marking ....................................................................................................................................................... 2
2. Introduction: .............................................................................................................................................. 3
3. Lifting & Tilting:........................................................................................................................................ 3
4. Battery & Charging: .................................................................................................................................. 4
5. Emergency-Stop: ....................................................................................................................................... 4
6. Interference: ............................................................................................................................................... 5
7. Safety:.......................................................................................................................................................... 5
8. Daily Inspection and Annual Service Inspection: ................................................................................ 5
9. Water Ingression:....................................................................................................................................... 5
10. Cleaning: ...................................................................................................................................................... 5
11. Recycling:..................................................................................................................................................... 6
12. Measurements & Weight:.......................................................................................................................... 6
13. Optional Equipment:................................................................................................................................. 7
14. Warranty: ..................................................................................................................................................... 7
15. Expected Product Life Span:.................................................................................................................... 7
16. Trouble Shooting: ...................................................................................................................................... 9
17. Appendix 1, Annual Serice Inspection.................................................................................................. 10
Dealer (Stamp):
Autho
rise
d Service (Stamp):
Manufacturer
Borringia A/S Hejreskovvej 18 B-C DK-3490 Kvistgård Phone: +45 49 13 88 55

2
1. Marking
Your Rise IIIS-130 chair carries marking, according to the applicable technical standards. For your
information each of the 2 labels on the Rise IIIS-130 are explained below:
Marking, Production, Frame- & Batch number for the Rise IIIS-130:
Explanation:
12C10S-130: Identifies the product, i.e. a Rise IIIS-130 Shower-, Commode- & Care Chair
12: Identifies the year of production, i.e. 2012
02: Identifies the month of production, i.e. February
200: Identifies the specific frame number of this one unit
Max. 130 kg Informs of the limit of the allowed weight of the user
Marking, Transport Packaging:
Rise III-130
Shower-, Commode-& Care Chair
Frame & Batch No. :
12C10S-130 12 02 200
Storage:
Keep Dry, 25-90% relative humidity,
0-50º Celcius, 250-1200 hPa
Explanation:
Borringia: The manufacturer
Rise III-130: Brand name of the product
Shower-, Commode- & Care Chair: Product description
Frame- & Batch No.: Identification of the specific product
Storage: Information about how to store the product
Keep dry: To be stored away from rain and water
25-90% Relative humidity: Indication of safe humidity-level
0-50º Celsius: Safe storage temperature
250-1200 hPa Safe air pressure
12C10S-130 12 02 200
Patented, EP 1286638
Max 130 kg

3
2. Introduction:
The Rise IIIS-130 Shower-, Commode- & Care Chair accommodates the needs of both the user and the helper.
The electrical elevation (Seat height 53,5 cm – 95 cm) makes it possible to avoid lifting the user with a traditional
sling. By elevating the seat to match the height of the bed or wheelchair, most users can transfer themselves
sideways to/from the Rise IIIS-130. For weak patients use of a transfer board may be advantageous.
The electrical elevating function enables the helper to obtain an ideal working position during toileting,
showering and personal hygiene routines. This allows the user to enjoy his/her daily shower, etc. without feeling
that they are a burden to the helper, thus contributing to less stressful care routines.
The elevating function combined with the seat, which can be tilted forward and backward, also enables the Rise
IIIS-130 to help the user to a standing position. This is possible because the chair not only lifts the seat up to
41,5 cm (16 in.) but, at the end of the elevating cycle, the user may be tilted forwards about 20 cm (8 in.). In this
way Rise IIIS-130, in the same way as typical catapult seats, may help the user to move their body’s centre of
gravity to a point directly above their feet. Getting to this vertical position is often one of the most difficult
problems for persons with limited walking or standing ability.
If you have any questions, comments or suggestions for improvements, please do not hesitate to call your local
dealer or contact Borringia directly by phone at TEL +45 49 13 88 55 or on the Internet: www.borringia.com.
Rise IIIS-130 Shower-, Commode & Care Chair is simple & logical to use. However, you must read the
following instructions carefully prior to using the chair:
3. Lifting & Tilting:
Both the lifting and the seat tilting functions are electrical. Power is supplied by a 24 volt battery (NiMh) with a
capacity of 1,4 Ah, sufficient for 25-30 full lifting cycles when fully charged.
Pressing the corresponding button on the hand control activates lifting/lowering and seat tilting. As soon as the
button is released, movement stops.
This button to
lift
This button to
lower
This button to
tilt the seat
forwards
This button to
tilt the seat
backwards

4
4. Battery & Charging:
1. Rise IIIS-130 is equipped with a 24 volt battery (NiMh) with a capacity of 1,5 Ah. When the battery is
fully charged, it has sufficient capacity for 25-30 full lifting cycles.
2. The battery and electronics box are equipped with a number of safety and signalling functions, which
help prevent that you run out of power during a job
a. When the battery has been discharged to a point, where there is sufficient power left for 5
lifting cycles, pressing the buttons on the hand control will result in a ”beep”. This is an
indication that charging is needed.
b. If the battery is still used without being charged, the electronics will at a certain point only allow
downwards movement. This is an indication that charging is needed.
c. If usage of the battery is continued, the electronics will at a given point prohibit all movement.
d. All batteries loose performance when they are cold, i.e. at 0° Celsius (32° F) or colder. If you
have left your Rise IIIS-130 or the battery in your car in cold weather, you may have to re-
charge the batteries, before using the Rise IIIS-130.
e. Normal operational temperature should be 0°- 40° Celsius (32° - 104° F)
To charge the battery follow the steps below.
1.
Pull the battery out of the electronics box
2. Push the battery into the charging station and turn it clockwise until the magnets lock the battery in
place.
As long as the green lamp blinks, charging is ongoing. When the green lamp is lit permanently,
charging has been completed. The yellow lamp is lit, when the charging station is connected to the
mains supply.
5. Emergency-Stop:
The removable battery also functions as an ”Emergency-Stop” in the – unlikely – situation that movement
occurs without a button being pressed on the hand control. By pulling the battery out of the electronics box, all
movement is stopped. The “Emergency-Stop” is deactivated by pushing the battery into the electronics box and
turning it clockwise until the magnets lock the battery in place.
1 2
Contact your dealer IMMEDIATELY and DO NOT USE the Rise IIIS-130 chair until
service inspection and repair has been CARRIED OUT!

5
6. Interference:
7. Safety:
·Never leave person with insufficient body control alone, when lifted
·Make sure that the hands, legs and feet of the user are not squeezed between parts of the
Rise IIIS-130 and other aid products (e.g. wheelchair, bed, or the like)
·Never use the Rise IIIS-130 in hot surroundings, e.g. in Saunas or in steam baths
8. Daily Inspection and Annual Service Inspection:
Daily
Inspection: Before using the Rise IIIS-130, always check the following points:
·There should be no visual cracks, dents or deformation of any metal part on the Chair
·All retaining rings should be in place
·The actuator should not make jarring sounds when activated with load
Service: Rise IIIS-130 must undergo at least one service inspection annually by an authorized
technician. Service agreements can be made with your dealer. Borringia cannot be held
responsible for damage to persons caused by the failure to carry out annual service
inspections.
9. Water Ingression:
Humidity: Rise IIIS-130 is designed for use in the shower. However, neither the chair nor the
electronics should ever be submerged
The electronics box is ONLY splash proof if correctly mounted and the battery inserted. If
the battery is removed during showering / cleaning, water will enter into the box and the
electronics may be destroyed. Failure as a result of this is NOT covered by the warranty.
10. Cleaning:
The Rise IIIS-130 is designed to be easy to clean:
·For daily cleaning, we recommend using wet-wipes either based on ethanol or containing anti-bacterial
ingredients.
·For thorough cleaning, the Rise IIIS-130 may be cleaned in automated washing machines or with
saturated steam for up to 1 minute at 90° Celsius (194° Fahrenheit), for 5 minutes at 85° Celsius (185°
Fahrenheit) or 10 minutes at 80° Celsius (176° Fahrenheit).
Should the electronics on the Rise IIIS-130 interfere with other electrical equipment,
contact your dealer IMMEDIATELY and DO NOT USE the Rise IIIS-130 chair until
service inspection and repair has been CARRIED OUT!
Always follow the instructions below:

6
·Cleaning of the toilet seat and the frame underneath it is easily done. Just pull upwards at the front of
the seat and remove it.
·Depending on the hardness of the water in your region, we recommend that chalk stains be removed
every 2 – 4 weeks. For environmental reasons, we recommend that you refrain from using acids or the
like. Simply spray the chair with standard household vinegar and leave it to work for 2-3 hours, then
rinse with water. Dry the chair with a cloth. In extreme cases you may want to use a household chalk
stain remover. Dilute the product with 4 parts of water before applying to the areas to be cleaned.
Lubrication: Approx. once a month, apply a bit of Vaseline to the piston of the actuators.
11. Recycling:
When the Rise IIIS-130 is taken out of use, you should see to it that the chair is disposed off in an
environmentally friendly way. Follow the instructions below:
Metal parts: All metal parts are made of steel and can safely be recycled. The painting will
not result in any toxic components as a result of recycling.
Electronics: The electronics including the batteries should be delivered to a recycling plant
equipped to handle electronic waste safely. If in doubt as to whether such a
facility is available, please return the electronic components to your dealer, who
will then pass these on to the manufacturer for safe recycling.
12. Measurements & Weight:
Height: 108– 148 cm (43 – 59in.)
Width: 57 cm (22.5 in.)
Length: 88 – 105 cm (35 – 41in.)
Seat height: 53,5 – 95 cm (20 – 37in.)
Seat width: 45 cm (17.7 in.)
Seat depth: 43 cm (17.0 in.)
Armrest’s height over seat: 21 cm (8.3 in.)
Width between armrests: 47 cm (18.5in.)
Backrest, Height over seat: 56 cm (22.5in.)
Backrest upholstery, Height over seat: 43 cm (17.0 in.)
Backrest width: 43 cm (17.0 in.)
Height of carriage: 15 cm (5.9 in.)
Clearance, carriage: 7 cm (2.8 in.)
Turning circle: 98 cm (38.6 in.)
Weight: 32,0 kg (70.6 lbs.)
Lifting range: 40 cm (15.7 in.)
Lifting & Tilting mechanism: Electrical actuator
Never expose the battery to temperatures above 90° Celcius (194° Fahrenheit) for
more than 1 minute, 85° Celcius (185° Fahrenheit) for more than 5 minutes or at
80° Celcius (176° Fahrenheit) for more than 10 minutes as this can cause danger
of explosion
Always leave the battery in the control box during cleaning as the battery forms an
integral part of the system, which prevents ingression of water

7
Seat Tilt: From +6 to – 31 degrees
Battery: 24 V (NiMh)
Battery Capacity: 25 – 30 lifting cycles
Charging: Via external charging station
Charging time: 4-5 hours
Max user weight: 130 kg 286 lbs.
13. Optional Equipment:
·Bedpan w. lid - stainless steel
·Holders for bedpan
·Holders for low hanging bedpan (+50 mm / 2”)
·Headrest, with soft adjustable PU-padding
·Headrest, Small
·Headrest, Medium
·Headrest, Large
·Backrest with built-in positioning system
·Anti-pressure pillow with toilet cut-out
·Locking mechanism for armrests
·Armrest for hemiplegics, left or right
·Crossbar between armrests
·Set of armrest with front closure
·Elevating footrests
·Calf belt between footrests
·Amputation largest, left or right
·Lid for PU-Toilet Seat
·PU-Toilet seat with hygiene cut-away at the front.
·Torso-, Hip- & Thigh Belt
·Swing-away Torso-Support with PU-Padding
·Seat-tilt limitation (Blocks forward tilt at horizontal level)
·Bracket for wall-mounting of the charging station
·Direction lock on rear wheel (Factory mounted)
14. Warranty:
Two years warranty on failure in metal construction. Warranty does not cover damage as a result of faulty use of
the chair. Casters, and the battery, which are subject to normal tear, are warranted for one year. Failure as a result
of misuse is not covered by the warranty.
15. Expected Product Life Span:
If the annual service inspections are carried out, the Rise IIIS-130 is expected to have a life time of 8 – 10 years.
Actuators, castors and battery may have to be replaced during the life span, depending on the intensity of use.

8
Removing the Seat for cleaning:
1. Pull upwards at the front of the seat and the seat can be removed completely for thorough cleaning of
both the seat and the seat frame.
2. In order replace the seat; place the seat at an angle as seen in picture 2. Slide the seat backwards.
3.
Push down on the front of the seat, until it clicks into position.
Footrests:
1. In order to remove the footrests, simply pull upwards
2. In order to put the footrests onto the chair, place the footrest over the holder as shown on the picture
and push downwards until the footrest clicks into place.
3. The height of the foot paltes may be adjusted by losening the thumb screw.
4.
The angle opf the foot plates may be adjusted by using the Allen key supplied with the chair.
1 2
1 2 3

9
16. Trouble Shooting:
Is the plug of the
hand control inserted in Insert the plug
the electronics box?
Are the plugs of the
actuators inserted in Insert the plug
the electronics box?
Yes
Does the
lift & tilt movement
work?
Is
the battery correctly' Push battery to the
incerted in the in bottom and turn until No
the electronics box?
magnets lock
Get in contact
with Borringia
Service representative
Does the Power Charge
Pack "beep" when the
activated?
battery
If MOVEMENT doesn't work
No
No
no
yes
3 4

10
17. Appendix 1, Annual Serice Inspection
Annual Service Inspection, Rise IIIS-130
Service Contract No.:
Product:
12C10, Rise IIIS-130, Shower & Commode
Chair
Frame
-
& Batch No.:
Client:
Departmen
t / User:
Contact person:
Valid from:
Automatic renewal date:
Telephone:
Fax:
Client Responsible:
Service Inspection performed by:
Date:
Control point:
Description:
All right:
Lubricated
Noise /
Irregularity
Tightened:
Comment Nr.:
Seat tilt Actuator runs both ways
Seat tilt Actuator lubricated
Cables for actuators Intact and secured with strips
Gas spring Securely tightened
Backrest Securely tightened
Seat Easy to remove and put back on the chair
Seat PU-surface is intact
Arm rests Securely fastened and move smoothly
Arm rests PU is intact
Foot plates on footrests Fold up/down, securely fastened
Footrests Height-adjustability works
Footrests Easy to remove
Emergency Stop Works and releases again
Lifting mechanism Lifts from bottom to top whit full load
Lifting Actuator Correctly mounted / Locking rings in place
Lifting Actuator O-Ring intact., greased & correctly inserted
Lifting Actuator Piston lubricated
Hand Control ”Hold to activate” works
Charging Station Intact, charges, lights work
Electronics Box No signs of water ingression
Electronics- & Control Box Charging- & safety cycles work correctly
Electronics- & Control Box Is correctly mounted
Castors Run easily, turn 360°and brakes work
Castors Do not drag to the side / securely fastened

11
Control point:
Description:
All right:
Lubricated
Noise /
Irregularity
Tightened:
Comment
No.:
Carrying axels Move smoothly / no noise
Carrying axels Screws tightened (Remember: Lock-Tight)
Carrying axels Lubricated
Frame No visual sign of deformation
Frame No visual sign of cracks
Frame Paint not scratched
Weldings No sign of cracks / deformation
Comment No: Proposed activity:
General Comments:
Written repair offer made?
Approved by contact
person? Repair carried out? Repair invoiced?

12
Offer for Repair
Explanation: Units
Price/Unit Price total
Sub-Total:
+ V.A.T:
TOTAL:
Offer Accepted:
Table of contents