Boxym bpw1 User manual

Wrist Type Blood Pressure Monitor
User Manual

Contents
Overview............................................................................................................................................. 1
Safety Precautions............................................................................................................................... 2
1 Product Introduction.........................................................................................................................4
1.1 Intended use...........................................................................................................................4
1.2 Contraindications.................................................................................................................. 4
1.3 Body...................................................................................................................................... 4
1.4 Display...................................................................................................................................4
1.5 Function description..............................................................................................................5
2 Battery Installation........................................................................................................................... 6
3 Initial Setup.......................................................................................................................................7
3.1 User Setup............................................................................................................................. 7
3.2 Time Setup.............................................................................................................................7
3.3 Sound Setup...........................................................................................................................8
3.4 Blood Pressure Unit Setting..................................................................................................8
4 Usage Methods of Wrist Strap......................................................................................................... 8
5 Blood Pressure Measurement...........................................................................................................9
5.1 Correct Measurement Methods.............................................................................................9
5.2 Operation Steps of Blood Pressure Measurement.............................................................. 10
5.3 Answers for Questions........................................................................................................ 11
6 Memory function............................................................................................................................12
6.1 Memory Recall....................................................................................................................12
6.1.1 Read Average Value......................................................................................................... 12
6.1.2 Read Memory Record...................................................................................................... 13
6.2 Memory Clear..................................................................................................................... 13
7.About Wrist Type Blood Pressure Monitor....................................................................................13
8 Manufacturer’s Declaration of the EUT........................................................................................ 14
9. Care and Maintenance...................................................................................................................17
9.1 Care Methods...................................................................................................................... 17
9.2 Maintenance Methods......................................................................................................... 18
Appendix............................................................................................................................................19
Ⅰ. Product Specifications.........................................................................................................19
Ⅱ. Blood pressure classification instructions...........................................................................20
Ⅲ. LCD Error message and the cause of error correspondence table......................................20
After-sale Services............................................................................................................................ 21
The latest revision date: 06-2017

V1.0
1
Overview
Before using this product, please read the user’s manual carefully and use it
accordingly. Please keep the user’s manual properly for reference at any time(The
pictures in this manual are for reference only).
Wrist Type Blood Pressure Monitor is with LCD digital display, which is used for measuring
blood pressure and pulse rate (suitable for crowd above 12 years old, not for neonate). This blood
pressure monitor applies to daily health care for home use or outward activities, you can measure
your blood pressure duly in order to make blood pressure health management. It is also applied for
blood pressure reference of clinical patients, but not as a diagnostic basis.
Features:
Small delicate design
Clear LCD digital display
It can store 99 groups of measuring results of two people and display the average reading of
measurement results of the latest three times
Automatic compression and decompression
Voice broadcast function(Optional)
Blood pressure classification function provides convenience for users to judge whether their
blood pressure value is normal or not
2 display units: kPa, mmHg
The product will automatically power-off in 1 minute after measurement
Pay attention to the following 5 points during measurement:
1)Take deep breath for relax before measurement
2)Take measurement after urination and defecation
3)Take measurement with proper posture
4)Take measurement at a fixed time everyday
5)Avoid taking measurement in extreme cold or hot room
BP value changes along with various factors
Keep stable mood when you measure at home, your measured value might be lower than that
from hospital. In order to take good management of your blood pressure, it is very important to
know the average blood pressure value with stability you measured at home.
The following conditions would cause BP value variation even you take
measurement at home.
Within 1 hour after meal
After having liquor, coffee and black tea
After smoking, sports and shower
Talking during measurement
When you are nervous and under unstable mood
Take measurement with the posture of abdomen being pressed
Moving during measurement

V1.0
2
Room temperature changes dramatically
When measuring places or environment changes frequently
Take measurement in moving vehicle
Under continuous measurement for a long time
Don’t rely on your own judgment towards measurement result, please follow the guidance of
doctor.
Safety Precautions
Warnings and graphic symbols in the user’s manual are aimed at enabling users to use the
product safely and correctly, so that avoid any damages to users and the others. Specific meanings
are as follows:
Symbol Descriptions
Warning: to remind you that misuse of the product may cause personnel damage,
measurement failure and article damage. (Article damage includes damage to
housing, property, pets and livestock.)
Enforcement:to represent compulsive behavior, namely matters must to be observed
when using the product
Prohibition: to stand for prohibited contents, namely matters not allowed to do when
using the product
Prohibition of Disassembly
This symbol means BF type equipment
Caution
Diseases such as diabetes, hyperlipidemia, high blood pressure may accelerate
atherosclerosis, even can cause obstacle in peripheral circulation. Blood
pressure value of wrist and upper arm may have large difference for patients
with such diseases and the aged. It is more appropriate for them to use arm-type
blood pressure monitor.
It is very dangerous for patients to make self-judgment and self-treatment
according to the measurement results. Please follow the guidance of the doctor.
Self- judgment is likely to deteriorate the patient’s condition.
There may be a risk, such as arm numbness or pain, that may be caused by over
inflation of the gasbag.
For the sake of safety, please keep the product in the place where the child is
not easy to touch.
Infants and children who can not express their thoughts are prohibited to use the
product, because they will be the cause of the accident or dispute.
Do not use the product for not measuring blood pressure, because it will be the
cause of the accident or dispute.
Please do not use chemical reagents such as thinner, alcohol, gasoline to clean

V1.0
3
this product
Do not beat the product heavily or fall from a height.
Please be sure to use the special arm wrist cuff of this product, otherwise it can
not be accurately measured
Do not use the mobile phone near the machine. There is a risk of making the
machine to malfunction.
Do not disassemble, repair or reconstruction of the machine and the arm belt of
the blood pressure meter, otherwise it can not be measured correctly.
Don’t crash and fall off the machine
Don’t bend or stretch wrist strap
Don't pressure the product before the wrist strap is twined around wrist
As the dry battery leakage can damage the product, so please pay attention to
the following points.
If you don’t use the product for a long time (more than 3 months), please take
out the battery.
After use, replace the battery which is in power shortage with a new one
immediately.
Do not intermix old and new batteries.
Install the positive and negative battery correctly.
Do not intermix different types of batteries.

V1.0
4
1 Product Introduction
1.1 Intended use
Wrist Type Blood Pressure Monitor is with LCD digital display, which is used for measuring
blood pressure and pulse rate (suitable for crowd above 12 years old, not for neonate). This blood
pressure monitor applies to daily health care for home use or outward activities, you can measure
your blood pressure duly in order to make blood pressure health management. It is also applied for
blood pressure reference of clinical patients, but not as a diagnostic basis.
1.2 Contraindications
Serious arrhythmia
1.3 Body
SET
ON/OFF
MEMORY
Display
Battery
cover
1.4 Display

V1.0
5
3
4
10
9
7
13
11
6
5
1
12
8
Fig 1 all symbols display screen
1. Low Power Signal
2. Date/Time display
3. User Symbol
4. Blood Pressure Classification Symbol
5. Average Symbol
6. Blood Pressure Unit Symbol
7. Sets of Memory
8. Memory Symbol
9. Systolic Pressure
10. Diastolic Pressure
11. Pulse Rate
12. Arrhythmia symbol
13. HR Symbol
1.5 Function description
1)Measurement method: decompression-type measurement
2)Result display: SBP/DBP/PR
3)Unit switch: Kpa/mmHg switch (default unit is mmHg)
4)Sets of memory: Double sets of memory, each set has 99 groups memory recall
5)Clock function: set the year, month, date, hour and minute.
6)Low power inspection: LCD will display such a signal to indicate low power,
please check the low power in any state of work
7)Blood pressure classification symbol: Please refer to appendix for
detailed information.
8)Error indication: Please refer to appendix Ⅱfor detailed information
9)Protective function for excessive pressure: If the air pressure is over 295mmHg

V1.0
6
(20ms), the product will exhaust quickly and automatically.
10)Power-off function: Automatically power-off if there is no operation within 1
minute.
11)Voice broadcast function(Optional)
2 Battery Installation
1)Take down the battery cover as the direction of the arrow.
2)Install two No.7 batteries, please pay attention to the positive and negative electrode.
3)Cover the battery
Attention
Signal “ ”: indicates that the batteries are using up, please change two new batteries in
time. (No.7 battery)
If you take out the batteries when power is on, the timer will be stopped. But the stored

V1.0
7
memory will not be removed.
Please reset the time after changing the batteries. Please refer to Chapter 3.2 Time Setup for
detailed information.
Dispose the used batteries according to local regulations about environmental protection.
3 Initial Setup
This blood pressure monitor can automatically store 99 groups of measured values and can
calculate the average value of recent three groups of measuring results. Set the right user
information and time, otherwise, the product cannot store the correct user information and
measuring time.
If you want to use the memory and average function, please make the initial setup before the
primary measurement. Initial setup includes user, time, sound and blood pressure unit setup.
3.1 User Setup
In the shutdown state,long press SET key until the user symbol " " on screen starts
flashing up , press the MEMORY key to switch" "or" " for choosing user, then press SET
key for confirming and switch to year setting mode at the same time, if not to choose the system
defaults user " ".
Press“MEMORY”key to switch
Press “SET”key to confirm
3.2 Time Setup
We can get known from 3.1, press SET key under user mode and it switches to year setting
mode; the screen displays”2016” and flashes, press MEMORY key, value increment can switch
between 2016-2035, when the year to be set appears, press SET key to confirm and it switches
to”month”,”date” setting mode at the same time, it means the value stands for month is in flashing
state, press MEMORY key, value increment can switch between 1-12, when the month to be set
appears, press SET key to confirm, the value stands for date is now in flashing state, it is the time
to set date, press MEMORY key, value increment can switch between 01-31, press SET key to
confirm and it switches to”hour”, “minute” setting mode at the same time, value stands for hour is
in flashing state now, press SET key, value increment switches between 0 to 23 . When the hour to
be set appears, press SET key to confirm, the value stands for minute is now in flashing state, it is
the time to set minute, press MEMORY key, value increment can switch between 00-59, when
minute to be set appears, press SET key to confirm and it switches to”volume” setting mode at the
same time, namely”SP” setting mode.

V1.0
8
Press
“MEMORY” key
to switch year
Press “SET” key
to confirm
Press “MEMORY”
key to switch month
Press “SET” key to
confirm
Press “MEMORY” key to switch hour
Press “SET” key to confirm
Note: Keep on pressing the MEMORY key,the number will increase fast.
3.3 Sound Setup
We can get known from 3.2, after setting up time, press SET key to confirm and switch to
"SP" mode at the same time, the "ON" or "OFF" is in a flashing state now. Press MEMORY key to
choose, when set to "ON", there is a voice prompt when BP monitor is under measurement and
viewing memory; when set to “OFF”, there is no voice prompt. After choosing it, press SET key
for confirming and switch to”BP Unit” mode at the same time. The system defaults “ON” if you
didn’t choose anything.
Press “MEMORY” key to switch
Press “SET” key to confirm
Note: If the instrument is optional without voice, no need to set the sound.
3.4 Blood Pressure Unit Setting
We can get known from 3.3, after setting up volume, press SET key to confirm and switch
to "BP Unit" mode at the same time, the "mmHg" or "KPa" is in a flashing state now. the system
defaults mmHg as blood pressure unit if there is no switch. After choosing it, press SET key for
confirming and it is shutdown at the same time.
Press “MEMORY” key to switch
Press “SET” key to confirm
4 Usage Methods of Wrist Strap
1)During measurement, the palm Should face upward.
Roll up sleeve, wrap the cuff directly against Your skin.
Do not apply over clothing.
2)Position the cuff leaving a clearance of approximately

V1.0
9
1-1.5cm between the cuff and the bottom of your palm.
The measured range of the wrist is 13.5-19.5cm.
3)Grasp the ends of the wrist cuff, pull it and make It close
to the wrist, then fold it, make sure the cuff is not too tight
or too low with your skin.
4)Fold the remaining part of the wrist cuff back out of the way.
Attention
Before fixing the wrist cuff, don’t press the switch button.
If the wrist cuff can not be used on your left wrist, it can be used on your right wrist.
If the cloth of the wrist cuff not closed with your skin, during the measurement, it will loose,
make sure the wrist cuff is securely fastened.
5 Blood Pressure Measurement
5.1 Correct Measurement Methods
1)Sit in a chair as shown in pic 1, don’t lean forward or bend your body.
2)Before measurement, take a deep breath three or four times, recover to the normal breath,
relax your shoulders and arms. Make sure your body relaxed.
3)During measurement, the middle of the wrist cuff should be the same height with the
heart. The body and your wrist can not move. We recommend that you put a soft cushion under the
elbow. As shown in pic 1.
Pic 1 measuring posture
Attention
During measurement, make sure your finger in the natural extensions, otherwise the
measured value will be high although the position of the wrist cuff is right.
Do not hold the wrist cuff by the other hand, otherwise it will cause the error
measurement.
Don’t press your arm during measurement, otherwise the value will be increased.
Before measurement, keep yourself in a relax, comfortable mood. During
约1~1.5cm

V1.0
10
measurement, don’t move the muscle on this arm, if you need, you can put one soft
cushion under your arm.
In every measurement, you should use same posture and wrist.
If the position of the wrist cuff is lower or higher than heart, the reading will be higher
or lower.
make sure the wrist cuff has worn well and with the same height of the heart.
During measurement, don’t talk. Please relax your hand and arm.
During measurement, don’t walk , don’t move your shoulder and arm, don’t bend your
finger.
5.2 Operation Steps of Blood Pressure Measurement
The machine has 2 kinds of BP units,which is mmHg and KPa,please refer to 3.4 part for the
specific method of setting.
Take the mmHg value as an example in the following explanation.
1)Press the “ON/OFF” key, the machine will automatically return to zero, arm with automatic
start filling with air, the pump started to inflate the wristband,
2)In the process of inflation, the screen will display the pressure changes in the wristband.
When the machine detects the pulse, the heart icon will flash.
3) After measurement, the machine will simultaneously exhaust and displays your blood
pressure value and pulse number.
Note: the machine will automatically save the blood pressure and the pulse number.
Attention
During measurement, keep quite until the measurement is finished.
Make sure the machine and the heart at the same height until the measurement is finished.

V1.0
11
If the buckle is off during measurement, please re-wrap the wrist cuff and fix it tightly and
measure it again.
Do not repeatedly measured within a short time, it may result in wrist gore, so that the
measurement value is not correct. In order to make the arterial blood pressure returned to its
previous state, relax your wrist for 2 or 3 minutes before the next measurement.
If Irregular Heartbeat(IHB) brought by common arrhythmias is detected in the procedure of
blood pressure measurement, a signal of “ ”will be displayed.Under this condition, the
blood pressure monitor can keep function,but the results may not be accurate,it’s suggested
that you consult with your physician for accurate assessment.
Interrupt measurement:
If you need to interrupt the measurement for some reason(for example, you feel not good), at
any time. Press “on/off” button. The machine will automatically reduce the air pressure of the
wrist cuff, and interrupt the measurement. If the switch button can not be used, please remove the
wrist cuff and stop the measurement.
5.3 Answers for Questions
Anomalies
Cause
Correction
No display after
press the start button
Whether the power exhausted
Change a new battery
Whether the polarity position
anti-access
Install the battery in proper
placement
Measurement failure
occurs frequently or
measured blood pres
sure value goes extre
mely high or low.
Wristband does not hold with the
heart position relative level
Measuring with correct posture
Wristbands don’t roll up
Roll your wristbands correctly
Whether speak or move the arm
when Measure
Keep quiet, relax body, re-measure
Whether Clothes rolled
oppression arm
Take off clothes that Oppression arm,
re-measure
Pump work, but the
pressure does not
rise
Whether Wrist strap is poor
connections , there is leakage
Ask for repair
Whether Wristbands damaged due
to long years
Please buy a new one
Wrist Cuff leak in a
short time
Whether the cuff wrap too loose
Please wrap the cuff correctly
Even if you press the ON / OFF button , can not enter the
measurement
Please re-power, measured again
During
measurement
suddenly power off
When not use for a long time, the
temperature variation may make
the battery power exhausted
Change a new battery

V1.0
12
The machine is in
good working
condition, but the
results of each
measurement is
different
Mental state and the measured time
will cause measurement values
changes in blood pressure
Before measurement ,Please take a
deep breath and keep in a quiet state
Note: Blood pressure is dynamic, so
must be some difference between the
measured value every time
The measured
values of arm and
wrist is different
Please refer to the "relevant Wrist Blood Pressure Monitor"
The measured
values of blood
pressure is different
with doctors
measured, even
serious small
Record results of your daily blood pressure test, using which to consult with
a doctor
PS: Frequent consultation with a doctor may cause Psychological stress, wh
ich will let your blood pressure test value higher than value under a relax sta
te
Other anomalies
Please press the start button and
check again or change a new battery,
if still can’t measure pleasure connect
us.
6 Memory function
6.1 Memory Recall
This blood pressure storing 2×99 set memory value, each measurement it will automatically
stored. When it stored 2×99 set memory value, the old value will be covered by the new one. It
can show the latest 3 times average value.
Attention
In order to make a note of measurement correctly, before measurement, please make sure set
the right user and time. Under one user, it has stored 99 set data, when you save the 100th set
date, the earliest one date will be deleted.
6.1.1 Read Average Value
Short press the MEMORY key it will show the latest 3 times average value. It will appear
“AVG” in the lower-left corner;
Attention

V1.0
13
When the number of memory value less than 3 sets, such as there are 2 sets of memory value ,
after pressing the memory button it will show the average value of these 2 sets. If there is
only 1 set memory value, after pressing the memory button it will show the average value of
this 1 set , that is the only 1 set measure value.
6.1.2 Read Memory Record
Press the MEMORY key in average mode, you can read the latest 1 time measurement
value,repeat press the MEMORY key to cyclic switch in the previous measurements.
6.2 Memory Clear
Long press the MEMORY key in the memory mode,all memory values will be deleted.
Warning
This operation deletes all memory values, users need to operate carefully.
7.About Wrist Type Blood Pressure Monitor
Blood pressure wrist and upper arm blood pressure value when disease
Diabetes and high cholesterol and high blood pressure will accelerate atherosclerosis.It can
cause dangerous diseases such as stroke or myocardial infarction, or arterial stenosis or peripheral
circulatory disorders when this symptom turns serious. Wrist blood pressure values in these
patients with upper measure blood pressure may have a greater difference. [Even healthy people
due to inappropriate test conditions, there may be a difference about 20mmHg (2.6kPa), so do not
be self-judgment, be sure to accept the doctor's guidance. ] In addition, the difference between the
wrist and upper arm measured blood pressure values, will much affected due to the physiological
state, but the changes of the wrist and upper arm measured blood pressure values are the
same.Therefore examining trends of change in blood pressure can be based on the measurement of
wrist blood pressure .
Wrist and upper arm blood pressure
Normally, blood pressure measurement is used to measure the brachial artery blood
pressure of the upper arm. Because the brachial artery in the wrist is connected with the upper arm,
so the measured blood pressure value is very closed to that one of the upper arm, and it can also
reflect changes of the arm blood pressure. There is a large difference of blood pressure value
between wrist and upper arm of those patients who have circulatory system diseases. Please
consult experts and make health management in accordance with the upper arm blood pressure

V1.0
14
value.
Blood pressure difference between wrist and upper arm
Normally, the measured blood pressure value will appear deviation between wrist and upper
arm of a healthy person in a quiet state. [The difference is within ±10mmHg ( ±1.3kPa)].
However, if measured immediately after exercise or bathing, the change of the difference in blood
pressure between wrist and upper arm is large due to large changes in blood circulation.
Correct the measurement position
When the height position of the wrist and the heart is not the same, due to the weight of blood,
the blood pressure value will change. When the wrist is positioned higher than the heart, the
measured blood pressure value is low; when the wrist position is lower than the heart, the
measured value will be high [each 10cm height difference, about 8mmHg (1kPa)] . But because
blood pressure measuring are closely related to he posture, so it does not necessarily change in
accordance with this value. Be sure to consult the correct measurement posture (see section 5.1).
Daily life and wrist blood pressure
Usually around diet, the upper arm and wrist blood pressure does not have a huge d
ifference. When bath in the high temperature water, due to the heated skin caused blood
vessels shrinkage, the blood pressure rises. But if bathing in warm water, it will expand p
eripheral vascular, decreased blood pressure, this time, the wrist blood pressure will decrea
sed bigger than the arm blood pressure.
8 Manufacturer’s Declaration of the EUT
Guidance and manufacturer´s declaration – electromagnetic emission –
for all EQUIPMENT AND SYSTEMS
1
Guidance and manufacturer´s declaration – electromagnetic emission
2
The product is intended for use in the electromagnetic environment specified below. The customer or the user
of product should assure that it is used in such an environment.
3
Emissions test
Compliance
Electromagnetic environment - guidance
4
RF emissions
CISPR 11
Group 1
The product uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
5
RF emissions
CISPR 11
Class B
The product is suitable for use in all establishments, includin
g domestic establishments and those directly connected to the p
ublic low-voltage power supply network that supplies building
s used for domestic purposes.
6
Harmonic emissions
IEC 61000-3-2
N/A
7
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
N/A
Guidance and manufacturer's declaration – electromagnetic immunity –
for all EQUIPMENT and SYSTEMS
Guidance and manufacturer´s declaration – electromagnetic immunity
The product is intended for use in the electromagnetic environment specified below. The customer or the user of the
product should assure that it is used in such an environment.

V1.0
15
Immunity test
EN 60601
test level
Compliance level
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30 %.
Electrostatic
transient / burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/output
lines
N/A
Mains power quality should be that of
a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
± 1 kV differential
mode
± 2 kV common mode
N/A
Mains power quality should be that of
a typical commercial or hospital
environment.
Voltage dips,
short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
< 5 % UT
(>95 % dip in UT)
for 0.5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
< 5 % UT
(>95 % dip in UT)
for 5 sec
N/A
Mains power quality should be that of
a typical commercial or hospital
environment. If the user of the
product requires continued operation
during power mains interruptions, it is
recommended that the product be
powered from an uninterruptible
power supply or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical
commercial or hospital environment.
NOTE UTis the a. c. mains voltage prior to application of the test level.
Guidance and manufacturer´s declaration – electromagnetic immunity –
for EQUIPMENT and SYSTEM that are not LIFE-SUPPORTING
Guidance and manufacturer´s declaration – electromagnetic immunity
The product is intended for use in the electromagnetic environment specified below. The customer or the user of the
product should assure that it is used in such an environment.

V1.0
16
Immunity test
EN 60601 test level
Compliance level
Electromagnetic environment - guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
N/A
3 V/m
Portable and mobile RF communications
equipment should be used no closer to any part of
the product including cables, than the
recommended separation distance calculated from
the equation applicable to the frequency of the
transmitter.
Recommended separation distance
P
V
d]
5.3
[
1
P
E
d]
5.3
[
1
80 MHz to 800 MHz
P
E
d]
7
[
1
800 MHz to 2.5 GHz
where pis the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and dis the
recommended separation distance in metres (m).b
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,a
should be less than the compliance level in each
frequency range.b
Interference may occur in the vicinity of
equipment marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic is affected by absorption and reflection
from structures, objects and people.
aField strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the product is used exceeds the
applicable RF compliance level above, the product should be observed to verify normal operation. If abnormal

V1.0
17
performance is observed, additional measures may be necessary, such as reorienting or relocating the product.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.
Recommended separation distances between portable and mobile
RF communications equipment and the EQUIPMENT or SYSTEM -
for EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING
Recommended separation distances between
portable and mobile RF communications equipment and the product
The product is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the product can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment (transmitters) and the product as
recommended below, according to the maximum output power of the communications equipment
Separation distance according to frequency of transmitter
m
Rated maximum
output of
transmitter
W
150 kHz to 80 MHz
P
V
d]
5.3
[
1
80 MHz to 800 MHz
P
E
d]
5.3
[
1
800 MHz to 2.5 GHz
P
E
d]
7
[
1
0.01
/
0.12
0.23
0.1
/
0.38
0.73
1
/
1.2
2.3
10
/
3.8
7.3
100
/
12
23
For transmitters rated at a maximum output power not listed above the recommended separation
distance d in metres (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
9. Care and Maintenance
9.1 Care Methods

V1.0
18
Keep the blood pressure clean frequently
Using a soft dry cloth towel clean this blood pressure monitor, if it is very filthy you can wet
the towel with water or neutral detergent, wring out it and wipe the monitor.
Don’t clean the body with naphtha, thinner or gasoline etc.
Don’t wet the cuff or attempt to clean the cuff with water
Attention
Please prevent water or other liquid seep into the blood pressure monitor
9.2 Maintenance Methods
Do not place the unit in the following places:
Easy splashing
Direct sunlight, high temperatures, humidity, dust, full of corrosive gas places
Some slant, vibrative, appulsive place .
Some place storing chemicals or corrosive gas.
Attention
Please Storage after taking out the battery if not use the unit long time.(more than three months)
Do not attempt to replace the parts without authorization.
Do not disassemble and repair this blood pressure monitor.
If you do not comply with all notes and the proper usage of the monitor, our
company does not assume responsibility for the quality.