manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Brüel & Kjær
  6. •
  7. Measuring Instrument
  8. •
  9. Brüel & Kjær 2230 Assembly instructions

Brüel & Kjær 2230 Assembly instructions

Other Brüel & Kjær Measuring Instrument manuals

Brüel & Kjær 2210 Original operating manual

Brüel & Kjær

Brüel & Kjær 2210 Original operating manual

Brüel & Kjær 2236 Guide

Brüel & Kjær

Brüel & Kjær 2236 Guide

Brüel & Kjær 7001 User guide

Brüel & Kjær

Brüel & Kjær 7001 User guide

Brüel & Kjær 2113 User guide

Brüel & Kjær

Brüel & Kjær 2113 User guide

Brüel & Kjær 2425 User guide

Brüel & Kjær

Brüel & Kjær 2425 User guide

Brüel & Kjær i714 Operation manual

Brüel & Kjær

Brüel & Kjær i714 Operation manual

Brüel & Kjær 2218 User manual

Brüel & Kjær

Brüel & Kjær 2218 User manual

Brüel & Kjær 2107 User guide

Brüel & Kjær

Brüel & Kjær 2107 User guide

Brüel & Kjær 4911 Operating instructions

Brüel & Kjær

Brüel & Kjær 4911 Operating instructions

Brüel & Kjær 4423 User guide

Brüel & Kjær

Brüel & Kjær 4423 User guide

Brüel & Kjær B&K 2245 User manual

Brüel & Kjær

Brüel & Kjær B&K 2245 User manual

Brüel & Kjær 4448 Series User manual

Brüel & Kjær

Brüel & Kjær 4448 Series User manual

Brüel & Kjær 2239 B User manual

Brüel & Kjær

Brüel & Kjær 2239 B User manual

Brüel & Kjær 2245 User manual

Brüel & Kjær

Brüel & Kjær 2245 User manual

Brüel & Kjær 2204 Operating instructions

Brüel & Kjær

Brüel & Kjær 2204 Operating instructions

Brüel & Kjær 2007 User guide

Brüel & Kjær

Brüel & Kjær 2007 User guide

Brüel & Kjær 1014 User guide

Brüel & Kjær

Brüel & Kjær 1014 User guide

Brüel & Kjær 2131 User manual

Brüel & Kjær

Brüel & Kjær 2131 User manual

Popular Measuring Instrument manuals by other brands

Powerfix Profi 278296 Operation and safety notes

Powerfix Profi

Powerfix Profi 278296 Operation and safety notes

Test Equipment Depot GVT-427B user manual

Test Equipment Depot

Test Equipment Depot GVT-427B user manual

Fieldpiece ACH Operator's manual

Fieldpiece

Fieldpiece ACH Operator's manual

FLYSURFER VIRON3 user manual

FLYSURFER

FLYSURFER VIRON3 user manual

GMW TG uni 1 operating manual

GMW

GMW TG uni 1 operating manual

Downeaster Wind & Weather Medallion Series instruction manual

Downeaster

Downeaster Wind & Weather Medallion Series instruction manual

Hanna Instruments HI96725C instruction manual

Hanna Instruments

Hanna Instruments HI96725C instruction manual

Nokeval KMR260 quick guide

Nokeval

Nokeval KMR260 quick guide

HOKUYO AUTOMATIC UBG-05LN instruction manual

HOKUYO AUTOMATIC

HOKUYO AUTOMATIC UBG-05LN instruction manual

Fluke 96000 Series Operator's manual

Fluke

Fluke 96000 Series Operator's manual

Test Products International SP565 user manual

Test Products International

Test Products International SP565 user manual

General Sleep Zmachine Insight+ DT-200 Service manual

General Sleep

General Sleep Zmachine Insight+ DT-200 Service manual

Sensa Core Lacto Spark user manual

Sensa Core

Sensa Core Lacto Spark user manual

VOLTCRAFT UTS-1980 operating instructions

VOLTCRAFT

VOLTCRAFT UTS-1980 operating instructions

EBCHQ 94915 Operation manual

EBCHQ

EBCHQ 94915 Operation manual

Balluff BIP LD2-T017-01-EP-S4 Series user guide

Balluff

Balluff BIP LD2-T017-01-EP-S4 Series user guide

Pulsar IMP Lite Series instruction manual

Pulsar

Pulsar IMP Lite Series instruction manual

ETCR 6800 user manual

ETCR

ETCR 6800 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

Brilel
&
Kjcer
037-0539
Precision Integrating
Sound Level Meter
Type 2230
Valid
from
serial no. 1028 340
.
.,
J,
Briiel
&
Kjcer
037-0539
Precision Integrating
Sound Level Meter
Type 2230
Valid
from
serial no. 1028340
BRUEL &
KJ.LER
Nrerum-Denmark Service Instruction 2230
Precision Integrating
Sound Level Meter
Type 2230
Valid
from
serial
no. 1 028
340
037-0539
Consisting of:
Service
Instruction
Checking
Procedure
Analog
Circuit
Digital
Circuit
ZC
0020
Preamplifier
ZF 0020
Attenuator
ZZ
0187
Main
Board
Mechanical
Parts
Circuit
Diagram
Updating
page
0-1
0-3
0-4
1-1
1-2
2-1
2-3
3-1
2
1
3
4
4-1
5-1
date
4.83
4.83
4.83
4.83
4.83
4.83
4.83
4.83
10.82
10.82
10.82
10.82
10.82
10.82
4.83
4.83
Due
to
the
constant
progress
in
technology
the
instrument
will
from
time
to
time
be
brought
up
to
date
in
order
to
provide
continuously
improved
performance.
For
this
reason
there
may
be
small
variations
between
the
instru-
ment
and
the
Service
Instruction.
However,
the
local
Service
Representative
is
in
possession
of
all
informations
regarding
any
change
that
has
been
made.
4.83
Trouble Shooting
If
any
fault
should
occur,
check
the
instrument
according
to
the
Checking
Procedure.
The
Checking
Procedure
is
divided
up
into
sections,
where
each
section
will
refer
to
a
subsection
in
the
Service
Instruction
corre-
sponding
to
a
specific
part
of
the
instrument.
The
subsection
in
the
Service
Instruction
will
normally
consist
of
a
Block
Diagram,
a
Technical
Description,
an
Adjustment
Proce-
dure
and a
Simplified
Diagram
where
the
test
points
are
identi-
fied.
Before
correcting
any
apparent
deviation
make
sure
that
the
measuring
instrument
has
tolerances
small
enough
not
to
affect
the
measurement.
When
a
fault
has
been
traced
and
corrected,
the
voltages
and
adjustments
influenced
by
the
correction
must
be
rechecked.
The
complete
instrument
should
then
be
tested
to
make
sure
that
all
basic
functions
are
operative.
The
tolerances
given
in
these
notes
are
intended
for
use
as
guide
for
adjustments.
Many
instruments
have
various
aids
to
be
used
when
trouble
shooting
or
adjustment
is
carried
out.
They
are
not
necessarily
mentioned
in
the
Service
Instruction.
The
local
Service
Representative
is
in
possession
of
information
regarding
these
aids.
Spare
Parts
Please
state
type
and
serial
number
of
the
instrument
when
ordering
spare
parts.
0-1
BRUEL
&
KJJER
Nrerum-Denmark
4.83
~
''
...
'''
.t
.•...
'
..
':''
.•.•
'
......
',.
·.···
·
·····
.····
..
·.·
·
....
·
.....
·.
·
...
·"
.
·······
·~
~~:
·
::
=-
·MQ,
;
:,
.,.
.
...
...
•
.•
::.
• .
1!101,
lfditlj
·
!lii!Jt
"'"~230
.....,..~.
:=I
tr
+,
'"~.~
H
..
·.
•
.. ··
.•
••
•
•.
•
•..
••.••.•
•••.••..
·
••
j•
..
:•,.·.··.··
••"·
••
·
..
f1
.
~
..
,.
.
..
·
;lfto!
lifO
·•·
.. •
•fHIIJ
!10
.. .
'•ti•
OIIt
10·
..
~
-)
411
.
..
C
tm
.
f"'\
,..,.;$
1
..
m;;filiiJO"""''~
Valid
from
serial
no
. 1028 340
Service
Instruction
2230
0-3
0
I
~
·
~
Co
w
,-·-·-·-·
-·-·-·
-·-·
-·-·--,
Ext.
Filters
AC
Output
DC
Output
1 I
Weighting
RMS
I I
Input
Amp.
Filters
Output
Amp.
Circuit
DC
Amp
.
Calcula-
tion
RAM
r
Program
ROM
~
Display
ci >
~
>
......,.._-+---+---.t
AID
~
Conv.
t------"'
Micro
Processor
~
Display
1----J Cl
L/
/l
Dr
1v
er
t--1
_I
l.l_/
Ref
.
Gen
.
A
I
I
I
I
I
I
•
I
I
II
--I
r--
1
Adder
: +
20dB
FSO
Switch
•
I
I
I
I
I
I
I
I
I
! j
......___.,__,
D
I ·i
I i
I
I .
I I
"---I
Decoder
Latch
rr==>.
But
fer
Display
Switch Error
Detector
rr
Control-
!
er
~)-
I
I •
:1
Overload
I i I :
-J
:£)>1--------'
. 1
Detector
1
I
I I :
I I
-·
-.-+----.
-+-·
--.
I I I I
....L_.
--·
__
.
__
Analog
Circuit 2230.2
.
--·--·--
·--
·-
: I
I
I
I
I
Ref.~
Test
Operate
dB
I
120~
110
100
90
80
70
FS
D
liJ~
LinD
FREa
WEIGHT.
--·-
~
-·--·
--·--·-·-·r--
1
I
SEL
~
Leq
MAX
MIN
SPL
DISPLAY
SWITCH
I
~
R
~SET
Digital Circuit 2230.31
1\)
1\)
w
0
CJ)
CD
<
c:;·
CD
::::l
~
2
~
a·
::::l
BRUEL
&
KJ.LER
Nrerum-Denmark
Sensitivity
POWER:
"On
"
REF., TEST, OPERATE:
"Operate"
FSD:
"100"
FREQ. WEIGHTING:
"Lin
1\
"
EXT. FILTEA:
"Out"
SOUND INCIDENCE:
"Frontal
"
DISPLAY:
"SPL"
TIME WEIGHTING:
"Fast"
DETECTOR:
"RMS"
ALL-MAX/MIN-PAUSE:
"All"
Linearity
FSD
to
"120"
Frequency Response
FSD
to
"100"
Weighting Filters
FREQ. WEIGHTING
to
"A
"
or
"C"
4.83
Checking Procedure 2230.1
Input
Voltage
DC
Output
Voltage
Input
signal
through
a 15 pF
capacitor
'
(Adaptor
Type
WA
0302)
to
Preamplifier
ZC
0020 input: 50
mV
,
1kHz.
Check
the Display reading: 94,
0.
If
necessary
adjust
"Sens. Adj." on
the
right
side
of
the
instrument.
If
not
consult
2230.2 Analog
Circuit.
L
....L..III:il..
l•
""'-....,
~
~
OdB
-10dB
-20dB
-30dB
-
40dB
-SO
dB
3,000V
2,500V
2,
000V
1,500v 1,000 v
O,SOOV
Input
signal 1kHz
corresponding
to
3,000 V
at
"DC
Output".
Decrease
the Input
voltage
in 10dB
step
and
check
that
the
DC
Output
de-
creases
by
500 mV.
Tolerance: ±
30
mV.
Display
reading
Freq.
Hz
unn
Lin
10 87,9 ± 2,0 93,4 ± 0,5
20 93,1 ± 0,5 93,7 ± 0,3
40 93,9 ± 0,3 93,9 ± 0,3
80 94,0 ± 0,3 94,0 ± 0,3
1,0k 94,0 94,0
2,5 k 94,0 ± 0,3 94,0 ± 0,3
S,Ok 94,0 ± 0,3 94,0 ± 0,3
10 k 94,0 ± 0,3 94,0 ± 0,3
20k
93,9 ± 0,3 93,5 ± 0,5
Input
signal 1kHz
corresponding
to
94,0 on
the
Display.
Check
the
frequency
response
according
to
the
table
above.
Display
reading
Freq.
Tolerance
Hz A c +
dB
-
10 79,7 1,5 2,0
20
435
87,8 1,0 1,0
40 59,4 92,0 0,5 0,5
80 71,5 93,5 0,5 0,5
160 80,6 93,9 0,3 0,3
315 87,4 94,0 0,3 0,3
1,0k 94,0 94,0 0 0
2,5 k 95,3 93,7 0,3 0,3
S,Ok 94,5 92,7 0,5 0,7
10k 91,5 89,6 1,0 1,5
20k
84,7 82,8 1,0 1,5
Input
signal 1kHz
corresponding
to
94,0 on
the
Display
.
Check
the
frequency
response
according
to
the
table
above
.
If
not
consult
2230.2 Analog
Circuit.
Valid
from
serial no. 1028 340
1-1
2230.1
Checking
Procedure
Random Freq. Response
SOUND INCIDENCE
to
"Random"
FREQ. WEIGHTING
to
"Lin
1\
"
Leq.
SEL
SOUND INCIDENCE
to
"Frontal"
FSDto"110"
DISPLAY
to
"Leq"
DISPLAY
to
"SEL"
TIME WEIGHTING
to
"Imp
or
Slow"
Overload
DISPLAY
to
"SPL"
TIME WEIGHTING
to
"Fast"
Test
REF., TEST, OPERATE
to
"Test"
REF.
REF. TEST, OPERATE
to
"Ref."
Calibration
REF. TEST. OPERATE
to
"Operate"
1-2
Freq.
Display
kHz
reading
1,0 94,0
2,5 94,3 ± 0,3
5,0 95,2 ± 0,5
10 96,8 ± 0,5
20 98,2 ± 0,5
Input signal 1kHz
corresponding
to 94,0 on the Display.
Check
the
frequency
response
according
to
the
table
above.
Input signal 1kHz
corresponding
to
94,0 on
the
Display.
Activate
"Reset"
Check
the Display reading: 94,0 ± 0,
1.
Activate
"
Reset"
and let the
Leq
run
for
exactly
10 s.
Check
that
the Display reads: 103,9 and is
increasing
every
second
the next 10
seconds.
Check
that
an
arrow
occurs
to
the
right
of
the Display.
Notice: This
arrow
can be
removed
by
removing
the
jumper
"1"
near the
"Peak
-
RMS"
switch.
If the
above
cannot
by
fulfilled
consult
item
2230.3
Digital
Circuit.
Input signal: 1,05
V,
1kHz.
Check
that
the Overload
indication
on the
Display
does
not
light.
Increase the Input
voltage
to: 1,12V.
Check
that
the
Overload
lights
: ± 0,05 V
If
not
consult
2230.2
Analog
Circuit.
Check
that
all
digits
and Signs on the
Display
lights.
Check
the Display reading: 94,0 ±
0,
1.
Mount
the
Preamplifier
ZC 0020 and
.;,
the
Microphone
Type 4155.
Check
by
means
of
a Sound Level
Calibrator
or
a
Pistonphone
that
the
calibra-
tion is
correct.
If
necessary
adjust
"Sens.
Adj.
".
4.83
BROEL &
KJIER
Nrerum-
Denmark
Detector Balance
POWER:
"On"
REF., TEST, OPERATE:
"Operate"
FSD: "110
dB
"
FREQ. WEIGHTING:
"Lin
1\
"
EXT. FILTER:
"In"
SOUND INCIDENCE:
"Frontal"
DISPLAY:
"SPL"
TIME WEIGHTING:
"Fast"
DETECTOR:
"RMS"
ALL-MAX/MIN-PAUSE
:
"All"
Pr
eamp
Offset
FSD
to
"
70
dB
"
EXT. FILTER
to
"
Out
"
FSDto
"110"
Weighting
Network
FREQ. WEIGHTING
to
"
C"
FREQ. WEIGHTING
to
"A"
Overload
EXT. FILTER
to
"In"
FREQ. WEIGHTING
to
"Lin
1\
"
Reference Generator
4.83
REF., TEST, OPERATE
to
"
Ref."
EXT. FILTER
to
"Out
"
Analog Circuit
2230.2
ADJUSTMENT
PROCEDURE
Input
signal
to
"Ext.
Filter"
socket
J201
pin
1: 633
mV
,
1kHz
.
Check
at
ZE 0410 pin 15
that
the
curves
are
equal.
If
necessary
adjust
P205.
Change
the
input
voltage
to
2mV.
Set
P206
to
middle
position.
Check
the Display reading:
"60,0".
If
necessary
adjust P207.
Check
the
"DC
Output"
voltage: 500 ± 2 mV.
If
necessary
adjust
P209.
Change
the
input
voltage
to
: 633
mV
.
Check
the "DC
Output
"
voltage
: 3000 ± 2 mV.
If
necessary
adjust
P208.
Check
the
Display reading: "11
0"
.
If
necessary
adjust
P1
03. (Ref.
to
Digital
Circuit
2230.
3)
.
Repeat
the
above
adjustments
until all
are
within
the
tolerances.
Short
circuit
the
Preamplifier
ZC 0020
input
to
ground
through
a 15pF
capacitor
(Adaptor
Type
WA 0302).
Check
the
voltage
at
the
output
of
the
Preamplifier
on
the
point
C205, 206:
0 ±
20mV
DC.
If
necessary
adjust P201.
Input
signal through the
adaptor
to
the
ZC
0020
input:
50
mV
,
1kHz.
Check
the
Display r.eading: "94,0".
If
necessary
adjust P206.
Check
the Display reading:
"94,0".
If
necessary
adjust P202.
Check
the
Display reading:
"94
,
0".
If
necessary
adjust P203.
Input
signal
to
"Ext.
Filter"
socket
J201
pin
1: 2,2
V,
1kHz.
Check
that
the
Overload
arrow
lights
within
± 0,2 dB.
If
necessary
adjust P204.
Check
the
frequency
at
the Ref.
input
to
ZE 0399
pin
3:
1000 ± 1 Hz.
If
necessary
adjust
P1
01
.
Check
the Display reading: "94,0".
If
necessary
adjust
P1
02.
Valid
from
serial no. 1028 340 2-1