BROEL
&
KJ~R
Nrerum
-
Denmark
1.
1.
Electrical
Calibration
REF
. 1:
"In"
INPUT LEVEL:
"Low"
GAIN CONTROL MODE:
"Auto"
LMS
DETECTOR
:
"Fast"
READOUT CONTROL: "Clock"
EXT.
FILTER
: "
Out
"
WEIGHTING SELECTOR:
"Lin."
REC
./
PLAYBACK: "Recording"
(Located on the left hand side
of
the
221 0)
REF
. 1. to
"Out"
LMS
DETECTOR
to
"Slow"
dB
+ 1
0
10
8.78
+0
,5
Checking Procedure 2210.1
31.
6
Check the
Meter
deflection: Approx.
-1
6 dB.
If no deflection consult
the
Preamplifier and
Log
Peak Detector
2210
.3.
Check the voltage at "To Ext.
Filter"
: 1 V RMS
Check the voltaQe at
"AC
Output
":
3.
16
V RMS.
If
this
voltage is not correct consult the VCA and
Weighting
Network
2210
.4.
Check the voltage at
"DC
Output"
:
4200
mV.
If necessary adjust
REF
. 1.
Press
REF
. 2 and activate
RESET
.
Check the voltage at
"DC
Output
":
2200
mV.
If necessary adjust
REF.
2.
The adjustment of
REF.
and
REF
. 2 should be repeated
until
the
reading is exactly
4200
mV and
2200
mV.
If
the voltages
for
REF
. 1 and
REF
. 2 cannot
be
obtained consult
the
LMS and Hold Circuit
2210
.5.
Check the reading on the Display:
84
,0 and
44,0
respectively.
If the Display reading is not correct consult the Logic and Display
2210
.6.
Activate MIC. SENS .
CORR
. and adjust MIC. SENS. CORR. potm:
to
O,OdB correction
when
REF
.':1' is pushed in.
Note
In order to proceed
with
the
Checking Procedure
it
is imperative
that
the above mentioned ad-
justment
of
REF
. 1.
REF
. 2 and
the
MIC. SENS. CORR. is correct.
In order to obtain the correct results of
the
following
measurements the signal
must
be ap-
plied to
the
221 0 via an Adaptor
JJ
2614
Input Signal:
50mV
251,2Hz
.
Check the Display reading:
94,0
±0,1 dB.
Check the frequency response according to
the
table
below
.
+
0.3
-0
.3
251.2
Ref
.
+0,2
o.z
1k
+
0.3
-o.3
-0
,7
4k
Bk
20
k
Frequency
in
·
Hz
If the Display reading is out of tolerance consult
the
VCA and
Weighting
Network Section
221
0.4.
Check the
Meter
reading: Approx.
-6
dB.
If
the
meter is
off
consult the Preamplifier and Log. Peak Detector
2210
.3.
Decrease the
Input
Signal to give a Display reading
of
84,0
dB
.
Check the Linearity of the Dynamic Range in 10 dB increments.
Tolerance±
0.4dB
. Ref.
84
,
0dB
.
If
the Linearity is
out
of tolerance consult the VCA and
Weighting
Network
2210
.4.
Note:
In order
to
maintain
a
sufficient
accuracy
it
is suggested
to
use a Precision
Attenuator
having an accuracy of 0,
05
dB
or
better.
Valid from serial no.
65
98
67
1-1