BRUEL & KJAER 1617 User manual

Instruction
Manual
Band Pass Filter Type
1617
'
--
--
-
,-
::::::
-
~:
..
•
r-
'-
C
-"'-
"
..
T
..
.._.
..
± .
::-
l:
::
•
033-0143
1617
The Type 161 7
has
50
third
-octave
filter
bands
with
centre
frequencies
2
Hz
to
160
kHz, 41
overlapping
octave
filter
bands
from
2 Hz
to
20
kHz, A, 6 , C, D-
weighting
networks
, and a
Linear
pass band 1 Hz
to
200
kHz. The
filter
pass bands
fulfil
the
most
severe
standards,
IEC
225
-
1966
, DIN
45651
(1
/ 1-oct) and
45652
(1
/ 3-oct).
and
ANSI 51 .
11
-
1966
Class
II
(1
/1-oct)
and Class III
(1
/
3-oct)
.
In
addition
to
facilities
for
filter
scanning
controlled
by
6 & K Level Recorders,
there
is a
built-in
interface
to
the
proposed
IEC
standard
via
which
full
control
can
be
exer-
cised.
An
internal
control
system
is able to
programme
the
averaging
times
of
the
mea-
suring
i
nstrument
that
is
connected
to
the
Type
161
7 so
that
the
optimum
6T
product
is
maintained
during
a
frequency
scan.
~
Bruel &
Kjcer

BAND
PASS FILTER
TYPE
1617
September
1977


CONTENTS
1.
INTRODUCTION
AND
SPECIFICATIONS
(PRODUCT
DATA)
..........
............
........
....
.................
1
2.
CONTROlS
.............
..........
.......
..
...........
..
....
..
...................................
...............
.....................
..
9
2.1 . FRONT PANEL
..................
.......
..........
.....................
.......
..
...
........
........
..
.... 9
2.2. REAR PANEL
......
....
..
....
..
..
...
.............................
.....
.. ..
....
.......
......
..............
10
3.
OPERATION
.....................
..
......................
...........
...................
...............
...........
..
.........
..
........
13
3.1. PRELIMINARY ADJUSTMENTS .....
........
.....
.............
..
...
..
.
........
..
....
..
..
..
........
13
Selection
of
Line Voltage
...............................................................
:............ 13
Fuse Rating
......
.
..........................
.
....................................
......
....
............
..
13
Line Connections
........
..
......
......
....
..
...........
...
..
.........
..
.......
..
.......
............
...
13
Grounding
Considerations
...................
.
..........................
..
....
...........
...........
14
3.2. OPERATION WITH MEASURING AMPLIFIERS ALONE
...................
..
...............
14
Measuring
Procedure ....
..................
.. ..
.............
......
..
..
............
..
........
.. ..
.
..
...
15
Calibration
..
....
..
..........
......................................
................
.
...................
....
16
3.
3.
OPERATION WITH MEASURING AMPLIFIER
AND
LEVEL RECORDER
..........
....
16
Fixed averaging
time
analysis
with
Level Recorder Type
2306
..
..........
........
....
....................
...........
..
..........
....
..
16
Analysis
with
variable averaging
time
using
the
Level Recorder Type
2306
........
......
..
..........
......
.......
.
............................
20
Fixed averaging
time
analysis
with
the
Level Recorder Type
2307
............................................................
.............
22
Analysis
with
variable averaging
time
using
the
Level Recorder
2307
................................
....
........
..........
.
.......
...
..
........
25
Analysis
with
variable averaging
time
using an X-Y Recorder
..............
.
..
..
..
.. ..
27
3.4
. OPERATION WITH OTHER
Bruel
& Kjrer INSTRUMENTS
........
......
..
......
..
........
28
Operation
with
Sound
Level
Meter
Type
2203
and Type
2209
.....
..
.
..
.
..
.
......................
...
..
........
..
.......
..
.
..
........
....
........
......
28
Operation
with
Vibration
Meter
Type
2511
.........................
............
...........
..
29
3.5. DIGITAL
DATA
TRANSFER AND REMOTE CONTROL
VIA
THE
IEC
INTERFACE BUS
................
.. ..
..
...........
..
......
.......
......
...........
.. ..
30
IEC
Functions
Implemented
.........
.......
..
......................
........
........................
31
Selection
of
Listener
Address
........
.
......
..
..
.....
..
.........
.. ..
.........
.......
..
............
31
Programming
Sequence
....................
.
..
.........
..
.........
....
........
....
..
.......
...
..
....
31
Parallel Poll ....
......
.........
.....
..........
..
.....
.....
.......
..
.........
......
....
..
........
.........
..
33
Indication
of
busy .....
.........
...........
.........
..
.......
........
.....
..
........
..........
..
.. .. .. ..
33
Interruption
of
readout
.................
.......
.........
..
.
..........
.......
...........
..
.
............
34
3.6. DIGITAL PROGRAMMING EXAMPLES .........
..
..................
..
......
.
...................
34
General
Programming
..............
....
.........
.
...........
.........
........
.....
..
..........
.......
34
Programming
for
operation
with
the
Hewlett
-Packard
9825A
....
............
.........
35
Programming
for
operation
with
the
Tektronix
4051
........
......
..
............
.........
36

4. SPECIAL
CONSIDERATIONS
IN
FREQUENCY
ANALYSIS
....
................
..
....................
..
........
..
38
4.1. FILTER DEFINITIONS ................
........
...
......
.
..
..
..
....
.. .. ..
..
.......
..
....................
38
The Ideal Filter
..
.
..............
.......
..
..
..
........
......
.................
................
..
.....
..
...
38
Bandwidth
.......
..
..
..
..
............
..........
..
..
..
.
..
........
.
..
.................
..
....
................
38
Octaves and Third Octave ......
..
.
.. ..
.
.. ..
....
..
......
.:...........
..
.
.. ..
...............
..
....
..
..
39
Specified Filter Frequencies
..
..
....
..
.......
.
..
.........
..
.
..
.
..
....
..
.........
.. ..
....
.. ..
..
.. ..
..
39
4.2
. RMS MEASUREMENT AND STATISTICAL ACCURACY ...
....................
......
..
...
.40
4.3. PRACTICAL ANALYSIS
OF
STATIONARY SIGNALS ..
..
..............
..
.........
..
..
.......41
Averaging
Time .
..
..
..
......
..
..
........
.............
..
........
.. .. .. .... ..
.
............
..
....
.. ..
..
....41
DC
or
AC Recording ...
..
..
..
.
......
.
.. ..
....
..............
.................
......
.................
..
.
.42
Writing
Speed .....
..
........
.. ..
..
......
............
..
..
..
..................
.....
.................
..
..
..
42
Paper Speed .
..
.
..
.
..
.
...........
......
..
........
..............
.........
.........
..
....
.. ..
..
...........
.42
4.4
. USE
OF
VARIABLE AVERAGING TIME PROGRAMS ...
..
.......
........
.....
....
..
.......
44

GENERAL FEATURES:
• .41
third-octave
filter
bands,
centre
frequencies
2
Hz
to
20
kHz
• 41 overlapping octave
filter
bands,
centre
frequencies
2 Hz
to
20
kHz
•
Third
-octave and octave
filters
in accordance
with
IEC
225
-
1966
, DIN
45651
and
45652,
and
ANSI S1.
11-1966
•
Digital
display
of
selected
centre
frequency
and
bandwidth
•
Built-in
A-weighting
network
•
Electronic
control
of
filter
switching
•
Manual
or
Level Recorder
controlled
filter
scanning
• Selectable scan
start
position
to
save
time,
eliminates
low
frequencies
when
required
•
Input
overload
warning
circuit
• Floating
or
grounded
Signal
input
17
-
094
types
161 7
and
161 8
Third-Octave
and
Octave Band Pass Filters
-
e -
-
e _
~
~
'.1
....
_ _
-::-::
I
.,
---
;
~
~=
.
,~
.
1618
ADDITIONAL
FEATURES
TYPE
1617
:
'
'''
'''
''I\l'''''
I
...
·
'<)
.
...
.
~
..
_.
f··
....
• 9
third-octave
filter
bands,
centre
frequencies
25
kHz
to
160
kHz
•
Built-in
B, C,
D-weighting
networks
•
Comprehensive
connection
facilities
for
use
with
other
instruments
•
Operation
of
front
panel
controls
via
built-in
IEC
Interface
•
Selection
and
control
of
averaging
time
program
for
the
measuring
instrument
USES:
_-.....e-no
__
---
'
....
,
...
(With
suitable
Measuring
Amplif
iers)
•
Third-octave
and octave
frequency
analysis of
sound
and
vibration
•
Measurement
of
frequency
response
•
Measurement
of
sound
transmission
loss
•
Measurement
bf
vibration
isolation
•
Testing
acoustic
materials
• Recording
of
spectrograms
with
constant
confidence
level
(1
61 7)

The
Third-Octave
and Octave
Band Pass
Filters
Type
1617
and
Type
1618
are
key
links
in
any
in-
strumentation
chain
for
measure-
ment
and
recording
of
the
fre-
quency
spectra
of
sound,
vibration,
and
electro-acoustic
signals
. Sev-
eral novel
features
have been
in-
corporated,
including
electronic
fil-
ter
switching
and
digital
indication
of
the
selected
filter
.
The
Band
Pass
Filter
Type
1618
is
the
basic
instru-
ment,
while
the
Type
1617
com-
bines
an
extended
frequency
range
with
comprehensive
control
and
syn-
chronisation
facilities
for
use
with
other
instruments.
The
Type 161 8 has fiIter
band
centre
frequencies
from
2
Hz
to
20
kHz
that
can
be
divided
into
41
third-octave
bands,
or
41
overlap-
ping octave
bands
covering
the
14
octave
frequency
range.
It
has
a
built-in
A-weighting
network,
and
there
is
indication
given
when
the
level
of
the
signal
being analyzed
overloads
the
input
section.
The
Band Pass
Filter
has
provision
for
fil-
ter
scanning
controlled
by B & K Le-
vel Recorders.
The Type
1617
has
filter
band
centre
frequencies
from
2 Hz
to
160
kHz
that
can
be
divided
into
50
third-octave
bands.
It
has 41
over-
lapping octave bands
covering
14
octaves
from
2 Hz
to
20
kHz. In
addi-
tion
to
the
A-weighting,
it
includes
B, C, and
D-weighting
networks,
and
there
is also
an
input
overload
indicator
lamp.
Filter
scanning
can
be
controlled
by a Level Recorder,
and
there
is a
built-in
interface
to
the
proposed
IEC
Digital
Bus
Stand-
ard to
permit
direct
control
by
other
instruments
and
systems
using
this
standard. A DC
ramp
output
can
control
the
X-axis
of
an X-V Recor-
der,
and a
control
circuit
is
included
to
select
averaging
time
programs
for
use by
the
measuring
instru-
ment.
Both
Band
Pass
Filters
will
gener-
ally
be used
with
one
of
the
Measur-
ing
Amplifiers
Type
2606,
Type
2607,
or
Type
2608,
but
other
in-
struments
such
as
the
Frequency
Analyzers
Type
2120
and
Type
2121
,
or
the
Heterodyne
Analyzer
Type
2010
can be
employed
.
Con-
nected
to
a B & K
Condenser
M
icro-
phone,
these
measuring
instru-
ments
all
form
a
sound
measure-
2
ment
system
that
satisfies
IEC
179
for
precision
sound
level
meters,
while
Type
2606,
Type
2607,
and
Type
2120
satisfy
IEC
179
A
for
Im-
pulse
measurement
.
sic
system
of
Type
1618
and Type
2608
is
for
RMS
measurement
in
the
audio
frequency
range,
with
fil-
ter
selection
by Level Recorder.
Opti-
mum
utilisation
of
all
measuring
modes
and
control
possibilities
is ob-
tained
with
a
combination
of
Type
1617
and Type
2607,
which
gives
the
ability
to
measure
and
analyze
in
the
frequency
range
from
2 Hz
to
160
kHz,
obtaining
RMS
and Peak
values
from
all
commonly
encoun-
tered
random,
quasi-random,
peri-
odic, and
complex
waveforms
.
This
combination
can be used
together
with
an X-V Recorder.
All
three
Measuring
Amplifiers
can be
employed
with
either
Band
Pass
Filter
, so
that
selection
of
which
Measuring
Amplifier
to
use
will
largely
depend
on
the
particular
measurement
application.
Table
1
shows
the
various
combinations
available,
together
with
brief
specifi-
cations,
and
the
type
of
measure-
ment
and
analysis
possible.
The
ba-
Measuring
Amplifiers
Band
Pass
Filter
Type
1618
Band
Pass
Filter
Type
1617
General
:
General:
General:
Freq. Range: 2
Hz
-
200
kHz Freq. Rang
e:
1.4
Hz -
22
kHz Freq. Range:
1.4
Hz -
180
kHz
Dyn
. Range: 1
OIN
-
300V
41
third-oct.
bands
50
third-oct.
bands
Fast and
Slow
averaging 41
overlapping
octave bands 41 overlapping octave bands
Input
overload
warning
A-weighting
filter
-A, B,
C,
D-weighting
filters
Averaging
time
programs
Control
via
built-in
IEC
interface
•
DC
ramp
for
X-V Recorder
control
Type
2608
and
2121
RMS
measurement
RMS
measurement
Crest
factors
up to 5
20Hz
-
22
kHz
20
Hz -
180
kHz
AC
output
to Level Recorders
(Audio
frequencies)
A-weighting
network
Manual
or
remote
control
of
filter
scanning
Digital
bus
control
of
filter
scanning
Measurements
can be
made
at
lower
frequen
cies
if
th
e
averaging
ability
of
a Level Recorder
2307
is used
Type
2606
Comprehensive
RMS
measurement
at
high
crest factors
RMS and Peak
measurement
Impulse
and
Hold modes
20Hz
-
22
kHz
20Hz-
180kHz
Crest
factors
up to
40
(Audio
frequencies)
AC
or
DC
outputs
to
Recorders
A, B, C,
D-weighting
networks
Manual
or
remote
control
of
filter
scanning
Range
settings
shown
on
Digital
bus
control
of
filt
er
meter
scale
scanning
Measurement
in Peak
or
Impulse
Modes
1.4
Hz -
22
kHz
1.4
Hz - 1
80
kHz
RMS measu
rement
s
ca
n be made at l
ower
fre
quencies
if
th
e
averaging
ability
of a Level Recorder
2307
is used
Type
2010
(RMS
only)
,
Measurement
of
RMS and Peak values
of
all
commonly
encount-
I
2607,
&
2120
ered
random
. quasi-
random
. periodic. and
complex
waveforms
RMS and Peak
measurement
Impulse
and Hold modes
1.4
Hz -'22 kHz 1
.4
Hz - 1
80
kHz
Crest
factors
up
to 5
Manual
or
remote
control
of
filter
scanning
AC
or
DC
outputs
to
Recorders
50
dB
logarithmic
DC
output
Manual
control
of
avo
time
fur
2607
,
2120
,
m2010
proportional
to
RMS Level
A, B, C,
D-weighting
networks
Digital
bus
control
of
filter
Range
settings
shown
on
meter
scanning
and
averaging
times
scale
for
2607
and
2120
This
combination
gives
optimum
use
of
all meas
urem
e
nt
and
control
facilities
,
including
X-V recordin!'l
(not
2010)
Table
1.
Band
Pass
Filter
and
Measuring
Amplifier
Combinations

Description
Both Band Pass Filters are
de-
signed
to
operate on
the
signals ob-
tained
from
the
External
Filter
termi-
nals
of
the
Measuring
Amplifiers
,
but
any
input
signal
of approxi-
mately
1 V RMS
with
a crest
factor
less
than
5 can be accepted
(the
higher
crest
factor
capability
of
the
Type
2606
can be used
at
lower
sig-
nal levels).
Input
and
output
are via
coaxial B & K sockets
on
the
front
panel, paralled by BNC
terminals
on
the
rear.
Filter I Filter
II
11 3
Octave
Centre
Centre
Bandwidth
Frequency
Frequency
at
3.7 dB
Hz Hz Hz
Approx
.
2
0,46
2.5 0 .
50
3.
15
0 .
73
4
0.92
5 1.
16
6.3
1,45
B 1.
83
10
2.
30
12.5
2.
90
16
3.
70
20
4.60
25
5.8
31 .5 7
.3
40
9.2
50
11 .6
63
14
.5
80
18
.3
100
23
125
29
160
37
200
46
250
58
315
73
400
92
500
116
630
145
800
183
1000
230
1250
290
1600
370
2000
460
2500
580
3150
730
4000
920
5000
1160
6300
1450
8000
1830
10000
2300
12500
2900
16000
3700
20000
4600
25000
5800
31500
7300
40000
Additional
9200
50000
1/
3-octav
e
11600
63000
filters
14500
80000
in
1617
18300
100000
23000
125000
29000
160000
37000
Table
2. Filter
centre
frequencies
in pre-
ferred
series
.
Heavy
type
denotes
preferred
series
for
full-octaves
Input
Warning
Front
P
anel'
Controls
Amplifier
Low
Pass
F
ilter
Output
Amplifi
er
Digital .
t::=+.;;
Controller
t-
Ou
tput
761039
Fig.l . Block
diagram
of
the
Band
Pass
Filters Type
1617
and
Type
1618
Fig.1 is a block
diagram
valid
for
both
versions
of Band Pass Filter,
where
shaded areas
represent
fea-
tures
found
only
in
the
Type
1617
.
Third-Octave
Band
Pass
Filters
The active
Filter
sections
consist
of a
matched
pair
of
variable
fre-
quency
six-pole
Butterworth
filters
which
can be
electronically
switched
to
yield
third-octave
or
full
-
octave
bandwidths
.
Between
them
the
two
Filters cover
the
frequency
range
from
2 Hz to
20
kHz
centre
frequencies
,
with
each
Filter
being
switched
to
alternate
third-octave
centre
frequencies
in
the
preferred
series (see Table 2).
In
the
Band Pass Filter Type
1617
there
is an
additional
set
of
third-octave
Filters
with
centre
fre-
quencies
between
25
kHz and
160
kHz,
to
extend
the
high
fre-
quency
range of
the
instrument.
The
filter
characteristics
of
the
in-
dividual
third-octave
Filters
used
in
these
instruments
fulfil
the
require-
ments
of
IEC
225
-
1966
, DIN
45652,
and ANSI S1.
11-1966
Class III,
which
are
the
most
rigor-
o
I,
/,
1
~
/:
'.\
/ :
20
\\
/ I
1\
/,'
3 '. \
/ I I \
/ "\
/ " \
/ \ \
/
50
\ \ IEC
/'
I "
I
/
I
/
0.2 0,5
00
~
60
.g
..
"
~
70
< 75
2
\
........
\
\
Typical B & K
1/3 Octave Filter
'.
ANSI
Class
III
.lOIN
45652)
'--
4
Frequency relative
to
centre frequency fl
fm
110103
Fig.2
. Typical
third
-
octave
filter
response
3

ous standards applying
to
third-
octave
filters
.The response
curve
for
a typical
third-octave
filter
is
shown
in Fig.2, and
the
top of
the
curve
in
the
enlarged
view
in Fig .3.
The
IEC,
DIN, and ANSI
limitations
are also
indicated in both Figures. Peak-to-
valley ripple in
the
pass
band
is less
than
0,5
dB
with
attenuation
within
±
0,5
dB
.
Attenuation
of
frequencies
outside 5 x and 1/ 5
of
the
band
centre
frequency
is
better
than
75dB
.
Octave
Band
Pass
Filters
Octave
Filters
are
formed
in both
Band Pass
Filters
by
electronically
altering
the
components
and
charac-
t~stics
of
the
Filter
circuits.
This
ar-
rangement
gives a
flat
crest
to
the
characteristic
curve, and
low
peak-
to-valley
ripple
.
The
octave Filters
cover
the
frequency
range
from
2 Hz to
20
kHz
centre
frequencies,
selectable at
third-octaves
in
the
preferred series.
There
is
no
provi-
sion for
connection
of
full-octave
fil-
ters at
higher
frequencies
in
either
Filter Set.
All
octave
Filters
contained
in
the
Type 161 7 and Type 161 8
conform
to
IEC
225
-
1966,
DIN
45651
,
and ANSI S1.
11-1966
Class II,
which
are
the
strictest
standards
for
octave
filters.
Fig.4
shows
the
top
of
a typical octave
filter
characteris-
tic,
attenuation
outside
8 x and 1/ 8
of
the
band
centre
frequency
is bet-
ter
than
60
dB. Peak-to-valley ripple
is less
than
0,5
dB,
while
attenua-
tion
in
the
pass
band
is
within
±
0,5
dB
.
The
total
integrated
random
(white) noise
power
passed by
the
practical octave and
third-octave
Fil-
+5
a Linear 1
Hz
to
200 kHz
/"
!-
/V
V
C
-5
-10
-15
'g
- 2
c
o
.
~
:J
C
e
;;.
o
2
4
6
8
~;
I
16
18
20
22
L-
__
~~~L_~~~~~~~
0,4 0,6 0,8 1,0 1,4 1,8 2,5
0,5 0,7 0,9 1,2 1,6 2,0
Frequency relative
to
centre frequency flfm
170124/1
Fig.3.
Top
of a
third-octave
filter
characteris-
tic
ters
in
the
Type
1617
and Type
1618
is equal to
that
which
would
be passed by an ideal octave
or
third-octave
filter.
Weighting
Networks
In
addition
to
the
Linear
re-
sponse, and
A-weighting
network
in-
cluded in
the
Type
1618,
the
Type
1617
contains
B- and
C-weighting
networks,
plus
the
D-weighting
net-
work
specified in
IEC
537
for
meas-
urement
of
aircraft
noise. The
fre-
quency
responses of
the
four
weighting
networks
are
shown
in
Fig.5
. The Figure also
indicates
the
Linear
range 1 Hz
to
200
kHz
obtain-
able
from
the
Type
1617,
and
the
Linear
range 1 Hz
to
40
kHz
from
the
Type
1618.
Filter
Selection
and
Scanning
Ranges
Filter
switching
is
accomplished
electronically
by
FET
switches
in
the
Filter
Selectors
that
are
regu-
~~
~i!all
Fre~Ue~cy
Char~ctJri
!
ti~sl
V
lI::
,.......f-'
b=
I:?""
~
./
B
/"
V
0
~
~
/,
~
/
-20
-2
5
-30
0,2 V
a,s
V V
10
,/
V
/20 50 100 200 500 1 k 2 k
:-
Fig.5
.
Weighting
networks
and
linear
functions
4
'"
-2
"0
_,
§ 0
'';::; I
..
:J
2
a;
3
:::
.
« 5
0.5 0.6 0.7 Q.8
0.9
1.
0
1.2
1.-4
1.6
1
.8
2
.0
2.5
Frequency relative
to
centre frequency flfm
170125/2
Fig.4.
Top
of a
typical
octave
filter
charac-
teristic
lated by
the
Digital
Controller
acting
on
instructions
from
internal
or
re-
mote
control
settings
.
Bandwidth
of
the
Filter
in
use is selected by
the
three
position
Selectivity
switch,
giv-
ing a choice
of
third-octave
band-
width
scanning
in
third
-octave
steps, octave
bandwidth
scanning
in
third-octave
steps
(with
adjacent
bands
partially
overlapping).
or
oc-
tave
bandwidth
scanning
in
full-
octave steps.
Manual
selection
of
any
particular
Filter
band is
made
by
turning
the
Manual
Filter
Selector
control
to
the
required
position
.
On
the
Band Pass
Filter Type
1618
the
Manual
Filter
Selector
has a scale graduated in
third-octave
centre
frequencies
from
2 Hz
to
20
kHz,
plus
the
two
positions
"A-weighting"
and
"lin-
ear"
.
There
are
two
measuring
ranges on
the
Band Pass Filter Type
161
7,
selected by
the
Ra
nge
switch
, and
hence
two
frequency
0
I'
"'-
~
"'-
"-
"'"
~
B+C
r\
5 k 10 k 20 k
Linear 1
Hz
to
200 kHz
1'.
~
1'\
\
1161
1
S1
I'"
1\
'"
\ \
~l\
"-
~
"
i
~
1\
50 k lOa k 200 k
500
k
770105

scales.
One
covers
the
full
fre-
quency
range
with
graduations
in
third-octaves
with
centre
frequen-
cies
from
2 Hz to
160
kHz. The
other
range covers
third-octave
cen-
tre
frequencies
between
2 Hz and
40
kHz, and
includes
A-
,
B-
, C-, D-
weighting,
and a
Linear
position
.
Both
models
have a
"Linear"
mode
available
on
the
Range
switch,
which
permits
a
Linear
output
to
be
obtained
at any
point
in a scan,
without
moving
the
Manual
Filter
Selector
.
The
centre
frequency
of
the
se-
lected
Filter
band
is
indicated
on
a
half
-i
nch
digital
display
that
also
shows
whether
the
Filter
is
function
-
ing as a
third-octave
or
as an oc-
tave
filter
,
or
whether
a
weighting
network
has
been
selected.
Filter
Scanning
In
addition
to its
control
of
the
Fil-
ter
switching
as
directed
by
the
Manual
Filter
Selector
,
the
Digital
Controller
can
operate
the
Filter
scan
on
the
command
s
of
an
exter
-
nal
source
.
Control
by a Level Recor-
der
Type
2306
or
Type
2307
is pos-
sible
with
both Band Pass
Filters
.
On
the
Type
1617
control
can also
be exercised via
the
built
-in IEC In-
terface
,
or
an X-V Recorder
can
be
operated
in
synchronism
with
Filter
scanning
(possible
with
Type
2607
Measuring
Amplifier
or
equivalent).
Selection
of
manual
or
recorder
control
is
made
by
the
Filter
Control
Mode
Manual
/Recorder
switch,
while
the
Stop/Run
switch
enables
the
Recorder
in
use
to
be
controlled
from
the
Band
Pass Filter.
This
con-
trol
facility
is blocked
when
"Man-
ual" is selected,
or
when
the
Type
1617
is being
controlled
via
the
In-
terface
.
When
operating
in
any
remote
control
mode
,
the
progress
of
the
scan can be
followed
on
the
Digital
Display
, as
the
Manual
Filter
Selec-
tor
does
not
rotate
during
an
auto-
matic
scan.
The
Filter
frequency
sweep
always
starts
from
the
band
in
which
the
Manual
Filter
Selector
is
standing,
it
sweeps
through
the
selected range, and being
internally
actuated,
it
returns
instantly
to its
starting
point.
This saves
analysis
time
when
low
frequencies
(with
correspondingly
long
averaging
times)
will
not
be
required
while
operating
the
Type
1617
with
an X-
Y Recorder,
or
under
digital
control
via
the
Interface
bus.
Similarly,
when
recording
sound
on
a Level
Recorder,
unwanted
low
frequency
bands can be excluded
from
the
trace
to
yield a
clear
audio
fre
-
quency
spectrogram
.
Connections
for
synchronisation
with
a Level Recorder Type
2307
is
made via a
7-pin
DIN socket on
the
rear panel (see Fig.6). Connection
to a Level Recorder Type
2306
is
achieved via an 8-pin DIN socket.
which
with
the
Type
1617
is also
used
for
X-V recording.
--
0 @
-I ,--,
• •
IIIIIII.
~
.IIII
11111111111111111111111111111111111111
-
@
-,-
, ,
• •
.~,
11111111111111111111111111111111111111
Fig
.
6.
Rear
panels
of
the
Band
Pass
Filters
CO(/)'f'ighI
AGF
2005
Fig
.7 .
Conditions
where
longer
averaging
times
are
required
for
high
frequency
signals
.
The
lower
curve
was
made
with
too
short
averaging
time.
while
for
the
upper
curve
.
averag-
ing
was
correct
5

Averaging
Time
Control
This
ability,
allows
the
Type
1617
to vary
the
averaging
time
of
the
Measuring
Amplifier
Type
2607
or
the
Narrow
Band
Analyzer
Type
2120
during
the
course of
the
fre-
quency
scan. The advantage is
that
the
averaging
time
can be kept as
short
as
possible
to
maintain
an ac-
ceptable
analysis
time
,
while
being
long enough at
lower
frequencies
to
achieve
optimum
measuring
accu-
racy, i.e.
to
maintain
a good
confi-
dence level. The best analysis
condi-
tions
are obtained
when
the
product
of
bandwidth
B (in
Hz)
and averag-
ing
time
T (in
s)
is held
constant
throughout
the
scan.
Therefore
the
Type
1617
is equipped
with
averag-
ing
time
programs
which
can be se-
lected according
to
test
conditions
to
maintain
the
BT
product
as
near
constant
as possible and give a con-
stant
confidence level.
At
higher
frequencies,
if
a
con-
stant
BT
product
is
to
be
main-
tained
,
the
averaging
time
becomes
so
short
that
the
writing
speed
of
the
Recorder
will
be
the
practical li-
mitation
on
the
speed of
the
analy-
sis.
Therefore
the
shortest
averag-
ing
time
that
can be selected by
the
programs in
the
Type
1617
is
0,1 s.
However
, some
high
fre-
quency
signals
like
the
noise or vi-
bration
from
a
slowly
rotating
ma-
chine
may have a
low
repetition
fre
-
quency
that
causes
low
frequency
modulation
as
shown
in
Fig.7,
where
0,1 s averaging
time
is
too
short
. To
allow
satisfactory
averag-
ing
times
to
be selected
for
this
type of signal, a choice
of
three
min
-
imum
times
(0
,1 s, 1,0 s, and
10
s)
is provided on
the
Type
1617
. Fig.8
is a
chart
indicating
the
averaging
times
and
change-over
frequencies
in
the
nine
programs
available,
to-
gether
with
the
minimum
and
maxi-
mum
averaging
times.
When
operating
with
variable av-
eraging
times
, selection
of
the
re-
quired
program
is made
with
the
two
Averaging
Control
switches
on
the
front
panel, and
the
Recorder
Averaging
Time
switch
on
the
rear
panel
must
also be set
to
"Variable
Averaging
Time".
Table 3
shows
how
the
programs
can be selected
by suitable
combination
of
the
Pro-
gram
and
Minimum
Time
switch
po-
sitions. The
programs
are
obtained
6
A
ve
raging
Time
(S)
100 C,F.J Slow Random
30 10 S
G,H,J
10
3
0,3
0,1
L-----~-
___
D
__
-,~,-----------·.------------'B=~--=~F~~~-,~~~~~~~~--C--D~;~:~.:
A,B,C
, ! , ! , , I ! ,
2 5 10 20
50
100 200 500
lk
2k
5 k 10 k 20 k
Sw
itchover Frequency (Hz)
Fig
.S.
Averaging
times
and
changeover
frequencies
Minimum
Av
.
Ti
me(s) 0,1 1,0
10
Slow
Rand
om
C F J
Fa
st Ra
nd
om
B E H
Sin
e A 0 (G)
Table
3.
Selection
of
the
required
averaging
time
program
via
the
15
lead AVERAGING TIME
CONTROL socket on
the
rear panel.
With
a Level Recorder,
automatic
averaging
time
control
functions
as
follows
. The scan
starts
with
the
Re-
corder being held
for
a period
of
five
times
the
programmed
averag-
ing
time
for
the
band,
during
which
time
the
pen comes up
to
the
cor-
rectly averaged level.
Then
a long
averaging
time
is selected
that
keeps
the
output
constant
while
the
Recorder
draws
the
trace
by advan-
cing
the
paper at
high
speed
to
the
next
frequency
band.
The
Filter
steps
to
the
next band in
the
se-
quence
, and
the
programmed
aver-
aging
time
is selected,
after
which
the
Recorder hold period begins
again.
Using an X-V Recorder,
where
the
X-axis is
controlled
by
the
DC
ramp
output
from
the
Type 161 7,
the
Recorder pauses
for
five
times
the
averaging
time,
and
then
takes
eight
steps
through
each
third-
octave band.
Figs.9
,
10
, and
11
show
analysis
of
the
same
input
signal recorded
using
three
different
methods
to il-
lustrate
their
relative
merits
.
Built
-
in
IEC
Interface
The
Type
1617
is
fully
program
-
mable
via a
built
-
in
interface
that
sat-
isfies
the
requirements
of
the
pro-
posed
IEC
Standard
Digital
Interface
for
Programmable
Instrumentation
.
This
permits
the
filter
frequency
bandwidths,
and averaging
time
to
be varied in an
arbitary
sequence,
and recording
initiated
to
suit
any
particular
measuring
program. It
also
permits
on-
line
changes to be
made to accomodate
new
events as
they
occur
.
-
Br
uel & Kj
ae
r
Br
uet &
Kjm
r
Sru
et &
Kje
r
DO
OOOOO
DOaOOQOoaOOOOODOOOOOODDOOOOODOODDOOOODDODD
~
-
ruel
&
Kjmr
P
otentiomet
er
R
a
n
ge
:
~
dB
R
ec
t
i
f
ier
:
~
l
ower
lim
.
Freq
.:
~
Hz
Wr.
Speed
:~
mm
/
5ec.
Pap
e r
~
e
d
:-L
mm
/ .e c
C
open
hag
en
50
2
1075
dBdS
40
2050
~
eas
u'ing
Ob
j.:
Fi
xed
---
~~;~ag;ng
-
3 rs
40
1/3
octave-
Time
used
250 s
20
f,
0 30
~~
~.'
B
~
~~
~~f"
---
1 20
~
Date
-==
5
ign
.:
__
_
10
20
'
'tt
z
50
1
00
200
OP
1124
Multipl
v Fre
qu
ency
Scal
e
by
Ze
ro
l
eve
l:
1612
/ 2
112
Fig.9
.
Frequency
analysis
with
Level
Recorder
and
fixed
averaging
time
d dB
60
3C
15
ABC
76
1042

The
following
interface
functions
are available on
the
Type
1617:
Acceptor
Handshake
(AH 1)
Listener
Function
(L2)
Remote Local
Function
(RL2)
Parallel Poll
Function
(PP2)
Busy
is
indicated
via NRFD (not
ready
for
data) hold
method
when
the
Band Pass
Filter
is
feeding
a Re-
corder.
A
Listen
Address
switch
on
the
rear panel
allows
the
instrument
Listener
address
to
be
set
for
use
with
remote
digital
control.
There
is
a
PPR
Selector
for
setting
the
Paral-
lel Poll Response, used
here
to
re-
port
whether
there
has
been
an
overload in
the
Input
Amplifier
si
nce
the
previous
interrogation
.
This
is used
to
assess
the
validity
of
the
levels
measured
in
the
filter
bands.
Connection
to
the
digital
bus is
made via
the
25-lead
plug
on
the
rear panel
of
the
instrument
.
Operation
with
other
instruments
Fig.12
shows
the
connection
pos-
sibilities
with
other
Bruel
& Kjcer
in-
struments,
including
the
cables re-
quired.
The
Band
Pass Filters can also be
employed as
external
filters
for
B & K
Sound
Level
Meters
Type
2203
and Type
2209
, and
for
Vibra
-
tion
Meters
Type
2510
and Type
2511
.
,-
.,1
~I
~I~
o
~
~Io
81~
.,1
~I
", I
'
~
I
~I
~I
Input
I
Input
Input
Band
Pass
Filter
1618
Measuring
Amplifier
2608
Measuring
Amplifier
2606
Mea
suring
Amplifier
2607
Br
ue
l & Kj
le
r
Bru
el &
Kj
se
r
Bt
uel &
Kj
e ,
000000000000000000000000
000
0000000000000000000000
rue
I&
Kjrer
Potent
iome
te:
R
ange
'~
dB
R
ec
tifier· DC l ow
er
li
m
Fr
eq '
200
Hz
Wr
Sp
eed"
800
mm/ sec Pap
er
Speed·
.!Q..
mm
/
sec
.- -
..
-.
--
10 7
Co
penhagen
50
2 5
dBd
B
40
2
aas
ur
ing
Obj
.:
Variable -
Averaging -
Time
--
3 1
1/3 octave-
Fast Random/D,1
Time
used
90
~
---
21
0
50
540
s.
0
30
~~
~~
;
~~
~
~
~f'
:
R
ae
. No
.:
0 8 18
__
Sign
.:
__
00
20
10
20
H,
50
100
QP
1124
Mul
t
ipl
y Fre
qu
enc
y Scale by
200
d d
6
4
3
1
50
0
1000
2000
5000
1
0000
200
00
40
0000
A 8 e li
Zero l
eve
l:
161
2 /
211
2
AB
C 161043
Fig.
10
.
Frequency
analysis
with
level
Recorder
and
variable
averaging
times
Ab
l(iue
4 dekader I
62
,S mm. Copyright
AG
F 1008
Fig.11
.
Frequency
analysis
with
X-V
Recorder
and
variable
averaging
times
Control
Cable
AQ
0034
-------------
-l~
Digital Conn
ec
tion
I 18
to
IEC I
nterf
ac
e
bus
I I
~
AO
0129 I
<{
I 1
:3
~~~-~~-~~~~~~~~~~~~~~~~-J~
2306
with
8 - pin Control I I
~
Socket:
AQ
0034
Level
Recorder 2306 1 0
2306
with
7 - pin Control +
~
Socket:.
AQ
0182 I
~
Signal Output, B & K
to
BNC Connection Cable AO 0064
Signal
Output
, B & K Connection
Cable
AO 0014
Level
Recorder 2307
1£
I.;,
I
.
~
1'=
__
...I
x- Y Recorder
AO 0064
Tape
Re
corder 7003 or 7004
76 1040
Fig.12.
Connection
possibilities
with
other
B & K
instrumentation
7

Specifications
l61
7
and
1
61
8
Band
Pass
Filters
:
In accordance
with
IEC
225
1966
, DIN
45651
and
45652
, and ANSI
Sl
.
ll-
1966
best classes
The total integrated r
andom
white
noise
power passed by
the
filters
in
these
in
-
struments
is e
qual
to
that
which
would
be
passed by an ideal
filter
Centre
Frequencies
1617
:
1/ 3
oct:
2
Hz
to
160
kHz
(50
bands)
1/ 1
oct
: 2
Hz
to
20
kHz (41 overlapping
bands at 1/ 3 octave
interv
als covering
14
octaves)
Centre
Frequencies
1618
:
1/ 3
oct:
2
Hz
to
20kHz
(41 bands)
1/ 1
oct
: 2
Hz
to
20
kHz (41 overlapping
bands at 1/ 3 octave intervals covering
14
octave
s)
Attenuation
Outside
Pass
Band
:
1/ 3
oct
: >
75
dB at 5 x and 115
ce
n-
tre
fr
e
quency
1/ 1
oct:
>
60
dB at Sx and l / S centre
frequency
Attenuation
at
Centre
Frequency
(fm)
:
1/ 3
oct
and
1/ 1
oct
:
OdB
±O
,
5dB
Maximum
Peak-to-Valley
Ripple
:
1/ 3
oct:
O,
5dB
1/ 1
oct
: O,
5dB
Overall
Selective
Frequency
Range
:
1617
:
l.4Hz
to
lS0kH
z
1618
:
l.4H
z to
22kH
z
Linear Pass
Band
(Available
from
Range
switch
or Ma
nual
Filter Selector):
1617:
l,6Hz
to
160kHz
attenuati
on is OdS
±
0,3
dB
1
Hz
to
200
kHz
att
e
nuation
is 0 dB
±0,5 dB
1618:
1,6
Hz
to 31 ,5 kHz a
tt
e
nuati
on is 0 dB
±0,3 dB
1
Hz
to
40
kHz atte
nuati
on is 0 dB
±0,5 dB
Filter
Selection
:
Type
1617
:
2
Hz
to 1
60
kHz
2
Hz
to
40
kHz, D, A, B, C, Linear
Type
1618
:
2
Hz
to
20
kHz, A,
Linear
Switching
control
:
Manual
:
from
"
Manual
Filter Se
lector"
Automatic
:
from
a Level Recorder
Automatic
: to control an X-V Recor
der
(1617
only)
(When sc
anning
octave filters, e
ither
full
-
octave or
third
-octave stepping can be se-
lected)
Automatic
: via
th
e
IEC
interface
bus
(1617
only)
Weighting
Networks:
Curves A, B, and C are in accordance
with
IEC
179
for
pr
ecision
sound
level
meters
(1
61
S has
A-w
e
ighting
only)
Curve D is
in
a
cc
ordance
with
IEC
537
Averaging
Time
Programmes
(1617
only):
Used
with
Measuring
Amplifi
ers
that
fea-
ture
remote controlled averaging
tim
es
(Type
2607)
Programmes
Available
:
See Table 3 and Fig.S
Input:
Via B & K coaxial socket on
front
panel,
paralleled by
BNC
socket on rear
Impedance
:
1
MOl
l
l00pF
Voltage
:
1V RMS
nominal
5 Vpeak
maximum
5,
6V
(± O,
3V)
overload
warning
lamp
lights
Distortion
:
Band
Pass
Filters
:
< 0,1%
with
1V signal level
< 0,3%
with
3,6 V signal level
Linear
Range:
< 0,1%
with
1V signal level
<
0,3
%
with
3,6 V signal level
Noise
:
<
150j1V
(typ.
100)
Band Pass Filters
<
11
Oj1V (typ.
SO)
A, B, and C-w eighting
network
s
<
250j1V
(typ.
lS0)
D-weig
hting
net-
work
< 1
OOj1V
(typ.
SO)
Linear range
Output
:
Via B & K coaxial socket on
front
panel,
paralleled by
BNC
socket on rear
Impedance
:
<
500
Minimum
Load
Impedance
:
5
kO
11
1 nF
for
less
than
± 0,2% reading
error
DC
Ramp
Output
(1617
only):
Used for
controlling
the
X-axis
of
an X-V
Recorder
oV
at
the
starting
fr
e
quency
O,
20S
V per 1/ 3 octave increase rate
10.4
V
maximum
output
Load impedance >
10
kO
Digital
Bus
Interface
(1617
only):
Conforms
with
the
proposed
IEC
Stand-
ard Digital Interface
for
Progra
mmable
In-
strum
e
ntation
Contains
the
following
function
s:
Acceptor
Handshake (AH1) .
Listener
Function
(L2)
Remote Local Function (RL2)
Parallel Poll
Function
(PP2)
Environmental
Conditions
:
Operating
Temperature
Range
:
5°C to
40
°C (41 °F to
104
°
F)
giv
e
no
ef
-
fect
Storage
Temperature
:
-25
°C to +
70
°C (-
13
°F to +
15S
0
F)
Permitted
Humidity
Range
:
The
outputs
from
the
i
nstruments
are ac-
curate
between 0 and
90
% relative
hu-
mid
i
ty
, provided
there
is
no
condensation
Power
Requirements
:
100
,
115
,
127
,
200
,
220
,
240
V AC
±1
0%
50
to
400Hz
1617
: appro
x.
30VA
1618
: approx.
2SVA
Comply
with
IEC
34S
Class I
Cabinet
:
Suppli
ed as model A
(light
-
weight
metal
cabinet), B (model A in a
mahogany
cab-
inet), or C (as A
but
with
flanges for
standard
19
"
ra
ck
mounting)
Dimensions
:
Height
:
133
mm
(5,
25
in)
Width
:
430
mm
(16
,9 in)
Depth
:
200mm
(7,
9in)
Weight:
1617
: 6,
5kg(14,3Ib)
1618
: 6,3
kg
(13
,
9Ib)
Accessories
Included
:
1 AN
0010
Power cable
2
JB
0002
Banana plug
2
JP
0035
BNC
plug
2
JP
0101
B & K coaxial plug
1
JP
0703
7-pin DIN plug
1
JP
OS02 S-
pin
DIN plug
2
VF
0012
Fu
se
Accessories
Available
:
See Fig.
12
.

2.
1.
FRONT
PANEL
Se
lected Fi
lt
er
-~
-
r r
'H
,I-'
,.
_
L,~
. •_
.....
.-
2. CONTROLS
Ma
n
ua
l Filter
Se
lector Averaging Control
.
--
~
-
Power
On
r _
._c..
....
...
_
-,
In
pu
t
Se
l
ectivity
Ra
n
ge
Filter
Co
n
tro
l Mode
Ou
tp
ut
770353
Fig
.2.1.
Front
panel
controls
of
the
Type
1617
POWER ON:
AVERAGING
CONTROL:
MANUAL
FILTER SELECTOR:
SELECTED FILTER:
INPUT:
Power
is
on
with
the
switch
in
the
up
position
.
Warm
-up
time
is ap-
proximately
five
seconds.
Two
switches
used
to
select
averaging
time
programs
,
when
the
Band
Pass
Filter
is
connected
to
a
Measuring
Amplif
i
er
such
as
the
Type
2607,
which
has
provision
for
remote
control
of
its
averaging
time
.
Program:
Selects
an averaging
time
program
suitable
for
"
Sine
",
"Fast
Random",
or
"Slow
Random"
signals
.
Minimum
Time:
Selects
minimum
averaging
times
for
the
averaging
time
programs.
A
rotary
switch
that
selects
the
filters
and
we
i
ghting
networks
when
operating
in
the
"Manual"
control
mode
.
The
two
frequency
scales
cover
the
measuring
ranges
that
can be
chosen
by
the
"Range"
sw
i
tch
and
indicate
the
selected
frequency
band.
A
digital
display
that
shows
the
centre
frequency
of
the
selected
filter
.
The
display
also
indicates
the
bandwidth
chosen,
and
whether
a
linear
or
weighted
response has been selected.
Standard
B & K coaxial
socket,
fitting
JP
0101
Plugs,
paralleled
by a
BNC socket
on
the
rear
panel.
There
is
an
i
nput
"
Overload"
lamp
, and
9

SELECTIVITY:
RANGE:
FILTER CONTROL
MODE
:
OUTPUT:
2.2.
REAR
PANEL
CHASSIS-SIGNAL
GROUND
:
10
a
separate
ground
terminal.
The
maximum
input
voltage
is 5 V,
and
the
impedance
is 1
MO
in
parallel
with
100
pF.
A
three-position
switch
that
selects
the
bandwidth
of
the
filters
,
and
the
centre
frequency
step
length
for
full-octave
scanning
.
Functions
are
as
follows:
Upper
position:
third-octave
bandwidth
with
scanning
in
third-octave
steps.
Middle
position:
full-octave
bandwidth
with
scanning
in
third-octave
steps,
with
overlapping
.
Lower
position:
full-octave
bandwidth
with
scanning
in
full-octave
steps.
A
three-position
switch
to
select
the
narrow
band
ranges,
or
"Linear",
for
both
manual
and
recorder
controlled
operation
.
Functions
are as
fol-
lows
:
Upper
position:
"Linear"
range
from
1 Hz to 1
80
kHz,
independent
of
the
position
of
the
MANUAL
FILTER SELECTOR.
Middle
position:
"2
Hz
to
160
kHz"
,
filters
are
selected
automatically
,
or
by
the
MANUAL
FILTER SELECTOR
from
the
inner
scale on
the
MANUAL
FILTER SELECTOR.
Lower
position:
"2
Hz
to
40
kHz-D-A-B-C-Lin
.",
filters
are
selected
au-
tomatically,
or
by
the
MANUAL
FILTER SELECTOR
from
the
outer
scale
on
the
MANUAL
FILTER SELECTOR.
Two
switches
that
control
how
the
filters
are
selected,
plus
two
LED
in-
dicators
,
one
to
show
whether
the
instrument
is
in
operation
under
manual
control,
and
the
other
indicating
that
all
front
panel
controls
are
inactive
while
the
i
nstrument
is
being
controlled
via
the
Digital
In-
terface
.
The
left-hand
switch
can
select
"Manual
" so
that
the
filters
are
chosen
by
the
MANUAL
FILTER SELECTOR and
start
of
Recorder read-
out
is
impossible,
or
it
can
select
"Recorder"
mode
allowing
filter
selec-
tion
to
be
made
by
extern
aI
control
.
The
right
-
hand
switch
enables
the
Recorder
in
use
to
be
controlled
from
the
Band
Pass
Filter
.
This
facility
is blocked
when
"Manual"
has
been
selected,
or
when
control
is
being
exercised via
the
Digital
Inter-
face.
Two
LEOs
light
to
indicate
whether
"Stop"
or
"Run"
is
selected
.
Standard
B & K coaxial
socket,
fitting
JP
0101
Plugs,
paralleled
by a
BNC socket on
the
rear
panel.
There
is a
separate
ground
terminal.
Im-
pedance less
than
200,
minimum
load
impedance
50
in
parallel
with
1 nF.
A
pair
of
clamping
terminals
with
a
shorting
strap
permits
the
signal
ground
to
be
connected
to
the
instrument
chassis
.

Line Voltage
Se
lector
Ch
assis
Sign
al
Gr
ound
A
ve
raging
Tim
e
F
uses
PPR
Se
l
ec
t
or
L
is
ten Add
ress
Output
r;
III
Mains
Input
Di
g
it
al
Bu
s (I
EC
I A
v.
Tim
e Control
Fig.2.2 .
Rear
panel
connections
on
the
Type
1617
I
nput
R
eco
rder
Control II
LINE VOLTAGE SELECTOR:
Allows
the
Filter
to
be operated on
mains
voltages
of
100
,
115, 127,
150
,
220
, and
240V
(± 10%), selected by
turg
the
selector
with
a
small
coin
or
a w ide-bladed
screwdriver
to
indicate
the
available vol-
tage
value
.
Any
mains
frequency
from
50
to
400
Hz
may
be used.
LINE INPUT:
[
FUSES:
DIGITAL BUS (IEC):
LISTEN ADDRESS:
PPR
SELECTOR:
A
three
-
pin
socket
for
connecting
line
and
ground
to
the
i
nstrument
.
The
Line
connection
lead
AN
0010
(supplied) is delivered
without
a
plug
to
permit
fitting
an appropriate plug
suiting
local standards
Note,
the
colour
code is given in Fig .2.3. Before
connecting
a
supply
the
line
voltage
setting
and
fuse
checks given in section 3.1.1 and
3.1 .2
should
be carried
out
.
Blue Neutral
AC Power
"oJ
Brown Supply
Liv
I J
Gr
een
and
Ye
llow Ground
172246
Fig.2.3.
Line
input
connections
Two
0,2
A
line
fuses
.
With
line
voltages
100,
115,
and
127
V, both
fuses
are used,
while
with
200,
220
, and
240
V
only
one
fuse is
used, see section 3.1 .2.
25
-pole D-socket
for
connecting
to
the
IEC
Interface
Bus.
5-pole
switch
for
selecting
the
listening
address
of
the
instrument
when
under
remote
control
via
the
IEC
Interface
Bus
.
8-pole
switch
for
selecting
the
Parallel Pole Response (only
one
pole
may
be set
to
1).
11

AVERAGING TIME
FIXED VARIABLE: A two-pOSition
sliding
switch
to select one
of
two
operation modes
with
a Level Recorder, as
further
explained in sections 3.3.3/
24
and
3.3.4/ 9 .
AVERAGING TIME CONTROL: A
15-pin
outlet
for
the
averaging
time
control
program
signals
from
the
Band Pass Filter
to
the
Measuring
Amplifier.
The socket
fits
the
15
-
pin
Plug
JP
1501
. For
connections
see section 3.3.2.
OUTPUT:
INPUT:
RECORDER CONTROL I:
RECORDER CONTROL
II
:
BNC socket in parallel
with
front
panel
output
socket.
BNC socket in parallel
with
front
panel
input
socket.
7-pin DIN socket
for
remote
switching
of
the
filters
by a Level Recorder
Type
2307,
and
for
remote
control
of
the
Level Recorder by
the
Type
1617
. For
connections
see Fig.2.
4.
8-pin
DIN socket
for
remote
switching
of
the
filters
by a Level Recorder
Type
2306,
and
for
remote
start
and stop
of
the
Recorder by
the
Type
1617
. X-ramp and pen-
down
signals
for
an X-V Recorder are also sup-
plied via
this
socket. For
connections
see Fig .2.5.
Input
for
fil
te
r
shi
ft
pu
lses
n + 32
J Lov
+
32
V
to
se
lect
--t-t-
Au
tomatic
Stop
Not
u
sed
+
32
V DC
from
Record
er
lindi
ca
t
es)
Recor
de
r on)
+ 32 V
to
-,:-t---
stop Recorder
immediately
+ 32 V
to
se
lect
pe
n I
itt
77
0
774
Fig.2.4. RECORDER CONTROL I
viewed
from
outside
the
socket
12
X
ra
mp ou
tp
ut wh
en
no L
eve
l
Recorder
is
connected. W
it
h
L
evel
Recorder + 6 V
to
se
lect
1/3
oct
. interval between pul
se
s
Fi
lte
r
sh
i
ft
pu
lses
oV to
se
lect
Au
to
matic Stop
oV
to
stop
Recorder
imm
ediately
Not
used
+ 6 V from
Level Recorder
(in
dicat
es recorder
on)
oV f
or
pen d
ow
n
770
3
55
Fig. 2.5. RECORDER CONTROL /I
viewed
from
outside
the
socket

3.
OPERATION
3.1.
PRELIMINARY
ADJUSTMENTS
Before
connecting
a
mains
supply
the
following
checks and
adjustments
should
be
car-
ried
out
to
ensure
safe operation
of
the
apparatus.
3.
1.1.
Selection
of
Line
Voltage
The
instrument
can be operated on
single
phase AC
line
voltage
from
100
V
to
240
V,
with
a
tolerance
of
±
10%
,
at
frequencies
between
50
and
400
Hz.
Set
the
LINE VOLTAGE SELECTOR
to
the
voltage
to
be used by
turning
it
with
a
small
coin
or
a broad-bladed
screwdriver
.
3.1.2.
Fuse
Rating
For
operation
on
100,
115
and
127
V,
two
0,2
A
fuses
(VF
0012)
are
required, one in
each
holder
.
Operation
with
200,
220
and
240
V
requires
only
one
0,2
A fuse, and
when
these
voltages
are
selected as described
in
3.1.
1,
the
left-hand
fuse is
electrically
by-passed.
Make
sure
that
only
fuses
with
the
required rated
current
and
of
the
specified type are
used
for
replacement.
The use
of
mended
fuses
and
the
short-circuiting
of
fuse
holders
shall
be avoided.
3.1.3
.
Line
Connections
Blue Neutral
AC Power
'"
B
row
n Supply
Li
ve
[ •
I~
I J
Gr
ee
n and Yell
ow
Gr
ound
172246
Fig.3. T.
Line
input
connections
13

Once
the
line
voltage
setting
and
fuse
have been checked,
the
line
supply
may
be con-
nected
to
the
LINE
input
socket
using
the
Power
Cable
AN
0010
supplied. To
fit
a
suit-
able
mains
plug
to
the
cable see
Fig.3
.1.
For
maximum
operating
safety
it
is
recommended
that
the
protective (green and
yellow)
conductor
of
the
cable be
connected
to
a
suitable
ground,
such
as
the
protective
ground
contact
of
a
line
socket
outlet.
The use
of
an
extension
cable
without
a protective
ground
conductor
should
be avoided.
3.1.4.
Grounding
Considerations
When
several
instruments
are used
together,
ground
loops and
their
resulting
problems
can easily be
formed
.
Therefore,
the
following
procedure
is necessary
to
ensure
that
the
system is
free
from
ground
loops, by
connecting
each
instrument
in
the
arrangement
to
ground
via
the
same
path.
1.
Connect
the
signal
grounds
of
all
the
instruments
together
(this
automatically
occurs
through
the
screens
of
the
signal
cables).
However,
if
two
instruments
are
con-
nected in parallel,
it
may
be necessary
to
disconnect
one
of
the
screens
in
the
signal
cables
to
prevent
formation
of a loop.
2.
Connect
the
signal
ground
to
the
chassis, and
the
chassis
to
the
mains
ground
of
one
instrument
only
,
preferably
the
one closest
to
the
measuring
point.
3. Check each of
the
other
instru
ments
in
the
system
to
verify
that
its chassis is
con-
nected
to
one and
only
one
of
the
following
:
mains
ground,
si
gnal
ground
,
or
the
chassis
of
another
instrument
.
The signal
ground
and chassis
of
the
Type
1617
are
normally
connected
by
the
metal
strap
between
the
ground
terminals
on
the
rear panel. The strap
must
be removed
to
iso-
late
the
signal
ground
from
chassis
ground
.
3.2.
OPERATION WITH
MEASURING
AMPLIFIERS ALONE
14
The
Measuring
Amplifiers
Type
2606,
Type
2607,
and Type
2608
are
amplifiers
cover-
ing
the
frequency
range 2
Hz
to
200
kHz. They
are
designed
for
sound
and
vibration
measurement.
accepting
inputs
from
a
wide
range
of
transducers
and
preamplifiers
.
They
can
supply
12
volts
DC
for
preamplifier
power
supplies,
or
the
200
volt
polarization
voltage
and
150
volt anode
supply
for
B & K
condenser
microphones.
They
contain
selec-
table
frequency
-
weighted
networks
for
sound
level
measurement.
and
high
and
low
pass
filters
to
limit
the
frequency
range
for
audio
frequency
measurements
.
An
external
filter
such
as
the
Type
1617
can be
connected
in series
with,
or
in place
of,
the
internal
weighting
networks
and
filters
to
permit
frequency
analysis.
Such
an ar-
rangement
is
shown
in Fig.3.2.
Indication
of
the
signal level can be
obtained
either
from
the
built-in
meter
,
or
from
a
permanent
record
obtained
on a Level
or
X-V Recorder
con-
nected
to
the
measuring
amplifier
's
output.
see
section
3.3 .
For
further
information
on
measur
ing
amplifiers
to
help
in
deciding
the
best
for
a
particu-
lar
application
, a
more
comprehensive
review
of
the
capabilities
of
the
Measuring
Ampli-
fier
is
gi
~
en
in section 1. See also
the
relevant
Product
Data
sheets
.

3.
2.1.
Measuring Procedures
Input
Signal
Measuring Ampl
if
ie
r 2606,
2607, 2608 Band
Pa
ss
Filter 1617
770356
Fig.3,2.
Basic
instrumentation
arrangement
1. Connect
the
instrumentation
as
shown
in
Fig.3.2.
2. Perform
the
preliminary
adjustments
for
the
Measuring
Amplifier
as described in the
relevant
Instruction
Manual
.
3. Set
the
Measuring
Amplifier
controls
as
follows
:
INPUT
"Direct"
when
the
B & K coaxial
input
is used, or
"
Preamplifier
Input"
when
the
7-pin B &K socket
is used
METER FUNCTION
As
appropriate:
for
further
information
see the
Measuring
Amplifier
Instruction
Manual
GAIN CONTROL
"Calibrated"
INPUT ATIENUATOR
"300
V"
OUTPUT
ATIENUATOR
"x
1"
FILTERS "
External",
plus
any
weighted
network
required;
if
only
audio
frequencies
are
to
be measured,
the
high pass
22.4
Hz
filter
and the
low
pass
22.4
kHz
fil!er
(not on
2608)
can be used to
limit
the
frequency
range
of
the
Filter
output
signal.
REMOTE/
MANUAL
Averaging
.
Time
Switch
(rear panel
of
2607
only)
"Manual"
POWER
"On"
4. Perform
the
preliminary
adjustments
for
the
Band Pass Filter as described in section
3.1.
5. Set
the
Band Pass Filter controls
as
follows:
SELECTIVITY
RANGE
AVERAGING CONTROL
FILTER
CONTROL MODE
Bandwidth
(and stepping) as required
As
appropriate
Switch
positions
immaterial
Left-hand
switch
"Manual"
Right-
hand
switch
in
neutral
position
15
This manual suits for next models
1
Table of contents
Other BRUEL & KJAER Laboratory Equipment manuals
Popular Laboratory Equipment manuals by other brands

Agilent Technologies
Agilent Technologies TriScroll Exhaust Silencer Kit instruction manual

Idexx
Idexx Quanti-Tray 2X user manual

Analytik Jena
Analytik Jena InnuPure C96 user manual

Piuro
Piuro 900410 operating instructions

Laarmann
Laarmann LM2000 user manual

BANDELIN
BANDELIN SONOPLUS mini20 Instructions for use