Cardionics E-Scope II User manual

Operator’s Manual
E-Scope®II
Electronic Stethoscope
Models:
718-7700, 718-7710, 718-7712, 718-7740, 718-7750
US Patent No. 5,025,809
Manual P/N: 014-7700, Revision F
Date: June 2009
® E-Scope is a registered trademark of Cardionics, Inc., Webster, Texas USA
™
Specialist is a trademark of Cardionics, Inc.

Note: U.S. Federal law restricts this device to sale on the order
of a physician.
IMPORTANT NOTE for Hearing Impaired Users.
Please be aware that using the E-Scope with your hearing aid
for auscultation of the heart, lungs or Korotkoff (blood pressure)
sounds may be limited by the capability of your hearing aid or
the interface between your hearing aid and the E-Scope
Electronic Stethoscope.
Heart Sounds have a frequency range of 20-650Hz while
breath sound range form 70 to 2000Hz.
Hearing aids are designed for enhancement of voice sounds
that have a frequency range from approximately 500 to
4000Hz. Some hearing aids can be programmed for lower
frequencies. However, even with this additional low frequency
response, the hearing aid may still not be able to reproduce the
heart and lung sounds adequately for diagnosis. Also, some
individuals may have hearing loss in the range for heart and
lung sounds that is not compensated by their hearing aid. For
these reasons, Cardionics recommends that you consult with
your audiologist for advice covering your hearing loss and the
response of your hearing aid.
We will be pleased to assist you or your audiologist.
b

Quick Setup Instructions:
• Turning the E-Scope ON/OFF:
Press any button on the head of the E-Scope to turn it on.
There is not an “OFF” button on the unit. To turn the E-
Scope off, simply wait approximately 1.5 minutes and the
unit will automatically turn OFF. When restarted, the E-
Scope will return to the last volume automatically.
• Adjusting the Volume:
There are 64 steps between the highest and lowest volume
positions on the E-Scope II. The best way to adjust the
volume is to press and hold the plus (+) or minus (-)
buttons on the stethoscope head until the desired volume
is achieved.
• Switching between Heart and Breath Sounds.
To switch between Heart/Breath Sounds, press the “n”
key on the head of the E-Scope. The default mode is
Heart Sounds. By blowing on the stethoscope head as you
press the “n”, you can easily determine if you are in the
heart or breath sound frequency mode. Breath sounds will
sound higher.
• Changing the Battery in your E-Scope
Simply snap the “open” panel on the back of your unit.
Replace battery with a fresh “AAA.”
c

Safety Precautions
The following symbols are applicable to this
device:
IPX0 No degree of protection against ingress of
water.
Attention: Read and understand all warnings
and cautions before use.
Type BF Equipment. This equipment
provides a high degree of protection against
electric shock.
Caution:
The E-Scope should be used by qualified medical
personnel only.
Observe best practice. The instructions in this
manual do not, in any way, supersede established
medical procedures or staff preference concerning
patient care. "Best Practice" as determined by the
medical community is always to be observed.
Do not operate this device in the presence of
flammable anesthetics or other flammable gases.
As a general safety precaution, this instrument should
be turned off and battery disconnected before
performing any cleaning procedures.
Use only AAA alkaline batteries. Follow all warnings
and instructions on the battery package including
proper disposal of depleted batteries.
d

Cautions continued:
At the end of the E-Scope useful life dispose or
recycle in accordance with your local, state and
governmental regulations.
The E-Scope complies with electromagnetic
compatibility (EMC) EN60601-1-2:2001. However,
close proximity to a radio transmitter may result in
interference heard through the E-Scope Binaural or
headset. If this occurs, move away from the radio’s
transmitting antenna.
Do not immerse any part of the E-Scope in any liquid
or subject it to any sterilization processes.
Important note for hearing aid users: Hearing
deficiencies vary considerably from person to person.
Your audiologist is the best source of information on
your particular hearing problem. You should consult
with your audiologist if you have any specific
questions about the use of your hearing aid with the
E-Scope Electronic Stethoscope or any similar
device. The information supplied by Cardionics for
hearing impaired users should be considered as a
general guideline only.
The E-Scope has significant volume output. Do not
turn on higher than required for normal listening.
No serviceable parts. Do not attempt to repair or
service the E-Scope. Return to Cardionics for any
and all repairs. Call 281-488-5901 for service
directions.
e

Contents
Page No.
1.0 Product Description ................... ......................................1
1.1 Models and Accessories ...................... ...........4
2.0 Using the E-Scope................. ..........................................8
2.1 general clinical use.................. ........................8
2.2 for teaching.......................................................8
2.3 recording....................................... ...................8
2.4 in telemedicine .............. ..................................9
2.5 other applications ................ ............................9
3.0 Operation..........................................................................9
3.1 Volume Adjust.............................................. .....9
3.2 Filter Select ......................................................9
3.3 Power ..................... ..........................................9
3.4 Battery……………………………………………..10
3.5 Sounds Output…………………………………… 10
4.0 Cleaning ................................ ..........................................12
5.0 Transport and Storage........... ..........................................12
6.0 Troubleshooting guide ..................... ................................12
7.0 Warranty .......... ................................................................13
8.0 Appendix................................................................. .........15
8.1 Recording Techniques ........................... .........15
9.0 Index....................................................................... .........17
List of Figures
Figure 1 E-Scope clinical model...................... ......................2
Figure 2 E-Scope belt model ....................... ..........................2
Figure 3 Wearing the E-Scope ........................ ......................3
Figure 4 Using the E-Scope ...................... ............................3
Figure 5 Headphone, Over the head style. ............................5
Figure 6 Headphone, Behind the head style ..........................5
Figure 7 Specialist Diaphragm .........................................….. 7
Figure 8 Specialist Bell…………………………………………..7
Figure 9 Specialist Pediatric Bell………………..…………...... 7
Figure 10 Accessory Pack……………………………………… 7
Figure 11 E-Scope Stethoscope head...................………….. 9
Figure 12 E-Scope Controls……..........................……………11
f

1.0 Product Description
The E-SCOPE II is a second generation, patented, CE certified,
electronic stethoscope. It allows the user to amplify or record heart
and breath sounds. The E-Scope II is an electronic stethoscope
featuring a slim-line design with the volume and filter controls on the
head of the stethoscope. The slim design allows it to be draped
comfortably around the neck. With the controls on the stethoscope
head, the user will find the E-Scope easy and convenient to use. The
sound quality is exceptional with 64 gain positions.
There are many reasons for using an electronic stethoscope rather
than an acoustic stethoscope:
• The ability to hear soft sounds in a noisy environment,
• The ability to record, edit and play back any sound,
• The ability to send sounds over telephone lines for diagnosis
by a specialist,
• The ability to send sounds by email for diagnosis by a
cardiologist,
• The ability to compare a patient with him or herself at
differing time intervals,
• The ability to filter the sounds so that only certain portions of
The sound complex can be heard,
• The ability to teach students by having more that one
simultaneous listener with a single patient,
The E-Scope is supplied with a Specialist Adult Diaphragm that may
be changed to a Specialist Adult or Specialist Pediatric bell. . Also
supplied are plastic adult, infant and pediatric size bells and a
selection of ear tips for maximum user comfort.
The E-Scope is equipped with a timer that will turn the instrument off
after approximately 1.5 to 2 minutes to conserve battery life.
The E-Scope has an auxiliary output for connecting to a second
listener or recording type device.
1

The E-Scope is designed to maximize hearing of heart and lung
sounds. A sound selector switch is provided to allow listening in the
proper frequency response for either heart sounds or breath sounds.
There are several models of the E-Scope II. The clinical E-Scope
(cat. no. 718-7740, 718-7700) and the Belt model E-Scope for hearing
impaired users (cat. no. 718-7750, 718-7710).
Figure 1
E-Scope II electronic stethoscope
Clinical model
Figure 2
E-Scope belt model for hearing impaired users
The Belt model E-Scope is designed for use with an ITE (in-the-ear)
type hearing aid. With the Belt model E-Scope, the normal earpieces
have been removed. A pair of headphones are then inserted into the
upper output of the E-Scope and placed over the hearing aids. Figure
2 shows the detail of the Belt model E-Scope and Figure 4 shows the
E-Scope in use.
2

Figure 3 Wearing the E-Scope
Figure 4 – Using the E-Scope
3

1.1 Models and Accessories
Cat. no. Description
Clinical Models
718-7700 E-Scope Electronic Stethoscope
Clinical model with Specialist Adult Diaphragm
718-7740 E-Scope Electronic Stethoscope with Specialist
Adult Diaphragm
and Specialist Adult Bell
Pediatric Model
718-7745 E-Scope Electronic Stethoscope (same as cat.
no. 718-7740 above except also supplied with
Specialist Pediatric Bell.
Ambulance Model for High Noise Areas
718-7800 E-Scope EMS Electronic Stethoscope
For use in an ambulance or high noise area.
A separate Operator’s Manual is provided for the
E-Scope EMS
Hearing Impaired Models ITE (in-the-ear) or “open fit” BTE
718-7710 E-Scope Hearing Impaired (Belt) model.
Supplied with Specialist Adult Diaphragm. This is
the same as the E-Scope, cat. no. 718-7700, except
that the ear pieces have been removed and a belt
clip added. This model is designed for use with
headphones (cat. no. 718-0405 or 718-0415). A
headphone is included.
718-7750 E-Scope Belt model. This is the same as above
cat. no. 718-7710 except it is supplied with both
Specialist Adult Diaphragm and Specialist Adult Bell.
4

Accessories
Cat. no. Description
718-0405 Over the head style headphone
These headphones are designed for use
with the E-Scope Belt model, cat. no. 718-7710 or
cat. no 718-7750.
Figure 5. Over the head style headphone.
718-0415 Behind the head style headphone
These headphones are designed for use with E-
Scope model 718-7710 and 718-7750. This style
has a single cord and behind the head style fit.
Figure 6. Behind the head style headphone.
718-0410 Stethophone
The stethophone is designed for teaching
to allow a second listener to hear simultaneously.
5

360-7740 Specialist™Adult Diaphragm
The performance of the Specialist Adult Diaphragm
is, on average, about 3 dB or 40% better than a
stamped metal diaphragm.
360-7750 Specialist Adult Bell
Like the above Specialist Adult Diaphragm, the
Specialist Adult Bell is machined from high-grade
stainless steel. Its performance is superior allowing
the listener to use the split finger technique by
placing the bell between the index and middle finger
thereby permitted gentle, controlled pressure to be
applied.
360-7755 Specialist Pediatric Bell.
The Pediatric Specialist Bell output is, on average,
about 60% higher than a stamped metal pediatric
bell over 50-200 Hz frequency range for both heart
and breath sound filter settings.
6

Accessory Pack. Each E-Scope is supplied with an Accessory pack
that contains a pediatric diaphragm, plastic adult bell, plastic pediatric
bell and plastic infant bell. These chest pieces can be used by
replacing the existing chest piece by unscrewing. Additionally, this
pack contains both mushroom and hard plastic ear tips. Inserting he
mushroom ear tips will require a little force. The hard plastic eartips
are self-threading. Place this eartip on the stethoscope binaural,
press and with a twisting motion, screw them on.
Figure 7 Figure 8
Specialist Adult Diaphragm Specialist Adult Bell
Figure 9 Figure 10
Specialist Pediatric Bell E-Scope Accessory kit
Supplied with all E-Scopes
Upper row left – adult plastic diaphragm, upper middle infant bell,
upper right, mushroom eartips;
Lower row – Pediatric diaphragm, pediatric plastic bell, hard plastic
eartips.
7

Cable Connections
Cat. no. Description
711-9000 Interface Cable. Connects to a PC or recording device that
accepts an 1/8th inch phone plug.
2.0 Using the E-Scope
2.1 To use the E-Scope, place the stethoscope on the patient in the
normal manner as shown in Figure 4. Press any button on the head
of the stethoscope to turn it ON. Press and hold the + button until the
volume is at a desired level. The filter is automatically set for heart
sounds. If you wish to change it to breath sounds, press the ●button
on the stethoscope head (Note: If the ●button is pressed to activate
the E-Scope then the filter will be in the breath mode).
The E-Scope will turn OFF automatically after 1.5 to 2 minutes. To
restart, press any button on the stethoscope head. The E-Scope
retains the last volume setting and reverts to that setting when
restarted.
2.2 For Teaching
For teaching purposes, it is possible to have a second person listen to
a patient simultaneously. Connect the stethophone (cat. no. 718-
0500) to the “sounds output” of the E-Scope. The student will hear
what the instructor hears.
2.3 Recording Sounds
We suggest you also review Appendix A in this manual for recording
techniques.
To record you will need a connecting cable (cat. no. 711-7128).
8

2.4 For Telemedicine Use
The E-Scope may be used for sending heart and/or lung sounds over
telephone lines or the internet using a recognized telemedicine system
such as PolyCom or Tandberg systems.
2.5 Other Applications
a. The E-Scope can be used for general clinical
use in place of an ordinary acoustic stethoscope.
b. Physicians or nurses with hearing impairment.
c. Ambulances
d. Emergency rooms
e. Pediatric wards
3.0 Operation
3.1 Volume Adjust. The E-Scope volume may be increased or
decreased by momentarily pressing the + or – button located on
the stethoscope head. Pressing and holding one of these buttons
for more than one second causes the volume to move up or
down. It takes about seven seconds for this automatic
adjustment to cover the full volume range of 64 steps. The E-
Scope retains the last volume setting and reverts to that setting
when the power is turned on.
Figure 11
E-Scope Stethoscope head
3.2 Filter Select. The ●button on the stethoscope head
alternately selects the heart or breath sound filter each time it is
momentarily pressed. Pressing the ●button to power on the E-
Scope initially selects the breath sound filter. Conversely,
pressing either the + or – button to start power initially selects the
heart sound filter.
3.3 Power. Pressing any of the buttons on the stethoscope head
turns ON power to the E-Scope. The E-Scope has a timer that
9

shuts off power after approximately 1.5 to 2 minutes. The timer is
restarted each time one of the switches is pressed.
3.4 Battery
A single AAA battery powers the E-Scope. This battery will operate
the E-Scope for approximately 4.5 months assuming it is used 30
times per day, 5 days per week.
To change the battery, remove the battery cover by sliding it outward.
Remove the old battery and install the new in the normal manner.
3.5 Sounds Output. This connector, located on the body of the E-
Scope, allows accessory units to be attached to the E-Scope. See
Figure 12.
4.0 Cleaning
The E-Scope can be cleaned using a standard alcohol swab or wipe
(70% Isopropyl Alcohol). Do not use excessive liquid or immerse any
part of the E-Scope. Doing so may result in moisture getting inside
the unit.
5.0 Transport and Storage
The E-Scope can be safely stored or transported with the following
environmental conditions: -20°C – 65°C (-4°F – 149°F), <95%
relative humidity.
10

Figure 12 - E-Scope Controls
-
S
C
O
P
E
E
+
-
SOUNDS
OUT
Electronic
Stethoscope
E-SCOPER
Cardionics
Output for either second
listener or accessory.
Sets the ideal frequency for
heart or lung sounds.
Pressing the dot while the E-
Scope is off will activate the
unit and adjust for lung
sounds. Press the dot while
the E-Scope is active and it will
toggle between lung and heart
sound modes.
Increase or decreases the
volume. Press either button
while the unit is off and it will
power on in the heart sound
mode by default and remain on
for 1.5 minutes. The timer
resets itself after every press
of a button. When the E-Scope
restarts, it returns to the last
volume setting.
Output Jack for Cardionics Headphone,
Model 718-0405 and 718-0415.
11

6.0 Troubleshooting Guide
Possible Problem Cause / Solution(s)
No power The E-Scope will turn off automatically
after 1.5 to 2 minutes of use. Press any
button on the stethoscope head to restart.
If this does not restart the E-Scope, check
the battery and replace if necessary. If the
E-Scope still does not restart, please call
Cardionics for assistance.
Not enough volume Press and hold the + button on the head of
the E-Scope. If using a hearing aid, the
volume should be increased gradually
while tapping the diaphragm lightly. Be
sure that diaphragm is securely tightened
on the threaded housing.
Clicking noise or static Speaker driver is clipping. (1) reduce
volume until clicking sound disappears. (2)
battery may be depleted. Replace, if
necessary (3) Use the bell of the
stethoscope. (4) Using the – button on the
stethoscope head, press and hold for 7
seconds. This takes the volume to zero.
Then, press and hold the + button for 2-3
seconds. There are 64 gain positions on
the E-Scope volume control. The above
maneuver has adjusted the E-Scope to
mid-range. Some additional adjustment
may be necessary. For static, place the
chest piece lightly on the chest. There
may be an initial static sound that will
quickly disappear.
12

7.0 Limited Warranty
What this warranty covers:
1. Cardionics, Inc. warrants this product to be free of
manufacturing defects in material and workmanship for a
period of one (1) year from the date of original consumer
purchase from Cardionics or an authorized dealer.
2. This written warranty is limited to the original consumer
purchaser, transferable only by written authorization of
Cardionics, Inc.
3. All warranties, expressed or implied, made by Cardionics,
including warranties of merchantability and fitness are limited
to the one (1) year period of this warranty. Some states do
not allow limitations on how long an implied warranty lasts,
so the above limitations may not apply to you.
4. This warranty is limited to repair of the product or
replacement thereof, at the discretion of Cardionics, Inc.
Calibration is considered normal maintenance and are not
included in the one year warranty.
What this warranty does not cover:
1. This warranty does not cover defects or damage resulting
from use of the product other than its normal, intended and
customary use. This warranty does not cover defects or
damages from abnormal use, abnormal conditions, improper
storage, exposure to moisture or liquid, unauthorized
modifications, repairs made by unauthorized personnel,
unauthorized connections (those not described in this
manual), misuse, neglect, abuse, accident, alternation,
improper installation or other acts which are not the fault of
Cardionics, including damage caused by shipping.
2. E-Scopes which have had the serial number removed or
made illegible.
3. Damage resulting from use of non-Cardionics approved
accessories.
13

Policies concerning service.
Shipping charges. The customer shall bear the cost of shipping the
product to Cardionics. We suggest that you insure the package
Authorization for return. Before sending equipment for repair, please
call Cardionics, at 281-488-5901 or in the U.S. or Canada 1-800-364-
5901 for a return memo authorization number (RMA#). This will help
us track your repair.
Reason for return. Please include a note with your name, address,
telephone and email address and what you believe the problem to be.
Packing for return. Equipment requiring repair should be suitably
packaged for shipping.
Where to obtain service. Ship package prepaid to Cardionics, Inc.,
910 Bay Star Blvd, Webster, Texas 77598 USA. We will endeavor to
complete repairs within fifteen (15) working days from date of receipt
of product by Cardionics.
Out-of-warranty service. . When a product is returned for service that
is out-of-warranty, Cardionics will call you to obtain your authorization
prior to making the repair. We will ask for a credit card payment at the
time of the call. If the repair is from an authorized Cardionics dealer or
institution, Cardionics will ask for a purchase order number.
Consequential Damages
CARDIONICS, INC. SHALL NOT BE LIABLE FOR INCIDENTAL OR
CONSEQUENTIAL DAMAGES ARISING FROM THE USE OR FAILURE OF
THIS PRODUCT, INCLUDING INJURY TO PERSONS OR PROPERTY.
Some states do not allow the exclusion or limitation of incidental or
consequential damages, so the above limitations or exclusions may
not apply to you. This warranty gives you specific legal rights, and
you may also have other rights that may vary from state to state.
14
This manual suits for next models
5
Table of contents
Other Cardionics Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Sizewise
Sizewise Platinum 6000 Quick reference guide

Onpharma
Onpharma Onset Mixing Pen reference guide

Itamar Medical
Itamar Medical WatchPAT 200 Unified Operation manual

ADC
ADC Satin Mac 0 Instructions for use

Bard
Bard QuickLink 70310QCA1 quick start guide

Stiegelmeyer
Stiegelmeyer Deka instruction manual