Cellink BIO X 10-mL User manual

BIO X 10-mL Pneumac Printhead
User Manual


Contents
1.Package contents 1
2.Technical specicaons 3
3.Safety informaon 9
3.1 BIO X system warnings 10
3.2 Pneumac Printhead warnings 10
4.Geng started 11
4.1 Unpacking and installaon 12
4.2 Raising and lowering the printhead 13
4.3 Loading the bioink cartridge 14
4.4 Your rst bioprint 16
4.5 Removing the printhead 23
4.6 Prinng opmizaon 25
5.Relevant G-code commands 27
6.Frequently asked quesons 31
7.Maintenance 35
Support informaon 37


1
Package contents
01

2
01 Package contents
Item Part number Quanty
BIO X 10 mL Pneumac Printhead 000000020356 1
Air adapter connector 10 mL 000000610041 1

3
Technical specicaons
02

4
02 Technical specicaons
• Compable with 10-mL plasc cartridge.
• Dimensions (height x width x depth):
• With screw: 63.5 x 30.5 x 51 mm.
• Without screw: 63.5 x 30.5 x 46.2 mm.
• Weight: 102 g.
• Maximum recommended pressure: 700 kPa.
• Material composion:
• External surfaces: Powder-coated aluminum.
• Internal surfaces: Aluminum.
• Recommended operang temperature: 20-23 degrees Celsius.
NOTE: We recommend operang the BIO X in temperatures between 20 and 23 degrees. In normal operang condions,
the chamber temperature of the BIO X will rise 2-3 degrees above room temperature, and the 10-mL Pneumac Printhead
temperature will rise above the chamber temperature.

5
02 Technical specicaons
• Running temperature (in recommended condions): 26-32 degrees.
• Heang capacity and rate:
• Total heang range: 30-65 degrees.
• Heang rate: Room temperature-40 degrees (10 minutes).
• Heang rate: 40-65 degrees (12 minutes).
• Cooling capacity: N/A.
• Build volume (X, Y, Z): 128 x 85 x (83.4–A) mm; A represents p length.

6
02 Technical specicaons
Figure 1:

7
02 Technical specicaons
Table 1:
Tip Type Tip Length Nozzle/Needle Length X Y Z Build Volume
Conical
Blunt
Blunt
Blunt
Micron-S

8

9
Safety informaon
03

10
03 Safety informaon
3.1 BIO X system warnings
• Please consult the BIO X manual for BIO X-specic and general warnings and safety procedures.
3.2 10-mL Pneumac Printhead warnings
• Do not clean the printhead by submerging it in liquid or using excessive spraying. Liquid inside the printhead can
damage the circuitry and motor.
• The LED on the printhead will turn yellow to indicate a warm printhead and red to indicate a hot printhead. Do not
touch the printhead when the LED is red. Touching a hot printhead risks serious injury.
• Do not manually move the printhead mount. Moving the printhead mount manually will damage the motor. If the
mount needs to be moved, place the printhead in the loading posion described in the Geng Started secon.
• Use the proper method to load and unload the 10-mL Pneumac Printhead onto the BIO X printhead mounts.
Instrucons are outlined in the Geng Started secon.
• Do not use a jerking moon to remove the printhead. Using a jerking moon risks hing and damaging the HEPA
lter.
• Tighten the locking screw before prinng to ensure that the cartridge remains in place during the prinng process.
Improper ghtening can cause the cartridge to move and material to dispense improperly.
• Do not touch a moving printhead. Fingers can be crushed between the printhead base and motor arm and can result
in serious injury.

11
Geng started
04

12
04 Geng started
4.1 Unpacking and installaon
NOTE: The prinng parameters displayed on this manual might not be accurate for your specic protocol. For
recommended parameters please consult the documentaon for the bioink being used.
1. Open the package. Remove the 10-mL Pneumac Printhead.
2. Aach the 10-mL Pneumac Printhead to the BIO X printbox. Align the printhead above the desired printhead mount (Figure 2A) and push
downward using one hand. Secure the boom of the prinhead mount with your opposite hand to avoid straining the motor (Figure 2B,C,D).
ABC D
Figure 2:

13
04 Geng started
4.2 Raising and lowering the printhead
1. On the user interface, navigate to the Ulies menu and then go to the Tools submenu.
2. Lower the desired printhead mount into the acve or loading posion by pressing the down arrow (Figure 3).
3. Press the up arrow to return the printhead to the inacve posion.
Figure 3:

14
04 Geng started
4.3 Loading the bioink cartridge
1. Make sure that the printhead is installed and in the acve or loading posion (Secon 4.1 and 4.2)
(Figure 2).
2. Fill a cartridge with bioink or obtain a prelled cartridge. Remove the boom cap if using a prelled
cartridge (Figure 4A). Aach a nozzle (Figure 4B) and remove the end cap (Figure 4C).
3. Align the cartridge adapter to the loaded cartridge (Figure 4D). Twist unl secure (Figure 4E).
ABC
Figure 4:
D E

15
04 Geng started
4. Load the cartridge into the printhead from the top (Figure 5B). Tighten the locking screw. To make this step easier, we
recommend placing the printhead in the acve posion (lowered) as explained in the Raising and lowering the printhead secon
(page 13) (Figure 5A).
5. Connect the air tubing to the air inlet on the printbox above the respecve printhead (Figure 5C).
6. Raise the printhead to the inacve posion (Secon 4.2).
Figure 5:
ABC

16
04 Geng started
4.4 Your rst bioprint
1. Prepare the printhead and load the bioink as detailed in Secon 4.3.
2. Select Bioprint from the Start menu (Figure 6).
Figure 6:.
This manual suits for next models
1
Table of contents
Other Cellink Printer Accessories manuals
Popular Printer Accessories manuals by other brands

Pitney Bowes
Pitney Bowes SendPro Desktop SendKit Installation instruction

Epson
Epson T033120 Product information sheet

Epson
Epson C823642 - Enet Print Server Type B Multi Protocol... Administrator's manual

Star Micronics
Star Micronics TSP700 Series Attachment instructions

VASTEX
VASTEX E-200 series Assembly and operation instructions

TonerRefillKits
TonerRefillKits brother ReChargX RX350 instructions