  9. Chamber BHC-NEB-P User manual

Chamber BHC-NEB-P User manual

Warranty
e-chamber Portable
Mesh Nebuliser
Product Code: BHC-NEB-P
Date of purchase*
Customer name*
Contact number*
Where Purchased
Presentation of this card will be required when a warranty service
is requested. A new card will not be issued if this card is lost
Please keep the packaging of this product so that it can be
used to return the device when maintenance is required.
The warranty period of the body of the device is 12 months.
The warranty peroid of the medication cup is 3 months.
The masks and mouthpiece are not covered under warranty.
Warranty claims will be reviewed on a case by case basis
and we will replace or repair parts if deemed necessary.
For all warranty claims, contact Bird Healthcare directly
at
sales@birdhealthcare.com or call us at TOLL FREE 1300 365 561
with the following information and a copy of your receipt/invoice
Damage by human behaviours: touching the mesh &
soaking/wetting the device
Damage caused by not correctly operating the
product according to the Quick Start Guide
Damage caused by accident
User disassembly without authorisation
Bird Healthcare Pty Ltd
18 Corporate Blvd
Bayswater VIC Australia 3153
Submit to sales@birdhealthcare.com
or call us at TOLL FREE 1300 365 561
Quick Start Guide
active
mesh
technology
Proudly
Supporting
Portable Nebuliser
with travel case,
masks & mouthpiece
www.birdhealthcare.com
Issues not covered by warranty include:
READ BEFORE USE
Safety notes
Warning
Only use parts and accessories provided. Parts and acces-
sories from unauthorised sellers may cause damage and
will be not be covered under warranty.
THIS INCLUDES the charging cable (output 5.0V, 1.0A).
Only abide by the troubleshooting guide. If the device
needs servicing beyond the troubleshooting guide, stop
using and contact Bird Healthcare immediately.
DO NOT attempt to service yourself unless specified.
DO NOT fill the medication cup beyond 10ml –
this can cause electric shock and/or damage to the device.
DO NOT use without liquid.
DO NOTusewiththick/sticky(viscous) liquid as this can block
the mesh. ONLY use with Salbutamol and saline solution.
Avoid getting the device wet – DO NOT expose the device
to running water or submersion.
DO NOT use the device in hot environments (>40°C).
DO NOT use the device near flammable gas or near
oxygen and anaesthetic mixtures.
DO NOT expose the device to corrosive liquids and gases.
DO NOT wrap the power cord around the device.
DO NOT subject the device and accessories to strong
vibrations and impacts.
DO NOT touch the centre of the spray mesh by hand or
with any other objects.
For the type, dose, and regime of your medication,
be sure to follow the instructions from your doctor.
For children and persons with special needs, use
with supervision. Use the provided mask if required.
This device is to be used for the specified purpose
of medication nebulisation. DO NOT use the device
for any other purpose.
Clean and disinfect the medication cup and
accessories before using for the first time.
Keep the medication cup empty
and dry when not in use.
This device is intended for single patient use to
avoid cross infection.
For the type, dose, and regime of your medication,
be sure to follow the instructions from your doctor.
For children and persons with special needs, use
with supervision. Use the provided mask if required.
This device is to be used for the specified purpose
of medication nebulisation. DO NOT use the device
for any other purpose.
Clean and disinfect the medication cup and
accessories before using for the first time.
Keep the medication cup empty
and dry when not in use.
This device is intended for single patient use to
avoid cross infection.
Not recommended for people with:
Pulmonary Embolism
Pulmonary Infarction or Odema
If using for acute asthma and you feel you are not
improving, seek urgent medical advice.
DO NOT use without liquid
ONLY use with Salbutamol and/or Saline solution
Replace Medication Cup every 3-6 months
Trouble Cause
Doesn’t
turn on
and
light
isn’t on
Atom-
isation
is too
small
Device isn’t on
There isn’t enough
medication
There is a
strange
noise
There is a fault in
the product
The spray mesh
is blocked
The wrong medication
is being used
Button may
be jammed
This is normal
USB Charging
Cable may
be broken
Device isn’t charging
See Warranty or use a cable with
the same voltage and amp
readings (this may void
your warranty)
Solution
Try turning on the device and
ensure the light comes on
Refill no further than the 10ml line
See Warranty
Place 3 drops of white vinegar and
3-6ml of water inside the
medication cup, then fully atomise
(DO NOT INHALE), then clean the
medication cup
See Warnings
Try pressing ON/OFF switch to
dislodge
See Charging
There is
medication
residual
e-chamber Portable Nebuliser
a
e
g
d
c
f
b
Instructions For Use
Use is for liquid medication nebulisation only
Clean, disinfect and dry all of the parts (See Cleaning).
Charge the device (See Charging).
First use
1.
2.
Charge the device using the provided USB charging
cable – if using a standard adapter, make sure it
complies with IEC 60601-1, 5.0V.
The device is usable whilst charging.
The device takes approximately 1 hour to fully charge.
When the device is on low battery it will flash slowly.
Please charge the device at least once per month,
even when not in use.
Charging
Remove the lid
on the medication cup.
Close the lid.
Click the green
ON/OFF switch to start
(pull the slider down if
concealing the switch).
1
Fill to below the max
10ml capacity line.
2
3
5
Hold the nebuliser vertical
to keep the liquid in
contact with mesh.
6
Attach the mouth piece
or mask if desired.
4
Take slow deep breathes to inhale the medication.
7
After 10 minutes, the device will automatically shut
o. If longer use is desired, simply switch the device
on again ensuring there is enough liquid medication.
8
NB: When finished, nebulise with 3ml of warm tap
water. Once a week, nebuliser with 3ml of warm tap
water and 1ml of vinegar. Press ON/OFF switch to turn o.
9
Detach the medication cup by pushing the ‘press’
button behind the device and pull the medication
cup upwards. Dispose of any medication.
10
Clean any accessories that were used during
nebulisation (See Cleaning ).
11
1.
2.
3.
Remove the medication cup, lid, mouthpiece or
mask from the device.
Submerge and hand wash in luke warm water
(no more than 50˚C) for about 5 minutes.
Set aside to dry in a clean place
and if required then disinfect (See Disinfect ).
Cleaning Troubleshooting
The temperature is between -2 0˚C and 55˚C.
The relative humidity is less than 80%.
Atmospheric pressure is between 70 and 106 kPa.
The device is not in storage for 5+ years unused.
Storage: Store the device in the provided carry case
ensuring the following:
Contents
a. Lid x 1
b. Device x 1
c. Medication cup x 1
d. Mouthpiece x 1
e. Mask x 2
f. Travel case x 1
g. USB charging cable x 1
Place cleaned medication cup, lid and mouthpiece in
disinfectant solution (such as a Milton per instructions).
Or steam sterilise in a standard baby steam device as
per device instructions.
Air dry completely after disinfecting
Disinfect (if required)
Product Code: BHC-NEB-P

Popular Medical Equipment manuals by other brands

WY'EAST MEDICAL TOTALIFT Instructions for use

WY'EAST MEDICAL

WY'EAST MEDICAL TOTALIFT Instructions for use

College Park Capital hydraulic knee user guide

College Park

College Park Capital hydraulic knee user guide

Nonin 8500 Operator's manual

Nonin

Nonin 8500 Operator's manual

Nonin 3230 Instructions for use

Nonin

Nonin 3230 Instructions for use

bort medical EpiContur Sport Instructions for use

bort medical

bort medical EpiContur Sport Instructions for use

biosignalsplux SENSPRO-EOG user manual

biosignalsplux

biosignalsplux SENSPRO-EOG user manual

Otto Bock 17B66 Series Instructions for use

Otto Bock

Otto Bock 17B66 Series Instructions for use

Invacare Rectangular Pan Holder and Pan 9610-1 Installation

Invacare

Invacare Rectangular Pan Holder and Pan 9610-1 Installation

Molnlycke Z-flex Heel boot manual

Molnlycke

Molnlycke Z-flex Heel boot manual

Siemens MAMMOMAT 300 Repair instructions

Siemens

Siemens MAMMOMAT 300 Repair instructions

Flaem Wi.Neb P0209EM Instructions for use

Flaem

Flaem Wi.Neb P0209EM Instructions for use

Limbs & Things 60100 user guide

Limbs & Things

Limbs & Things 60100 user guide

Sharper Image 206604 user guide

Sharper Image

Sharper Image 206604 user guide

Manamed MANASPORT instructions

Manamed

Manamed MANASPORT instructions

Gima 27841 manual

Gima

Gima 27841 manual

ZAMST FILMISTA WRIST Instructions for use

ZAMST

ZAMST FILMISTA WRIST Instructions for use

BEYES DuraMax X1 Instructions for use

BEYES

BEYES DuraMax X1 Instructions for use

Heraeus PALAMIX Instructions for use

Heraeus

Heraeus PALAMIX Instructions for use

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.