Charder MA601 User manual

Please keep the instruction manual at hand all the time for future reference.
Body Composition Analyzer
USER MANUAL
MA601

1
Explanation of the graphic symbols
Charder Electronic Co., Ltd.
No.103, Guozhong Rd., Dali Dist.,
Taichung City, 412 Taiwan
SN-T13000001 Designation of the serial number of every device,
applied at the device.(Number as an example)
“Please note the accompanying documents“
or “Observe operating instructions”
Identification of manufacturer of medical product
including address
“Electro-medical appliance“ with attachment for type BF
Dispose of old appliances separately from your
household waste !
Instead, take them to designated collection points for
electronics.
Carefully read this operation manual before setup and
commissioning, even if you are already familiar with
Charder scales.
Manufacturing date
Lot number or batch number

2
During measurement, this machine will send a low level imperceptible electrical current through-
out the body. Individuals with implanted medical devices, such as:
1. Pacemakers
2. Electronic lungs and other electronic medical life support equipment
3. ECG devices
must not use this machine, as the electric current may affect the implanted device, endangering
lives.
Warning: To avoid electric shock, this device should be plugged into a grounded electrical outlet
Copyright Notice
Charder Electronic Co., Ltd.
No.103, Guozhong Rd., Dali Dist.,
Taichung City 41262 Taiwan
Tel: +886-4-2406 3766
Fax: +886-4-2406 5612
Website: www.chardermedical.com
E-mail: [email protected]
Copyright© Charder Electronic Co., Ltd. All rights reserved.
This user manual is protected by international copyright law. All content is licensed, and usage
is subject to written authorization from Charder Electronic Co., Ltd. (hereinafter Charder)
Charder is not liable for any damage caused by a failure to adhere to requirements stated in this
manual. Charder reserves the right to correct misprints in the manual without prior notice, and
modify the exterior of the device for quality purposes without customer consent.
Warning
After the MA601 has been turned on, the screen will remain dark for about 10 seconds.
This is normal, and the device will continue with self-calibration process.
NOTE

3
CONTENTS
4
4
6
7
11
12
12
13
14
17
20
20
21
23
25
25
27
28
29
35
35
36
45
55
55
55
56
58
59
62
I. SAFETY NOTES
A. General Information
B. Precaution Symbols
EMC guidance and manufacturer's declaration
II. INTRODUCTION TO THE MA601 BODY COMPOSITION ANALYZER
III. INSTALLATION
A. Product Parts and Assembly
B. Environment
C. Installation Instructions
IV. EXTERIOR AND PANEL DEFINITION
V. GETTING STARTED
A. Power Supply
B. Start Screens
VI. INSTRUCTIONS FOR OPERATION
VII. MEASURING INSTRUCTIONS
A. Measuring Posture
B. Proper Measurement Posture (feet)
C. Proper Measurement Posture (hands)
D. Measuring Procedure
VIII. ABOUT RESULTS
A. Result Sheet
B. Result Sheet Explanation
IX. SYSTEM SETTINGS
X. PRINTING
A. Printer Compatibility
B. Connecting Printer
C. Configure Printer Settings in the device
XI. TROUBLESHOOTING
XII. FREQUENTLY ASKED QUESTIONS(FAQs)
XIII. PRODUCT SPECIFICATIONS

A.General Information
Thank you for choosing this Charder Medical device. It is designed to be easy and
straightforward to operate, but if you encounter any problems not addressed in this
manual, please contact your local Charder service partner.
Before beginning operation of the device, please read this user manual carefully, and
keep it in a safe place for reference. It contains important instructions regarding installa-
tion, proper usage, and maintenance.
General Handling
■This device is intended for indoor use only.
■Do not place the device on slippery surfaces.
■Ensure all parts are properly locked and tightened before operating the device.
Electric Shock
■Do not touch the power supply with wet hands.
■Do not crimp the power cable, and avoid sharp edges.
■Do not overload extension cables connected to the device.
■Route the network and power cable carefully, to avoid tripping.
■Keep the device away from liquids.
Injuries and Infections
■Ensure that subjects do not have wounds or contagious diseases on the palms of their
hands or the soles of their feet.
■For hygiene purposes, Charder recommends cleaning the measuring platform after
each measurement with a soft cloth and alcohol.
■Ensure that the measuring platform is dry before usage.
4
Preventing Device Damage
■Please contact your local Charder distributor for regular maintenance and calibration.
■This device does not contain any user-maintained parts. All maintenance, technical
inspections, and repairs should be conducted by an authorized Charder service
partner, using original Charder accessories and spare parts. Charder is not liable for
any damages arising from improper maintenance or usage. Dismantlement of the
device will void the warranty.
■Take care to make sure fluids do not enter the device, as they may damage the
internal electronics.
■Switch off the device before disconnecting the power supply.
■Do not place the device in direct sunlight, or in close proximity to an intense heat
source. Excessively high temperatures may damage the internal electronics.
■Strong cleaning agents can damage the measuring platform's surface.
I. SAFETY NOTES
Caution
Caution

A.General Information
Thank you for choosing this Charder Medical device. It is designed to be easy and
straightforward to operate, but if you encounter any problems not addressed in this
manual, please contact your local Charder service partner.
Before beginning operation of the device, please read this user manual carefully, and
keep it in a safe place for reference. It contains important instructions regarding installa-
tion, proper usage, and maintenance.
General Handling
■This device is intended for indoor use only.
■Do not place the device on slippery surfaces.
■Ensure all parts are properly locked and tightened before operating the device.
Electric Shock
■Do not touch the power supply with wet hands.
■Do not crimp the power cable, and avoid sharp edges.
■Do not overload extension cables connected to the device.
■Route the network and power cable carefully, to avoid tripping.
■Keep the device away from liquids.
Injuries and Infections
■Ensure that subjects do not have wounds or contagious diseases on the palms of their
hands or the soles of their feet.
■For hygiene purposes, Charder recommends cleaning the measuring platform after
each measurement with a soft cloth and alcohol.
■Ensure that the measuring platform is dry before usage.
5
I. SAFETY NOTES
Preventing Device Damage
■Please contact your local Charder distributor for regular maintenance and calibration.
■This device does not contain any user-maintained parts. All maintenance, technical
inspections, and repairs should be conducted by an authorized Charder service
partner, using original Charder accessories and spare parts. Charder is not liable for
any damages arising from improper maintenance or usage. Dismantlement of the
device will void the warranty.
■Take care to make sure fluids do not enter the device, as they may damage the
internal electronics.
■Switch off the device before disconnecting the power supply.
■Do not place the device in direct sunlight, or in close proximity to an intense heat
source. Excessively high temperatures may damage the internal electronics.
■Strong cleaning agents can damage the measuring platform's surface.
Caution
Caution
Alcohol wipes can be used to clean the electrodes and weighing platform.
Alcohol-based cleaning solutions should not be used on the touch screen.
Usage of Results
■The MA601 is not a diagnostic device. Results should be interpreted with assistance
from a professional.
■BIA results are calculated based on impedance values validated with representative
population studies and statistical analysis. As such, the technique is best suited for
tracking progress for an individual over a period of time, or for categorizing large
groups of people, rather than used as a one-time analysis. For more information on
getting the best results, please see Chapter VI. (INSTRUCTIONS FOR OPERA-
TION)

6
I. SAFETY NOTES
Additional information regarding the operating environment,
conditions for installation, or special conditions in usage.
Indicates helpful hints and supplementary information.
Indicates actions that should not be performed.
Bold text identifies buttons on the display panel or computer screen.
Hazard icon warning against possible electric shock.
NOTE
Bold
Identifies the possibility of serious injury or death for the
user if the device is mishandled, or safety instructions are
not followed.
Identifies the possibility of physical injury or device damage
if the device is mishandled, or safety instructions are not
followed.
The caution symbol indicates general precautions that should
be taken when using the device.
Warning
Caution
B. Precaution Symbols

7
I. SAFETY NOTES
EMC guidance and manufacturer's declaration
The Body Composition Analyzer
MA601 is suitable for use in all
stablishments, including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic
purposes.
Guidance and manufacturer’s declaration-electromagnetic emissions
The Body Composition Analyzer MA601 is intended for use in the electromagnetic environ-
ment specified below. The customer or the user of the Body Composition Analyzer MA601
should assure that it is used in such an environment.
Emission test Compliance Electromagnetic
environment-guidance
RF emissions
CISPR 11
Group 1
The Body Composition Analyzer
MA601 uses RF
energy only for its internal unction.
Therefore, its RF emissions are
very low and are not likely to cause
any interference in nearby electron-
ic equipment.
RF emissions
CISPR 11
Harmonic
emissions
IEC 61000-3-2
Voltage
fluctuations
/flicker emissions IEC
61000-3-3
Class B
Class A
Compliance

8
I. SAFETY NOTES
Guidance and manufacturer’s declaration-electromagnetic immunity
The Body Composition Analyzer MA601 is intended for use in the electromagnetic environ-
ment specified below. The customer or the user of the Body Composition Analyzer MA601
should assure that it is used in such an environment.
Immunity test IEC 60601
test level Compliance level Electromagnetic
environment-guidance
Electrostatic
discharge(ESD)
IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV,
± 8 kV, ± 15 kV air
+ 6 kV contact
+ 8 kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should be
at least 30%
Electrical fast
transient/burst IEC
61000-4-4
±2kV for power
supply lines
±1kV for input/output
lines
+ 2kV for power
supply lines Not
applicable
Mains power quality should
be that of a typical
commercial or hospital
environment.
+ 1kV line(s) to
line(s)
+ 2kV line(s) to
earth
+ 1kV differen-
tial mode
Not applicable
Mains power quality
should be that of a
typical commercial or
hospital environment.
Voltage Dips,
short interruptions
and voltage
variations on
power supply input
lines IEC
61000-4-11
0% UT for 0,5
cycle
0% UT for 1 cycle
70% UT(30% dip
in UT) for 25
cycles
0% UT for 5 s
<5% UT(>95% dip in
UT) for 0,5 cycle
40% UT(60% dip in
UT) for 5 cycles
70% UT(30% dip in
UT) for 25 cycles
<5% UT(>95% dip in
UT) for 5 s
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user of
the Body Composition
Analyzer MA601 requires
continued operation
during power mains
interruptions, it is recom-
mended that the Body
Composition Analyzer
MA601 be powered from
an uninterruptible power
supply or a battery.
Power frequen-
cy(50, 60 Hz)
magnetic field
IEC 61000-4-8
30 A/m 3 A/m
The Body Composition
Analyzer MA601 power
frequency magnetic fields
should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.
Surge IEC
61000-4-5

9
I. SAFETY NOTES
NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Guidance and manufacturer’s declaration-electromagnetic immunity
The MA601 is intended for use in the electromagnetic environment specified below.
The customer or the user of the MA601 should assure that is used in such and environment.
Immunity test IEC 60601 test level Compliance
level
Electromagnetic environment-
guidance
Conducted RF
IEC 61000-4-6
3 Vrms
150 KHz to 80
MHz
6 V in ISM bands
between 0,15 MHz
and 80 MHz 80 %
AM at 1 kHz
3 Vrms
Portable and mobile RF
communications equipment
should be used no closer to
any part of the Body
Composition Analyzer MA601
including cables, then the
recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Radiated RF
IEC 61000-4-3
3 V/m
80MHz to 2,7 GHz 3 V/m
Recommended separation
distance:
d = 1,2 √P
d = 1,2 √P 80MHz to 800
MHz d = 2,3 √P 800MHz to 2,7
GHz
Where P is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined
by an electromagnetic site
survey,ashould be less than
the compliance level in each
frequency range.b
Interference may
occur in the vicinity
of equipment
marked with the
following symbol:

10
I. SAFETY NOTES
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be consid-
ered. If the measured field strength in the location in which the Body Composition Analyzer
MA601 is used exceeds the applicable RF compliance level above, the Body Composition
Analyzer MA601 should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as re-orienting or relocating the
Body Composition Analyzer MA601
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
The Body Composition Analyzer MA601 is intended for use in an electromagnetic environment
in which radiated RF disturbances are controlled. The customer or the user of the Body Compo-
sition Analyzer MA601 can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the
Body Composition Analyzer MA601 as recommended below, according to the maximum output
power of the communications equipment.
Separation distance according to frequency of transmitter m
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2,7 GHz
d =1,2√P d =1,2√P d =2,3√P
W
0,01
0,1
1
10
100
0,12
0,38
1,2
3,8
12
0,12
0,38
1,2
3,8
12
0,23
0,73
2,3
7,3
23
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in metres (m) can be estimated using the equation applicable to the
frequency of the transmitter, where p is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer.
NOTE1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Rated maximum output
power of transmitter
Recommended separation distance between
portable and mobile RF communications equipment and the MA601

11
Trainers and athletes understand that fitness is much more than how you look and how much
you weigh. Quantifying where the muscle is going. Seeing if the loss in weight is from burning
through fat, or insufficient hydration. Tracking where progress has been made, and where to
focus your efforts. The fitness community has asked for precise tools and data to keep up with
increasingly advanced needs, and Charder is proud to present the MA601 Body Composition
Analyzer, designed to assist professionals in elevating program quality and progress analysis.
Body composition analysis was originally used primarily in the field for quantifying and measuring
the fundamental makeup of the body. Bioelectrical Impedance Analysis (BIA) is a fast, simple,
and non-invasive assessment of body composition with accurate results validated with widely
acknowledged and accepted gold standards such as DXA.
The MA601 provides the relevant measurement values and data that you need to bring your
program to the next level. Boasting multiple measurement frequencies and sophisticated
algorithms, Charder stands by our devices with clinical trials and over ten years of original
peer-reviewed scientific research, for results you can trust.
II. INTRODUCTION TO THE MA601
BODY COMPOSITION ANALYZER

12
III. INSTALLATION
Hand electrodes
Column
Measuring Platform
A. Contents
Unboxing accessories
MA601 body composition analyzer
Power adaptor DC 12V, 3A, 60VA
Power cord
User manual
Power cord
User manual
Power adaptor

13
III. INSTALLATION
B. Environment
The device should be placed on a flat and hard surface. Usage on carpet may result in
static electricity, which may damage the equipment and cause inaccuracies in measurement.
placed on hard surface
clear surroundings
keep space between the wall
flat surface
placed on carpet
uneven surface
placed against the wall
objects placed around the device

14
III. INSTALLATION
C. Installation Instructions
3. Remove polyethylene foam from box
4. Raise display column up into upright position
1. Open MA601 box
2. Remove user manual and power adaptor from box
Power adaptor
user manual

15
III. INSTALLATION
5. Take hand electrodes out from box and place them on hand electrodes holder on the
display
6. Remove polyethylene foam from box
NOTE:
At least two people are needed to remove the MA601 from its box

16
III. INSTALLATION
Use fastener to tighten column and base platform
Plug 12V Charder power adaptor in the jack.
Turn power switch ON to start the device
Bubble level adjustment instruction

17
IV. EXTERIOR AND PANEL DEFINITION
Control panel
Left hand electrode
Foldable joint
Foot electrode
Adjustable feet
Touch panel display
Right hand electrode
Column
Bubble level
Sole electrode

IV. EXTERIOR AND PANEL DEFINITION
18
Power switch
12V Power input
Touch display panel
Print button
Enter
Delete/Backspace
Numeral keys
Male
Female
Function button
Setup button
LAN port (Ethernet)
USB port x 2
Hand electrode holder

19
III. INSTALLATION
For connecting to a printer or USB flash drive
For connecting the MA601 to a network
For connecting to a power adaptor
For switching the MA601 on and off
USB port
LAN port
Power jack plug
Power switch
Rear panel definition
Rear panel Instruction
LAN port (Ethernet) x 1
USB port x2
Power switch
12V Power input
Table of contents
Other Charder Medical Equipment manuals