Title: Blanketroll II Rev: 3-04-2010
Model 222, 222A by Cincinnati
S b Zero Prod cts
DISCLAIMER: THIS PROCEDURE PROVIDED "AS IS" AND
WITH POSSIBLE FAULTS. USER MUST VERIFY BEFORE USE.
NEITHER PROVIDER NOR WEBSITE ASSUMES ANY
RESPONSIBILITY FOR ITS USE.
File = blanketroll.pdf
1. General
This nit heats and cools a water filled blanket between 39
and 107 degrees F. Typical inspection interval: 365 days.
2. Reference Doc ments
Operation and Technical Man al, Cincinnati-S b-Zero
P/N B2OTM circa 1984
See photo of label instr ctions.
Note: Monitor mode refers to a mode used only to
measure/display patient temperature via an optional
external temperature probe (YSI 400 Series).
utomatic Control refers to a feedback mode where
control is based on a patient connected probe.
3. Tools / Fixt res
Temperat re probe (ex: Fl ke ________)
Ad lt Hypothermia blanket – see photo.
Flowmeter: Model TFR Tester
Safety Analyzer
Distilled water.
4. Set p/Preliminaries
1.. Check the nit and accessories for cleanliness and damage.
2.. Remove panel and blow o t d st.
3.. Check the date strip on the back of the nit to
determine if the water needs to be drained from the
nit.
4.. If the water was drained, refill the reservoir to the
screen level with distilled water; if screen is missing,
the minim m level is abo t 3/16” above the
plate/s rface visible below the filling hole.
5.. Optional: Add the recommended disinfectant to the
water and place a new strip, dated one month ahead,
on the back of the nit.
6.. Check the castors for free movement; spray if
necessary.
7.. Check the hose o tlet connectors.
8.. Check each of the switches.
9.. Check the control panel by pressing <Test
Indicators>.
10.. Check the man fact rer's labels.
11.. Check the power cord, strain relief, power pl g, and
circ it breaker for operation.