G25-S / G25-E
Disinfection Soak Stations
Operator's Manual

Table of Contents
Principle of Operation 2
Safety Precautions 2
Environmental Conditions 2
Supply Ratings 2
Warning Symbols 3
Signal Words 3
Quick Troubleshoot 4
General Information 5
Main Power / Ratings Plate 6
Specifications for GUS Disinfection Soak Stations 6
Installation and Setup 6
Upon Delivery 6
Installation 6
Countertop models 6
Mobile models (Assembly of Cart Required) 7
Filling the High-Level Disinfection Tray 7
Operation 8
Placing Instruments into the System 8
Removing and Rinsing Instruments 8
Disposal of Disinfectant 8
Built-in Disposal System 8
Filter Life 10
Filter Replacement 10
Cleaning the System 11
Rinse Water 11
Limited Warranty 12
Registration Card 12
M0035_E 1 G25-S / G25-E
GUS Disinfection Soak Station™

Principle of Operation
The GUS® Disinfection Soak Station provides for the safe soaking of heat sensitive instruments in high-level
disinfectant solutions. The system’s blower forces air through a carbon filter which removes toxic vapor from the
high-level disinfectants.
Do not use the system for any purpose other than its intended use.
Safety Precautions
Disconnect power before replacing filter or when servicing the unit.
Refer to the MSDS of the high-level disinfectant that you are using for additional safety precautions.
Environmental Conditions
lFor indoor use only.
lAltitude up to 2,000 meters.
lTemperature range: 5°C to 40°C.
lMaximum relative humidity: 80% for temperatures up to 31°C, decreasing linearly to 50% relative humidity
at 40°C.
lMain supply voltage fluctuations not to exceed +/- 10% of the nominal voltage.
lInsulation over voltage category II.
lPollution degree 2.
Supply Ratings
AC: 110-120V0.5A - 2.0A50/60 Hz
Fuse: 250V3.0A5X20 FSF Panel Mounted
M0035_E 2 G25-S / G25-E
GUS Disinfection Soak Station™

Warning Symbols
The meaning of the labels and symbols that appear on the packaging and/or the system are as follows:
Dangerous electrical voltage
Hazardous area
Protective Grounding
Signal Words
The following signal words are used throughout this manual:
WARNING: Indicates a potentially hazardous situation which, if not avoided, could result in death or
serious injury.
CAUTION: Indicates a potentially hazardous situation which, if not avoided, may result in minor or
moderate injury. It may also be used to alert against unsafe practices or potential equipment damage.
NOTE: Indicates additional helpful information.
M0035_E 3 G25-S / G25-E
GUS Disinfection Soak Station™

Quick Troubleshoot
If the GUS system is not running, check the following:
1. Ensure that the unit is plugged into a live wall outlet.
2. Make sure the green On/Off switch is on. The switch will be illuminated when on.
3. Push the white ‘test’ button on the GFCI, located next to the On/Off switch. Then push the white ‘reset’
button. This should restart the system and the green On/Off switch should now be illuminated.
4. In the event that the system is still not working, you will need to check the fuse, which is externally mounted.
5. Unscrew fuse holder from the unit. The fuse holder is located to the right of the faceplate next to the fuse
label. If damaged, replace with a standard 3-amp fuse.
6. If the unit is still not working, call CIVCO Medical Solutions at 1-800-445-6741.
If you can still smell fumes from your high-level disinfectant while using the GUS system, check the following:
1. Both the soak and rinse trays are inside the disinfection soak station. It is very important that there is always a
rinse tray containing clean water for the initial rinse. Failure to do so will result in exposure to vapor when
transferring the instrument to the sink for rinsing.
2. The system is left on at all times while disinfectant is in the soak tray.
3. Lids are placed on the soak tray when not in use and during the soaking cycle.
4. Fresh disinfectant is being activated and poured into your soak tray while INSIDE the system, with it running
on high speed.
5. Disinfectant is being neutralized INSIDE the system prior to disposal.
6. Filters have been changed as recommended (every six months with daily use).
7. Spills outside of the system are immediately neutralized and cleaned up.
8. No other disinfectants are being used in the room, including inside an Automated Reprocessing System.
9. All personnel using the system have been fully trained on its use.
10. Verify that the system switches to the high-speed setting when you open the front acrylic door. You should
hear a marked increase in the noise level.
If the disposal pump is not working, check the following:
1. Check to make sure that the in-line filter on the suction hose is not clogged. If it is, you must replace the
filter. Contact CIVCO Medical Solutions for replacement in-line filters.
2. If the disposal pump is still not working, call CIVCO Medical Solutions at 1-800-445-6741.
M0035_E 4 G25-S / G25-E
GUS Disinfection Soak Station™

General Information
The GUS Disinfection Soak Station eliminates vapors from high-level disinfectants and contains splashes and spills.
Blowers pull fumes into a patented carbon filter, where they are adsorbed and eliminated. Clean air is returned to
the room through the top louver panel. These systems are available as Standard Countertop models (see picture),
or as Space Saver models which are designed to fit under an upper wall cabinet.
The GUS System with the standard filter is designed to remove fumes from:
lGlutaraldehyde
lOrtho-phthalaldehyde (OPA)
Optional filters are available for:
lPeracetic Acid
lHydrogen Peroxide
lFormaldehyde
WARNING: Failure to properly follow the instructions in this manual may result in damage to the
system and/or scopes.
M0035_E 5 G25-S / G25-E
GUS Disinfection Soak Station™

Main Power / Ratings Plate
The main power switch is located on the inside front panel of the system. The switch is illuminated when the
system is ON. The power cord plugs into a regular wall outlet. The fuse is a 250V 3.0A 5X20 FSF, panel mounted.
The ratings plate and serial number are located on the right side of the system.
Specifications for GUS Disinfection Soak Stations
UL & CE Approved.
Model
Air
Treated
(CFM)
FPM w/door
closed low
FPM w/door
open high Voltage
Current
w/door
closed
Current
w/door
open
Dimensions
(outside)
H x W x D
Dimensions
(inside)
H x W x D
G25-E 210 120 58 110V
50/60Hz 40 watts 75 watts
26.5"H
27"W
27.5"D
22"H
25.6"W
22"D
G25-S 210 120 58 110V
50/60Hz 40 watts 75 watts
16.5"H
27"W
27.5"D
14"H
25.6"W
22"D
Installation and Setup
Upon Delivery
Match all items in the package with those listed on your packing list. Inspect each item for damage. Fill trays (if
provided) with water and check for leakage. If any components are damaged or missing, or if you have any
questions, call CIVCO Medical Solutions at 1-800-445-6741. Do not use the system.
Installation
COUNTERTOP MODELS
1. Place the white polypropylene base on the countertop with the grooves facing up. Note that the grooves
should be on the sides and back.
2. Gently lift the system by the two side straps and place into the grooves on the base. There must be a
clearance of 1” above the unit.
3. Plug the system into a standard wall outlet.
4. Remove the protective covering from the acrylic doors.
5. Open the acrylic doors and place the soak and rinse trays inside.
M0035_E 6 G25-S / G25-E
GUS Disinfection Soak Station™

MOBILE MODELS (ASSEMBLY OF CART REQUIRED)
1. Remove the cart from the top of the box and place onto a level surface. Assemble as directed.
2. Gently lift the system out of the packing box and, with someone holding the cart, place the system on top.
CAUTION: It may require two people to lift the system and place it onto the cart.
3. Plug the system into a standard wall outlet.
4. Remove the protective covering from the acrylic doors.
5. Open the acrylic doors and place the soak and rinse trays inside.
WARNING: Extreme caution should be taken if moving the mobile system while it contains high-level
disinfectant.
Filling the High-Level Disinfection Tray
CAUTION: Make sure to turn the system ON before pouring high-level disinfectant into the soak tray.
CAUTION: Always leave the system running when high-level disinfectant is in the soak tray.
CAUTION: Keep the soak tray covered as often as possible, to minimize evaporation of the disinfectant.
CAUTION: Change the rinse water at least once per day.
1. Switch system ON using the GREEN lighted power switch located on the inside front of the system.
2. Open the front acrylic door. This automatically increases the fan speed, drawing more air through the
system.
3. One tray is for high-level disinfectant and the other is for clean rinse water. The rinse tray must be inside the
system with the soak tray.
NOTE: Not all trays are autoclavable. Check before autoclaving.
4. Remove the lid off the high-level disinfectant tray and set aside.
5. Prior to pouring the high-level disinfectant, read carefully and follow all label instructions.
6. Activation of the high-level disinfectant (if required) should be done inside the unit with the system turned
ON. Once activated, pour the solution into the soak tray. The rinse tray is used for initial rinsing of
instruments prior to removal from the system. Fill rinse tray with clean rinse water – change rinse water at
least daily. Refer to your instrument’s instruction manual regarding use of sterile water for rinsing.
CAUTION: It is important that you keep the soak tray containing high-level disinfectant covered at all
times to prevent evaporation.
M0035_E 7 G25-S / G25-E
GUS Disinfection Soak Station™

Operation
Placing Instruments into the System
1. All scopes and instruments should be pre-cleaned and dry prior to soaking in high-level disinfectant.
2. Open the front acrylic doors of the system and remove the lid on the soak tray.
3. Gently place your instruments into the tray containing high-level disinfectant.
4. Follow instrument manufacturer’s instructions for flushing lumens and channels of scopes.
5. Replace lids and close acrylic door.
Removing and Rinsing Instruments
1. When done soaking, open the front acrylic doors and remove the lid from the soak tray.
2. Gently lift instruments from the soak tray and shake off excess disinfectant.
3. Place instruments into the rinse tray inside the GUS system to remove residual chemical.
4. Remove from rinse tray and conduct final rinsing and wiping procedures as you normally do.
5. Refer to your instrument’s instruction manual for proper rinsing and storage procedures.
6. You must leave the system running at all times when the soak tray contains high-level disinfectant.
Disposal of Disinfectant
In some states and municipalities, it is illegal to dispose of used disinfectants without first deactivating them.
Deactivation also protects the user from exposure to splashes and spills. CIVCO Medical Solutions manufactures
Glute-Out®, a glycine-based neutralizer (part # 610-1045).
After the disinfectant is deactivated, it is safe to dispose of down the sink/drain with cold water running.
Thoroughly clean and rinse out the container. With the unit turned on, put the soaking container back into place
and refill it with fresh high-level disinfectant.
Built-in Disposal System
NOTE: Laboratory (L) Models do not include a disposal system.
The built-in disposal system is ideal for emptying the soak tray from inside the system, without the risk of exposure,
splashes or spills.
A disposable in-line filter is attached to the suction hose for the disposal system, which will prevent lint and small
debris from reaching the pump. If the filter becomes clogged, you will notice a decrease in water pressure.
Replacement in-line filters can be purchased in packs of five from CIVCO Medical Solutions.
1. Verify that the main power switch is ON before disposing of high-level disinfectant.
M0035_E 8 G25-S / G25-E
GUS Disinfection Soak Station™

2. We strongly recommend the use of neutralizer (Glute-Out®) when using OPA or glutaraldehyde to prevent
splashes, spills, and exposure to vapor. Add the required amount of neutralizer into the soak tray containing
high-level disinfectant and allow sufficient time for neutralization to take place.
3. Insert the clear plastic suction hose (shorter hose with in-line filter) into the quick disconnect fitting located
next to the disposal On/Off switch on the front inside of the system. Place the other end of the hose into the
soak tray.
4. Insert the discharge hose into the quick disconnect fitting located on the left side of the system. Place the
other end of the hose into the sink or hopper drain and turn on the cold water (DO NOT USE HOT WATER).
5. Switch on the pump.
6. After all the disinfectant is pumped out, add clean water to the soak bowl, rinse and pump out. Use the
suction hose to ‘vacuum’ all liquid from the bowl.
WARNING: Failure to fully rinse out disposal system after each use may damage the system and void
your warranty.
7. Switch off the pump and disconnect the suction and discharge hoses from the disinfection soak station. Store
until needed again.
WARNING: The disposal system on your GUS system is not designed to pump solids. A disposable in-line
filter on the suction hose must be in place at all times. Failure to do so may damage the pump and void
your system warranty.
M0035_E 9 G25-S / G25-E
GUS Disinfection Soak Station™

Filter Life
GUS® has a patented filter that, in normal everyday use, will have a six-month life. Call CIVCO Medical Solutions at
1-800-445-6741 to purchase additional filters.
Filter Replacement
WARNING: Turn the main power switch OFF and disconnect the power cord from the electrical outlet.
NOTE: It is recommended that the filter change take place at the time of replacing the disinfectant.
1. Remove the top louver plate located at the top of the machine by unscrewing the acorn nut with a 5/16 nut
driver. Note the G25 has a single top louver plate.
2. Remove the old filter. Put the used filter into a plastic bag and dispose of it with other non-hazardous waste.
This does not need to be red tagged.
3. Install new filter and replace louver panel.
Top Louver Plate
CAUTION: The system’s carbon filter must be replaced every six (6) months.
CAUTION: Prior to changing the filter, neutralize and dispose of any high-level disinfectant in the
soaking trays, and remove all equipment from the system.
CAUTION: Disconnect the system’s power supply before removing the filter.
CAUTION: An F251-A filter is required for use with hydrogen peroxide or peracetic acid based high-level
disinfectants.
M0035_E 10 G25-S / G25-E
GUS Disinfection Soak Station™

Cleaning the System
WARNING: When cleaning the system, always wear appropriate personal protective equipment.
Potentially infectious material adhering to the unit could pose an infection control risk.
CAUTION: Do not use solvents to clean the acrylic door. Doing so may damage the doors.
CAUTION: Do not wipe surfaces with hard or abrasive materials. Doing so will scratch the surfaces.
1. Turn the power switch OFF and disconnect the power cord from the wall outlet.
2. To clean metal surfaces: Use a clean, lint-free cloth moistened with a mild detergent solution and water. A
mild solvent may be used on these surfaces for difficult-to-clean areas.
3. To clean the HDPE plastic (side panels), use mild detergent solution such as dish washing liquid and water.
Hard to clean spots may be cleaned with a mild solvent if extra care is taken.
4. Make sure that the unit is completely dry before use.
Rinse Water
Once removed from the disinfectant solution, the instruments/scopes must be thoroughly rinsed with water to
remove any disinfectant residue. The initial rinsing must be conducted in the GUS system in order to eliminate risk
of exposure to residual chemicals left on the instrument.
Refer to your instrument or scope’s manufacturer instruction manual regarding the use of sterile water.
M0035_E 11 G25-S / G25-E
GUS Disinfection Soak Station™

Limited Warranty
CIVCO Medical Solutions warrants the CIVCO GUS Disinfection Soak Station to be free of defects in materials and
workmanship under normal use, for a period of one year from the date of delivery. This warranty does not include
filters, acrylic, or disposal pump unless it can be determined that the failure is due to defects in material or
workmanship.
If the system is not working correctly or is defective, contact CIVCO Medical Solutions and we will repair or replace
the device at our option during the warranty period.
CIVCOMedical Solutions assumes no liability for consequential damages of any kind as a result of the use or misuse
of the system by the purchaser, the purchaser's employees, or any others.
THIS WARRANTY IS EXPRESSLY IN LIEU OF ALL OTHER WARRANTIES, GUARANTEES, OBLIGATIONS OR OTHER
LIABILITIES EXPRESSED OR IMPLIED, AND THERE ARE NO WARRANTIES WHICH EXTEND BEYOND THE
DESCRIPTION ON THE FACE HEREOF. WARRANTIES OF MERCHANTABILITY AND OF FITNESS FOR A PARTICULAR
PURPOSE ARISING FROM A COURSE OF DEALING OR USAGE ARE SPECIFICALLY EXCLUDED.
This warranty does not cover conditions and damages resulting from any of the following:
lImproper installation, delivery within the facility, or maintenance.
lAny repair, modification, alteration or adjustment not authorized by the manufacturer.
lMisuse, abuse, accidents or unreasonable use.
lIncorrect electric current, voltage or supply.
lFailure to follow proper procedures for use outlined in this manual.
lConsequential or incidental damages sustained by any person as a result of any breach of this warranty.
Some states do not allow the exclusion or limitation of consequential or incidental damages, so the above
exclusion may not apply.
Registration Card
Please complete and return the attached Warranty Card to CIVCO Medical Solutions. Call 1-800-445-6741 to
purchase additional filters.
M0035_E 12 G25-S / G25-E