See Instructions for Use for full instructions,
warnings, precautions & contraindication
The following chemicals are compatible with the Koala cable: Alcohol-free
hand soap, Sodium hypochlorite (bleach) solution 10% in water, 70-80% Iso-
propanol, Ethanol 70%, Dilution of formaldehyde 3-6%, Cidex and Cidex plus.
Swab
Circumferential Sensor for
o
non-occluded, 360 readings
20 20
40 40
60 60
80 80
100 100
0 0
20 20
40 40
60 60
80 80
100 100
0 0
Proper Placement Improper Placement
Studies indicate that IUPCs are inadvertently placed outside
the amniotic membranes (extraovular - between the
chorion and the decidua-endometrial lining).
A catheter placed extraovular will still provide a reading,
but not the reading of absolute intrauterine pressure.
Deliberate and careful steps must be taken to ensure
proper catheter placement in the intra-amniotic space with
all IUPCs. Koala features a clear amniolumen to conrm
proper placement in the amniotic space.
Proper placement is conrmed with ashback of amniotic
uid, crisp waveforms and a good baseline.
Troubleshooting
Re-Zeroing the Cable Throughout
Labor (after initial insertion)
1Disconnect cable from catheter
(Do not disconnect cable from monitor)
2. Zero UA monitor according to manufacturer’s instructions
3. Wait 15 seconds
4. Reconnect cable to catheter and verify functionality
clinicalinnovations.com
747 West 4170 South | Murray, Utah USA 84123
Phone +1.801.268.8200 | Fax +1.801.266.7373
Toll Free +1.888.268.6222
© Clinical Innovations, LLC. All rights reserved. | ART-0144 Rev.01
Reusable Cable Cleaning Instructions
• Wipe the monitor plug and cable with water or alcohol
• Clean the outside of the catheter connector with soap and
water
• Apply water or alcohol to the swab that is provided. If
swabs other than those included are used, take care to
avoid disturbing the transducer protecting gel inside the
connector.
• Insert the swab inside the cable connector cone
• Rotate the swab and move it around the inside of the cone
Dampened Waveforms/Oligohydramnios
Action: If proper catheter placement is conrmed with
amniotic uid ashback, disconnect catheter from cable and
ush with 10 ml to 20 ml through Koala’s amnioport. The
catheter can also be rotated, advanced or retracted slightly.
Replace blue vented cap with the clear tethered cap after
ushing.
Local Distributor:
Reusable Connector
AMNIO
Endometrial Lining
Amniotic
Fluid
Space
Placenta
Circumferential Sensor for
o
non-occluded, 360 readings
20 20
40 40
60 60
80 80
100 100
0 0
20 20
40 40
60 60
80 80
100 100
0 0
Swab
forMOM. forBABY. forLIFE.™
Extraovular Placement
If amniotic uid is not present in catheter
lumen, question placement of catheter.
Presence of blood in catheter, high
baseline and dampened waveforms,
negative readings, or endometrial
tissue on catheter tip are all suggestive
of extraovular placement.
Action: Do not advance catheter; withdraw
and redirect. Pull back catheter tip to introducer
and alter catheter direction by changing angle of introducer,
or determine alternate position for placement and proceed
with insertion. Repeat insertion steps until proper placement
can be conrmed.
Good Baseline & Crisp Waveforms