4. Press and hold the Mode button until the
CALIBRATING icon flashes (about 10
seconds). Let go. Calibration will begin.
FRESH AIR IS REQUIRED. DO NOT
BREATHE ON UNIT.
6. If the LCD displays PASS press ENTER.
Calibration is complete.
5. After 10 minutes the LCD will display
PASS or FAIL.
7. If it displays “FAIL” repeat the process.
WhentoCalibrate:
The CO2 Safety Alarm should be
calibrated at least once a year, or after
an alarm, using normal, outdoor air
(approximately 400ppm). If you KNOW
the room is filled with fresh air, you can
calibrate the unit in-place. Otherwise you
should move the unit and power supply
outdoors and calibrate. Do not breathe
toward the unit during the calibration
process, as your breath’s CO2 will affect
the calibration settings.