Gather the following items:
 • CoaguchekXSProMeter
 • ContainerofTestStrips
 • TestStripCodeChip
 • Safe-T-ProPlusLancet
 • AlcoholWipe
 • CottonBallorTissues
1.
Makesureyouhavethecodechip
readytoinsertifyouareopening
anewboxofstrips.Thecodekey
needstobeinsertedintothemeter
everytimeyouuseanewlotofstrips.
2.
Cleanthengertipwithanalcohol
wipe.Orhavethepatientwashhisor
herhandsinwarmsoapywater.
Drythoroughly.
3.
Before continuing, review these
tips for getting a good drop of
blood.
Increasingthebloodowinthe
ngerwillhelpyoutogetagood
dropofbloodsokeepinmindthese
tips:
Warmthehand.Havethepatient•
holditunderhisorherarm,and/
orwashwithwarmwater.
Havethepatientletthatarm•
hangbyhisorherside.
Massagethengerfromitsbase.•
Use these techniques until the
fingertip has increased in colour.
4.
Takeateststripoutofthecontainer.
Close the container tightly.
You have 10 minutes to use the test
strip once you remove it from the
container.
5.
Slidetheteststripintotheteststrip
guideinthedirectionofthearrows
untilitstops.
The meter turns on and the main
menu appears.
If you insert a strip from a new lot
number for the first time, the meter
will display the number of the test
strip code chip required. Remove
the old code chip and insert the new
one.
6.
TouchtheScan buttonandholdthe
operatorIDbarcodeapproximately
10cmfromthebuilt-inscanner.The
meterbeepsoncethebarcodehas
beensuccessfullyread.
7.
PressPatient Test onthemainmenu
screen.
8.
TouchtheScanbuttonandholdthe
patientIDbarcodeapproximately
10cmfromthebuilt-inbarcode
scanner.Themeterbeepsoncethe
barcodehasreadsuccessfully.
An hour glass symbol appears.
The test strip symbol appears after a
few seconds with a flashing drop of
blood. You have 180 seconds to apply
the sample.
9.
Twisttheprotectivecapoffthe
lancet.
Massage the finger until you see
increased colour in the fingertip.
Keepingthehanddown,pressthetip
ofthelancetrmlyagainstthesideof
thengertip.Pressthepurpletrigger
button.
Gently squeeze from the base of the
finger to develop a drop of blood.
10.
Within15secondsoflancingthengertip,
applythebloodtothetargetareaonthestrip.
Findthecleartargetareaontheteststrip.You
candosefromthesideorthetop.
Hold the blood drop to the test strip until you
hear a beep. The flashing blood drop symbol
will disappear and an hourglass symbol appears
and the test begins.
Do not add more blood to the test strip.
Do not touch the test strip while the test
is in progress.
Themeterperformsanautomaticquality
controlcheckonthestripbeforeitdisplays
theresult.
QC √appears in the display briefly following a
successful QC check. The result appears in less
than a minute.
11.
Recordyourresultasinstructedby
yourfacility.
Note: If during testing the meter
displays an error message, refer to
the Error Messages section of the
user manual for an explanation and
steps on how to proceed.
12.
Quick Reference Guide
CoaguChek XS Pro
®
POINT OF CARE
TESTING
Operator’s Manual
CoaguChek XS Pro
®
POINT OF CARE
TESTING
Operator’s Manual
ForAssistancePhone0508 MY LIFE | 0508 69 54 33
CoaguChek XS Pro
®
POINT OF CARE
TESTING
Operator’s Manual
CoaguChek XS Pro
®
POINT OF CARE
TESTING
Operator’s Manual
Running Liquid QC - Quick Reference Guide
1. 2. 3. 4. 5. 6.
7. 8. 9. 10. 11. 12.
ForAssistancePhone0508 MY LIFE | 0508 69 54 33
Protection against infections: Thereisapotentialriskofinfection.MedicalstaffandotherpersonsusingtheCoaguChekXSPro
instrumenttoperformmeasurementsformorethanonepatientmustbeawarethatanyobjectcomingintocontactwithhumanbloodisa
potentialsourceofinfection.
FromtheCoaguChekXSPTControls
Kit(storedinthefridge),takeouta
bottleofcontrolcontainingthedried
powderandadiluentlleddropper.
Checktheexpirydate.
Carefullyremovethescrewcap
andrubberstopperfromthequality
controlbottleandmakesurethat
noneofthedriedpowderisremoved
withtherubberstopper.
Holdingthedropperatthebottomof
thestemusescissorstocutoffthe
tipofthedropperattheupperendof
thestem.Donotsqueezethebulbof
thedropper.
Insertthedroppertiprst,into
thecontrolsolutionbottle.Gently
squeezethebulbtodispensethe
entirecontentsofthedropperonto
thedriedpowder.Donottouchthe
driedmaterialandthedropper.
Gentlyswirl2-3times.
Minimumreconstitutiontimeone
minute.Use within 30 minutes of
making up.
Slidetheteststripintotheteststrip
guideinthedirectionofthearrows
untilitstops.
The meter turns on and the main
menu appears.
If you insert a strip from a new lot
number for the first time, the meter
will display the number of the test
strip code chip required. Remove
the old code chip and insert the new
one.
TouchtheScan buttonandholdthe
operatorIDbarcodeapproximately
10cmfromthebuilt-inscanner.The
meterbeepsoncethebarcodehas
beensuccessfullyread.
PressControl Test on the main
menuscreen.
Selectthecorrespondingcodeforthe
QCtestdisplayedonthescreenora
newcodeifusinganewlotofcontrol
andinsertqualitycontrolcodechip
whenprompted.
The meter stores the code chip
information in the memory for future
use.
Thedroppericonashestoindicate
thatthemeterisreadytoperformthe
test.
Atthesametimea180second
countdownbegins.Youmustapply
thesamplewithinthistime.
Mixthecontrolagainbygently
swirling.Drawthesolutionintothe
dropperandapplyonehangingdrop
tothecleartargetareaandthemeter
willbeepwhenthecontrolsampleis
detectedandtheINRmeasurement
begins.
TheresultwillappearandtheQC
rangewillbeunderneathinbrackets.
Ifthisisoutofrange,indicatedby
anupordownarrow,repeatthetest
mixingafreshvialofQCmaterial.if
stilloutofrange contact 0508 MY
LIFE | 0508 69 54 33.

Other Coaguchek Medical Equipment manuals

Coaguchek XS Plus System User manual

Coaguchek

Coaguchek XS Plus System User manual

Coaguchek INRange User manual

Coaguchek

Coaguchek INRange User manual

Coaguchek XS System User manual

Coaguchek

Coaguchek XS System User manual

Coaguchek XS System User manual

Coaguchek

Coaguchek XS System User manual

Coaguchek XS System User manual

Coaguchek

Coaguchek XS System User manual

Popular Medical Equipment manuals by other brands

Flyte MTI-1 quick start guide

Flyte

Flyte MTI-1 quick start guide

Codonics SLS-WAVE Technical brief

Codonics

Codonics SLS-WAVE Technical brief

Poskom VET-20BT product manual

Poskom

Poskom VET-20BT product manual

Okuvision OkuStim Instructions for use

Okuvision

Okuvision OkuStim Instructions for use

Firefly Medical ivea 600S user manual

Firefly Medical

Firefly Medical ivea 600S user manual

Braun Acculan 3Ti Instructions for use/Technical description

Braun

Braun Acculan 3Ti Instructions for use/Technical description

Gambro Prisma System Calibration Instruction

Gambro

Gambro Prisma System Calibration Instruction

HEINE OMEGA 200 instructions

HEINE

HEINE OMEGA 200 instructions

Mizuho MES-CK08-070-16EN-0 Instructions for use

Mizuho

Mizuho MES-CK08-070-16EN-0 Instructions for use

Stryker SV2 Operation manual

Stryker

Stryker SV2 Operation manual

Orliman Lumbitron Elite Use and maintenance instructions

Orliman

Orliman Lumbitron Elite Use and maintenance instructions

R82 Mustang user guide

R82

R82 Mustang user guide

biodex NxSTEP 950-485 Instructions for use and assembly

biodex

biodex NxSTEP 950-485 Instructions for use and assembly

Otto Bock 17AD100 10 Instructions for use

Otto Bock

Otto Bock 17AD100 10 Instructions for use

Mindray BeneView T1 Service manual

Mindray

Mindray BeneView T1 Service manual

SOMNICS iNAP R07-A Instructions for use

SOMNICS

SOMNICS iNAP R07-A Instructions for use

Nidek Medical GYC-1000 Operator's manual

Nidek Medical

Nidek Medical GYC-1000 Operator's manual

novos Bililed Maxi+ user manual

novos

novos Bililed Maxi+ user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.