Kanso®2SoundProcessor
QuickGuidefor Clinicians
5D1674149 V4 2020-12
Cochlear Ltd (ABN 96 002 618 073), Australia, Tel: +61 2 9428 6555
Cochlear Americas, USA, Recipient Customer Service, Tel: (800) 483 3123
Cochlear Canada Inc, Canada, Tel: +1 416 972 5082
www.cochlear.com
The Aqua+ should support your recipient’s normal programs.
However, if their hearing performance deteriorates, try these
solutions:
• Create custom MAPs that don’t use SCAN
• Use fixed directionality only (Standard or Zoom, with WNR
off).
ProgrammingwithKanso2Aqua+
For wireless programming, use the Cochlear Wireless
Programming Pod. The Wireless Programming Pod connects to a
compatible rechargeable battery module and a processing unit
to perform a sound fitting. Refer to the Wireless Programming
Pod User Guide.
For wired programming, use the Cochlear Kanso 2
Programming Adaptor Cable with the Cochlear Programming
Cable and Freedom™Programming Pod. Refer to the Freedom
Programming Pod User Guide.
WARNING
Programming Adaptor Cables can be a strangulation hazard. Keep out of
reach of children.
NOTE
Wired programming can only be carried out in clinical environments.
Wirelessandwiredprogramming EnablingForwardFocus
If you believe the recipient may benefit from bass and treble
control, you can enable it.
WARNING
Only enable bass and treble control for recipients who can provide reliable
feedback on sound quality. Give special consideration to young children,
older adults or difficult-to-test recipients.
Enablingbassandtreble
Programmingtips
If you believe the recipient may benefit from master volume
control, you can enable it.
WARNING
Only enable master volume limit control for recipients who can provide
reliable feedback on sound quality. Give special consideration to young
children, older adults or difficult-to-test recipients.
Enablingmastervolumelimit
Cochlear, Hear now. And always, Nucleus, and the elliptical logo are either trademarks or registered
trademarks of Cochlear Limited. App Store, iPhone and iPod touch are trademarks of Apple Inc,
registered in the U.S. and other countries. Android is a trademark of Google Inc. Google Play is a
trademark of Google LLC. Roger is a trademark of Sonova AG.
© Cochlear Limited 2020
ForwardFocus is a clinician enabled and user-controlled feature
designed to help recipients hear better in face-to-face
conversations by reducing distracting noise occuring behind the
recipient.
IMPORTANT
ForwardFocus should only be enabled for users 12 years and older who are
able to reliably provide feedback on sound quality and understand how to
use the feature when moving to different or changing environments. It
may be possible to have decreased speech understanding when using
ForwardFocus in a quiet environment.
Before enabling FowardFocus, you should ensure that recipients
can correctly identify quiet and noisy environments. Because
the algorithm for ForwardFocus assumes the sound signal of
interest is in front of the recipient, and the competing noise is to
the sides and/or behind the recipient, you should counsel
recipients to orient themselves so that distracting sounds are
behind them, and the sound they want to hear is in front of
them.
Advise caregivers of children on the importance of feedback
from the recipient to ensure the most appropriate program is
being used.
If you enable ForwardFocus for a recipient, they turn it on and
off using the Nucleus Smart App. ForwardFocus will be turned
off when the sound processor is turned off.
Advise the recipient to turn FowardFocus off when they move to
a different listening situation, for example, when moving from a
noisy café to a quiet room, or if they notice difficulty in
understanding speech in a quiet environment.
Clinicians should provide recipients with several listening
scenarios such as a noisy play ground, a noisy classroom, or
listening in a café, to ensure that the recipient understands how
to orient themselves in front of the noise to ensure maximum
benefit of the ForwardFocus feature.
Cochlear Ltd (ABN 96 002 618 073) 1 University Avenue, Macquarie University, NSW 2109, Australia
Tel: +61 2 9428 6555 Fax: +61 2 9428 6352
Cochlear Ltd (ABN 96 002 618 073) 14 Mars Road, Lane Cove, NSW 2066, Australia Tel: +61 2 9428 6555
Fax: +61 2 9428 6352
Cochlear Deutschland GmbH &Co. KG Karl-Wiechert-Allee 76A, 30625 Hannover, Germany
Tel: +49 511 542 770 Fax: +49 11 542 7770