Confycare EL12 User manual

EL12 / EL12C / EL12D / EL12M
Electrical Tilt Table


EN
1
SAFETY PRECAUTIONS
1. During the adjustment of the table, do not put any
part of the body into the gap of any moving parts.
2. Please operate with both hands when adjusting the
leaning angle of back, foot or head sections.
3. When the patient is on the table, it is not recommended
to adjust the angle of the table top.
4. Before using the product, ensure the four wheels of
the table frame touch the ground at the same time and
check the stability of the table frame.
5. Before connecting to the power supply, please check
whether the incoming line to the controller is intact.
6. A regular inspection (at least once a year) is needed
to the treatment table.
7. Do not exceed the table weight capacity of 250 kg.
The safe load of the arm rest is 15 kg.
8. Always clean the table after each use with a soft cloth
dampened with water and a mild antibacterial detergent.
Do not use abrasive cleansers..
9. Disconnect the table from the power source before
attempting any maintenance, installation, removal or
replacement procedures to prevent electrical shock and
possible damage to the table.
10. Please DO NOT make any changes to the table.
Replacement parts or labor furnished can be operated
by Coinfycare, the selling dealer or a service technician
certified by the manufacturer..
CONTENTS OF CARTON
Electrical Tilt Table: 1 unit
Instruction Manual: 1 unit
NORMAL OPERATING CONDITIONS
Temperature: 5 ℃ - 40 ℃
Relative humidity: 30% - 75%
Air pressure range: 860-1060hpa
Power supply: AC 220V ± 10%, 50Hz
REGULATION AND PRODUCT
STANDARDS
This table complies with the European regulation of
medical devices (EU) 2017/745 and with the following
standards:
- EN 60601-1:2006+A1:2013+AC:2014
Medical electrical equipment. Part 1: General
Requirements for Basic Safety and Essential
Performance.
- IEC/EN 60601-2-52:2009+A1:2015 clause
201.9.8.3.2
Medical electrical equipment. Part 2-52: Particular
requirements for basic safety and essential
operation of medical tables.
- EN 60601-1-6:2010+A1:2015
Medical electrical equipment. Part 1-6: General
Requirements for basic safety and essential
performance. Collateral standard: Suitability for
use
- EN 60601-1-2:2015
Medical electrical equipment. Part 1-2: General
requirements for basic safety and essential
performance. Collateral standard:
Electromagnetic disturbances. Requirements and
tests.
- EN ISO 13485:2016
Medical devices - Quality management systems -
Requirements for regulatory purposes
(ISO 13485:2016)
DIN EN ISO 13485:2016
PRECAUTIONS AND SAFETY
INDICATIONS
Coinfycare is not liable for any material or personal
damage due to misuse or incorrect use of this table
or for any purpose other than described in this user
manual.
Inspect the condition of the table before using it for
the first time and on a recurring basis as indicated in
the “MAINTENANCE” section.
Keep table out of high moisture environments.
Do not transport the table over rough or outdoor
surfaces. The retractable caster system could
become damaged and inoperable.
This table should be operated, transported, and
stored in temperatures between 5° C to + 45° C,
with max humidity of 75%.
DO NOT operate this table in an environment
where other devices are being used that
intentionally radiate electromagnetic energy in an
unshielded manner. Portable and mobile
communications equipment can affect Medical
Electrical Equipment.

EN
2
This table generates, uses, and can radiate radio
frequency energy and, if not installed and used in
accordance with the instructions, may cause harmful
interference to other devices in the vicinity. However,
there is no guarantee that interference will not occur in
a particular installation. Harmful interference to other
devices can be determined by turning this table on
and off. Try to correct the interference using one or
more of the following: reorient or relocate the receiving
device, increase the separation between the
equipment, connect the table to an outlet on a
different circuit from that which the other device(s) are
connected, and consult Coinfycare for help.
Unplug the table from the power supply before starting
cleaning or any types of maintenance operations.
The table should be kept out of the reach of children
when there is no adult supervision.
The table is designed for discontinuous interval
operation, so the continuous running time of the motor
should never exceed 2 minutes. Should this happen,
damage may occur due to overheating of the
electronics.
Always use original accessories and spare parts
provided by your dealer.
Always cover the table with a towel, disposable paper
or a cover and change it for each user, thus ensuring
correct hygiene and extending the useful life of the
upholstery.
Only connect the table to a properly installed and
grounded outlet. The local mains voltage must
correspond to the technical data of the device.
The power outlet should be accessible in case
the table needs to be quickly unplugged at any
given time. Always pull the plug and not the mains
cable.
Unplug the table from the power supply before starting
cleaning or any types of maintenance operations.
TECHNICAL SPECIFICATIONS
Product dimensions
EL12 - Electrical Hi-Lo & Tilt Table
Length: 1950mm
Width: 640mm
Lifting range of bed surface: 580-1040mm
Tilting angle: ≤85°
Size of activity table: 300x730cm
EL12C - Electrical Child Tilt Table
Length: 1550mm
Width: 560mm
Lifting range of bed surface: 480-840mm
Tilting angle: ≤85°
Size of activity table: 300x650cm
EL12D - Electrical Tilt Table
Length: 2040mm
Width: 640mm
Tilting angle: ≤85°
Size of activity table: 300x730cm
EL12M - Electrical Dynamic Tilt Table
Length: 2040mm
Width: 870mm
Lifting range of bed surface: 580-940mm
Tilting angle: ≤85°
Size of activity table: 300x730cm
If the electrical supply cable is defective,
disconnect the table immediately and inform your
dealer or a specialized technician.
Position the power cord properly so that it cannot be
tripped over.
Unplug the power cord from the mains if the table is
not going to be used for a long period of time.
If you detect any abnormality in the operation of the
table, consult your dealer or a specialized technician.
Never manipulate any electrical part or open or
repair any part of the table.
Only your dealer or a specialized technician should
repair the table.
Never touch the table or the power cord with wet
hands.
Never place your hands or any other part of your
body in any hole or void on the table while it is in
operation.
Make sure there are no objects in the space under
the table that could cause collisions or entrapment
during operation.
Make sure that your feet are not resting on the
peripheral foot bar or foot control during treatment
and especially when sitting near the table. This
could create a dangerous situation where your leg
is caught between the section and the peripheral
foot bar or floor surface.
Take into account the weight distribution on the
table when working. The weight must be distributed
along the table without concentrating on its ends to
maintain the correct stability.
You should not sit on the head and foot sections. If
you do, there is a danger of tipping and table damage.
PLACEMENT IN THE
WORKPLACE
The table has four wheels to facilitate movement
to the place where it will be used.
Once correctly positioned, the table has two
adjustable legs that prevents it from moving during
use (except the EL-12M)

EN
3
1
2
3
EL12 Electrical Hi-Lo & Tilt Table
OPERATION
Figure 1: Table in working position.
Locked
Locked / release
Figure 1
Red side down
When installing the activity table do not relax
your hands before the activity table is fixed
to avoid accidental slippage and cause
injury.
● insert the square rod of activity table into the
Figure 2: Table in a moving position. square tube , adjusting the square rod to
the right distance then tighten the knob to fix.
● To adjust the height of the activity table
tighten the knob on the bracket to fix it
in place.
Release
Release
Figure 2
Flat Adjusting the footboard plate
Foot
Connect the table to an electrical outlet using the
power cord.
Motor functions
● Hand controller for adjusting height & tilt.
● Before adjusting, make sure there are no objects
in the space below the table that could cause
collisions or entrapment.
● Press LINE 1 on the controller panel to
adjust the height of the table.
Before adjusting make sure the table is in the
non-tilting position.
● release the bolt 2 then adjusting the
supporting rod 3 .
● the footboard can be adjusted to suit the desired
angle Increment depends on the patient’s need.
Fixation straps
● Press LINE 2 on the controller panel to
adjust the tilt of the table.
Chest strap
Waist strap
Leg straps
Fix and adjust the activity table
Tighten the belts by fastening the straps.
The position of the straps can be moved up and
down along the square tube.
Knob
1
Square Tube 3
Knob 2
board
1
Bolt
2
Supporting rod
3
Height adjust of the table
Tilting adjust of the table

EN
4
1
2
EL12C Electrical Child Tilt Table
OPERATION
Fixing and adjusting the head/waist pad
Figure 1: Table in working position.
Locked
Locked / release
Figure 1
Red side down
● Pushing the pole on the pad into the metal
adapter adjust to the appropriate size then tighten
the black knob.
Figure 2: Table in a in moving
position.
Fixing and adjusting the knee support pad
● Step 1: Pushing the pole on the knee pad
frame then tighten the black knob.
Black knob
Release
Release
Figure 2
Flat
Connect the table to an electrical outlet using the
power cord.
Motor functions
● Hand controller for adjusting height & tilt.
● Before adjusting, make sure there are no objects
in the space below the table that could cause
collisions or entrapment.
● Press LINE 1 on the controller panel to
adjust the height of the table.
● Step 2: Put the knee pad into the frame adjusting
to the tube is used to adjust the gap between
the knee pad to knee.
The tube is used to adjust the distance
between the two knees. Then tightening all black
knobs.
Tube
● Press LINE 2 on the controller panel to
adjust the tilt of the table.
Tube
Black knob
2
1
Height adjust of the table
Tilting adjust of the table

EN
5
1
2
3
Adjusting the foot board plate Fix and adjust the activity table
Foot
Before adjusting make sure the table is in the
non-tilting position.
● release the bolt 2 then adjusting the
supporting rod 3 .
● the foot board can be adjusted to suit the desired
angle Increment depends on the patient’s need.
Fixation straps
When installing the activity table, do not
relax your hands before the activity table is
fixed to avoid accidental slippage and cause
injury.
● insert the square rod of activity table into
the square tube , adjusting the square rod to
the right distance then tighten the knob to fix.
● To adjust the height of the activity table
tighten the knob on the bracket to fix it
in place.
Chest strap
Waist strap
Leg straps
Fix and adjusting the leg support
Tighten the belts by fastening the straps.
The position of the straps can be moved up and
down along the square tube.
● Fix the leg support to the channel with the black
knob.
● Adjusting the height up and down then it will lock
into place
Knob
1
Square Tube 3 Knob 2
board
1
Bolt
2
Supporting rod
3

EN
6
1
2
3
EL12D Electrical Tilt Table
OPERATION
Figure 1: Table in working position.
Locked
Locked / release
Figure 1
Red side down
When installing the activity table do not relax
your hands before the activity table is fixed
to avoid accidental slippage and cause
injury.
● insert the square rod of activity table into the
Figure 2: Table in a in moving
position.
square tube , adjusting the square rod to
the right distance then tighten the knob to fix.
● To adjust the height of the activity table
tighten the knob on the bracket to fix it
in place.
Release
Release
Figure 2
Flat Adjusting the foot board plate
Foot
Connect the table to an electrical outlet using the
power cord.
Motor functions
● Hand controller for adjusting height & tilt.
● Before adjusting, make sure there are no objects
in the space below the table that could cause
collisions or entrapment.
● Press LINE 1 on the controller panel to
adjust the height of the table.
Before adjusting make sure the table is in the
non-tilting position.
● release the bolt 2 then adjusting the
supporting rod 3 .
● the foot board can be adjusted to suit the desired
angle Increment depends on the patient’s need.
Fixation straps
Chest strap
● Press LINE 2 on the controller panel to
adjust the tilt of the table.
Leg straps
Waist strap
Fix and adjust the activity table
Tighten the belts by fastening the straps.
The position of the straps can be moved up and
down along the square tube.
Knob
1
Square Tube 3
Knob 2
board
1
Bolt
2
Supporting rod
3
Height adjust of the table
Tilting adjust of the table

EN
7
EL12 Electrical Dynamic Tilt Table
OPERATION
Figure 1: Table in working position.
Lowest
Locked
Locked / release
Figure 1
Red side down
Figure 2: Table in a in moving position.
Release
Release
Figure 2
Flat Highest
Connect the table to an electrical outlet using the
power cord.
Motor functions
● Hand controller for adjusting height & tilt.
● Before adjusting, make sure there are no objects
in the space below the table that could cause
collisions or entrapment.
● Press LINE 1 on the controller panel to
adjust the height of the table.
● Press LINE 2 on the controller panel to
adjust the tilt of the table.
Tilting Position
Height adjust of the table
Tilting adjust of the table

EN
8
Dynamic Position
Adjusting the back section
Bolt Handle
● Step 2: Turn the black handle into the
unlock position & the back section can then
slide on the track.
Single leg training
● Step 1: Remove the leg upholstery section
from the frame
● Step 2: Take the single leg pad from under the
backrest & insert into the square tube to fix it in
place.
● Step 3: Turn the black handle to the unlock
position
● Before adjusting make sure the table is in a flat
position.
● When using the tilt function, make sure the Bolt
on the back section is on “LOCK”
Lift the back section:
Release the bolt then push the handle and adjust
the angle.
Lower the back section:
Push the handle down then lower the backrest
to the flat position and put the bolt back to the
hole.
Dynamic status
Double Leg Training:
● Step 1: Take out the leg section from the table
frame
Before training make sure to tighten the belts.
Press your foot up to the top of the foot plate,
then release your legs to slide on the track.
Adjusting the training stroke
Set the leg training stroke according to each
patient. This should always be set according
to patient strength.
Leg pad

EN
9
Black bolt
late
1
Bolt
2
Supporting rod
3
1
2
3
Pull the black bolt to adjust the stroke, then lock
again
Black handle to unlock sliding & lock tilting
function
When installing the activity table do not relax
your hands before the activity table is fixed
to avoid accidental slippage and cause
injury.
● insert the square rod of activity table into
the square tube , adjusting the square rod to
the right distance Tighten the knob to fix.
● To adjust the height of the activity table
tighten the knob on the bracket to fix it
in place.
Adjusting the footboard plate
Chest strap
Waist strap
Leg straps
Adjusting the split foot plate
P
Tighten the belts by fastening the straps.
The position of the straps can be moved up and
down along the square tube.
Split-leg design
Chest strap
Before adjusting make sure the table is in the flat
position.
● release the bolt 2 then adjusting the
supporting rod 3 .
● the foot plate can be adjusted to suit the desired
angle Increment.
Waist strap
Split-leg straps
Fix and adjust the activity table
Knob
1
Square Tube 3
Knob 2

EN
10
Legend to symbols and information on the product
label.
- Model: Product identification.
- Model Num.: Product version.
- : CE conformity marking.
- : Unique serial number identifying the product.
- : Date of manufacture.
- Product Net Weight : Table weight in kg.
- INPUT : Necessary values of the electrical mains supply.
- ON: 2 min / OFF: 18 min: Indicates the discontinuous
operation of the device according to IEC/EN 60601-1,
which indicates that 2 minutes of continuous operation
of the motor or motors should not be exceeded.
- : Consult the instructions for use or the user
manual.
- : Maximal load capacity that the table can
support.
- : Waste from Electrical and Electronic
Equipment. Selective waste collection
of electrical and electronic equipment is
required.
- : Classification of medical electrical
equipment - Double Insulation.
- : Electrical Safety Classification - Type B.
- : Protection against splashing water or liquids.
- : Manufacturer identification.
- : Authorized representative in the EU.
- : Importer.
TROUBLESHOOTING
If, during the life of the table you detect any malfunction
or need a repair, contact your dealer and do not use the
table until you have received a diagnosis or the table has
been professionally repaired.
MAINTENANCE
Carry out a visual inspection at least once a month
to check the general conditions of the table. Moreover,
you must do the following:
- Check the correct adjustment of the different
sections.
- Check the proper functioning of the motors and
their movements.
- Check and clean the gas struts
- Check and tighten screws (if necessary).
- Grease the parts with friction.
- Check the conditions of the electrical wiring and
the socket.
- Check the wheel immobilization system.
RECYCLING
If you need to dispose of the table or any of its
parts or accessories, consult and respect the
laws and regulations applicable to waste
management and protection of the environment.
WARRANTY
This warranty shall remain in effect for 5 years
(60 months) for the structure and 3 years
(36 months) for the base frame and motor from the
date of original consumer purchase. An original
sales receipt or proof of purchase may be required
in order to redeem the warranty.
All repairs to the Product must be performed by a
service center certified by Coinfycare. Any
modifications or repairs performed by unauthorized
centers or groups will void this warranty. This
statement of guarantee prevails over any other
statement specified or implied by the seller.
CLEANING AND DISINFECTION
Before cleaning the table, unplug it from the mains.
Clean the table after each use.
Dust the table with a soft, dry cloth.
To clean the upholstery, apply a solution of warm
water and a mild detergent (for example, neutral
soap) with a well wrung soft cloth. Wipe the
table over once cleaned.
Disinfect upholstery and other surfaces with a
suitable bactericide for this purpose.
Apply the disinfectant on a soft cloth, never directly
on the table.
Never use a mixture of disinfectants or a mixture of
disinfectant and soap.
Wipe the table upholstery over once cleaned &
disinfected with a cloth and water only. This
will remove any residue from the upholstery.
Avoid the accumulation of moisture on the moving
parts of the metal structure, as they could rust.
Since we do not know which disinfectant will be
used, we cannot guarantee that it will not cause
damage to the upholstery. Read the instructions
and warnings for use of the disinfectant product
carefully before its purchase or use.

EN
11
TRANSPORTATION AND STORAGE
CONDITION
Transport condition
The products can be transported by general means
of transportation. They should be protected from the
sun and rain, and should not be mixed with toxic,
harmful and corrosive materials.
Storage
The packed electric multifunctional physiotherapy
bed should be stored in a dry and well ventilated
environment without corrosive gas.
SHUTDOWN PROCEDURE
Adjust the bed to a suitable position to ensure that
the patient is safely left, then turn off the power.
Contraindication
Please note that our beds are not sterilized.
Please do not use them according to aseptic
standards.
CONTACT
Applicant / company name / after sales service unit:
Huangshan Jinfu Medical Instrument Co., Ltd
Domicile / production address: No. 17,
Qiliting, Qimen County, Huangshan City,
Anhui Province China
Tel: 0086 571 87006218
www.coinfycare.com
European legal representative:
LOTUS NL B.V.
Koningin Julianaplein 10, 1e Verd 2595AA,
The Hague, Netherlands



JF-EMN-2022V.01
www.coinfycare.com
This manual suits for next models
3
Table of contents
Other Confycare Medical Equipment manuals

Confycare
Confycare FIX-MT2 User manual

Confycare
Confycare JF-D39 User manual

Confycare
Confycare EL3603 User manual

Confycare
Confycare EL03EM3 User manual

Confycare
Confycare EL05 User manual

Confycare
Confycare EL04 User manual

Confycare
Confycare EL3605 User manual

Confycare
Confycare EL08 User manual

Confycare
Confycare EL35 User manual