- Back lifting; - Leg lifting; - Vertical lifting; - Emergency release.
1.3 Product function
- Back angle (0-75°±2°), knee angle (0-35°±2°), leg angle (0-10°±1°), height adjustment (440mm-740mm (±50mm)), the
posture of the bed is realized by electric operation, and it is equipped with indicator lights to display the execution status
of the function.
- An emergency stop button is installed on the right side of the patient's bed to stop the movement of the bed in an
emergency.
- Equipped with a battery to deal with emergencies such as power outages.
- A circular degree display sticker is designed on the guardrail. The circular display sticker is engraved with the degree
of inclination, and there are steel balls in it. When the bed panel is inclined, the degree of circularity. The steel ball in
the digital display will scroll to the corresponding position with the lifting angle, and has the function of displaying the
degree of lifting.
1.4 Product feature
The drive motor in the electric push rod is decelerated by the worm gear, and then the rotary motion of the motor is
changed into linear motion through the screw and nut structure.
Reverse to complete the putter action. The push rod realizes various actions of the electric hospital bed through the
movement of various levers, rockers or connecting rods.
1.5 Working principle
The intended use of this product is in hospitals, nursing homes and similar care facilities.
The expected usage environment requirements are as follows:
a) Ambient temperature range: 5℃~40℃;
b) Relative humidity range: 30%-75%;
c) Atmospheric pressure range: 700hpa~1060hpa;
d) Power supply: AC 100-240V, 50Hz±1Hz.
e) Internal working voltage: DC 24V
1.6 Working condition
Used during diagnosis, treatment, or monitoring of an adult patient under medical supervision to support the patient's
body in a clinically required position.
1.7 Scope of application&contraindication
1.7.1 Scope of application
It is suitable for adult patients who need nursing and recuperation.
1.7.2 Target user
There have been no reports.
1.7.3 Contraindication
2. Precautions
- Please be careful not to put your head, hands, feet, etc. into the gap between the guardrail and the headboard and
foot board;
- Please do not put your hands (fingers) and feet between the bed body and the bed panel;
- Please do not sneak under the bed, or put your hands and feet under the bed;
- Please do not sit on the side rails of the bed and on the head and tail boards;
- Please do not sit on the rising bed panel;
- Please do not climb, play, jump on the bed or use the bed as a climbing platform;
- Please do not clamp the power cable, and do not be caught by the power cable on the casters!
A safety hazard can result if the power cord is not handled properly, such as rollers running over cables or wires,
pinching between the head/foot plate assembly and moving parts. Therefore, patients and medical staff should not use
the roller to press the cable through the cable,pay attention to the squeezing between the head/foot plate assembly and
the movable parts.
- When it is necessary to go under the bed for cleaning and other work, please unplug the power plug.