Cylon CBT-STAT User manual

SMART ENERGY CONTROL
CBT-STAT and CBT-STAT-H
USER GUIDE

2© Cylon Controls Ltd. MAN0120US rev 16
CBT-STAT and CBT-STAT-H
© Cylon Controls Ltd. 2019. All Rights Reserved.
CBT-STAT and CBT-STAT-H (MAN0120US rev 16)
Information in this document is subject to change without notice. The software described in this document is furnished under a
license agreement or nondisclosure agreement. The software may be used only in accordance with the terms of those agreements.
No part of this publication may be reproduced, stored in a retrieval system or transmitted in any form or any means electronic or
mechanical, including photocopying and recording for any purpose other than the purchaser’s personal use without the written
permission of Cylon Controls Ltd.
TRADEMARKS
All trademarks and trade names used in this document are acknowledged to be the copyright of their respective
holders.
UniPut™ technology is covered by UK Patent GB 2 400 991 and Irish Patent 84413
www.cylon.com
www.cylon-automatrix.com

MAN0120US rev 16 © Cylon Controls Ltd. 3
Contents
Style conventions used in this document:
UI Text: Text that represents elemets of the UI such as button names, menu
options etc is presented with a grey background and border, in
Tahoma font which is traditionally used in Windows UIs. For example:
Ok
Standard Terms (Jargon): Text that is not English Language but instead refers to
industry standard concepts such s Strategy, BACnet, or Analog Input
is represents in slightly condensed font. For example:
BACnet
Code: Text that represents File paths, Code snippets or text file
configuration settings is presented in fixed-width font, with a grey
background and border. For example:
$config_file = c:\CYLON\settings\config.txt
Parameter values: Text that represents values to be entered into UI fields, or
displayed in dialogs is represented in fixed-width font with a shaded
background. For example
10°C
Product Names: Text that represents a product name is represented in bold
coloured text. For example
CBX-8R8-H
PC Keyboard keys: Text representing an instruction to press a particular key on
the keyboard is enclosed in square brackets and in bold folt. For
example:
[Ctrl]+[1]

4© Cylon Controls Ltd. MAN0120US rev 16
CBT-STAT and CBT-STAT-H
CONTENTS
SECTION 1 : Introduction
What are CBT-STAT and CBT-STAT-H? 6
SECTION 2 : User Mode
User Mode 8
SECTION 3 : Engineering Mode
Engineering Mode 12
Using the Engineering Mode menus 13
Available parameters 14
Balancing Procedure - CBV-2U4-3T(-N) 18
Balancing Procedure –CBT12iVAV 19

SECTION 1 : INTRODUCTION

6© Cylon Controls Ltd. MAN0120US rev 16
CBT-STAT and CBT-STAT-H
WHAT ARE CBT-STAT AND CBT-STAT-H?
The CBT-STAT and CBT-STAT-H provide dedicated, visually
appealing Room Control displays for use with CBV and CBT
Unitary Controllers.
Both displays allow the user to view and adjust selected
7parameters within the controller to which they are connected,
and each has an integrated temperature sensor. The CBT-STAT-
Halso has an integrated humidity sensor.
While the display can be used for local control, the CBT12iVAV
and CBT14 Unitary Controllers can be easily integrated into the
Cylon Controls BACnet system architecture.
By default the CBT-STAT and CBT-STAT-H operate in User mode
–see User Mode on page 8 for details.
The CBT-STAT or CBT-STAT-H can be put into Engineering Mode,
allowing the display to be used as a commissioning tool,
adjusting preset points in the Unitary Controller’s default
strategy. See Engineering Mode on page 12.

SECTION 2 : USER MODE

8© Cylon Controls Ltd. MAN0120US rev 16
CBT-STAT and CBT-STAT-H
USER MODE
User mode is the default behavior for the CBT-STAT and CBT-STAT-H keypads.
Keypad and Display
1st line (large digits)
2nd line (small digits)
Operation Mode indicators
Keypad
Left button
No function
.
Up button
Increase setpoint
.
Right button
Toggle occupancy override (if Permit
Occupancy Override is enabled)
.
Down button
Decrease setpoint
.
c
a
b
d
Temperature
In User mode, the 1st line (large digits) on the display shows the current Temperature Sensor reading.
Humidity
On the CBT-STAT-H, the 2nd line (small digits) shows the current Humidity Sensor reading.
Temperature Setpoint
On the CBT-STAT, the 2nd line (small digits) shows the current value of the Temperature Setpoint.
On the CBT-STAT-H, pressing any of the buttons or changes the 2nd line (small digits) to show the
Temperature setpoint.
a
b
c
d
a
b
b
b

MAN0120US rev 16 © Cylon Controls Ltd. 9
How to adjust the Temperature Setpoint
If the Controller’s Strategy has been configured to permit it, the user can adjust the Temperature setpoint or occupancy status.
•Press any button so that the temperature setpoint is displayed on the second line, with a flashing unit symbol.
•Press the up button or down button to adjust the setpoint value by the span defined in the Controller’s configuration,
until the desired temperature setpoint is displayed.
•The right button can also be used to request the Strategy to override the schedule and force occupancy mode. Permit
Occupancy Override must be enabled in the controller configuration.
Operation Mode
At the bottom of the display there is a line of icons that give information about the current operation of the
CBT12iVAV or CBT14 Unitary Controller’s Strategy:
Indicates that the controller is operating in ‘heating’
mode.
Indicates that the controller is operating in ‘cooling’
mode.
Indicates that the controller strategy is currently
operating in ‘occupancy’ mode.
Indicates that functions of the Controller Strategy
are in manual override mode.
Indicates that an alarm state is detected in the
controller strategy.
Indicates that the fan is operating.
Indicates that the keypad is operating in
Engineering mode.
c


SECTION 3 : ENGINEERING
MODE

12 © Cylon Controls Ltd. MAN0120US rev 16
CBT-STAT and CBT-STAT-H
ENGINEERING MODE
Keypad and Display
1st line (large digits)
2nd line (small digits)
Operation Mode indicators
Keypad
Left button
Change menu page.
Up button
Change menu line, Increase value.
Right button
Change menu page,
Start parameter editing, Accept changes.
Down button
Change menu line, Decrease value.
c
a
b
d
How to enter Engineering Mode
1. Press and hold (“long-press”) both the Up button and Down button together for at least 3
seconds until the 1st line (large digits) displays the text and the 2nd line
displayswith the right-most digit flashing.
2. Enter the Engineering Mode password by changing the number on the 2nd line as follows:
•Press the up button or down button to change the value of the flashing digit.
•Use the left and right buttons to move between digits.
•The default password is which can be changed over the network.
•To reset the entered password to the default, press and hold the left button
3. When the correct password is displayed, press and hold (“long-press”) the right button to accept it
and to open the first menu page.
How to exit the Engineering Mode
•If you wish to exit Engineering mode at any point without entering the password, press and hold (“long-
press”) both the up and down buttons together.
a
b
c
d
a
b
b

MAN0120US rev 16 © Cylon Controls Ltd. 13
USING THE ENGINEERING MODE MENUS
When the correct password is entered, the Engineering Menu page line will be displayed. This is indicated
in the display as follows:
The 1st line (large digits) shows the Page number and Line number of the commissioning parameters (see
Available parameters on page 14).
The 2nd line (small digits) shows the label for the parameter addressed by the current line.
Navigating through the Engineering Menus
To scroll through the lines on the current page, use the the up and down buttons (short press).
To navigate to a different page, use the right and left buttons (short press).
Editing a parameter
To view or edit the value of the parameter on the current line, press and hold (“long-press”) the right button .
The 2nd line (small digits) will change to show the current value for that parameter, with either the right-
most character (when the value is a number) or the whole text (when the value is an option label) flashing.
•Press the up button or down button to change the value of the flashing digit or label.
Note: In the case of digits, this is not a simple change to the digit itself, but an addition or subtraction
(from the full parameter value) of the value represented by that specific digit.
For example, if the value is +127, then the left-most digit (“1”) represents 100. If that digit is
selected, then the down button will first change it to +027 (subtract 100 from 127), then
another press of the Down button will change it to -073 (subtract 100 from 27).
•Use the left and right buttons to move between digits
•To reset the parameter to its default value, press and hold the left button
•To apply the edited value, press and hold the right button
a
b
b

14 © Cylon Controls Ltd. MAN0120US rev 16
CBT-STAT and CBT-STAT-H
AVAILABLE PARAMETERS
The parameters that can be adjusted are defined by setpoints in the CBT12iVAV or CBT14 Unitary Controller’s
strategy, with the exception of the device parameters on Menu Page 1.
This section lists all of the editable parameters by Menu Page, which can be accessed as detailed in Using the
Engineering Mode menus on page 13.
Refer to MAN0113US CBT12iVAV User Guide, MAN0139US CBV-2U4-3T & CBV-2U4-2T-N, MAN0128US_CBTCBT-
3T6-5R_HeatPump or MAN0130US_CBT-3T6-5R_RoofTopUnit for a detailed explanation of the Unitary Controller
setup and sequencing.
Menu Page 1: Device Setup
Line
Display
Parameter
Range
01
Device Instance
02
MSTP Station
03
Baud Rate
= 9,600 baud,
= 19,200 baud
= 38,400 baud
= 76,800 baud
Menu Page 2: Configuration
Line
Display
Parameter
Range
01
InputConfigA
-128 to +128
02
InputConfigB
-128 to +128
03
OutputConfigA
-128 to +128
04
OutputConfigB
-128 to +128
05
OutputConfigC_Bsbd
-128 to +128
06
AOConfigD
-128 to +128

MAN0120US rev 16 © Cylon Controls Ltd. 15
Menu Page 3: Setpoints
Line
Display
Parameter
Range
01
ZoneOccCoolStpt
02
ZoneOccHeatStpt
03
ZoneUnoccCoolStpt
04
ZoneUnoccHeatStpt
05
SetptSliderSpan
06
TABPriAirMAXStpt
07
TABPriDbandStpt
08
TABPriAirMaxHgtStpt
10
TABFanDbandStpt
11
TABFanMaxStpt
12
TABFanHtgMaxStpt
13
TABFanStandbyStpt
14
TABParallelDisableFanSP
15
SeriesFanStartLckOutSP

16 © Cylon Controls Ltd. MAN0120US rev 16
CBT-STAT and CBT-STAT-H
Menu Page 4: Primary Balancing
Line
Display
Parameter
Range
01
DampClosedMan
02
DamperReverse
(“FWD”“REV”)
03
TABPriAirZero
04
TABPriDuctDia (round equivalent) [ CBT12iVAV only ]
05
KFactorOut [ CBV-2U4-3T(-N) only) ]
TABPriPickupK [ CBT12iVAV only ]
06
PriAirFlowCV
07
EffPriAirFlowSTPt
08
TABPRiKFactor [ CBT12iVAV only ]
09
KFactorSelect [ CBV-2U4-3T(-N) only) ]
TABPRiAirManMode [ CBT12iVAV only ]
10
MeasuredFlow/K [ CBV-2U4-3T(-N) only) ]
TABPRiManStPt [ CBT12iVAV only ]
11
TABPriHoodAtMax [ CBT12iVAV only ]
12
TABAutoCalKFactor [ CBT12iVAV only ]
13
DampMaxMan
14
DampMinMan
15
DampOpenMan

MAN0120US rev 16 © Cylon Controls Ltd. 17
Menu Page 5: Fan Configuration
Line
Display
Parameter
Range
01
DampAuxMan
02
ElecHeatCFMTestStpt
03
EnableManualHeatCmd
04
ManualSlaveHeatCmd
05
FanCmdAnimation
06
EffFanSpeedCmd
07
EnableManualCoolCmd
08
ManualSlaveCoolCmd

18 © Cylon Controls Ltd. MAN0120US rev 16
CBT-STAT and CBT-STAT-H
BALANCING PROCEDURE - CBV-2U4-3T(-N)
Zeroing Airflow sensor
The CBV-2U4-3T(-N) comes from the factory with the airflow sensor zeroed. To compensate for possible air
leakage around the damper, users are able to zero the airflow sensor in the field. To zero the airflow sensor,
users will need to do the following:
1. Drive damper to the closed position by enabling DampCloseMan (P4L01).
2. After damper has fully shut, set TABPriAirZero (P4L03) to on.
After 10 seconds the TABPriAirZero (P4L03) will be automatically set back to off and the airflow sensor will now
be zeroed.
Be sure to disable DampCloseMan (P4L01) when finished with this step.
Airflow Calibration Procedure
This section provides information on the Airflow sensor calibration procedure of the CBV-2U4-3T airflow sensor
using both Flow/K and K select in the VAV Flow Calculation firmware block
Using Flow/K
1. Set KfactSelect (P4L09) to
TRUE
.
2. Set Max cfm flow at TABPriAirMaxStpt (P3L06).
3. Set to DamperMaxMan (P4L13) to
TRUE
.
4. Allow airflow PriAirFlowCV (P4L06) to reach TABPriAirMaxStpt (P3L06). Let damper modulate until
airflow balances out.
5. Enter balancers airflow reading in CFM in MeasuredFlow/K (P4L10) FactorSelect.
6. Again, allow airflow PriAirFlowCV (P4L06) to reach TABPriAirMaxStpt (P3L06). Let damper
modulate until airflow balances.
7. Repeat as necessary.
8. Calculated K factor can be read at KFactorOut(P4L05)
9. Set DampMaxMan (P4L13) to
FALSE
when complete.
2 point balancing with minimum flow can also be achieved by doing the following:
10. After balancing system to the maximum air flow, Set min cfm flow at TABPriDbandStpt(P3L07)
11. Set DampMinMan (P4L14) to
TRUE
12. Allow airflow PriAirFlowCV (P4L06) to reach TABPriDbandStpt (P3L07)
13. Enter balancers airflow reading in CFM in MeasuredFlow/K (P4L10)
14. Again, allow airflow PriAirFlowCV (P4L06) to reach TABPriDbandStpt (P3L07)
15. Calculated K factor can be read at KFactorOut (P4L05)
16. Set DampMinMan (P4L14) to
FALSE
when complete.
Using K Select
1. Set KfactSelect (P4L09) to
FALSE
.
2. Set MeasuredFlow/K (P4L10) to unit manufacturer K-FACTOR
3. Calculated flow can be read at PriAirFLowCV (P4L06)

MAN0120US rev 16 © Cylon Controls Ltd. 19
BALANCING PROCEDURE –CBT12iVAV
This method will auto-calculate and set the K factor P4L08 TABPriKFactor (A214).
To utilize this feature, perform the following steps on Menu Page 4: Primary Balancing.
Note: Each parameter is shown below with its Menu page and line number, its Parameter Name and the
associated (Setpoint Number)
1. After determining what the maximum CFM setpoint is going to be for the VAV box, set
TABPriAirManMode (D204) to .
2. Note the value in TABPriManStpt (A218).
3. Wait for the value of PriAirFlowCV (A194) to stabilize, then:
•If the value of PriAirFlowCV (A194) reaches the value noted in step 2, get a hood
reading at the diffusers to get the actual CFM of the VAV box.
•If the value of PriAirFlowCV (A194) doesn’t reach the value noted in step 2, set
TABPriManStpt (A218) to a value below the value reached by PriAirFlowCV (A194) as a
reference point.
4. Set the TABPriHoodAtMax (A245) to the actual CFM reading that was recorded utilizing the
hood.
5. Set the TABAutoCalKfactor (D151) to . It will stay for 4 seconds then automatically be
set back to . This will automatically set the TABPriKFactor (A214) internally, through the
Strategy calculating the required K factor. It will also set the TABPriAirManMode (D204) to
, and the calculated EffPriAirFlowStpt (A197) will go back to normal operation.
Note: When setting the TABPriAirManMode (D204) to , it will stay on until the time entered
on point TABPriAirManModeTimer (A217) has expired, unless TABAutoCalKfactor (D151) is
set to . The default for TABPriAirManModeTimer (A217) is 600 seconds or 10 minutes.

UNITED STATES
ONE TECHNOLOGY LANE
EXPORT, PA 15632
PHONE: (724) 733-2000
FAX: (724) 327-6124
1 SUNDIAL AVE - SUITE 219 N
MANCHESTER, NH 03103
75 WADE GREEN BUSINESS PARK
1301 SHILOH ROAD NW, SUITE 1411
KENNESAW, GA 30144
IRELAND
CYLON CONTROLS LTD
CLONSHAUGH BUSINESS &
TECHNOLOGY PARK
CLONSHAUGH
DUBLIN 17
IRELAND
T + 353 1 245 0500
F + 353 1 245 0501
UK
CYLON CONTROLS UK LTD
ENDEAVOUR HOUSE
COOPERS END ROAD
ST ANSTED
ESSEX CM24 1SJ
UNITED KINGDOM
T + 44 870 178 1800
F + 44 870 403 6570
WWW.CYLON-AUTOMATRIX.COM
This manual suits for next models
1
Table of contents
Other Cylon Control System manuals
Popular Control System manuals by other brands

McIntosh
McIntosh CR16 - SERVICE Service manual

Adtech
Adtech CNC9 Series Maintenance manual

VIA Technologies
VIA Technologies ARTiGO-A600 Development guide

StarLogixs
StarLogixs CybaMICRO 2011 quick start guide

NPI
NPI INR-01-T Operating instructions and system description

HBX
HBX HydroBloc 0110 installation manual