Dabi Atlante MICRO MOTOR INTRA 32MM C SPRAY TB User manual


2
PRESENTATION OF MANUAL
INSTRUCTIONS FOR USE
ATTENTION
Technical Name: Dental Hand Instruments
Trade Name: Dental Hand Instrument for Low Speed
Model:
• MICRO MOTOR INTRA 32MM - C SPRAY TB
• MICRO MOTOR INTRA 32MM - C SPRAY TM
• MICRO MOTOR INTRA 32MM - S SPRAY TB
• MICRO MOTOR INTRA SL 30 - C SPRAY TB
• MICRO MOTOR INTRA SL 30 - C SPRAY TM
• MICRO MOTOR INTRA SL 30 - S SPRAY TB
• MICRO MOTOR INTRA SL 30 TB SI
• MICRO MOTOR INTRA SL 30 TM SI
Brand: Dabi Atlante
Manufacturer/ Distribuitor:
Alliage S/A Indústrias Médico Odontológica
C.N.P.J. 55.979.736/0001-45 - Insc. Est. 582.002.897.114
Rod. Abrão Assed , Km 53+450m - Cx. Postal 782 CEP 14097-500
Ribeirão Preto - S.P. - Brasil
Phone +55 (16) 2102-5000
Technical Duties: Ricardo J. Ravaneli
CREA-SP: 5060714523
Registro ANVISA nº: 10069210073
For greater safety:
Read and understand all the instructions contained in these instructions for use before installing
or operating this equipment.
Note: These instructions for use must be read by all the operators of this equipment.

3
INDEX
02 PRESENTATION OF MANUAL
04 IDENTIFICATION OF EQUIPMENT
04 Indication of Equipment
04 Purpose of the equipment
05 Description of the Equipment
06 MODULES, ACCESSORIES, OPTIONS AND MATERIALS OF CONSUMPTION
08 TECHNICAL SPECIFICATIONS
08 General features
08 Dimensions
09 Packing symbols
09 Product symbols
10 Abbreviations
11 OPERATION OF EQUIPMENT
12 PRECAUTIONS, RESTRICTIONS AND WARNINGS
12 Recommendations for preserving the equipment
12 Transportation, storage and operation
12 Sensibility to predictable environmental conditions in regular use conditions
12 Precautions and warnings “during the use” of equipment
13 Precautions and warnings “after” the use of equipment
13 Precautions in case of alteration in the functioning of equipment
13 Precautions to be adopted against foreseeable or uncommon risks, related to the deactivation
and abandoning of equipment
14 CORRECTIVE AND PREVENTIVE MAINTENANCE AND PRESERVATION
14 Maintenance of the MIDWEST / BORDEN Type Lining
14 Maintenance of O-Ring
15 Additional procedures for reuse
15 Cleaning and Disinfection
15 Lubrication in spray
16 Sterilization
16 Preventive Maintenance
16 Corrective Maintenance
17 UNFORESEEN EVENTS – SOLUTION OF PROBLEMS
17 EQUIPMENT’S WARRANTY
17 FINAL CONSIDERATIONS

4
IDENTIFICATION OF EQUIPMENT
Dear Customer
Congratulations. You have made a good choice when you decided to buy a Dabi Atlante product
comparable to the best products available in the World. This manual is a general presentation of
your product and it will give you important details to help you to solve possible problems.
Please, read it and keep this with you.
IDENTIFICATION
Technical Name: Dental Hand Instruments
Trade Name: Dental Hand Instrument for Low Speed
Model:
• MICRO MOTOR INTRA 32MM - C SPRAY TB
• MICRO MOTOR INTRA 32MM - C SPRAY TM
• MICRO MOTOR INTRA 32MM - S SPRAY TB
• MICRO MOTOR INTRA SL 30 - C SPRAY TB
• MICRO MOTOR INTRA SL 30 - C SPRAY TM
• MICRO MOTOR INTRA SL 30 - S SPRAY TB
• MICRO MOTOR INTRA SL 30 TB SI
• MICRO MOTOR INTRA SL 30 TM SI
Brand: Dabi Atlante
Indication of Equipment
This equipment is for dental use use only. It must be operated and utilized by specialized professional
(certied professional, according to the legislation of the country) and following the instructions of
the manual. The operation of the equipment required, for the professional, the utilization of correct
instruments and it should to be in perfect conditions of the use, and to protect the professional,
the patients and others, in the eventual danger situation.
Purpose of the equipment
The Dabi Atlante handpieces are intended to remove caries, remove restorations and perform tooth
sectioning , as an aid in extracting teeth.

5
Description of Equipment
Compact handpiece, ergonomic design, body made of metal with surface treatment. Rotor supported
on bearings providing a low noise level, speed adjustment in clockwise and counterclockwise
directions, adapts to any contra angle or straight piece handpiece in the INTRA system.
The Dabi Atlante handpieces have been planned to oer maximum performance in the most severe
clinical procedures. With a design which caters for ergonomics, torque, sensitivity and lightness.
Produced with technology (MADE IN BRAZIL) and material of the utmost durability.
Internal irrigation System (according to model acquired): without the need for adaptations, the
internal irrigation system allows direct passage of water ow to the counter angle or straight piece
with the power required for the formation of the spray.
To ensure safe operation of your equipment, use only the mounting assembly (chair, equip, water
unit and reector) provided by Dabi Atlante Authorized Reseller Service.
IDENTIFICATION OF EQUIPMENT

6
The contents of this page are of an informative nature, the equipment being able to dier
from that illustrated. So, upon acquiring the product check the technical compatibilty
between equipment, coupling and accessories.
* Optional
MODULES, ACCESSORIES, OPTIONS AND MATERIALS OF CONSUMPTION
External Spray
Irrigation System
Terminal type
*BORDEN
Terminal type
*MIDWEST
Internal Spray
Irrigation System
01
c c c
02 03
* Available in BORDEN and MIDWEST versions
c - Control ring
01 - Micromotor Intra 32MM / SL 30 - C Spray TB / TM
02 - Micromotor Intra 32MM / SL 30 - S Spray TB / TM
03 - Micromotor Intra SL 30 TB / TM SI (internal spray)

7
MODULES, ACCESSORIES, OPTIONS AND MATERIALS OF CONSUMPTION
01 02 03
* Optional
01 - Coupling lining (TB/TM)
02 - O-ring
03 - Lubricant Lubs
*
Attention
-The use of any part, accessory or material not
specied or foreseen in these instructions for use is
entirely the user’s responsibility.

8
Work pressure
Maximum 275,79 kpa = 40 psi = 2,75 bar
Air consumption
54 l/min
Speed
De 5.000 à 20.000 rpm
Fitting terminal
Borden Type 02 holes
Midwest Type 04 holes
Coupling
via the INTRA system
Straight Piece and Contra Angle
Sterilization
Autoclavable up to 135ºC
TECHNICAL SPECIFICATIONS
General features
Weight
Model Net Weight Gross Weight
32 MM 77g 137g
SL 30 80g 140g
ABBREVIATIONS
MM - Micromotor
S - Spray
C Spray - With Spray
S Spray - Without Spray
SI - Internal Spray
TM - Midwest Type
TB - Borden Type

9
Packing symbols
Product symbols
TECHNICAL SPECIFICATIONS
Indicates the ideal temperature for
sterilization in autoclave. Recyclable.
General warning:
If the instructions are not properly
followed, the use can lead to
damage to the product, user or
patient.
Note:
It indicates useful information for
operation of the product.
Warning
See the manual. Borden Type Terminal.
It determines the maximum
quantity of boxes which can be
stacked during transportation and
storage “as per packaging”.
Packing to be transported and /
or stored avoiding humidity, rains
and wet oor.
Packing to be transported and /
or stored with the harrows up.
The packing must be stored and
transported away from direct sun
light exposure.
Packing to be transported and / or
stored with care (should not suer
drop and neither receive impact).
Temperature limit for the packing
to be stored or transported.

10
Midwest Type Terminal. Midwest type Terminal - Optical
Fiber (LED / UV LED lighting).
TECHNICAL SPECIFICATIONS
Dimensions (mm)
BORDEN TYPE TERMINAL - WITH SPRAY
MIDWEST TYPE TERMINAL - WITH SPRAY

11
OPERATION OF EQUIPMENT
Terminal Coupling
Make sure that the instrument connection is compatible with the equipment (type Borden or type
Midwest as gures (a) and (b). Connect Micro Motor to equip terminal, thread the ring of the terminal
turning it clockwise until it is fastened.
Before using it for the rst time, follow the lubrication procedures.
Warning
It is indispensable that the driving air is dry, clean and uncontaminated. Dirty and moist
air lead to premature wear of the bearings.
For the handpiece to function well it is necessary to have a driving pressure of 275.79 kpa = 40
psi = 2.75 bar.
Ring lock/unlock Ring lock/unlock
g. a
g. d
g. c
Regulator Ring
g. b
Type BORDEN Type MIDWEST
Coupling of Micro Motor with handpieces (Straight Piece and Contra Angle)
Attaching the Contra Angle or Straight Part in the coupling of Micro Motor (g. c). To disengage by
pressing, pull the handpiece.
Adjusting speed and direction of rotation
To adjust the speed and direction of rotation, turn the regulator ring clockwise or counterclockwise
as required (g. d).

12
Recommendations for preserving the equipment
Your Dabi Atlante equipment has been designed and developed according to the standards of
modern techology. Similarly to other kinds of equipment, it requires special care, which is many times
neglected due to several reasons and circunstances. Therefore, here are some important reminders
for your daily routine. Try to follow these simple rules, which will save you a lot of time and will
avoid unnecessary expenses once they start making part of your working procedure.
Transportation, storage and operation
This equipment must be transported and stored observing the following directions:
- Avoid falls and impacts;
- Keep it dry, do not expose it to rain, water drops or wet oor;
- Keep it away from water and direct sunlight, and in it original wrapping;
- Don’t move it over irregular surfaces, protect it from rain and observe the maximum stack quantity
specied in the packaging;
Environmental condition for transportation or storage:
- Room temperature range for transportation or storage -29ºC to +60ºC.
- Relative humidity range for transportation or storage 20% to 90%.
- Atmospheric pressure range 500hPa to 1060 hPa (375 mmHg to 795 mmHg).
Environmental operation condition:
- Room temperature range for functioning +10ºC to +40ºC.
- Recommended room temperature range +21ºC to +26ºC.
- Relative humidity range for functioning 30% to 75%.
- Atmospheric pressure range 700 hPa to 1060 hPa (525 mmHg to 795 mmHg).
- Operation altitude: ≤2000m.
PRECAUTIONS, RESTRICTIONS AND WARNINGS
ATTENTION
The Equipment maintains its condition of safety and ecacy, provided that it is maintained (stored)
as mentioned in this instruction of use. Thus, the equipment will not lose or alter its physical and
dimensional features.
Sensibility to predictable environmental conditions in regular use conditions
The equipment has been planned not to be sensitive to interference such as magnetic elds,
external electrical factors, electrostatic discharge, pressure or variance of pressure, provided that
the equipment is installed, maintained, clean, preserved, transported and operated as per this
instruction for use.
Precautions and warnings “during the use” of equipment
- The equipment should only be operated by duly enabled and trained technicians (Dental Surgeons,
Capacitated Professionals).
- Before the rst use and/or after long interruptions from work such as vacations, clean and
disinfect the equipment.
- If any maintenance should be required, only use services of the Dabi Atlante Authorized Technical
Assistance.
- Owing to its delicate components, this handpiece must not be dropped or undergo impacts.
- Although this equipment has been planned in accordance with the standards of electromagnetic
compatibility, it can, in very extreme conditions, cause interference with other equipment. Do not
use this equipment together with other devices very sensitive to interference or with devices which

13
PRECAUTIONS, RESTRICTIONS AND WARNINGS
Precautions and warnings “after” the use of equipment
- Always maintain the equipment clean for the next operation.
- Do not modify any part of the equipment. Do not disconnect the cable or other connections
without need.
- After using the equipment, clean and disinfect all the parts which may be in contact with the patient.
Precautions in case of alteration in the functioning of equipment
- If the equipment has any abnormality, check if the problem is related to any item listed in the topic of
unforeseen events (failures, causes and solutions). If it is not possible to resolve the problem, contact
your representative (Dabi Atlante).
Precautions to be adopted against foreseeable or uncommon risks, related to
the deactivation and abandoning of equipment
In order to avoid environmental contamination or undue use of the Equipment after it has become
useless, it should be discarded in the suitable place (as per the local legislation of the country).
- Pay attention to the local legislation of the country for the conditions of installation and disposal
of residue.
create high electromagnetic disturbance.
- Do not expose the plastic parts to contact with chemical substances, use in the routines of dental
treatment, such as: acids, mercury, acrylic liquids, amalgams, etc.
- The use of pneumatic or electric tools to perform any type of recording on handpieces, could lead
to damage to the surface treatment of the product, causing the loss of warranty.
Dabi Atlante shall not be responsible for:
- Use of the equipment diering from that for which it is intended.
- Damages caused to the equipment, the professional and/or the patient by the incorrect installation
and erroneous procedures of maintenance, diering from those described in these Instructions for
use which come with the equipment or by the incorrect operation of it.

14
CORRECTIVE AND PREVENTIVE MAINTENANCE AND PRESERVATION
Maintenance of the BORDEN Type Lining
Insert a screwdriver of 1.2mm in the face of the lining as per (g. a), using it as a lever until removing
it from the hole.
Remove the damaged lining (g. b).
Place the new lining in the hole leaving the holes aligned, an with a screwdriver push the lining
carefully downwards (g. c).
Maintenance of the MIDWEST Type Lining
Use the same procedure as the BORDEN type.
Maintenance of O-Ring
Change:
Insert a screwdriver of 1.2mm in the channel of the O-ring. 1, moving it outwards (g. a), rst
removing the ring from channel 1 and then from channel 2, (g. b).
In order to fasten the new ring, execute the opposite process without the screwdriver (g. c).
g. a
g. a
g. a
g. b
g. b
g. b
g. c
g. c
g. c

15
CORRECTIVE AND PREVENTIVE MAINTENANCE AND PRESERVATION
Additional procedures for reuse
The equipment can be reused in undetermined, i.e. unlimited, quantities, only needing to be cleaned
and disinfected.
Cleaning and Disinfection
These processes should be used in the handpieces.
Uncouple the piece from the INTRA t terminal.
By using a clean, soft cloth, clean the outside by employing alcohol at 70% or neutral soap.
Do not use microabrasive material or steel wool when cleaning, or employ organic solvents
ordetergents which contain solvents such as ether, stain remover, etc.
Spray lubrication
To carry out the lubrication of your equipment, we recommend the use of the Lubs product.
Lubrication must be carried out by the posterior hole of micromotor as illustrated in g. (a). Firmly
t the oil applicator nozzle into the hole indicated in g. (a) and apply the spray for 1 or 2 seconds
(g. b) by holding the vial in vertical position.
Perform these operations (spray lubrication) with the regulating ring positioned clockwise and
counterclockwise.
Before of use, activate a handpiece during some seconds to eliminate the excess of oil. Cover the
piece with absorbent paper to remove the excess of oil.
Warning
Never submerge the instrument in a disinfection bath.
Never use product peracetic acid for cleaning, since it is corrosive.
The lubricant doesn’t follow the product.
Attention:
The non-use of the lubricant will result
In the loss of product warranty.
Terminal type
BORDEN
Channel To
Lubrication
(Entry of air)
Channel To
Lubrication
(Entry of air)
Terminal type
MIDWEST
g. a g. b
NOTE: Handpieces no submitted to process
of sterilization should be lubricated at
least twice a day.
control ring

16
CORRECTIVE AND PREVENTIVE MAINTENANCE AND PRESERVATION
Sterilization
Follow the instructions of the manufacturer of the appliance.
Only sterilize clean and lubricated handpieces.
Clean, grease and remove the overow oil in the apparatus according to the previous indicate, put
it in the surgical grade paper (not provided by Dabi Atlante) and sterilize it in autoclave until 135°C
(275°F) through a cycle.
The bit breaker can be placed into an autoclaved up to 135°C (275°F) during one cycle.
Preventive Maintenance
The equipment must suer routinely measurements, following the current legislation of the country.
But, never with a period superior to 3 years.
For protecting your equipment, look for a Dabi Atlante’ technical assistance for periodic reviews
as preventive maintenances.
Corrective Maintenance
Dabi Atlante states that the supplying of the circuits’ diagram, Part lists or any other information
that permits the technical assistance by the user, can be requested, since previously agreed between
the buyer and Dabi Atlante.
Warning
Sterilize before 1st use, product do not provided sterilized.
Not valid or recommend to the sterilization by ethilene oxyde.
Do not sterilize in an oven.
Before sterilization, connect it to the delivery unit and activate it for a few seconds to
eliminate the excess of oil.
If the excess of oil is not eliminated, it may cause burnt residue in the bearing.
Attention
In case of the equipment presents any abnormality; check if the problem is related to some
of the listed items under the item Unpredictable (situation, cause and solution).
If it’s not possible to solve the problem, shutdown the equipment and demand the presence
of a Dabi Atlante’ technician from the nearest resale, or ask through the Attendance Service
Dabi Atlante: + 55 (16) 3512-1212.

17
Problem Probable cause Solution
-The micromotor does not
function.
-The control ring is in the inter-
mediary position between the
counterclockwise and clockwi-
se directions of speed.
-Lack of lubrication or inade-
quatelubrication.
-The connection of the instru-
ment / terminal is loose.
- Bearing damaged.
-Turn the control ring in a
clockwise or counterclockwise
direction.
-Lubricate as per the indica-
tions.
-Tighten the connection.
-Send to the authorized ser-
vice.
-Insucient speed. -Lack of air or low air pressure.
-Lack of lubrication or Inade-
quate lubrication.
-Check the air pressure in the
network.
-Lubricate as per the indica-
tions.
-Failure and noise in the func-
tioning.
-Air leakage.
-Lack of lubrication.
-Examine the hoses and con-
nections.
-Lubricate as indicated.
Upon coming across any problem in operation, follow the instructions below to check and
repair the problem, and/or get in touch with your representative.
FINAL CONSIDERATIONS
Among the care you have to take with your equipment, the most important is regarding of the spare
parts replacement.
To ensure the lifetime of your device, only replace original spare parts from Dabi Atlante. They have the
assurance of the standards and technical specications required by the Dabi Atlante representative.
We call your attention to our authorized resellers’ chain. Only this chain will keep your equipment
constantly new, because it has trained technical assistant and specic tools for the correct maintenance
of your device.
Whenever you need, demand the presence of a Dabi Atlante’ technician from the nearest resale, or ask
through the Attendance Service Dabi Atlante: + 55 (16) 3512-1212.
EQUIPMENT’S WARRANTY
This equipment is covered by the warranty terms and norms contained in the Warranty Certicate
that accompany the product.
UNFORESEEN EVENTS – SOLUTION OF PROBLEMS

18

19

This manual suits for next models
7
Table of contents
Other Dabi Atlante Dental Equipment manuals

Dabi Atlante
Dabi Atlante Spectro 70X Series User manual

Dabi Atlante
Dabi Atlante PROFI NEO User manual

Dabi Atlante
Dabi Atlante Syncrus G3 User manual

Dabi Atlante
Dabi Atlante RAIOS X Spectro 70X User manual

Dabi Atlante
Dabi Atlante EAGLE 3D User manual

Dabi Atlante
Dabi Atlante EAGLE 3D User manual

Dabi Atlante
Dabi Atlante Bellagio User manual

Dabi Atlante
Dabi Atlante COBRA CONTRA ANGLE Series User manual

Dabi Atlante
Dabi Atlante COBRA CONTRA ANGLE INTRA 32 CA User manual

Dabi Atlante
Dabi Atlante EAGLE User manual